Journal of Clinical Microbiology, December 2002, p. 4757-4759, Vol. 40, No. 12
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.12.4757-4759.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Department of Pathology, Division of Laboratory Medicine, University of Louisville,1 Clinical Microbiology Laboratory, University of Louisville HospitalLouisville, Kentucky 40202,2 the Clinical Microbiology Laboratory, William Beaumont Hospital, Royal OakMichigan 480733
Received 18 April 2002/ Returned for modification 12 July 2002/ Accepted 18 September 2002
|
|
|---|
|
|
|---|
In an effort to improve workplace safety for personnel who are responsible for the collection of blood cultures, a plastic blood culture bottle has been developed (bioMérieux, Inc., Durham, N.C.) for the purpose of providing a safer product by minimizing the exposure to infectious agents associated with breakage of glass bottles. The plastic bottle has been designed for use in the BacT/ALERT microbial detection system and utilizes two materials in a multilayer configuration to provide transparency, gas impermeability, and durability to withstand terminal sterilization, features previously available in a glass bottle (A. S. Massey and E. B. Moser, Abstr. 101st Gen. Meet. Am. Soc. Microbiol., abstr. C-5, 2001). Additionally, the plastic bottle does not require venting, which reduces the risk of needle stick injury and eliminates a step in the processing of bottles. The plastic bottle also contains a liquid emulsion sensor and provides a larger headspace for the proper atmospheric conditions. Preliminary studies comparing the performance of the plastic and glass versions of the BacT/ALERT standard aerobic, standard anaerobic, and charcoal-containing media demonstrated equivalent performance and showed that the plastic bottle provides a practical alternative to the current glass bottle (Massey and Moser, Abstr. 101st Gen. Meet. Am. Soc. Microbiol.; J. Walsh, M. Chi, C. Ronsick, B. Robison, and T. Thorpe, Abstr. 101st Gen. Meet. Am Soc. Microbiol., abstr. C-4, 2001). The study described here was designed to compare and demonstrate that the performance of the BacT/ALERT plastic standard aerobic (PSA) system, when used with the BacT/ALERT microbial detection system, is at least equivalent to the vented glass BacT/ALERT standard aerobic (VA) culture bottle for recovery and speed of detection of microorganisms in suspected cases of bacteremia or fungemia. A previous study had shown that the performance of a nonvented aerobic glass bottle was equivalent to that of the standard aerobic glass bottle; the latter is no longer commercially available (5).
This study was conducted in two tertiary-care medical centers, the University of Louisville Hospital (ULH), Louisville, Ky., and the William Beaumont Hospital (WBH), Royal Oak, Mich. Blood samples were collected by the phlebotomy team and nursing personnel at ULH and by the phlebotomy team at WBH (2, 4, 5). An aliquot of ca. 20 ml was obtained from patients with clinically suspected bacteremia or fungemia. Equal aliquots of blood were aseptically inoculated into each bottle with the VA bottle being inoculated first. Blood volumes were measured in the laboratory by comparing each bottle to known volume standards. Only blood culture sets with differences in fill volumes between bottles of
30% of the larger volume were included in the analysis. Bottle pairs that did not meet these criteria were excluded from the study but were processed to maximize recovery of microorganisms from each culture. All VA bottles were transiently vented to air for at least 30 s prior to placement in the BacT/ALERT incubator module; PSA bottles were loaded directly. Cultures were incubated for a total of 5 days, with terminal subcultures performed at random on 20% of the instrument-negative PSA bottles to accurately establish the false-negative rate. When a bottle signaled positive, broth from the vial was Gram stained and subcultured onto sheep blood and chocolate agars, respectively. All isolates were identified by standard microbiologic procedures (3). False-positive bottles (i.e., those which had a positive instrument signal but were Gram stain negative and subculture negative) were reincubated until growth occurred, whereupon the bottles were reflagged as suspected positive, or until the original 5-day incubation period had expired. All bottles were processed independently of the other bottles in a given set, that is, a negative bottle was not examined when the other bottle in the set was flagged as a suspected positive. If these bottles remained negative for 5 days, they were terminally subcultured. Instrument-negative bottles that grew an organism on terminal subculture were categorized as false negative. The clinical significance of recovered microorganisms was assessed by chart review using published criteria (6).
To determine differences between the two aerobic bottles (PSA and VA) with respect to clinical sensitivity, McNemar's test for paired samples was applied to the positive yield data generated for compliant sets (1). A probability value of less than 0.05 would indicate a significant difference in the clinical sensitivities of the two bottles. The signed-rank test was used to determine the differences between the PSA and VA bottles with respect to time to detection (1). Differences were analyzed using a separate signed-rank test for each of the bacterial species. A probability value of less than 0.05 would indicate a significant difference in the time to detection between bottles. False-positive and false-negative rates were determined as the ratio of the number of false positives or false negatives to the total number of cultures tested.
Results for recovery and time to detection were compared for adequately filled bottle pairs, as defined above.
A total of 3,797 paired, PSA and VA samples were received for culture, of which 3,552 were adequately filled pairs. There were 407 positive cultures detected in one or both bottles. The PSA bottles detected 324 (9.1%) positives, 8 (0.23%) false positives, and no false negatives, while the VA bottles detected 291 (8.2%) positives, 4 (0.11%) false positives, and 1 (0.03%) false negative. The one false negative detected in the VA grew a Corynebacterium sp. on subculture. The corresponding PSA was positive for coagulase-negative staphylococci and Corynebacterium spp.
Of the 205 (201 monomicrobic and 4 polymicrobic) positive cultures with clinically significant organisms, 146 (71.2%) were detected by both bottles, 35 (17.1%) were detected by the PSA bottle only, and 24 (11.7%) were detected by the VA bottle only. The comparative yields of clinically significant bacteria and fungi from the two aerobic culture bottles are summarized in Table 1. There was no statistically significant difference between bottles in recovery of clinically significant organisms. The overall higher yield of microorganisms in the plastic bottle is most likely due to the increased headspace in the bottle (80 ml versus 73 ml in the vented bottle). The content of the plastic bottle headspace is 20% carbon dioxide in oxygen versus the same concentration of carbon dioxide in ambient air for the vented bottle. The latter also has a slightly higher vacuum.
|
View this table: [in a new window] |
TABLE 1. Comparative yields of clinically significant microorganisms in BacT/ALERT PSA and VA blood culture bottlesa
|
|
View this table: [in a new window] |
TABLE 2. Clinical isolates recovered from septic episodes of monomicrobic (143) and polymicrobic (2) bacteremia or fungemia detected by BacT/ALERT PSA and VA culture bottlesa
|
|
View this table: [in a new window] |
TABLE 3. Comparison of mean times to detection of growth when the same clinically significant microorganism was isolated from both bottlesa
|
This study was supported by a grant from bioMérieux, Inc.
|
|
|---|