Journal of Clinical Microbiology, February 2002, p. 461-465, Vol. 40, No. 2
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.40.2.461-465.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
The JONES Group/JMI Laboratories, North Liberty,1 CAST Laboratories of the University of Iowa College of Medicine, Iowa City, Iowa,3 The Tufts University School of Medicine, Boston, Massachusetts ,2 Boehringer Ingelheim Vetmedica, Inc., St. Joseph, Missouri4
Received 11 September 2001/ Returned for modification 28 October 2001/ Accepted 27 November 2001
|
|
|---|
4 µg/ml for susceptibility and
32 µg/ml for resistance. Correlate zone diameters for specific nonfastidious species were as follows: for Pasteurella multocida and staphylococci tested on Mueller-Hinton agar, susceptibility at
19 mm and resistance at
11 mm, and for Actinobacillus suis, Erysipelothrix rhusiopathiae, and Streptococcus suis tested on enriched chocolate Mueller-Hinton agar, susceptibility at
16 mm and resistance at
8 mm. When Actinobacillus pleuropneumoniae was tested, a susceptibility breakpoint of
16 µg/ml (
9 mm) was suggested for veterinary fastidious medium broth and enriched chocolate Mueller-Hinton agar. Absolute categorical agreement between NCCLS dilution and disk diffusion test results with these criteria ranged from 90.5 to 96.2%. Tiamulin susceptibility testing methods appear to be accurate in their categorical classification for indicated species, and their availability will allow immediate testing of animal isolates to guide therapy via appropriate levels of dosing and to monitor the development of resistance for agents in this unique class. |
|
|---|
1 µg/ml) against anaerobic bacterial species, intestinal spirochetes, and Mycoplasma spp. (3, 12, 13). The present study was initiated in order to document the sustained antimicrobial activity of tiamulin against indicated bacterial species and rapidly growing and fastidious animal pathogens and to establish choices of diagnostic reagents. MIC and disk diffusion zone diameter correlates were determined in order to ensure the precision and accuracy of the susceptibility testing procedures performed according to National Committee for Clinical Laboratory Standards (NCCLS) guidelines for susceptibility testing of veterinary antimicrobial agents (5, 6). |
|
|---|
|
View this table: [in a new window] |
TABLE 1. Tiamulin activity as demonstrated by MIC and zone diameter of inhibition results tested against 392 nonfastidious bacteria
|
|
View this table: [in a new window] |
TABLE 2. Tiamulin activity as demonstrated by MICs and zone diameters of inhibition for 244 fastidious veterinary pathogens with three broth medium formulations
|
The 30-µg tiamulin disks were produced by BD Microbiology Systems (Cockeysville, Md.), and these reagents were used following NCCLS methods M2-A7 (8) and M31-A (5). The nonfastidious isolates were tested on Mueller-Hinton agar (MHA), and the fastidious isolates were tested on chocolate MHA (CMHA). The fastidious organisms were tested using incubation conditions of 35°C in 5 to 7% CO2 for 20 to 24 h and the nonfastidious organisms were tested in ambient air at 35°C for 16 to 18 h. The MIC endpoints and zone diameters of inhibition were read as defined by the NCCLS documents (5, 7, 8). The results were compared by scattergram, regression statistics, and error-rate bounding analysis. Breakpoints for susceptibility were proposed based on available tiamulin pharmacokinetic data (Boehringer Ingelheim Vetmedica, Inc., data on file) and NCCLS (6, 9) guidelines. Quality control was performed with A. pleuropneumoniae ATCC 27090 and S. aureus ATCC 25923 tested against enrofloxacin; all results were within MIC and zone diameter limits established by Marshall et al. (4), NCCLS (5), and the reagent manufacturer.
In addition, preliminary experiments were performed to determine the optimal drug disk content by using investigator-prepared disks containing 5, 15, 30, 60, and 90 µg of tiamulin. Each disk was tested (10 replicates) against the NCCLS quality control strains: Enterococcus faecalis ATCC 29212 and S. aureus ATCC 25923, 29213, and 43300 (methicillin resistant). A 15-µg erythromycin disk was used as a control drug. Furthermore, Neo-Sensitabs (Rosco Diagnostica, Taastrup, Denmark) at 30 µg were tested against control strains with ranges 1 to 2 mm larger than those recorded for the corresponding disk diffusion method. This susceptibility testing method is often used in Europe.
|
|
|---|
Previous experiments in Denmark (1) using A. pleuropneumoniae strains (26 total) demonstrated average zone diameters of 20.3 mm at pH 8.0 and smaller zones at pH 7.2 to 7.4. The average MIC was 3.1 µg/ml (4 to 8 µg/ml at pH 7.4), and a susceptible breakpoint was suggested at
8 µg/ml, with a correlate zone diameter of
13 mm. However, these zones were based on technical or methodological details which did not conform to NCCLS documents for antimicrobial susceptibility tests with bacterial strains of animal origin, including (i) a tablet drug delivery system, (ii) semiconfluent growth inoculum (lighter than NCCLS-recommended confluent growth), and (iii) modified or elevated pH due to not using CO2 incubation, which favors fastidious-pathogen growth (5, 8). All of these modifications should result in reduced zone diameters of inhibition around 30-µg tiamulin disks used in the NCCLS procedure.
Susceptibility testing for tiamulin against nonfastidious pathogens. Organisms listed in Table 1 were also tested with the 30-µg tiamulin disk (5). Two distinct populations of susceptible organisms were identified based on relative tiamulin potency: staphylococci, for which the MICs at which 90% of strains were inhibited (MIC90s) were 1 or 2 µg/ml and zone diameters were >20 mm, and P. multocida strains, for which MICs were 8 to 32 µg/ml and zone diameters were 10 to 21 mm. All other organisms listed in Table 1, such as enterococci and enteric and nonfermentative gram-negative bacilli, were resistant to tiamulin, with MICs being >32 µg/ml and median zone diameters being 6 to 12 mm.
To accurately categorize these two groups of tiamulin-inhibited organisms, the staphylococci were declared susceptible (MIC breakpoint,
4 µg/ml) and P. multocida was categorized as moderately susceptible or intermediate (MIC range, 8 to 16 µg/ml). The correlate zone breakpoints were
19 mm for susceptibility and
11 mm for resistance (Fig. 1). These criteria produce an intermethod accuracy rate of 96.2%, and errors were as follows: very major (false susceptible) = 0.0%, major (false resistant) = 0.0%, and minor = 3.8%. All error rates were acceptable, and the results should indicate the need for possible dosage adjustments to address infections caused by strains for which tiamulin MICs are expected to be 8 or 16 µg/ml, such as P. multocida (see discussion of A. pleuropneumoniae below). The correlation coefficient was 0.98 (Fig. 1). Previously reported in vitro experience confirms our results (1, 2). This potency was similar to that noted for two fastidious species of Actinobacillus (Table 2).
![]() View larger version (15K): [in a new window] |
FIG. 1. Scattergram comparing reference tiamulin MICs to zone diameters around 30-µg tiamulin disks for 390 strains of nonfastidious, rapidly growing organisms. Horizontal lines show the proposed susceptibility ( 4 µg/ml) and resistance ( 32 µg/ml) MIC criteria. Vertical lines show the correlate interpretive zone diameters of 19 and 11 mm, respectively (96.2% intermethod agreement).
|
To establish susceptibility testing methods and remain consistent with the dose-related criteria suggested for nonfastidious species, the same MIC breakpoints (
4 µg/ml for susceptibility;
32 µg/ml for resistance) were applied to these species. However, two distinct scattergram populations were also identified, where E. rhusiopathiae produced markedly larger zones around 30-µg tiamulin disks (slow grower) than S. suis strains; MICs for the two species in VFM (5) broth were the same (Fig. 2). Regardless of this phenomenon, a single set of interpretive criteria could easily be established for the disk diffusion method:
16 mm for susceptibility and
8 mm for resistance.
![]() View larger version (18K): [in a new window] |
FIG. 2. Scattergram comparing reference tiamulin MICs in VFM broth compared to zone diameters around 30-µg tiamulin disks when testing the fastidious veterinary pathogens: A. pleuropneumoniae, A. suis, E. rhusiopathiae, and S. suis (226 total strains). Horizontal and vertical solid lines indicate the proposed MIC breakpoints of 4 µg/ml ( 16 mm) as susceptible and 32 µg/ml as resistant for non-Actinobacillus species. Broken horizontal and vertical lines show the NCCLS-approved interpretive breakpoints for A. pleuropneumoniae (5). Circled numbers are non-Actinobacillus results.
|
16 µg/ml (
9 mm) for susceptibility and
32 µg/ml (
8 mm) for resistance. Intermethod absolute categorical accuracy was 90.5% for the 170 A. pleuropneumoniae strains and 90.1% for all 191 Actinobacillus strains tested. A similar modification (no intermediate range) could be applied to P. multocida among the nonfastidious species.
H. parasuis isolates were tested by comparing reference tiamulin MICs (in HTM) (5) and the zone diameters around 30-µg disks (on CMHA). Wide variations were noted in each test (data not shown), but the criteria for other fastidious species allowed acceptable accuracy between methods (7, 8) for testing tiamulin against H. parasuis. The small number of isolates tested (18 total; less than target numbers of
100 strains), the wide zone range variations at each MIC, and the use of HTM (not VFM) as the broth medium require further test development before definitive interpretive criteria can be recommended for this agent or any other.
In conclusion, tiamulin has a reputation as an effective veterinary-pathogen-specific antimicrobial that remains active against anaerobes, intestinal spirochetes, Actinobacillus spp., Pasteurella spp., and many common isolates of staphylococci and streptococci of animal and human origin (1-3, 12, 13). The proposed interpretive criteria (accepted by the NCCLS for A. pleuropneumoniae) for testing tiamulin against key indicated pathogens by NCCLS methods (5) will allow expanded testing of this agent and the accumulation of susceptibility statistics to monitor emerging resistance to the pleuromutilin class. These interpretive criteria are supplemented by recently published guidelines for quality control of tiamulin dilution and disk diffusion tests (5, 10). Care must be exercised to test the veterinary pathogens on appropriate media (5, 6) and to apply the species-specific breakpoint criteria and quality control procedures.
The study was funded, in part, by an educational/research grant from Boehringer Ingelheim Vetmedica, Inc.
|
|
|---|