Journal of Clinical Microbiology, April 2002, p. 1518-1521, Vol. 40, No. 4
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.4.1518-1521.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Service of Infectious Diseases, Hospital Carlos III, Instituto de Salud Carlos III,1 Service of Microbiology, Hospital Doce de Octubre,2 Abbott Laboratories, Madrid, Spain3
Received 4 June 2001/ Returned for modification 23 October 2001/ Accepted 6 December 2001
|
|
|---|
|
|
|---|
In this study, we have evaluated the performance of a new viral load test, the Abbott LCx quantitative assay, for the measurement of HIV-1 in plasma from subjects carrying different HIV-1 variants. The inter- and intrarun reproducibilities of the assay, its specificity and linearity, and its ability to detect and accurately quantitate specimens from subjects infected with non-B subtype HIV-1, including group O viruses, were evaluated. Finally, a comparison with the results obtained by use of Cobas-Monitor version 1.5 (11) and Quantiplex version 3.0 (2) was performed. Both assays have a detection limit of 50 copies/ml.
Plasma samples were randomly collected from HIV-1-seropositive patients attending two clinical centers in Madrid, Spain. Ten specimens taken from HIV-1-infected patients and two samples taken from the BBI HIV-1 RNA quantification panel (Boston Biomedica Inc., West Bridgewater, Mass.) were tested in triplicate on four separate days. The estimated HIV RNA concentration spanned the dynamic range of LCx, as determined by using the 0.2-ml procedure (detection limit of 178 HIV RNA copies/ml). The intraassay, interassay, and total assay variabilities were assessed. The same lot of LCx reagent and the same controls were used throughout the complete reproducibility study. The same technician performed all the runs.
Human EDTA-treated plasma samples from 41 healthy donors found to be negative for antibodies to HIV-1 or HIV-2 by an enzyme immunoassay (Abbott) were obtained from a blood bank. All samples were tested by LCx by using the 1.0-ml procedure (detection limit of 50 HIV RNA copies/ml). Initially reactive samples were tested twice by LCx and another viral load test.
Samples obtained from an HIV-1-infected patient carrying high levels of plasma HIV RNA and culture supernatants from cells infected with the BRU and BLA strains of HIV-1 were serially diluted fivefold in seronegative plasma and in culture medium, respectively. Then, they were tested in parallel by using LCx and Quantiplex.
A total of 362 plasma samples were selected for the comparative analysis of LCx and the other viral load assays (all of them were tested with Quantiplex and 101 were tested with Cobas-Monitor). The 1.0-ml protocol of LCx was used for the comparative analyses. Moreover, 64 samples from 24 HIV-1-infected patients carrying group M non-B subtypes (15 A, 10 C, 2 F, 35 G, and 2 H) and 24 samples from 5 HIV-1 group O-infected patients were examined. Viruses from all individuals included in this study had been previously genetically characterized (3, 4, 7, 8).
Viral burden in samples from HIV-1 group O-infected patients was assessed by using a p24 antigen test which has a sensitivity of 2 to 5 pg/ml (Murex, Dartford, United Kingdom) and MUPROVAMA, a noncommercial RT-PCR assay based on the simultaneous amplification of overlapping fragments of a single HIV RNA target sequence in a single reaction (6).
Assay linearities for LCx and Quantiplex were assessed graphically by parallel-line analysis related to dilution factor. Averages were statistically compared by using the paired t test. Linear regression and correlation analyses were used to determine assay relationships. Qualitative analyses comparing the number of samples with fewer or more than 50 HIV RNA copies/ml in different methods (LCx, Quantiplex, and Cobas-Monitor) were carried out by using a concordance method based on the
index.
The inter-and intrarun reproducibilities were assessed by using specimens that were tested once in four different runs. Total variation ranged from 1.37% ± 0.09% to 13.06% ± 0.39% (Table 1). A greater variation was noted in the range of 2.62 to 2.95 log copies/ml, near the lower limit of quantitation of the LCx assay (2.25 log copies/ml).
|
View this table: [in a new window] |
TABLE 1. Precision of LCx using the 0.2-ml sample extraction procedure
|
Serially diluted culture supernatants were quantitated in parallel by using LCx and Quantiplex. Excellent parallel-line linearity was observed across all dilutions assayed (Fig. 1). A strong correlation between both tests was noted (R = 0.98). However, HIV RNA values determined by LCx were uniformly higher (0.21 log unit, on average) than those determined by Quantiplex, although the difference was not statistically significant.
![]() View larger version (23K): [in a new window] |
FIG. 1. Linear comparison between LCx and Quantiplex.
|
![]() View larger version (22K): [in a new window] |
FIG. 2. Correlation between LCx and Quantiplex (a) and Cobas-Monitor (b).
|
A qualitative comparison based on the lower limit (<50 HIV RNA copies/ml) showed a concordance of 88% between LCx and Quantiplex, with a
index of 72%. Thirty-four samples found positive by LCx were found negative by Quantiplex, and 9 samples found positive by Quantiplex were found negative by LCx. In both situations, positive values did not reach statistically significant differences. When we compared LCx with Cobas-Monitor, we obtained a concordance of 93%, with a
index of 79%. Only five samples found positive by Cobas-Monitor were found negative by LCx, and only two samples found positive by LCx were found negative by Cobas-Monitor.
A total of 88 specimens belonging to non-B subtypes were tested by using the LCx 0.2-ml protocol. The results obtained for 64 specimens carrying group M non-B subtypes were compared with those obtained by using Quantiplex. The results obtained for the remaining 24 specimens, obtained from group O-infected persons, were compared with those generated by using an in-house RT-PCR, MUPROVAMA (6), as well as by using a quantitative p24 antigen test.
Of the 64 non-B subtype specimens, 41 (67.2%) were quantitated by LCx, whereas Quantiplex quantitated 29 (45.3%) (P < 0.05). Of note, 19 samples below the limit of detection of LCx (178 copies/ml) were drawn from subjects on antiretroviral therapy, which might have truly suppressed viral replication in those subjects. In addition, viral loads in LCx tended to be higher than those in Quantiplex. Differences above 0.5 log unit were seen in 36% of the total specimens, 40% of the subtype C specimens, and about 26% of the subtype A and G specimens (Table 2).
|
View this table: [in a new window] |
TABLE 2. Results obtained for quantifying HIV-1 non-B subtypes using LCx and Quantiplex
|
|
View this table: [in a new window] |
TABLE 3. Positive results for testing HIV-1 group O specimens using LCx, a p24 antigen assay, and a noncommercial RT-PCR assay
|
We thank John Hackett for reviewing the manuscript.
|
|
|---|