Journal of Clinical Microbiology, May 2002, p. 1681-1686, Vol. 40, No. 5
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.5.1681-1686.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Sezione di Microbiologia, Dipartimento di Patologia, Università di Verona, 35100 Verona, Italy
Received 12 July 2001/ Returned for modification 27 September 2001/ Accepted 5 February 2002
|
|
|---|
|
|
|---|
In the last 20 years, a variety of automated systems for the identification and antimicrobial susceptibility testing (AST) of microorganisms has been developed based on automated interpretation of the results of biochemical tests or using microdilution trays following overnight incubation and photometric determination of growth (6, 18, 20, 21). Advances in technology that may provide rapid bacterial identification and AST are now recognized as having both clinical and financial benefits (2).
The VITEK system originated in the 1970s as an automated system for identification and AST and has evolved today into the VITEK 2 system, which automatically performs all of the steps required for identification and AST after a primary inoculum has been prepared and standardized (9). This system allows kinetic analysis by reading each test every 15 min. The optical system combines multichannel fluorimeter and photometer readings to record fluorescence, turbidity, and colorimetric signals.
The purpose of this study was to evaluate the VITEK 2 system (software version VT2-R01.02) for identification and AST of gram-positive cocci belonging to Staphylococcus spp ., Enterococcus spp ., Streptococcus agalactiae, and Streptococcus pneumoniae. Because of the increased incidence of diseases caused by these microorganisms and the emergence of resistance to several antimicrobial agents (3, 7, 8, 12, 15, 23), rapid and accurate identification as well as MIC evaluation for these pathogens has become increasingly important.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Performance of the VITEK 2 system for identification of gram-positive cocci compared to that of the API method
|
Comparative identification methods. All isolates were tested on appropriate API strips (bioMérieux), including API Staph for members of the Micrococceae family and API 20 Strep for members of the Streptococcaceae family. Other testing performed included serologic typing for beta-hemolytic streptococci, coagulase test for Staphylococcus aureus, motility test and yellow pigment production for enterococci, and optochin susceptibility for S. pneumoniae (5, 11, 17).
Identification with VITEK 2.
The test panels (ID-GPC) contained 46 fluorimetric tests that included pH change tests and derivatives to detect aminopeptidases and -osidases. Substrates used for detection of aminopeptidases are coupled with 7-amino-methylcoumarin (7AMC); substrates for the detection of -osidases are usually coupled with 4-methylumbelliferone (4MU). The 21 test substrates are as follows: 4MU-
-L-arabinofuranoside, 4MU-
-D-galactoside, 4MU-
-D-glucoside, 4MU-
-D-N-acetylneuraminic acid, 4MU-ß-D-galactoside, 4MU-ß-D-glucoside, 4MU-ß-D-glucuronide, 4MU-ß-D-mannoside, 4MU-n-acetyl-ß-D-glucosaminide, 4MU-phosphate, alanine-7AMC, arginine-7AMC, aurease (butiloxicarbonyl-Val-Pro-Arg-AMC), histidine-7AMC,
-glutamic acid-7AMC, threonine-7AMC, lysine-7AMC, phenylalanine-7AMC, proline-7AMC, pyroglutamic acid-7AMC, and tyrosine-7AMC. Furthermore, the ID-GPC card includes 16 fermentation tests (for D-raffinose, amygdaline, arbutine, D-galactose, glycerol, D-glucose, L-arabinose, lactose, D-maltose, D-mannitol, N-acetylglucosamine, salicin, D-sorbitol, D-trehalose, D-melibiose, and D-xylose), two decarboxylase tests (for ornithine and arginine), and six miscellaneous tests (for urease, pyruvate, optochin, novobiocin, polymyxin B sulfate, and 6% NaCl).
The card was automatically filled by a vacuum device, sealed and inserted into the VITEK 2 reader-incubator module (incubation temperature, 35.5°C), and subjected to a kinetic fluorescence measurement every 15 min. The results were interpreted by the ID-GPC database, and final results were obtained automatically. All cards used were automatically discarded into a waste container.
Agar dilution reference method. This method was performed according to the NCCLS guidelines (13, 14). A 0.5 McFarland bacterial suspension was used directly in this test. Antimicrobial concentration ranges used and breakpoints are shown in Table 2.
|
View this table: [in a new window] |
TABLE 2. Antimicrobial concentration ranges and breakpoints used in AST
|
High-concentration aminoglycoside testing. The agar screen reference method for gentamicin and streptomycin was performed according to NCCLS methods (13, 14) in brain heart infusion agar and using gentamicin at 500 µg/ml and streptomycin at 2,000 µg/ml.
Genetic methods. The presence of mecA (staphylococci) and vanA and vanB (enterococci) genes, as well as genes for high-level aminoglycoside resistance in enterococci, was confirmed by PCR as previously described (1, 4, 16, 22).
AST with VITEK 2. The 0.5 McFarland bacterial suspension was diluted to 1.5 x 107 CFU/ml in 0.45% saline. Cards were automatically filled, sealed, and loaded into the VITEK 2 instrument for incubation and reading. The AST-P515 card used for staphylococci contained benzylpenicillin, clindamycin, erythromycin, fusidic acid, gentamicin, kanamycin, lincomycin, minocycline, nitrofurantoin, norfloxacin, ofloxacin, oxacillin, pristinamycin, rifampin, teicoplanin, tetracycline, tobramycin, trimethoprim-sulfamethoxazole, and vancomycin; the AST-P516 card used for enterococci and S. agalactiae contained ampicillin, ampicillin-sulbactam, cefuroxime, ciprofloxacin, clindamycin, erythromycin, high-concentration (HC) gentamicin, imipenem, HC kanamycin, levofloxacin, nitrofurantoin, norfloxacin, ofloxacin, quinupristin-dalfopristin, HC streptomycin, teicoplanin, tetracycline, trimethoprim-sulfamethoxazole, and vancomycin; the AST-P506 card used for pneumococci contained amoxicillin, benzylpenicillin, cefotaxime, ceftriaxone, chloramphenicol, erythromycin, imipenem, ofloxacin, pristinamycin, tetracycline, trimethoprim-sulfamethoxazole, and vancomycin. The antibiotic-microorganism combinations selected for evaluating the performance of the VITEK 2 system in this study are shown in Table 2.
Analysis of identification results. There are four possibilities for analysis of identification results: (i) correct identification, in which strains are correctly identified to the species level or strains with low discrimination are resolved (two or more species are proposed by the VITEK 2 system [one being identical to the reference method], and the identification can be resolved by simple additional tests); (ii) low discrimination, in which strains with low discrimination are not resolved (two or more species are proposed by the VITEK 2 system [one being identical to the reference method], and the identification cannot be resolved by simple additional tests); (iii) misidentification, in which discrepant results are obtained for strains (one or more species proposed by the VITEK 2 system are different from those identified by the reference method); and (iv) no identification, in which no identification was provided. The mean time for result generation was also calculated for all identifications.
Analysis of susceptibility testing. There are two possibilities for analysis of susceptibility testing. (i) Category agreement (CA) may occur. In CA, the MICs determined by both methods identified the category of microbial susceptibility as susceptible, intermediate, or resistant according to NCCLS interpretative criteria (13, 14). (ii) Discrepancies may occur. Discrepancies were considered very major errors (VME) when the VITEK 2 system indicated susceptibility and the reference method indicated resistance, were considered major errors (ME) when the VITEK 2 system indicated resistance and the reference method indicated susceptibility, and were considered minor errors (mE) when the VITEK 2 system indicated intermediate susceptibility and the reference method indicated susceptibility or resistance or when the VITEK 2 system indicated susceptibility or resistance and the reference method indicated intermediate susceptibility.
Quality control. The quality control strains Enterococcus casseliflavus ATCC 700327, Kocuria rosea ATCC 186, S. aureus ATCC 29213, Staphylococcus lugdunensis ATCC 700328, Staphylococcus sciuri ATCC 29061, S. pneumoniae ATCC 6301, and Streptococcus uberis ATCC 9927 were included for identification; S. aureus ATCC 29213, Enterococcus faecalis ATCC 29212, E. faecalis ATCC 51299, Escherichia coli ATCC 35218, and S. pneumoniae ATCC 49619 were included for AST.
|
|
|---|
Low-discrimination identifications were frequent among coagulase-negative staphylococci (CNS) (10%), whereas only one misidentification occurred for S. aureus and one misidentification occurred for CNS. Three Staphylococcus epidermidis isolates were not identified. The only S. aureus isolate misidentified by the VITEK 2 system was identified as Staphylococcus warneri.
Five Enterococcus faecium isolates were not discriminated from other enterococcal species: four were not discriminated from Enterococcus hirae and Enterococcus durans, and one was not discriminated from E. casseliflavus-Enterococcus gallinarum. The latter discrepancy was resolved by motility test and yellow pigment production, whereas the others were not resolved since no simple test is available to discriminate among E. faecium, E. hirae, and E. durans. Three E. durans isolates were not discriminated from or were misidentified as E. hirae. Four strains (two E. faecalis strains and two E. faecium strains) were not identified.
One strain of S. agalactiae was not discriminated from Streptococcus acidominimus, but the discrepancy was resolved by immunological testing, and one strain was not identified.
For 14 S. pneumoniae isolates the correct identification was proposed by the VITEK 2 system as an alternative to other streptococcal species. All of the low discrimination identifications were resolved by the performance of the optochin test. Two strains were not identified.
Of the total gram-positive strains, discordant identifications were seen with 1.6% of the isolates and no identification was observed for 2.6% of the isolates. All quality control strains were correctly identified by the VITEK 2 system.
Time to obtain identification. The VITEK 2 system identified all S. agalactiae and all S. pneumoniae isolates within 2 h; 98% of S. aureus isolates and 71% of CNS were identified in 3 h. Six hours was required to identify the remaining isolates, i.e., two S. aureus isolates and 29 CNS.
AST. MICs of each antimicrobial agent generated by the VITEK 2 system were compared with each MIC determined by the NCCLS reference procedure (13, 14). A susceptibility category was also assigned to each MIC based on the current NCCLS breakpoint criteria (13, 14). Since there were a number of off-scale MICs that could not be compared, the agreement was defined by CA and not by essential agreement (VITEK 2 system MIC results within 1 doubling dilution from the reference result). Discrepant results for oxacillin (staphylococci), high-concentration gentamicin and streptomycin (enterococci), and vancomycin and teicoplanin (enterococci) were resolved by genetic methods which identified the gene(s) responsible for resistance. The final error rates were calculated using the values obtained when errors were resolved.
In total, 2,483 drug-organism combinations were tested. Performance compared to reference standard testing is shown in Table 3.
|
View this table: [in a new window] |
TABLE 3. Performance of the VITEK 2 system for AST of gram-positive cocci compared to that of the reference method
|
For enterococci, VME and mE were found with teicoplanin in particular. Table 4 shows that the VITEK 2 system correctly identified the glycopeptide resistance phenotype for the five reference strains and that most VME were clustered at the susceptible breakpoint.
|
View this table: [in a new window] |
TABLE 4. Glycopeptide MICs of reference strains and glycopeptide-resistant clinical isolates with discrepant results
|
Four penicillin-intermediate and two penicillin-susceptible strains of S. pneumoniae were identified by the VITEK 2 system as resistant and intermediate, respectively. Two cefotaxime-susceptible strains were identified as intermediate. The only penicillin-resistant strain in our collection was correctly identified.
Time to obtain MIC. The time to obtain MICs ranged from 6 h (20% of S. aureus, 5% of CNS, 49% of enterococci, 33% of S. agalactiae, and 0% of S. pneumoniae) to 17 h (time required to complete all CNS AST). Ninety percent of AST was completed in 8 h for S. aureus, in 11 h for CNS, in 9 h for enterococci, in 7 h for S. agalactiae, and in 9 h for S. pneumoniae.
|
|
|---|
In our study, the VITEK 2 system demonstrated similar rates of accuracy in identification of staphylococci. Taxa with low-discrimination identification results included a very limited number of strains to provide useful information for improvements in the VITEK 2 system, and none of the problematic reactions for the misidentified strains was significantly more frequently observed than others. The fact that CNS were not identified as well as S. aureus could be explained by the slow metabolism of some CNS strains, leading to ambiguous results in the reaction wells. This was also observed in evaluation of other automated identification systems with CNS (15).
A relatively low rate of correct identifications was observed in our study with E. faecium (71.4%). More recently Garcia-Garrote et al. (10) reported the accuracy of the VITEK 2 system as 98.3 and 76.3% in the identification of E. faecalis and E. faecium, respectively; in that study the latter species was mostly misidentified as E. casseliflavus-E. gallinarum. In our study most E. faecium isolates were misidentified as E. hirae or E. durans, and these discrepancies could not be resolved since simple tests are not available to discriminate among these species, whereas a simple motility test can discriminate E. faecium from E. casseliflavus-E. gallinarum. We have no explanations for accuracy shown by the VITEK 2 system in our study and for the prevalence of misidentification with E. hirae or E. durans. Nevertheless, in the clinical setting, reasons for species identification of enterococci are very limited (serious infections, such as endocarditis, or epidemiological surveillance within hospitals). In general, presumptive identification to the genus level together with determination of susceptibility is considered to be sufficient (12). Several taxonomy changes have been introduced in the Enterococcus genus, mainly involving species other than E. faecalis (5). The difficulties of commercial tests in the identification of species other than E. faecalis probably reflects problems with the new species definition or in the criteria available for identifying them.
Antimicrobial susceptibility test.
The Food and Drug Administration has established minimal performance characteristics to assess antimicrobial susceptibility tests (20). These guidelines indicate that CA should be >90%, ME should be <3%, and VME should be
1.5%. Our investigation indicates that the VITEK 2 system provides accurate susceptibility test results, since the agreement with the reference method was well above, and the rates of discrepancies (VME and ME) were far below, the limit for all the antibiotic-microorganism combinations.
The detection of oxacillin-resistant staphylococci with the VITEK 2 system was very accurate. Even if VME or ME were observed in CNS, the rates were below the recommended limit. Our results agreed with those reported in the few studies which have assessed the performance of the VITEK 2 system in AST of staphylococci, reporting for oxacillin resistance detection, with agreement rates ranging from 95 to 100% (M. E. Reverdy, L. Carret, and A. Quaglia, Abstr. 8th Eur. Congr. Clin. Microbiol. Infect. Dis., abstr. P284, 1997; R. Gross, B. Grünastel, K. Becker, and G. Peters, Abstr. 8th Eur. Congr. Clin. Microbiol. Infect. Dis., abstr. P657, 1997; C. Bradford, L. Meeh, D. Freiner, J. Rader, and J. Gerst, Abstr. 98th Gen. Met. Am. Soc. Microbiol., abstr. C479, 1998; C. Shubert, R. Griffith, W. McLaughlin, M. Ullery, and M. Peyret, Abstr. 98th Gen. Met. Am. Soc. Microbiol., abstr. C478, 1998; J. Krzyszton-Russjan, J. Walory, and K. Nowak., Abstr. 9th Eur. Congr. Clin. Microbiol. Infect. Dis., abstr. P827, 1999).
In our study the VITEK 2 system correctly detected all vancomycin-resistant enterococci but failed to identify 4 of the 22 teicoplanin-resistant isolates. Failures in detection of glycopeptide resistance were also reported by others (10), who, however, reported a VME rate for vancomycin higher than that for teicoplanin. In our study, vancomycin resistance was correctly identified in all strains, and the VME with teicoplanin would have as a consequence the misidentification of the precise glycopeptide resistance phenotype. Correct identification of the phenotype or genotype can be subsequently performed with supplementary tests. In addition, it should be stressed that the high VME rate observed for the teicoplanin-enterococci combination is due in part to the fact that the selection of organisms included in our study favored resistant strains. These problems may not be seen in the largely glycopeptide-susceptible enterococci population commonly encountered in clinical microbiology laboratories at this time in Italy (7).
Agreement among results for ampicillin, HC gentamicin, and HC streptomycin resistance was found to be better in our study than in a previous one (10), but we examined a lower number of resistant strains.
The results of VITEK 2 system AST for S. agalactiae correlated highly with those of the reference method and are similar to those reported by others (Ghanem et al., Abstr. 98th Gen. Met. Am. Soc. Microbiol., abstr. C484, 1998). In this species, the resistance usually encountered is that to tetracycline, and the VITEK 2 system correctly identified tetracycline resistance in all strains. One susceptible strain was identified as resistant. This ME may not be significant, since tetracycline is not the drug of choice for treatment of infections caused by S. agalactiae.
Finally, no VME or ME were observed in the AST results for S. pneumoniae, but higher MICs of penicillin, cefotaxime, and ofloxacin causing mE rates were seen. The VITEK 2 system performed very well for erythromycin-resistant strains, which are encountered in Italy more frequently than penicillin-resistant strains (3).
In conclusion, this study validates the VITEK 2 system as an easy-to-use system that provides a rapid (4 to 15 h) and accurate means for identification and AST of most commonly isolated species of gram-positive cocci.
We thank Cécile Casado, Jean-Pierre Marcel, and Randy Turner for useful discussions.
|
|
|---|