Journal of Clinical Microbiology, June 2002, p. 1989-1993, Vol. 40, No. 6
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.6.1989-1993.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Johns Hopkins University School of Hygiene and Public Health,1 Johns Hopkins University School of Medicine, Baltimore, Maryland,3 Unidade de Pesquisa em Tuberculose, Hospital Universitario Clementino Fraga Filho, Universidade Federal do Rio de Janeiro, Rio de Janeiro, Brazil2
Received 15 November 2001/ Returned for modification 17 January 2002/ Accepted 13 March 2002
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
An antibody-based serological test for TB has long been sought. Such a test would be attractive because of its potential technical simplicity, rapidity, low cost, and lack of reliance on sputum (which sometimes can be difficult to obtain or uninformative). Unfortunately, most serological tests for TB have had low sensitivities and low specificities (2). Studies that use recombinant culture filtrate antigens indicate that antibody responses (at least to the antigens studied) in TB patients may be heterogeneous from patient to patient (12). Incorporation of multiple antigens in a diagnostic test may therefore increase the sensitivity of an antibody-based test. For the antibody-based tests evaluated to date, the presence of anti-M. tuberculosis antibodies in persons with latent M. tuberculosis infection, prior active TB disease, vaccination with Mycobacterium bovis bacillus Calmette-Guerin (BCG), or infection with mycobacteria other than M. tuberculosis may have contributed to the low specificities of the tests for the diagnosis of active TB.
The ICT Tuberculosis test (AMRAD Corporation, Melbourne, Victoria, Australia) is a rapid card-based test for the detection of immunoglobulin G (IgG) antibodies directed against five purified M. tuberculosis antigens immobilized in four lanes on a test strip and uses an anti-human IgG labeled with colloidal gold. Since the target proteins are secreted by actively growing organisms, the test has the potential to have a high degree of specificity for active TB disease. Incorporation of multiple antigens may result in a high degree of sensitivity. After the introduction of the first ICT Tuberculosis test (ICT-1), a manufacturer-modified version (ICT-2) was introduced. Modifications included reductions in the concentrations of proteins in bands 1, 2, and 4; a reduction in the width of the nitrocellulose strip; and reductions in the specimen pad size, conjugate pad size, and conjugate volume. ICT-1 was labeled by the manufacturer for use with serum specimens, and ICT-2 was developed to potentially include whole-blood and plasma specimens.
The purpose of the study described here was to evaluate the performance of the ICT Tuberculosis tests for the diagnosis of PTB in a setting with a high prevalence of TB.
| MATERIALS AND METHODS |
|---|
|
|
|---|
10 mm). A chest X ray and a sputum specimen were not obtained solely for study purposes for the controls since all controls had a prior chest radiograph not compatible with active TB and no symptoms at the most recent annual evaluation or at study entry. Persons less than 16 years of age were excluded from the study.
Demographic and clinical data were collected for all study participants through a standardized questionnaire, interview, and physical examination. Venous blood was collected from all participants. Study participants underwent a two-step TST. An induration of
10 mm in the first or second step was considered a positive TST result. Patients suspected of having PTB were also given a postanterior and lateral chest X ray and were tested for HIV (by enzyme-linked immunosorbent assay with confirmation by Western blotting).
The sputa of all patients suspected of having PTB were tested for acid-fast bacilli (AFB) by the Ziehl-Neelsen technique and were cultured in Löwenstein-Jensen medium. Patients unable to expectorate sputum or for whom two consecutive specimens of expectorated sputum were smear negative for AFB underwent sputum induction with a 3% hypertonic saline solution generated with a DeVilbiss ultrasonic nebulizer (Ultra Neb 99; Home and Aid Healthcare Inc.). All specimens that were culture positive for mycobacteria were tested by standard biochemical methods to distinguish M. tuberculosis from other nontuberculous mycobacteria. The reference standard for the diagnosis of PTB was growth of M. tuberculosis in a culture of sputum. The Caracas clinical case definition (18) was used to establish a diagnosis of AIDS.
Venous blood was drawn from the study participants and placed into glass tubes without preservative or anticoagulant for separation of serum and into EDTA- or heparin-coated tubes for whole blood and separation of plasma. The serum remaining after testing by ICT-1 was frozen at -20°C for retesting at a later date by ICT-1 and ICT-2. Whole blood from seven patients suspected of having PTB was not retained for testing; plasma but not serum was obtained for seven patients suspected of having PTB, and serum but not plasma was obtained for seven other patients suspected of having TB. These participants were maintained in the study since the performance of each test with each type of specimen was evaluated separately.
ICT-1 and ICT-2 were performed and their results were interpreted according to the instructions of the manufacturer by an investigator masked to the clinical and microbiological status of the study subjects. For ICT-1, 50 µl of serum, whole blood, or plasma was tested as soon as possible on the day of collection. Each positive band was scored for location (bands 1 to 4, with band 1 at the bottom of the card) and for intensity (0, no signal; 1, weak signal; 2, strong signal). A positive signal for any antigen was considered an overall positive ICT-1 result. All whole-blood, serum, and plasma specimens were tested by ICT-1. In order to assess the impact of freezing and thawing of the serum samples on the test results, three fresh specimens were put through 1, 5, and 10 freeze-thaw cycles and then retested by ICT-1. At approximately 3 months after they had been frozen, the sera were retested by ICT-1 to assess whether the sera had deteriorated during freezing and were concomitantly tested by ICT-2 to assess whether the manufacturer's modifications of the ICT Tuberculosis test improved its performance. For ICT-2, 35 µl of serum was tested after the serum was thawed. The results were recorded and interpreted as described above.
All data were entered into a database by the investigators. Analyses were carried out with Stata Statistical Software (release 6.0, 1999; Stata Corporation, College Station, Tex.) and Epi Info (version 6.04b, 1997; Centers for Disease Control and Prevention, Atlanta, Ga.). The sensitivity and specificity of tests with each antigen were assessed separately, as were the overall sensitivities and specificities of ICT-1 and ICT-2. Yates' corrected P values were used for comparisons of the sensitivities and the specificities. Results were stratified by TST (skin test positivity), history of vaccination with BCG, history of previous TB, smear results for AFB, HIV infection, and AIDS.
| RESULTS |
|---|
|
|
|---|
|
ICT-1. Six whole-blood specimens from patients suspected of having PTB gave invalid results by ICT-1 because the specimens clotted before they ran down the nitrocellulose strip. One serum specimen from a patient suspected of having PTB yielded invalid results because the serum was viscous and did not run down the nitrocellulose strip within 2 min in two attempts.
The specificities of ICT-1 for controls and patients suspected of having PTB are shown in Table 2. Among the controls, the specificities ranged from 71 to 100%, while among patients suspected of having PTB the specificities were poorer, ranging from 46 to 67%. Although for all specimen types the specificities of ICT-1 were lower for the group B controls than for the group A controls, these differences were not statistically significant. In addition, the differences in the specificities for the group B controls with respect to TST status and vaccination with BCG were not significant. The specificities of ICT-1 for the different specimen types from patients suspected of having PTB were similar, and there were no significant differences in specificities for subgroups of patients of having PTB.
|
|
50% and the NPVs were 82 to 87%.
|
For three serum specimens tested, multiple freeze-thaw cycles did not affect the ICT-1 results (data not shown). After 3 months of frozen storage, all serum samples were retested by ICT-1, and the calculated sensitivities and specificities from the first to the second round of testing were similar (data not shown).
ICT-2. A total of 102 frozen serum specimens were tested by ICT-2, of which 60 were from patients suspected of having PTB, 21 were from the group A controls, and 21 were from the group B controls. There were no invalid ICT-2 results for the frozen sera.
The specificities of ICT-2 for controls and patients suspected of having PTB are shown in Table 2. Among the group B controls, the specificity of ICT-2 was not significantly different for TST-positive versus TST-negative patients or for patients with a history of vaccination with BCG versus those without a history of vaccination with BCG. For patients suspected of having PTB, the specificities of ICT-2 were not different among the subgroups indicated in Table 2.
The sensitivities of ICT-2 for patients with PTB are shown in Table 3. The sensitivity of ICT-2 was higher for patients with PTB whose sputum was smear positive for AFB than for patients with PTB whose sputum was smear negative for AFB, but the difference was not statistically significant. The sensitivities of ICT-2 were not significantly different for the subgroups of patients with PTB indicated in Table 3. The PPVs and NPVs of IVT-2 are shown in Table 4. Overall, the manufacturer's modifications to ICT-1 did not substantially improve its performance.
For the group B controls, patients with PTB, and patients suspected of having PTB but found not to have PTB, there were no significant differences in the distributions of positive test bands, the intensities of positive band signals, and the numbers of positive bands (data not shown).
| DISCUSSION |
|---|
|
|
|---|
The ICT Tuberculosis tests are designed to detect IgG antibodies to five antigens secreted by M. tuberculosis in a technically simple and inexpensive card format. Since antibody responses to at least some M. tuberculosis antigens may be heterogeneous from person to person (12), the use of multiple antigens would be predicted to increase the sensitivity of the test. In our study, the sensitivities of ICT-1 and ICT-2 were similar or slightly higher than those reported previously (3, 14). However, the predictive value of a positive test result (PPV) or a negative test result (NPV) was too low to be clinically useful for our study population.
In this study, the sensitivities of ICT-1 and ICT-2 were higher for patients with PTB whose sputum was smear positive for AFB than for patients with PTB whose sputum was smear negative for AFB, although the number of patients with PTB whose sputum was smear negative for AFB included in the study was small and the differences were not statistically significant. This trend has been observed in other studies of antibody-based serodiagnostic tests for TB (1, 2, 15, 16). Whether this simply reflects a quantitatively greater antibody response in patients whose sputum is smear positive for AFB and has a high bacillary burden (11, 17) or a qualitatively different immunological response to M. tuberculosis is not clear and warrants further study. The low sensitivity of the ICT Tuberculosis tests for patients with PTB included in the study whose sputum was smear negative for AFB is disappointing, because this group of patients would potentially derive benefit from a rapid diagnostic blood test. However, since the overall PPVs of the ICT Tuberculosis tests with serum were calculated to be 66 to 67% in settings with a very high prevalence of PTB (>50% prevalence of PTB among patients suspected of having PTB), further evaluation of the ICT Tuberculosis tests with serum from larger numbers of patients whose sputum is smear negative for AFB may be warranted in order to more definitively assess the tests' utility in settings with a very high prevalence of TB but without a laboratory able to perform mycobacterial culture.
The calculated high NPVs (94 to 96%) of the ICT Tuberculosis tests in a setting with a very low prevalence of PTB (a <10% observed prevalence of PTB among patients suspected of having PTB) raises the possibility that the ICT Tuberculosis tests may have some utility in excluding the diagnosis of PTB in such settings.
A major strength of this study is the evaluation under field conditions of the specificities of ICT-1 and ICT-2 for PTB in a population of patients who were suspected of having PTB and who had symptoms and/or clinical signs compatible with PTB. In this population, in which the prevalence of PTB among patients suspected of having PTB was 29%, the low specificities of ICT-1 and ICT-2 for PTB and the relatively low PPVs and NPVs limited the usefulness of the tests. An additional strength of this study is the evaluation of the patterns of positive test bands and the total number of positive test bands for the study subjects. Although neither of these was found to be useful in distinguishing subjects with PTB from those without PTB, to the best of our knowledge this type of analysis has not been performed previously for the ICT Tuberculosis tests.
A potential limitation of this study is that M. tuberculosis culture was used as the reference standard for the diagnosis of PTB. It is possible that some patients suspected of having PTB had culture-negative PTB disease and were improperly assigned to the group considered not to have PTB, thereby reducing the observed specificities of the ICT Tuberculosis tests. However, this is unlikely since none of the patients who were suspected of having PTB but who were assigned to the group considered not to have PTB were reported as having confirmed TB to the TB surveillance system of the state of Rio de Janeiro during the year following the conclusion of the study. In addition, relatively few cases of PTB were identified in this prospective study. This limited the subgroup analysis and contributed to wide 95% CIs. Finally, there was a significant age difference between the group B controls and both the group A controls and patients suspected of having PTB. This age difference reflected the initiation of mandatory vaccination of all Brazilian infants with BCG in 1976. While this age difference is not optimal, endemic controls were used because the inclusion of nonendemic controls would have severely confounded the results.
In conclusion, neither ICT-1 nor ICT-2 was sufficiently predictive for use as a routine diagnostic aid for PTB in a setting with a 29% observed prevalence of PTB among patients suspected of having PTB. However, further evaluation of the performance of the ICT Tuberculosis tests in selected settings may be warranted. A better understanding of the repertoire and dynamics of antibody responses in patients with M. tuberculosis infection and other mycobacterial infections may facilitate the development of more sensitive and specific antibody-based methods for the diagnosis of active PTB disease.