Journal of Clinical Microbiology, June 2002, p. 2314-2315, Vol. 40, No. 6
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.6.2314-2315.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
| LETTER TO THE EDITOR |
| LETTER |
|---|
|
|
|---|
A critical point is the use of collection tubes adapted for pneumatic transport systems. Limited vacuum aspiration is sufficient to destroy brittle structures, increasing the amount of debris and contributing to electronic background "noise" (1).
Two different UF-100 software versions (original up to version 00-12, revised up to version 00-14) are currently being used. The algorithms for the detection of bacteria ("BACT" channel) differ between the two versions. "H-BACT," a highly sensitive bacterial count, is displayed separately in the original version. Zaman et al. (9) used the revised version, in which the former H-BACT particles are added to the BACT channel of the instrument to generate one single result. The revised apparent BACT count is approximately 20 times higher than the original BACT count. Due to this different definition of the BACT channel, lowering of the electronic threshold in this channel makes the system more sensitive to noise. Thrombocytes and debris may account for false-positive bacterial counts in cerebrospinal fluid samples (8).
We have compared both software versions using a set of 46 samples. Data for bacteria obtained by the two algorithms are not comparable in UTI screening. By using the former version, the following relationship between the white blood cell (WBC) count (y, cells/microliter) and the BACT count (x, cells/microliter) was obtained: log y = 0.914 log x - 0.989 (with r = 0.737). For the new algorithm, the correlation was weaker: log y = 0.977 log x - 2.140 (with r = 0.598). The poorer performance of the new software was due to the fact that small debris particles were considered microorganisms.
Okada et al. (using the new software) calculated a sensitivity of 83% and a specificity of 76.4% (6). Differences in the diagnostic capabilities of flow cytometry in earlier studies (sensitivity, 55%; specificity, 90%) (2) can be explained by the different software versions used. Zaman et al. reported a low sensitivity for UTI screening. However, the consideration of flagging was ignored. The UF-100 analyzer is able to detect interference with erythrocyte, crystal, and yeast cell counts, which might lead to a misclassification of these elements and an underestimation of bacteria. This is then flagged appropriately. When no attention has been paid to flagging, some false-negative results may arise. Furthermore, "positive" or "low reliable" flags for UTI are generated by UF-100 based on a comprehensive judgment of the relevant data provided by bacterial count, bacterial size (forward scatter), and WBC count. In case a large number of nonbacterial particles are detected as BACT by UF-100, the judgment of this three-parameter-rule system is low reliable.
Manufacturer-set reference values should not be confused with cutoff values. Attempts to establish reference values by an international multicenter study have failed because of preanalytical differences. Very low cutoffs for pathogenicity were suggested by Zaman et al. compared to those found in other literature (3, 7).
Zaman et al. excluded samples based only on culture results. Definition of a positive culture does not comply with particle analysis but is a combination of microbiological colony counting and clinical validation. Samples with contamination had been excluded. But the UF-100 counts dead and viable bacteria (4), regardless of contamination (mixed growth) or infection (pure colony).
The preanalytical phase is extremely important in urinalysis (4, 5). The flow cytometric evaluation of additional results for various cell types offers great help in identifying samples of poor preanalytical quality that have been received. Both epithelial cell counts and the WBC/BACT ratio may be helpful. Moreover, flow cytometry offers the advantage of measuring conductivity as well (4, 5, 7), allowing us to correct the effect of sample dilution on cell counts.
Although urine flow cytometry is not a perfect technique, the additional data provided by this novel technology allow a more balanced interpretation of the apparent bacterial count. For better prediction of UTI, algorithms using a narrower gate for the bacterial channel are preferred.
| REFERENCES |
|---|
|
|
|---|
|
Joris R. Delanghe* Michel R. Langlois Birgitte Wuyts Marc L. De Buyzere Department of Clinical Chemistry Ghent University De Pintelaan 185 B 9000 Ghent, Belgium
|
||||||
|
* Phone: 32 9 240 29 56 Fax: 32 9 240 49 85 E-mail: joris.delanghe{at}rug.ac.be |
| LETTER |
|---|
|
|
|---|
4 m/s, this interference can be avoided. De Langhe et al. contend that we ignored "the consideration of flagging." It is not clear, even from the context, precisely to which "flagging" they refer. All the flags relevant to bacterial counts and UTI (e.g., WBC, leukocyte esterase positivity, nitrite positivity), including the presence of yeast cells, were taken into account.
As De Langhe et al. well know, for calculating diagnostic sensitivity and specificity, one needs a cutoff(s). It is in this context that different cutoffs for bacteria were used. This, we believe, was quite clear from the table.
In the clinical microbiological practice, commonly used pathogenicity cutoffs are 104 and 105 bacteria (CFU) per ml. For special cases, as noted in the article, 103 bacteria/ml is also used (1; J. A. Washington, Letter, Arch. Pathol. Lab. Med. 122:120-122, 1998).
Although we are well aware of the preanalytical and analytical problems associated with urine specimens and of the potentials of flow cytometry, we appreciate the highlighting of these by De Langhe and colleagues.
| REFERENCES |
|---|
|
|
|---|
|
Zahur Zaman* Jan Verhaegen Sylvie Roggeman Department of Laboratory Medicine University Hospital Leuven Herestraat 49 B-3000 Leuven, Belgium
|
||||||
|
* Phone: 32 9 240 29 56 Fax: 32 9 240 49 85 E-mail: joris.delanghe{at}rug.ac.be. |
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||