Received 10 December 2001/ Returned for modification 10 March 2002/ Accepted 28 April 2002
|
|
|---|
4°C. The optimal condition for storage was -70°C. A newly developed quantitative PCR based on monitoring the amplification by measuring the increase in fluorescence caused by the binding of SYBR green dye to double-stranded products was applied herein. DHBV genomic DNA cloned in a plasmid was used for the generation of standard DHBV DNA for quantitative PCR. It validated results from PCR in terms of the copy number of DHBV particles. The specificity of PCR was demonstrated by melting curve analysis, and the differentiation of two DHBV isolates amplified from dried serum was demonstrated based on their melting temperatures determined by GC contents and sequence. It was easier and simpler than other PCR-based DNA techniques. The use of serum dried on filters allows samples from distant field for which cold storage and transportation are a problem to be mailed to the diagnostic laboratory. Samples can be archived for comparison and used as a source of DNA for cloning and sequencing. |
|
|---|
Specimens of serum or blood dried on filter paper have been used for serologic and screening tests for human immunodeficiency virus (HIV), measles virus, human cytomegalovirus, phenylketonuria, and HBV surface antigens (7, 11, 15, 24, 27). This method provides a simple and powerful approach for large-scale screening and epidemiological studies. Dried samples are light, are easy to store or mail, cannot be broken or spilled, and take up little space. When the PCR was used to amplify a defined DNA sequence, it offered a quick and reliable diagnosis and analysis of DNA isolated from serum dried on filter papers (3, 6). In this study, a sensitive PCR assay was applied to detect DHBV from dried serum or blood on filters. The immobilization potential of different filter papers and the temperature effect of storage were compared by using this system. The precise quantification of the DHBV level from each experiment was first accomplished with the use of the newly developed SYBR green quantitative PCR. In conclusion, this technique not only facilitated the field investigation of DHBV in flocks but also extended the usefulness of DHBV as a model of HBV.
|
|
|---|
DHBV DNA isolation. Viral nucleic acid was prepared from serum or blood dried on filters by using a QiaAmp blood and tissue kit (Qiagen, Hilden, Germany) or phenol-chloroform extraction. For the QiaAmp kit, dried serum filters were incubated with 200 µl of distilled water for 1 h at 37°C and then used for the lysis reaction with chaotropic salts, detergents, and a final proteinase K concentration of 1 mg/ml. After the addition of 2-propanol to decrease solubility, DNA was bound to spin columns and washed twice before elution in 20 µl of distilled water. For the phenol-chloroform extraction, filter paper (Whatman paper no. 1) with dried viral serum (1 µl) was incubated for 1 h at 37°C in a 400 µl of extraction buffer containing 0.01 M Tris, 0.01 M EDTA, 0.01 M NaCl, 0.5% (wt/vol) sodium dodecyl sulfate, and 50 mg of proteinase K/ml. Following incubation, DHBV DNA was phenol-chloroform extracted, ethanol precipitated, and suspended in 10 µl of distilled water. Preparations served as the template for both conventional and quantitative PCR.
PCR amplification. Viral DNA sequences were amplified by using the oligonucleotide primers DCD 03 (5'-ACTAGAAAACCTCGTGGACT-3', bp 241 to 260) and DCD 04 (5'-GGGAGAGGGGAGCCCGCACG-3', bp 349 to 368). A single PCR sample containing 10 µl of DNA was amplified in a 50-µl reaction mixture containing 10 mM Tris-HCl (pH 8.3), 2.0 mM MgCl2, each deoxynucleoside triphosphate at 0.2 mM, each primer at 0.5 µM, and 50 U of Taq polymerase (Perkin-Elmer) per ml. PCRs were subjected to 30 cycles (denaturation, 30 s at 95°C; annealing, 30 s at 55°C; extension, 1 min at 72°C) and 1 final extension cycle at 72°C for 10 min. Primers mapped in the N-terminal third of the duck HBV pol gene. Ten microliters of the reaction mixtures was analyzed on 3.0% agarose/Tris-Cl-sodium acetate-EDTA (NuSieve, FMC) in the presence of 0.5 µg of ethidium bromide per ml and photographed under UV illumination. The results were compared to those for uninfected controls.
Construction of standard plasmids for SYBR green quantitative PCR. The plasmid (pDHBV) used as the standard DNA was constructed by ligation of a PCR fragment in pTrueBlue-VUII according to the instructions of the manufacturer (Genomics One, Quebec, Canada). The cloned fragment comprised a 128-bp region of the DHBV DNA polymerase gene amplified by primers DCD 03 and DCD 04. Plasmids were linearized with NcoI (Roche) and purified over a 0.8% agarose gel in 1x Cl-Tris sodium acetate-EDTA. The band of correct size was excised and extracted by using a QIAquick gel extraction kit (Qiagen) according to the manufacturer's instructions. The concentrations of linearized plasmid preparations were determined by measuring the optical density at 260 nm. Conversion to genome equivalents was done by using the following equations: an optical density at 260 nm of 1 = 50 µg/ml, and 1 bp = 660 g/mol. This resulted in a molecular mass for the plasmids (2,815 bp) of 1.82 x 106 g/mol. For the standard curve, a series of 10-fold-diluted plasmids pDHBV was included in each test.
SYBR green quantitative PCR. The SYBR green PCR kit (PE Applied Biosystems) was used with 50-µl reaction mixtures in a model 7700 sequence detector (PE Applied Biosystems). The PCR mixture contained 1x SYBR PCR buffer, 2.0 mM MgCl2, 0.2 mM (each) dATP, dCTP, and dGTP, 0.5 mM dUTP, 0.1 µM primers DCD 03 and DCD 04, 0.5 µl of AmpErase uracil-N-glycosylase (UNG) (1 U/µl), and 0.25 µl of AmpliTaq Gold polymerase (1 U/µl). After initial incubation at 50°C for 2 min to activate the UNG and 95°C for 10 min to activate the Taq polymerase, thermal conditions followed 40 cycles of 95°C for 30 s, 63°C for 30 s, and 72°C for 1 min. For each run, quadruplicates of standard plasmids pDHBV representing 104 to 107 copies of DHBV genome and DNA samples isolated from 1 µl of dried viral serum in different filters were used. This was done because all copy numbers of DHBV genome of tested samples fell in the 104 to 107 copies of DHBV genomes in pilot experiments. A copy number of DHBV genome was calculated according to the standard curve generated by the real-time PCR software of the model 7700 sequence detector (PE Applied Biosystems).
Comparison of sensitivities and specificities for DHBV DNA detection of fresh serum, dried serum, and dried blood. Ten microliters of 30 blood or serum samples randomly obtained from the ducklings was directly used or spotted onto Whatman paper no. 1 and allowed to dry before DNA extraction. Viral nucleic acid was prepared by using a QiaAmp blood and tissue kit (Qiagen), dissolved in 10 µl of distilled water, and subjected to PCR. The sensitivities and specificities of these assays were calculated.
Melting curve analysis and direct sequencing of PCR products. Melting curve analysis was performed to measure the specificity of the PCR product. It also was used to rapidly differentiate strains of DHBV based on melting temperature (Tm). After PCR, samples were heated to 95°C for 30 s and 63°C for 20 s and then heated to 95°C at a ramp rate of 0.2°C/s for 19 min 59 s. Results were analyzed with the melting curve analysis software of the model 7700 sequence detector (PE Applied Biosystems). The Tm of PCR products was calculated at the same time. PCR products were purified by using a Genelute PCR purification kit (Sigma) and then sequenced in both directions with primers DCD 03 and DCD 04 by using a PRISM ready reaction DyeDeoxy terminator cycle sequencing kit as recommended by the manufacturer on a UA373 DNA sequencing system (PE Applied Biosystems).
|
|
|---|
![]() View larger version (31K): [in a new window] |
FIG. 1. Detection of DHBV by PCR in dried serum samples spotted onto Whatman paper no. 1 (A), nylon membrane (B), and nitrocellulose membrane (C). Lanes M, 50- to 1,000-bp size markers; lanes 1, 50 µl of serum sample; lanes 2, 10 µl of serum sample; lanes 3, 1 µl of serum sample; lanes 4, 0.1 µl of serum sample; lanes 5, negative control.
|
![]() View larger version (89K): [in a new window] |
FIG. 2. Detection of DHBV by PCR in dried serum samples (1 µl) spotted into Whatman paper no. 1. In lanes 1 to 4, a study of storage temperature is shown. Lane M, 50- to 1,000-bp size markers; lane 1, -70°C; lane 2, -20°C; lane 3, 4°C; lane 4, 37°C. In lanes 5 to 7, a study of extraction method is shown. Lane 5, Qiagen column; lane 6, phenol-chloroform extraction; lane 7, negative control.
|
|
View this table: [in a new window] |
TABLE 1. Comparison of fresh serum, dried serum, and dried blood for detection of DHBV DNAa
|
![]() View larger version (114K): [in a new window] |
FIG. 3. Quantitative PCR quantification of DHBV in dried serum sample (1 µl) spotted onto Whatman paper no. 1 (arrow A), nylon membrane (arrow B), and nitrocellulose membrane (arrow C). (A) Amplification plot of extracted DNA and the standard DHBV plasmids starting from 104 to 107 molecules. The cycle number was plotted versus the change in fluorescent intensity. (B) Standard curve generated from the amplification plot. The starting copy number was plotted versus the threshold cycle.
|
![]() View larger version (119K): [in a new window] |
FIG. 4. Quantitative PCR quantification of DHBV in dried serum sample (10 µl) spotted into Whatman paper no. 1 stored at -70°C (arrow A), -20°C (arrow B), 4°C (arrow C), and 37°C (arrow D). (A) Amplification plot of extracted DNA and the standard DHBV plasmids starting from 104 to 107 molecules. The cycle number was plotted versus the change in fluorescent intensity. (B) Standard curve generated from the amplification plot. The starting copy number was plotted versus the threshold cycle.
|
![]() View larger version (29K): [in a new window] |
FIG. 5. Melting curve analysis of PCR products obtained following 40 cycles of amplification. Melting peaks were produced by plotting the change in the relative intensity of fluorescence for each sample over temperature. Note that the Tms of the PCR products of DHBV Indiana and California isolates were 81.2 and 80.6°C, respectively. No peaks of amplification products from negative control groups (arrow C) were observed.
|
|
|
|---|
Collection of serum on filter paper has been used to screen for metabolic disorders such as phenylketonuria and sickle cell disease (27). It enables large-scale, worldwide screening of viruses, evolutionary studies, and searches for mutations (22, 26). It allows samples from distant field or developing countries, in which cold storage and transportation are problems, to be concentrated into one laboratory (12). Samples can be mailed after collection, stored, and archived for comparison or used as a source of DNA for cloning and sequencing (19). Studies have shown the efficacy of collecting blood or serum on filter paper followed by PCR analysis for screening and surveillance of perinatal transmission of HIV and human cytomegalovirus and global distribution of malaria genotypes (5, 24). When PCR was applied for viral detection in infected ducks, dried or fresh serum gave fewer false negatives than did dried blood. This is due to a more complete removal of erythrocyte contaminants from serum preparations during drying and contaminants such as the porphyrin moiety of heme, which is known to be a strong inhibitor of PCR (17). This inhibition effect of dried blood on filters caused a decrease in sensitivity (82%). Although microwaving dried blood samples on filters prior to PCR was reported to enhance the sensitivity for detection of HBV, blood still had larger amounts of potential erythrocyte contaminants than did serum (14). Viral DNA was extracted and amplified by PCR from fresh serum or dried serum on filters. Due to the high levels of DHBV in the serum of infected ducks, 0.1 µl was sufficient for detection of virus on the filter paper. The highest yield of PCR products was observed when DHBV-infected serum was loaded into Whatman paper, because of its higher absorption and rehydration potentials to serum. Other advantages include its availability and price, which is lower than that of other filters.
No significant difference in DNA extraction yield was observed between silica gel-based columns (Qiagen) and phenol-chloroform. It was more convenient to use Qiagen columns in large-scale screening. Similar results were observed with other viruses (1, 26). Serum dried on filters was stable under different temperatures for 4 weeks, but the yields of PCR products decreased when the temperature was
4°C. A temperature of -70°C was optimal for storage. These findings are consistent with those of studies with HIV and hemorrhagic enteritis virus (6, 26). The rate of partial degradation of dried DNA on filters or the presence of PCR inhibitors may increase proportionately with storage temperature. Also, difficulty in removing viral DNA from filters increases with prolonged storage time and higher temperature (12).
To precisely estimate the amounts of virus in samples, a quantitative PCR was applied herein. The detection and quantification of target DNA take place simultaneously. Amplification is performed in a closed tube and requires no post-PCR electrophoresis. Therefore, contamination with amplicons can be avoided (22). This assay is more reproducible and convenient than other quantitative PCR tests. Competitive quantitative PCR for DHBV DNA requires an exogenous competitor for each target as a control which needs to be constructed and characterized (2). Fluorogenic quantitative PCR requires an efficient probe with double-labeled fluorescent reporter dyes to be incorporated in each reaction (4, 8, 28). Our system has advantages in sensitivity over the fluorogenic probe system, because multiple dyes can bind to a single amplified molecule. We cloned a plasmid containing viral genome for the generation of standard DHBV for quantitative PCR. The detection limit of this assay was 10 copies of DHBV DNA (data not shown), and it further validated results from PCR experiments in terms of the copy number of DHBV particles.
Steps were taken to avoid false positives in quantitative PCR, because SYBR green dye can bind to any double-stranded DNA, even nonspecific amplicons. A "hot start" AmpliTaq Gold polymerase, which requires incubation at 95°C to activate just before the first denaturation step of thermocycling, was applied. The use of UNG and the incorporation of dUTP in place of dTTP can prevent carryover contamination by PCR product, because illegitimate amplifications that incorporate dUTP can be cleaved by UNG. Any nonspecific amplicon would be eliminated before the amplification, with no subsequent contamination detected on further runs (20). The specificity of PCR was verified by the performance of melting curve analysis for the PCR product, which depends on its GC contents, length, and sequence (25, 29). Two batches of sera dried on filters containing different DHBV isolates (California and Indiana isolates) were amplified and examined by melting curve analysis. The melting temperatures of PCR products from California and Indiana isolates were 80.6 and 81.2°C, respectively. The difference between the melting temperatures was attributed to the nucleotide sequence variations in these isolates (sequence homology was 98.7%). Therefore, the differentiation of virus strains by melting curve analysis was successful. It was more applicable (with a difference of only four bases in this case) and faster (20 min) than PCR-restriction fragment length polymorphism or PCR-single-strand conformation polymorphism (13, 15, 16). No amplification from primer dimer or other sources was present in the controls. Sequence analysis further confirmed the specificity of PCR amplifications.
Although dried serum cannot provide live virus that can be grown and studied in culture, the loss of viral infectivity has the advantage of safety during shipping. This technique can be further refined and optimized for different applications, including the study of genetic diseases, forensics, the human genome, and persistent infections in humans and animals.
This work was supported partially by a grant from Southern Microbiological Associates of Birmingham, Ala.
|
|
|---|