Journal of Clinical Microbiology, July 2002, p. 2616-2617, Vol. 40, No. 7
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.7.2616-2617.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Department of Laboratory Medicine, National Institutes of Health, Bethesda, Maryland 20892-1508
Received 7 December 2001/ Returned for modification 14 March 2002/ Accepted 14 April 2002
|
|
|---|
|
|
|---|
We have observed that samples must occasionally be reanalyzed because of a weak signal from the RT-PCR of the QS internal control, particularly with new operators of the Roche HIV-1 viral load assay. Because only trace amounts of free RNA are circulating in plasma, typically no RNA pellet is visible after the ethanol precipitation step. We hypothesized that samples with low RT-PCR signals for the QS are due to the poor recovery of RNA because of accidental aspiration of the nonvisible RNA pellet. We therefore tested whether the addition of a coprecipitant, Pellet Paint (Novagen Inc., Madison, Wis.), would improve the rate of recovery of RNA during the RNA isolation step and thus reduce the number of samples that need to be retested. Pellet Paint is a pink dye-labeled polymer that improves the efficiency of RNA precipitation even when the RNA concentrations are very low (2 ng/ml) (1), and it makes the RNA readily visible after precipitation. According to the manufacturer, Pellet Paint does not interfere with various molecular biology techniques, such as RT-PCR (1). The use of Pellet Paint has recently been shown to enhance the sensitivity and reproducibility of a qualitative RT-PCR test for enterovirus (3).
HIV-1 viral load tests (COBAS AMPLICOR HIV-1 Monitor Test, version 1.5; Roche) were performed with samples from HIV-positive patients before and after the addition of 2 µl of Pellet Paint per sample. Figure 1 shows the results of the RT-PCR assay for the QS. Routinely, two dilutions (dilution 1 [1:1] and dilution 2 [1:9]) of the QS RT-PCR products are analyzed by the Roche assay. Specimens with optical density (OD) readings for the QS below 0.15 are invalid for calculation of the HIV viral load because it indicates that there is a problem either with the RNA isolation step or with the PCR amplification. The OD readings for 5 of the 38 samples tested without Pellet Paint were less than 0.15 when dilution 2 was analyzed, and the OD reading of 1 sample was less than 0.15 when dilution 1 was analyzed. In contrast, both dilutions of all of the samples processed in the presence of Pellet Paint had OD readings greater than 0.15. Furthermore, on the basis of the mean OD reading, more RT-PCR product was obtained for the samples treated with Pellet Paint. The average OD readings for dilution 2 were 0.44 ± 0.25 without Pellet Paint and 0.57 ± 0.15 with Pellet Paint (P < 0.01). The amount of PCR product was also less variable for the samples treated with Pellet Paint, as evidenced by the smaller range of values (Fig. 1). On the basis of the reported properties of Pellet Paint (1), the most likely mechanism for the higher OD values for the QS is that the addition of Pellet Paint increased the amount of RNA recovered from the samples.
![]() View larger version (17K): [in a new window] |
FIG. 1. Scatter plot of OD readings for the QS in the absence (-PP) and presence (+PP) of Pellet Paint. Thirty-six specimens were assayed in three separate runs. The dotted line represents the lower acceptable limit of the OD reading for the QS. OD readings for dilutions 1 (A) and 2 (B) of the QS are shown.
|
![]() View larger version (23K): [in a new window] |
FIG. 2. Correlation of AMPLICOR HIV-1 viral load test results with and without Pellet Paint. Twenty-eight specimens including patient samples and positive control samples were assayed for HIV viral load (24 specimens were assayed by standard assays [shaded circles]; 4 specimens were assayed by ultrasensitive assays [filled circles]). The results are graphed by use of the log of the HIV copy number (number of HIV RNA copies per milliliter).
|
It is relatively easy to add a coprecipitant to the AMPLICOR HIV-1 viral load assay, and the addition of a coprecipitant does not significantly add to the cost of the test. It costs less than 1% relative to the cost of the AMPLICOR HIV-1 viral load test. Overall, in fact, use of the coprecipitant should result in cost savings because, depending on the experience of the users, it significantly reduces the 5 to 15% retesting rate due to the better recovery of small amounts of RNA.
In conclusion, the two main advantages for the use of the coprecipitant is that it decreases the frequency of sample retesting because of problems associated with low readings for the QS and it improves the overall precision of the AMPLICOR HIV-1 viral load assay.
|
|
|---|