Journal of Clinical Microbiology, August 2002, p. 3064-3065, Vol. 40, No. 8
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.8.3064-3065.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Section of Infectious Diseases, Department of Medicine, School of Medicine, Hacettepe University, Ankara, Turkey,1 Program for Appropriate Technology in Health, Seattle, Washington,2 Department of Medicine, University of Alabama at Birmingham School of Medicine,3 Sexually Transmitted Diseases Control Program, Jefferson County Department of Health, Birmingham, Alabama4
Received 24 September 2001/ Returned for modification 21 December 2001/ Accepted 1 April 2002
|
|
|---|
|
|
|---|
Sera and patients. A total of 353 sera from 157 patients were evaluated. Two hundred one sera were collected from 30 patients at the time of syphilis diagnosis and up to 12 months following treatment. Fifteen additional sera from patients with false-positive RPR test results (RPR reactive and FTA-ABS nonreactive) were also evaluated, as were 137 consecutively collected sera from 112 patients that had been screened for syphilis while attending a sexually transmitted disease clinic. For comparison purposes, the Macro-Vue RPR (Becton Dickinson Microbiology Systems, Cockeysville, Md.) and the FTA-ABS indirect fluorescent-antibody tests (Zeus Scientific, Inc., Raritan, N.J.) were performed as described in the manufacturers' instructions (4). Syphilis diagnosis and interpretation of RPR and FTA-ABS test results were performed in accordance with recommended procedures (4). Two hundred sixteen sera had been stored at -70°C, and 137 were tested without freezing within 48 h of collection.
ICS test. The strip used in the ICS test is configured as a 5- by 75-mm strip. For the present study, recombinant 47-kDa antigen capture reagent was applied to nitrocellulose strips laminated to plastic backing and immobilized as a thin horizontal line across the strip. A second line of mouse anti-human immunoglobulin G, which served as a positive control for each test strip, was similarly applied a short distance above the reagent line. Polyester pads were then attached to the top and bottom of the backing. The lower pad was impregnated with colloidal gold signal reagent, and the upper filter paper pad served as a receptacle for excess serum which accumulated after wicking through the nitrocellulose strip by capillary action.
For testing with the ICS, sera at room temperature were vortexed and 65 µl of each serum specimen was transferred to a test tube. ICSs were placed in the tubes so that the sera could be absorbed by the lower pad of the test strip and could wick through the nitrocellulose strip. After 8 min, the strip results were read. A positive reaction was characterized by the appearance of two lines on the cellulose strip, while a negative reaction was characterized by only one visible line (the control) (Fig. 1). All sera were tested in blinded fashion with the RPR, FTA-ABS, and ICS tests. ICS tests were interpreted and classified independently by two observers.
![]() View larger version (42K): [in a new window] |
FIG. 1. ICS test for specific T. pallidum antibodies. Left, reactive test strip showing reactive control (upper) and test (lower) lines; right, nonreactive test strip showing single, control line.
|
All ICS tests were independently interpreted by two observers with 100% agreement. Of the 199 positive ICS test results for sera from known syphilis patients, 34 were weakly positive (i.e., the band intensity on the ICS was faint). All sera that were weakly positive by the ICS test were FTA-ABS test reactive. Eleven sera (33%) with weakly positive ICS test results were RPR nonreactive, while the other 23 sera which gave faint ICS test results were RPR positive with a range of titers from 1:1 to 1:512. Thus, in some cases, sera with high RPR titers were relatively weakly positive by the ICS test (but were nonetheless clearly positive). Conversely, some FTA-ABS-reactive specimens with low titers or negative results for the RPR test gave strongly positive ICS results. No patient serum specimen with a reactive FTA-ABS test result had a negative ICS test result.
The second group of specimens consisted of 15 sera collected from 15 patients with biologically false-positive RPR (RPR-reactive and FTA-ABS-nonreactive) serological tests for syphilis. All sera from these patients were negative by the ICS test.
The third group of specimens consisted of 137 sera collected from 112 patients attending the local Department of Health sexually transmitted disease clinic and who had been screened for syphilis as part of routine care. The FTA-ABS tests were reactive for 13 patient sera (8 of these were RPR nonreactive) and nonreactive for 124 (7 of these were RPR positive [biologically false positive]). Agreement between the ICS and FTA-ABS test results was 100%. Assuming that all 13 patients with reactive FTA-ABS tests had syphilis at the time of serum collection or in the past, the sensitivities, specificities, positive predictive values (PPV), and negative predictive values (NPV) of the ICS tests for this relatively small number of patients were all 100%.
In summary, the ICS test permits the evaluation of serological evidence of syphilis by an assay that could be readily performed in nonlaboratory settings. The ICS test in this pilot study utilized a recombinant 47-kDa antigen that is expressed and made as a complete protein in Escherichia coli and that is purified by conventional methods (2). The 47-kDa antigen has also been used in immunoblot assays with other T. pallidum recombinants, including the 44.5-, 17-, and 15-kDa antigens (1, 2, 3).
In the present study, we found that the ICS test provides accurate, qualitative detection of antibodies to T. pallidum. The test appears to be more sensitive than the RPR test, yielding positive test results for a number of treated syphilis patients who had nonreactive RPR tests but reactive FTA-ABS tests. The ICS test was also nonreactive for 22 patients with biologically false-positive (RPR-reactive and FTA-ABS-nonreactive) sera. The qualitative nature of the test was also evident in that ICS test reactivity did not appear to correlate with RPR titers.
We believe that the ICS test has the potential to contribute to serological screening that is performed as part of syphilis control efforts, particularly in settings where the prevalence of previously treated syphilis is low or where logistical constraints make a test which can give results rapidly and without laboratory equipment outweigh the need to avoid treatment of patients who have had syphilis and have been successfully treated in the past. For instance, the test might be helpful in settings where at-risk patients are not reliably available for subsequent follow-up (i.e., emergency departments or short-term correctional facilities) or where treatment delays might have serious consequences (i.e., for evaluation of pregnant women). Conversely, in some settings where the prevalence of treated syphilis is high or where the response to therapy is being assessed, the relatively sensitive, qualitative nature of the test might be less helpful to clinicians. Further study to examine both the performance and utility of this test is warranted.
|
|
|---|