Journal of Clinical Microbiology, September 2002, p. 3476-3477, Vol. 40, No. 9
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.9.3476-3477.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Department of Pathology, Stanford University School of Medicine,1 Clinical Microbiology and Virology Laboratory, Stanford Hospital and Clinics, Stanford, California2
Received 20 March 2002/ Returned for modification 9 May 2002/ Accepted 27 June 2002
| ABSTRACT |
|---|
|
|
|---|
| TEXT |
|---|
|
|
|---|
The Roche PCR assay (Roche Molecular Systems, Inc., Pleasanton, Calif.) utilizes reverse transcription-PCR (3). In brief, samples are prepared by precipitation of the viral RNA and addition of an HCV internal control RNA for assay quality control. Subsequently, the target RNA is reverse transcribed and then amplified with specific HCV primers which are designed from the highly conserved 5' untranslated region of the viral genome. Target-specific oligonucleotide probes hybridize with the amplified products. The presence of a hybridization product is established by colorimetric analysis on a COBAS AMPLICOR analyzer.
The qualitative Bayer assay (2) amplifies RNA by in vitro transcription of a DNA template. The viral nucleic acids in the specimen, as well as those of an internal control, are hybridized to capture oligonucleotides complementary to highly conserved HCV 5'-untranslated-region sequences. The RNA is subsequently captured on magnetic microbeads. After the addition of primers, reverse transcriptase, and T7 RNA polymerase, RNA is amplified from a DNA template during a 1-h incubation at 41.5°C. Single-stranded, fluorescently labeled probes are then hybridized to the RNA amplicons, whereas unhybridized probes are chemically degraded. A Leader HC luminometer (GenProbe, San Diego, Calif.) distinguishes the chemiluminescent signals of the target sequence in the samples and the internal control by differential light emission (Bayer Diagnostics Division, Norwood, Mass.).
We evaluated 182 patient sera by the TMA assay. The 182 samples included specimens sent to our laboratory for HCV enzyme immunoassay, HCV recombinant immunoblot assay (RIBA), and qualitative or quantitative HCV PCR by the Roche assay. We also included random specimens that were assessed for analytes not related to HCV, such as human immunodeficiency virus, hepatitis A and B viruses, cytomegalovirus, syphilis organisms, and rubella and varicella zoster virus. A side-by-side evaluation of the TMA assay and the PCR assays was performed with 172 of 182 samples, with TMA assay runs of 24, 54, and 94 patient samples in comparison with either 10 (one ring) or 22 (two rings) patient sample runs of the Roche system. We included one positive and one negative control sample in each Roche ring and three positive as well as three negative calibrators in each TMA assay run. Twelve of the 172 samples were duplicates for reasons not related to the assay. The remaining 10 (of 182) samples were not directly compared to the Roche method but were used to evaluate the possibility of performing INNO-LIPA genotyping (Innogenetics Inc., Norcross, Ga.) after the TMA assay.
Our study population consisted of 10 samples with indeterminate RIBA results and 65 with HCV antibody-positive results. Of the 170 samples, 25 were originally sent for testing by Roche PCR and an additional 10 were both positive by PCR and subsequently genotyped by the INNO-LIPA system (Innogenetics, Inc.) (6). Sixty were random samples (tested for HCV a priori in the TMA assay and PCR comparison).
A total of 65 of 170 samples were HCV positive by either qualitative or quantitative PCR (Roche Amplicor), and 66 of 170 were positive by the TMA assay. The sample newly identified as positive by TMA analysis remained negative by qualitative PCR on repeat testing. This finding may reflect either a false-positive result, higher sensitivity of the TMA assay, or PCR inhibition of this sample. The sample was subsequently determined to be negative by RIBA (Chiron Corporation, Emeryville, Calif.), but this is a less sensitive assay than PCR or the TMA assay. Therefore, the true status of this specimen could not be determined. Results were obtained for all TMA samples, whereas two PCR samples failed to undergo amplification after several attempts. Results for the 12 samples that were duplicated in the TMA assay were entirely equivalent.
The TMA assay requires a sample volume of 500 µl, versus 200 µl for PCR. Although the Roche amplified product can be used directly, subsequent genotyping using the line probe method (INNO LiPA HCV II; Innogenetics, Inc.) from the TMA product is possible only after an additional amplification. On the other hand, the TMA assay is very easy to perform, does not require repeated vortexing as does the Roche assay, and requires minimal instrument maintenance.
The time to results and hands-on time per sample were measured once in each individual assay setup (Fig. 1). Both are substantially better for TMA analysis, even when a large number of specimens is tested. Consequently, labor costs are lower for the TMA assay. The general cost of disposables is also lower for TMA assay, but these costs remain well below $2 per reportable result in both assays. Comparing list prices for both assays, the reagent cost of Roche PCR analysis is significantly lower than that of the TMA assay, although manufacturers' volume discounts certainly influence final pricing.
|