Journal of Clinical Microbiology, October 2003, p. 4714-4717, Vol. 41, No. 10
0095-1137/03/$08.00+0 DOI: 10.1128/JCM.41.10.4714-4717.2003
Copyright © 2003, American Society for Microbiology. All Rights Reserved.
Department of Medicine, Brooke Army Medical Center, Fort Sam Houston, Texas
Received 16 April 2003/ Returned for modification 26 June 2003/ Accepted 11 July 2003
|
|
|---|
24 h. However, the mean time to growth detection for C. glabrata isolates in anaerobic medium was 22.14 ± 2.47 h, but it was 120.89 ± 35.33 h in aerobic medium (P < 0.001). The BACTEC 9240 system detected growth of most Candida isolates; however, the delayed time to detection of C. glabrata is clinically significant. Given the high rate of false negatives, terminal subcultures may be helpful in certain situations. |
|
|---|
Episodes of candidemia are most commonly detected by standard blood culture media in automated blood culture systems. Multiple studies have demonstrated effective recovery of Candida spp. by use of the BACTEC 9240 automated blood culture systems (6, 8, 9, 11, 14). However, these earlier studies generally addressed C. albicans recovery; few NAC were recovered, and the performance of the system specifically for these species has not been well defined. Additionally, earlier studies were all conducted on clinical specimens, many without terminal subculture, and therefore the number of false negatives may not have been fully examined. In the course of performing a study of simulated candidemia, we noted some interesting observations in the overall detection of growth and time to detection among various Candida spp. We report our observations of the performance characteristics of the BACTEC 9240 system in the detection of growth, the time to detection, and the utility of terminal subcultures for a broad range of Candida spp., using a simulated candidemia model.
|
|
|---|
1 x 106 to 5 x 106 yeasts/ml (7). A 1:100 dilution of each yeast suspension was then performed to produce a density of 1 x 104 to 5 x 104 yeasts/ml. A 0.1-ml aliquot of this suspension was introduced into each set of blood cultures, one aerobic and one anaerobic bottle, to produce a final inoculum of
1,000 yeast cells. Inoculum densities were verified by colony counts of serial dilutions on solid medium. The blood culture bottles utilized in this study were the BACTEC Plus Aerobic/F and Anaerobic/F (Becton Dickinson). All bottles were inoculated with 10 ml of fresh, whole blood from healthy volunteers. The protocol was approved by the Brooke Army Medical Center institutional review board, and informed consent was obtained from all volunteers. After the addition of 0.1 ml of yeast suspension, the inoculated bottles were immediately placed in the BACTEC 9240 automated blood culture system (Becton Dickinson) for up to 21 days. This system incubates specimens at 35°C with continuous agitation and uses a fluorescence technology to indicate the quantity and rate of CO2 production, indicative of microbial growth.
Individual blood culture bottles were removed from the automated system when growth was detected, and the time to detection was recorded. A 0.1-ml aliquot was withdrawn from each positive bottle, plated on solid medium, incubated at 30°C, and read after 24 h.
Blood culture bottles in which no growth was detected were removed from the automated system after 21 days of incubation. Terminal subculture was performed; a 0.1-ml aliquot was removed from each bottle and then plated on solid medium. Each plate was incubated at 30°C and read daily for up to 72 h.
Statistical analysis of data was performed using SPSS software. One-way analysis of variance was followed by Student-Newman-Keuls tests. Additional tests included Fisher's exact test, McNemar tests, and Cohen's kappa test when necessary.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Detection of growth in simulated candidemia by the BACTEC 9240 using Aerobic Plus/F and Anaerobic Plus/F blood culture bottles
|
Terminal subculture was performed on all 100 bottles. A comparison of growth detection by the BACTEC 9240 to terminal subculture is presented in Table 2. Forty-five of 50 (90%) aerobic bottles had growth detected by the BACTEC 9240, while 49 of 50 (98%) had positive terminal subcultures. In contrast, only 11 of 50 (22%) anaerobic bottles had growth detected by the BACTEC 9240, while 48 of 50 (96%) had positive terminal subcultures (P < 0.001). All 56 bottles with growth detected by the BACTEC 9240 also had positive terminal subcultures. Of the 44 bottles with no growth detected by the BACTEC 9240, 41 of 44 (93%) had positive terminal subcultures. Four of five "negative" aerobic bottles (three C. glabrata and one C. rugosa) were subculture positive; one C. albicans bottle was subculture negative. Thirty-seven of 39 anaerobic bottles with no growth detected by the BACTEC 9240 were subculture positive (11 C. albicans isolates, 8 C. tropicalis isolates, 4 C. glabrata isolates, 4 C. krusei isolates, 3 C. parapsilosis isolates, 2 C. lusitaniae isolates, 2 C. guilliermondii isolates, 2 C. kefyr isolates, and 1 C. rugosa isolate); 1 C. albicans isolate and the C. firmetaria isolate were subculture negative. A total of three of the bottles were therefore true negatives, with no growth in the anaerobic bottle of the C. firmetaria isolate and both the aerobic and anaerobic bottles of one C. albicans isolate.
|
View this table: [in a new window] |
TABLE 2. Candida growth detection in the BACTEC 9240 automated blood culture system compared to growth detection by terminal subculture
|
The mean time to growth detection for each Candida spp. is listed in Table 3. The mean time to growth detection for most Candida spp., including C. albicans, C. tropicalis, and C. para psilosis, was
24 h. C. glabrata was the only species that required a substantially longer incubation period in aerobic medium for growth detection, with a mean time to growth detection of just over 5 days and a range of 3 to 8 days (69 to 178 h). The mean time to growth detection of C. glabrata in aerobic bottles was significantly longer than in the corresponding anaerobic bottles (121 versus 22 h; P < 0.001). Additionally, the mean time to growth detection of C. glabrata in aerobic bottles was significantly longer than the mean time to growth detection of all other Candida spp. in aerobic media (121 versus 18 h; P < 0.001).
|
View this table: [in a new window] |
TABLE 3. Incubation time to growth detection by the BACTEC 9240 using Aerobic Plus/F and Anaerobic Plus/F blood culture bottles
|
|
|
|---|
1,000 yeast cells and incubated for up to 21 days. The majority of bottles with no growth detected were anaerobic; more troubling was the fact that a full 10% of aerobic bottles had no growth detected despite a relatively high yeast inoculum and a prolonged incubation period. Forty-one of the 44 (93%) "negative" bottles demonstrated heavy growth of yeast when terminally subcultured, representing 41 false-negative cultures. A review of the literature revealed that other authors have described a similar phenomenon with regard to recovery of Candida spp. with clinical use of the BACTEC 9240 (8, 12, 14). Shigei et al. terminally subcultured Plus Aerobic/F and Anaerobic/F clinical blood culture specimens in which growth was not detected after 5 days of incubation in the BACTEC 9240. Forty-four Candida spp. were detected by the system, 42 in aerobic and 2 in anaerobic media. Terminal subculture detected an additional 21 Candida spp., 4 in aerobic and 17 in anaerobic media, all representing false negatives (12). Pohlman et al. and Ziegler et al. also performed terminal subculture on "negative" Plus Aerobic/F and Anaerobic/F bottles. Although both studies evaluated only a limited number of Candida spp., one-third of all false negatives recovered were Candida spp. (8, 14). Though terminal subculture of negative blood cultures is not routinely performed by most clinical microbiology laboratories, product labels from the manufacturer of the BACTEC 9240 and the Plus/F media do recommend terminal subculture of negative bottles for maximum yield of pathogens. Difficulties with the detection of Candida pathogens in automated blood culture systems do not seem to be limited to this system. Borst et al. also noted that terminal subculture of negative bottles from the BacT/ALERT (Biomérieux, Durham, N.C.) automated blood culture system may be helpful for the recovery of Candida spp. (2). The reason for the high rate of false negatives continues to be poorly defined. Thirty-nine of the 44 negative bottles were in anaerobic medium, with 37 of the 39 being false negatives. It is possible that the growth rate, and hence CO2 production, under anaerobic conditions might not exceed the threshold necessary for detection in the BACTEC 9240 automated blood culture system. However, this does not explain why growth was not detected in 5 of 50 aerobic bottles when Candida spp. are generally considered obligate aerobes (13). In this study, C. albicans does appear to be an obligate aerobe, with the BACTEC 9240 detecting no anaerobic growth of C. albicans. However, in contrast to C. albicans, some NAC clearly grow and are active enough in anaerobic media to be detected by the BACTEC 9240 system. Five of 9 (56%) C. glabrata and 4 of 12 (33%) C. tropicalis isolates in this study had growth detected in the Plus Anaerobic/F medium by the BACTEC 9240. Additionally, all "negative" bottles had growth detected on terminal subculture, except for one C. albicans and one C. firmetaria isolate, revealing that the yeasts remain viable in the anaerobic medium but are not detected by the BACTEC 9240.
Most clinical laboratories using automated blood culture systems incubate routine blood cultures for only 5 days and regard any organisms that do not grow fast enough to be detected before that time as likely contaminants. This may not be true for pathogens that possess slower growth characteristics or lower metabolic activity, such as yeasts. Debate continues over the required duration of incubation for yeasts. Prevost-Smith and Hutton evaluated the recovery of yeast from an older version of the BACTEC system in 1992 and concluded that 5 days of incubation should be sufficient for recovery of yeast pathogens (9). Sixty-one Candida spp. were isolated, usually within 48 h. Only two C. glabrata isolates were isolated, and these were recovered after an average of 96 h. Interestingly, this group continued incubation for up to 21 days and subcultured on days 7 and 21 but isolated no further Candida spp. Reisner and Woods looked at 88 Candida spp. recovered in the BACTEC 9240 Plus Aerobic/F medium (11). Most were detected within 48 h, except for two C. glabrata isolates on day 5 and one C. albicans isolate on day 6. These investigators ultimately recommended 6 days of incubation for optimal recovery of yeast, influenced somewhat by the delayed recovery of other yeasts, including Cryptococcus neoformans and Histoplasma capsulatum. More recently Bourbeau and Pohlman have suggested that only 3 days of incubation are required to isolate clinically relevant blood pathogens, including Candida spp. (3). Thirty-seven Candida spp. were isolated from BacT/ALERT FAN aerobic medium, and all were detected within 72 h, though no terminal subcultures were performed. An increasing number of clinical microbiology laboratories are now adopting this shorter incubation period.
Our microbiology laboratory currently employs a 5-day incubation period for all routine blood cultures; longer durations can be requested by the physician when clinically indicated. The delay in detection of C. glabrata growth in the aerobic media is of some concern even with 5 days of incubation. Six of the nine isolates had detectable growth in the aerobic medium, but the average time to positive was 120.89 h, just slightly longer than the time after which these specimens would have been flagged as negative. This would have led to one additional missed episode of C. glabrata candidemia, or two additional missed episodes if we employed a 3-day incubation period. This delayed time to recovery for C. glabrata isolates in the BACTEC 9240 system using Plus Aerobic/F bottles was also noted by Fricker-Hildago et al., who demonstrated a mean time to detection of 125.92 ± 48.87 h (5). One study did report a shorter time to detection; four C. glabrata isolates had a mean time to detection of 21.8 h using the same medium and system (6). In our own study, C. glabrata growth was detected on average 4 days faster in the anaerobic medium than in the aerobic medium, a difference that is both statistically and clinically significant. Thus, even with use of both aerobic and anaerobic blood culture media and a 5-day incubation period, three of nine episodes of simulated C. glabrata candidemia would have been missed.
The BACTEC 9240 automated blood culture system detects the majority of Candida pathogens in at least one bottle of a blood culture set. We demonstrate that this system, using a simulated candidemia model, is effective for the recovery of most C. albicans and NAC. The BACTEC 9240 automated blood culture system alone did not detect 4 of 50 episodes of simulated candidemia. However, it would not have detected five (10%) episodes of simulated candidemia if we had followed our standard protocol of incubating aerobic and anaerobic bottles for only 5 days; six episodes of simulated candidemia would have been missed with a 3-day incubation protocol. Overall, there continues to be difficulty with the detection of Candida spp. in Plus Anaerobic/F medium and a delay in C. glabrata recovery in Plus Aerobic/F medium using the BACTEC 9240 system. With the increasing prevalence of candidemia, more study is needed to optimize the recovery of Candida spp. from automated blood culture systems.