Journal of Clinical Microbiology, March 2003, p. 1266-1269, Vol. 41, No. 3
0095-1137/03/$08.00+0 DOI: 10.1128/JCM.41.3.1266-1269.2003
Copyright © 2003, American Society for Microbiology. All Rights Reserved.
Gen-Probe Incorporated, San Diego,2 Department of Pathology and Laboratory Medicine, UCLA Medical Center, Los Angeles, California1
Received 27 August 2002/ Returned for modification 16 November 2002/ Accepted 5 December 2002
|
|
|---|
|
|
|---|
In conjunction with advances in conventional blood culture methodologies, there have been several studies examining the use of nucleic acid tests for the rapid detection of positive blood cultures (4, 6, 7, 10, 11, 13, 14; D. Bruckner, L. Gibson, J. Hindler, J. Hogan, I. Andruszkiewicz, K. Clark-Dickey, and W. Weisburg, Abstr. 101st Gen. Meet. Am. Soc. Microbiol., abstr. C-1, p. 147, 2001). Specifically, the Gen-Probe Incorporated hybridization protection assay (HPA), which uses a DNA probe to detect rRNA, has been utilized to evaluate positive blood cultures. The chemiluminescent signal generated by the HPA has been found elsewhere to directly correlate with CFU (Bruckner et al., Abstr. 101st Gen. Meet. Am. Soc. Microbiol.; D. Fuller and T. Davis, Abstr. 39th Intersci. Conf. Antimicrob. Agents Chemother., abstr. 1558, 1999; J. F. Hindler, S. Kozen, and D. A. Bruckner, Abstr. 39th Intersci. Conf. Antimicrob. Agents Chemother., abstr. 1557, 1999).
Organon Teknika (Durham, N.C.) has replaced its standard vented BacT/Alert bottle with a redesigned nonvent (NV) blood culture bottle. As a result, laboratories have been confronted with performing a verification study of these new blood culture bottles. The objective of this study was to perform a seeded study to (i) evaluate the NV bottle for the ability to support growth in a timely manner by conventional methods and (ii) compare conventional methods with molecular probe detection as a possible rapid alternative method for blood culture bottle verification.
In this seeded study performed at UCLA Medical Center, BacT/Alert standard and NV (aerobic and anaerobic) bottles were evaluated. Bottles were tested in triplicate with approximately 50 organisms/bottle. The bacterial isolates used include Enterococcus faecium (UCLA 236), Staphylococcus aureus (ATCC 29213), Streptococcus viridans (UCLA 893 and patient sample 4457), Streptococcus pneumoniae (ATCC 49619 and patient sample 4452), Escherichia coli (ATCC 25922), Pseudomonas aeruginosa (ATCC 27853), Bacteroides fragilis (ATCC 25285), Clostridium perfringens (ATCC 13124), Listeria monocytogenes (UCLA 1174), Corynebacterium jeikeium (UCLA 1126), Haemophilus aphrophilus (UCLA 1146), and Candida albicans (ATCC 14053). Inoculum was prepared from an overnight culture, with diluted 0.5-McFarland-standard suspensions of each isolate. All bottles were incubated in a BacT/Alert3D continuous monitoring system (Organon Teknika Corp.) according to the manufacturer's recommendation. Bottles remained incubated until they flagged positive or until 7 days had elapsed. The time from loading to detection was recorded for each bottle. If isolates failed to grow, 5 to 10 ml of blood was added to bottles and retested in duplicate. Furthermore, isolates still failing to grow were replaced with a patient strain and evaluated with and without blood.
When bottles were triggered as positive, cell mass was measured by viable plate counts and probe detection was performed with the Gen-Probe Incorporated all-bacterial and all-fungal DNA probes directed toward rRNA targets. Bottles not triggered as positive were terminally subcultured and probed after 7 days. The all-bacterial and all-fungal probes were derived from the 23S and 18S regions of the rRNA genes, respectively. For probing a 1-ml sample was centrifuged to pellet the organism, washed twice with a saponin-based cocktail, lysed at 100°C in a succinate-buffered detergent cocktail, and frozen for further analysis. Frozen pellets were diluted when necessary to obtain relative light unit (RLU) levels within the linear range of the luminometer when run in an HPA (1, 9).
Overall, the BacT/Alert NV bottles were equivalent to the standard BacT/Alert bottles as determined by both conventional and molecular methods. These results agree with others with clinical samples (2, 12). The comparative mean times to detection are summarized in Table 1. In the bottles that did not require blood, the standard and NV bottles were equivalent in their mean times to positivity. Isolates failing to grow without blood were equivalent in their mean times to positivity for the standard and NV bottles after blood was added to the bottles (Table 1), excluding C. jeikeium.
|
View this table: [in a new window] |
TABLE 1. Mean time to positive blood bottle detection
|
The comparative mean CFU per milliliter in the positive blood culture bottles for the various representative organisms and bottles are summarized in Table 2. For bacteria, the positive bottles were detected in the late log phase of growth, within a range of 1.53 x 107 to 7.41 x 109 CFU/ml, excluding H. aphrophilus. For yeast, positive bottles were detected with approximately 4.5 x 106 CFU/ml. Interestingly, molecular probe detection was proportionally equivalent to conventional culture methods for verification of the positive bottles. Results of the mean comparative RLU/ml are summarized in Table 3. The CFU-per-milliliter numbers for H. aphrophilus may be underrepresented due to the organism's granular growth in broth. Despite clumping of cells, the RLU-per-milliliter signals for H. aphrophilus had substantially lower standard deviations than did the CFU per milliliter from culture. This phenomenon demonstrates that measurement of total amount of rRNA is not dependent on the organisms being monodispersed. This suggests that a nucleic acid detection system is a better measure of organisms when used for quality control of growth media. When bottles failed to flag as positive after 7 days, they were terminally subcultured and probed.
|
View this table: [in a new window] |
TABLE 2. Mean CFU/per milliliter in positive blood culture bottles
|
|
View this table: [in a new window] |
TABLE 3. Mean RLU/per milliliter in positive blood culture bottles
|
The 7-day subcultures of C. albicans from the standard aerobic bottles had a mean of 35 CFU/ml, while the 7-day subcultures of the standard and new NV anaerobic bottles had means of <30 and 2.3 x 103 CFU/ml, respectively. Either no probe or low probe signals were detected for the 7-day subcultures of C. albicans.
The ratio of the RLU to CFU for a given organism is dependent upon a number of variables. The growth state, plating efficiency, and maximum growth rate will affect this ratio for a given organism. Organisms such as E. coli with very rapid doubling times have much higher numbers of ribosomes per cell than do organisms with low growth rates, such as mycobacteria. Size is a factor in some mycoplasma organisms, which are so small that only a very limited number of ribosomes can be accommodated within the volume of the cell. For most organisms from bloodstream infections detected in routine blood culture systems, the ratios of RLU to CFU were found to be consistent for a variety of targets (Bruckner et al., Abstr. 101st Gen. Meet. Am. Soc. Microbiol.; Fuller and Davis, 39th ICAAC; Hindler et al., 39th ICAAC).
HPA probe detection requires less specimen processing than do conventional methods and provides an answer within 1 to 2 h of the bottle triggering as positive. It has the capability of detecting positive blood cultures much earlier than the automated systems due to a 103-CFU/ml difference in sensitivity. The HPA probe matrix methodologies have provided early, rapid, and precise detection of positive blood cultures (Bruckner et al., Abstr. 101st Gen. Meet. Am. Soc. Microbiol.; Fuller and Davis, 39th ICAAC; Hindler et al., 39th ICAAC). It has also been applied to blood components as a rapid means of screening for transfusion-associated bacterial contamination (3). HPA provided quick and accurate verification of blood culture bottle performance, thus further demonstrating the versatility of this technology.
|
|
|---|