Comparison of Diagnostic Sensitivities of Three Assays (Bartels Enzyme Immunoassay [EIA], Biotest EIA, and Binax NOW Immunochromatographic Test) for Detection of Legionella pneumophila Serogroup 1 Antigen in Urine
Carmen Guerrero, Carmen M. Toldos, Genoveva Yagüe, Cristobal Ramírez, Tomás Rodríguez, and Manuel Segovia*
Servicio de Microbiología, Hospital "J. M. Morales Meseguer," 30008 Murcia, Spain
Received 24 March 2003/
Returned for modification 28 May 2003/
Accepted 14 October 2003

ABSTRACT
The Bartels enzyme immunoassay (EIA), Biotest EIA, and Binax
NOW immunochromatographic test (ICT) urinary antigen kits for
the detection of
Legionella pneumophila serogroup 1 were compared
using 178 frozen urine samples. When nonconcentrated urine samples
were used, the sensitivity levels of both enzyme EIAs were significantly
higher than the sensitivity level of the ICT (Bartels EIA, 71.3%;
Biotest EIA, 65.1%; Binax NOW ICT, 37% [
P < 0.001]). After
concentration of the urine samples, no significant differences
in sensitivity were found among the three tests.

INTRODUCTION
Laboratory diagnosis of
Legionella pneumophila infection is
performed on the basis of the results of culture growth, direct
immunofluorescence, serologic testing, and antigen detection
in urine. Since antigen detection in urine has proved to be
a sensitive and rapid method for detecting
L. pneumophila serogroup
1, this technique has become one of the most used tools for
the diagnosis of Legionnaires' disease. The urinary antigen
test permits early diagnosis and initiation of appropriate antibiotic
therapy (
9), and it leads to the recognition of outbreaks of
Legionnaires' disease, allowing a rapid epidemiological investigation
(
10,
11,
13). Currently, there are several commercially available
tests (the Binax
Legionella urinary antigen enzyme immunoassay
[EIA], the Biotest
Legionella urine antigen EIA, and the Binax
NOW
Legionella urinary antigen immunochromatographic test [ICT])
for the detection of
L. pneumophila antigen in urine. All of
these tests have been previously evaluated, confirming their
utility for the diagnosis of Legionnaires' disease (
6,
12).
A wider range of sensitivity has been reported, however, although
no research has demonstrated significant differences in sensitivity
between these tests. Recently a new EIA (Bartels
Legionella Urinary Antigen) for the detection of
L. pneumophila serogroup
1 antigen in urine has been developed. We evaluated this new
test by comparing its sensitivity to that of two widely used
tests: the Biotest EIA and the Binax NOW ICT.
A total of 178 urine specimens (obtained from patients with Legionnaires' disease during an outbreak which occurred in Murcia, Spain, in July 2001) were included in the study. The samples were from patients who fulfilled the epidemiological criteria, suffered from symptoms compatible with pneumonia, and showed radiological signs of infiltration and for whom Legionnaires' disease was confirmed either by culture (defined as isolation of L. pneumophila serogroup 1 from respiratory specimens) or serology (defined as a fourfold increase in L. pneumophila serogroup 1 antibody titer in paired sera or, in patients with clinically compatible illness, a high [
256] stationary titer value).
During the outbreak, nonconcentrated urine samples were tested within a few hours after receipt by using Biotest EIA (182 were positive and 47 were negative); the samples were then frozen at -80°C. Samples were retested simultaneously using the Biotest Legionella urine antigen EIA (Biotest AG, Dreieich, Germany), Binax NOW Legionella urinary antigen test (Binax, Portland, Maine) and Bartels ELISA Legionella urinary antigen (Intracel, Frederick, Md.) after the samples were stored at -80°C for 3 months. All the tests were performed and the results were calculated following the manufacturers' instructions. Urine samples were tested once. First, they were tested in nonconcentrated form. Nonconcentrated urines that had previously tested negative were concentrated 25-fold by selective ultrafiltration (Minicon B15; Millipore Corp., Bedford, Mass.) as described previously (3).
Data were analyzed with the statistical program Winstat 3.0. Sensitivity levels of the three tests were compared using a chi-square test.
The results of the three tests using nonconcentrated samples are shown in Table 1. The sensitivity levels were as follows: Biotest EIA, 61.5%; Bartels EIA, 71.3%; and Binax NOW, 37%. The differences between the sensitivity levels of the two EIAs were not significant (P = 0.2). By contrast, both EIAs showed sensitivity levels significantly higher than that of the ICT (P < 0.001). The results obtained after concentration of the negative-testing urine samples are shown in Table 1. The increase in sensitivity was statistically significant when using Binax NOW: 37 versus 69.6% (P < 0.001). A clear, but not statistically significant, increase in sensitivity levels was found for the EIAs: 61.5 versus 76.4% (Biotest EIA) and 71.3 versus 75.8% (Bartels EIA). The differences in sensitivity levels for the three tests were not significant (P = 0.2). Multiple comparisons of the commercially available kits have been previously published (5, 7, 8). None of these previous studies were reported to have demonstrated significant differences in the abilities of the different tests to detect L. pneumophila antigen in urine. When nonconcentrated urine samples were tested in our study, there were no significant differences between the results for the two EIAs, although the Bartels EIA gave a slightly higher level of sensitivity (Table 1). By contrast, the Binax NOW ICT was significantly less sensitive than the EIAs, detecting only 37% of the cases. We did not expect to find such a large difference, since previous studies have shown that the sensitivity levels of both tests (the ICT and the EIAs) are approximately the same (8, 14). In spite of this, our findings are in agreement with those of Dominguez et al. (2), who observed that the sensitivity of the ICT assay using nonconcentrated samples was 55.5% relative to the EIA results; after concentration of the urine, the sensitivity levels of the two methods (EIA and ICT) were the same.
View this table:
[in this window]
[in a new window]
|
TABLE 1. Comparisons of sensitivity levels of the three Legionella urinary antigen assays of urine samples in unconcentrated form and after concentration
|
Although it seems clear that concentration of urine increases
the sensitivity of the tests, the reported increase shows great
variations in the different studies (
3,
4,
5,
14). Most of these
previous reports were retrospective and used frozen samples.
The different storage times of specimens and the different tests
used could explain these variations. Chang et al. (
1) reported
that long-term storage of urine leads to a decrease in sensitivity
for EIAs and radioimmunoassays. In our case, when nonconcentrated
urine specimens were retested using the Biotest EIA after storage
at -80°C for 3 months, antigen was not detected in 11% of
previously positive specimens. It is possible that failure to
detect specimens that had previously given positive results
could be related to the reproducibility of detection by the
Biotest EIA. Another explanation is the decay of the urinary
antigen in the stored frozen specimens. This apparent instability
of the antigen in frozen samples may partially decrease the
sensitivity of the tests, and it is possible that this decrease
is larger for the ICT assay. In our study the concentration
of negative-testing samples increased the sensitivity of all
three tests, but this increase (4, 11, and 32% for the Bartels
EIA, Biotest EIA, and Binax NOW ICT, respectively) was statistically
significant only for the Binax NOW ICT.
Although the lack of specificity data for the three assays represents a major limitation, our study provides relevant data related to the clinical sensitivity levels of different commercial kits for the detection of L. pneumophila antigen in urine. In conclusion, the Bartels EIA is comparable to the Biotest EIA in its ability to detect soluble antigen of L. pneumophila in urine. The EIAs showed a level of sensitivity significantly higher than that of the ICT with nonconcentrated frozen urine specimens, but is possible that the level of ICT sensitivity might be higher at the time of the clinical diagnosis. Until prospective studies prove this fact, however, we suggest that urine samples be concentrated when an ICT is used.

FOOTNOTES
* Corresponding author. Mailing address: Hospital "J. M. Morales Meseguer," Servicio de Microbiología, Avda. Marques de los velez s/n, 30008 Murcia, Spain. Phone: 34-968-360953. Fax: 34-968-360950. E-mail:
micromorales{at}hotmail.com.


REFERENCES
1 - Chang, F. Y., J. E. Stout, and V. L. Yu. 1996. Assessment of enzyme immunoassay versus radioimmunoassay for detection of Legionella pneumophila serogroup 1 antigen in frozen urine specimens. J. Clin. Microbiol. 34:2628-2629.[Abstract]
2 - Domínguez, J. A., N. Gali, L. Matas, P. Pedroso, A. Fernández, E. Padilla, and V. Ausina. 1999. Evaluation of a rapid immunochromatographic assay for the detection of Legionella antigen in urine samples. Eur. J. Clin. Microbiol. Infect. Dis. 18:896-898.[CrossRef][Medline]
3 - Domínguez, J. A., J. M. Manterola, R. Blavia, N. Sopena, F. J. Belda, E. Padilla, M. Gimenez, M. Sabria, J. Morera, and V. Ausina. 1996. Detection of Legionella pneumophila serogroup 1 antigen in nonconcentrated urine and urine concentrated by selective ultrafiltration. J. Clin. Microbiol. 34:2334-2336.[Abstract]
4 - Domínguez, J. A., L. Matas, J. M. Manterola, R. Blavia, N. Sopena, F. J. Belda, E. Padilla, M. Giménez, M. Sabria, J. Morera, and V. Ausina. 1997. Comparison of radioimmunoassay and enzyme immunoassay kits for detection of Legionella pneumophila serogroup 1 antigen in both concentrated and nonconcentrated urine samples. J. Clin. Microbiol. 35:1627-1629.[Abstract]
5 - Domínguez, J. A., N. Galí, P. Pedroso, A. Fargas, E. Padilla, J. M. Manterola, and L. Matas. 1998. Comparison of the Binax Legionella urinary antigen enzyme immunoassay (EIA) with the Biotest Legionella urine antigen EIA for detection of Legionella antigen in both concentrated and nonconcentrated urine samples. J. Clin. Microbiol. 36:2718-2722.[Abstract/Free Full Text]
6 - Formica, N., M. Yates, M. Beers, J. Carnie, G. Hogg, N. Ryan, and G. Tallis. 2001. The impact of diagnosis by Legionella urinary antigen test on the epidemiology and outcomes of Legionnaires' disease. Epidemiol. Infect. 127:275-280.[Medline]
7 - Franzin, L., and D. Cabodi. 2000. Comparative evaluation of two commercially available antigen enzyme immunoassays (EIA) for the detection of Legionella pneumophila urinary antigen in frozen non-concentrated urine samples. New Microbiol. 23:383-389.[Medline]
8 - Helbig, J. H., S. A. Uldum, P. C. Lück, and T. G. Harrison. 2001. Detection of Legionella pneumophila antigen in urine samples using BinaxNOW immunochromatographic assay and comparison with both Binax Legionella urinary antigen enzyme immunoassay (EIA) and Biotest Legionella urine antigen EIA. J. Med. Microbiol. 50:1-8.[Medline]
9 - Kashuba, A. D., and C. H. Ballow. 1996. Legionella urinary antigen testing: potential impact on diagnosis and antibiotic therapy. Diagn. Microbiol. Infect. Dis. 24:129-139.[CrossRef][Medline]
10 - Lepine, L. A., D. B. Jernigan, J. C. Butler, J. M. Pruckler, R. F. Benson, G. Kim, J. L. Hadler, M. L. Cartter, and B. S. Fields. 1998. A recurrent outbreak of nosocomial Legionnaires' disease detected by urinary antigen testing: evidence for long-term colonization of a hospital plumbing system. Infect. Control Hosp. Epidemiol. 19:905-910.[Medline]
11 - López, P., A. Chichilla, M. Andreu, C. Pelaz, and J. Sastre. 2001. El laboratorio de Microbiología clínica en el brote de Legionella spp. en la comarca de Alcoy: rentabilidad de las dos difrentes tecnicas diagnosticas. Enfer. Infecc. Microbiol. Clin. 19:435-438.
12 - Ruf, B., D. Schurmann, I. Horbach, F. J. Fehrencach, and H. D. Pohle. 1990. Prevalence and diagnosis of Legionella pneumophila: a 3-year prospective study with emphasis on application of urinary antigen detection. J. Infect. Dis. 162:1341-1348.[Medline]
13 - Wever, P. C., E. P. F. Yzerman, E. J. Kuijper, P. Speelman, and J. Dankert. 2000. Rapid diagnosis of Legionnaires' disease using an immunochromatographic assay for Legionella pneumophila serogroup 1 antigen in urine during an outbreak in the Netherlands. J. Clin. Microbiol. 38:2738-2739.[Abstract/Free Full Text]
14 - Yzerman, E. P. F., J. W. den Boer, K. D. Lettinga, J. Schellekens, J. Dankert, and M. Peeters. 2002. Sensitivity of three urinary antigen tests associated with clinical severity in a large outbreak of Legionnaires' disease in the Netherlands. J. Clin. Microbiol. 40:3232-3236.[Abstract/Free Full Text]
Journal of Clinical Microbiology, January 2004, p. 467-468, Vol. 42, No. 1
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.1.467-468.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Diederen, B. M. W., Peeters, M. F.
(2006). Evaluation of Two New Immunochromatographic Assays (Rapid U Legionella Antigen Test and SD Bioline Legionella Antigen Test) for Detection of Legionella pneumophila Serogroup 1 Antigen in Urine.. J. Clin. Microbiol.
44: 2991-2993
[Abstract]
[Full Text]
-
Rojas, A., Navarro, M. D., Fornes, F. E., Serra, E., Simarro, E., Rojas, J., Ruiz, J.
(2005). Value of Serological Testing for Diagnosis of Legionellosis in Outbreak Patients. J. Clin. Microbiol.
43: 4022-4025
[Abstract]
[Full Text]