Journal of Clinical Microbiology, October 2004, p. 4641-4648, Vol. 42, No. 10
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.10.4641-4648.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
Focus Technologies, Cypress, California
Received 11 February 2004/ Returned for modification 26 March 2004/ Accepted 23 June 2004
|
|
|---|
|
|
|---|
WNV is a member of the Flaviviridae family and is in the Japanese encephalitis serocomplex that includes Japanese encephalitis (JE) virus and St. Louis encephalitis (SLE) virus. Other closely related flaviviruses include yellow fever (YF) virus and dengue virus types 1 to 4. The flavivirus antibody response is predominantly generated against the highly immunogenic envelope protein that contains both flavivirus cross-reactive epitopes and virus-specific epitopes (8, 23). Hunt et al. (8) developed a recombinant protein composed of the JE virus E-protein gene and the preM gene that resulted in the formation of noninfectious particulate JE virus antigen. The particulate nature of the recombinant protein allowed the antigen to maintain a tertiary structure similar to that of the native virus envelope protein and thus was a good candidate for both vaccine and serologic applications. Further development efforts resulted in the production of other flavivirus recombinant particulate antigens, including WNV (6). Studies at the Centers for Disease Control and Prevention (CDC) demonstrated the utility of the WNV particulate antigen for the detection of WNV antibodies and the diagnosis of WNV infection in humans. The recombinant WNV antigen was used to develop the Focus Technologies immunoassay for the detection of IgG and IgM WNV antibodies that recently received clearance from the U.S. Food and Drug Administration. We investigated the utility and performance characteristics of the Focus Technologies IgG and IgM assays in various healthy and diseased patient populations, as well as the reactivity with sera from various flavivirus vaccinees and infected individuals.
|
|
|---|
Focus Technologies WNV IgM ELISA. The WNV IgM ELISA is a mu-capture assay format wherein patient and control samples were diluted 1:101, and 0.1-ml portions of the patient and control sera were added to microtiter wells containing goat anti-human IgM. The kit controls included a negative control, positive control, and a calibrator/cutoff control. After the IgM was captured on the microtiter well, the WNV preM/E recombinant protein was added to each well. After the wells were washed, the presence of WNV antigen bound to either the patient or control IgM was detected by using a peroxidase-labeled mouse anti-flavivirus monoclonal antibody. The chromogen TMB was added, and the final reaction product was measured in a spectrophotometer at a wavelength of 450 nm. Index values for the patient samples and controls were obtained by dividing the absorbance of the patient or control well by the absorbance of the calibrator (cutoff control). Index values of <0.9 were considered negative, values from 0.9 to 1.1 were considered equivocal, and values of >1.1 were considered positive for IgM antibody.
An additional step was evaluated to detect possible false-positive IgM results when the standard procedure described above was followed. Briefly, when a patient sample was positive for WNV IgM by using the standard assay procedure, the specimen was tested again by using the following background subtraction method. All calibrators and controls were run exactly as indicated in the package insert. The patient specimens were diluted 1:101 per the package insert, and 0.1 ml of the diluted serum was added to duplicate wells of the IgM-capture plate and incubated for 2 h at room temperature. After the wells were washed, 0.1 ml of specimen diluent was added to one well, and WNV antigen was added to the second well. After incubation for 1 h at room temperature and further washing, all wells received the peroxidase-labeled mouse monoclonal anti-flavivirus antibody; the remainder of the assay was performed as described above. The patient's final IgM index value was determined by using the net absorbance value, calculated by subtracting the absorbance value of the well receiving diluent only from the absorbance value of the well receiving the WNV antigen. The net absorbance was divided by the absorbance obtained for the kit's calibrator/cutoff control, and the index value was interpreted per the package insert (i.e., index values of <0.9 were considered negative, values from 0.9 to 1.1 were equivocal, and values of >1.1 were positive for WNV IgM antibodies).
CDC WNV IgG and IgM ELISA. The comparative methods used here were the CDC's WNV IgG and IgM ELISA procedures (9, 14). Sixteen separate state public health laboratories (SPHLs) all used the CDC procedure to generate the IgG and IgM results shown in the WNV (SPHL) panel outlined below. The comparative data generated at the New York Department of Health laboratory for the encephalitis/meningitis and SLE panels also used the CDC ELISA protocol. Briefly, the IgG method utilizes an anti-flavivirus monoclonal antibody immobilized to the microtiter plate well to capture WNV antigen. Next, the immobilized monoclonal antibody-antigen complex is incubated with diluted patient sera, and IgG antibodies to WNV in the patient sample are bound to the antigen. The presence of WNV IgG is detected by using a peroxidase-labeled anti-human IgG antibody. The IgM method is an antibody capture ELISA wherein the patient's IgM is captured in a microtiter plate. The antibody capture step is followed by the addition of WNV antigen derived from either suckling mouse brain or recombinant WNV envelope protein. A control antigen (either uninfected suckling mouse brain or recombinant protein control antigen, matched to WNV antigen used) well is included for all samples tested by both the IgG and the IgM ELISA methods. After the addition of WNV or control antigen to the sample wells, peroxidase-labeled flavivirus monoclonal antibody is added to each well, followed by the chromogen TMB (with appropriate wash steps in between). A negative serum or a panel of negative sera was run with each assay and a patient sample with an absorbance reading three times higher than the negative serum was considered positive for the presence of WNV antibody.
WNV PRNT. The plaque reduction neutralization test (PRNT) performed at Focus Technologies Reference Laboratory followed the protocol outlined by Blitvich et al. (2).
Serum samples. The following serum panels were utilized in the present study.
(i) WNV (SPHL) panel. Specimens from 583 possible WNV cases were included in this panel. The sera were initially tested at Focus Technologies Reference Laboratory using an in-house native WNV antigen-based ELISA (21). These in-house IgG and IgM assays were only similar to the Focus kit assays in that the IgG ELISA was an indirect ELISA format and the IgM assay was a mu-capture format incorporating the 6B6-C1 anti-flavivirus monoclonal antibody. The positive samples were then forwarded to the appropriate SPHL (n = 16), where the samples were tested by using the CDC WNV IgG and/or IgM ELISA protocols. A subset of the samples also had PRNT results provided (n = 126). SPHL IgM results were provided for all 583 samples and IgG results were provided for 103 samples. The 583 samples were then tested by using the Focus Technologies IgG and IgM ELISA kits.
(ii) Encephalitis/meningitis panel. Three hundred sequentially submitted serum samples forwarded to the New York State Department of Health laboratory for evaluation of an infectious source for encephalitis or meningitis were tested by using the Focus WNV ELISAs, as well as by using the CDC ELISA protocols. All New York samples were tested at the New York State Department of Health laboratory facility. An additional 50 sera submitted to the CDC, Ft. Collins, Colo., for evaluation for encephalitis/meningitis were also included in this panel. This panel was included to evaluate the Focus WNV kits for reactivity of sera from patients with clinical symptoms consistent with WNV infection; at the time the sera were submitted, the possible infectious source was unknown. The CDC WNV IgM ELISA protocol was performed on all 350 samples to determine the presence of WNV IgM antibodies, and the CDC WNV IgG ELISA protocol was performed on 327 samples.
(iii) Blood donor panel. Sera from 236 normal blood donors from Southern California were used in this panel. All samples were collected in 2001 prior to the identification of any WNV activity in Southern California; this population was thus considered naive for exposure to WNV.
(iv) JE panel. This panel consisted of 20 sera from subjects who had been vaccinated with JE virus (supplied by Acambis, plc) and 20 sera from naturally infected individuals from Southeast Asia (supplied by Jan Groen, Erasmus University, Rotterdam, The Netherlands).
(v) Dengue panel. Nineteen sera were kindly supplied by the CDC Dengue Reference Laboratory, San Juan, Puerto Rico. The samples were from secondary dengue infections from the Caribbean region.
(vi) SLE panel. A total of 32 SLE sera from naturally infected persons were included in the study. The New York SPHL provided 22 sera; all samples had measurable PRNT titers, confirming that the flavivirus antibodies present in these patients were due to exposure to SLE virus. A second group of 10 SLE sera from infected individuals was provided by Acambis.
(vii) YF panel. Forty sera from subjects who had been vaccinated with YF virus were provided by Acambis (20 sera) and Matthias Neidrig, Robert Koch Institute (20 sera).
(viii) Autoimmune serum panel. Sera containing either anti-nuclear antibody (ANA) (n = 20) or rheumatoid factor (RF) (n = 21) were obtained from Focus Technologies Reference Laboratory. The ANA sera had titers of 1:40 or higher, as determined by indirect immunofluorescence with HEp-2 cells. The RF sera had >20 IU of rheumatoid factor/ml as determined by nephelometry.
(ix) IgM serum panel. Sera with IgM antibodies to one of four different infectious agents (cytomegalovirus [CMV], Epstein-Barr virus [EBV], herpes simplex virus [HSV], and Borrelia burgdorferi) that are associated with clinical symptoms similar to those seen in individuals infected with WNV were obtained from Focus Technologies Reference Laboratory. A total of 79 sera were included. The presence of IgM to each infectious agent was determined as follows: CMV IgM (n = 14), determined by using the Diamedex CMV IgM ELISA; EBV IgM (n = 19), determined by the Focus Technologies EBV VCA RIFA IgM; HSV IgM (n = 20), measured by the Diamedix HSV IgM ELISA; and B. burgdorferi IgM (n = 20), measured by using the Focus Technologies Lyme IgM immunofluorescence assay.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Concordance of IgG ELISA results for the WNV (SPHL) panel and for the encephalitis/meningitis panela
|
The presence of WNV IgG antibody in a previously "naive" blood donor population is shown in Table 2. Two hundred and thirty-six healthy blood donor sera from Southern California were screened for the presence of WNV IgG; seven sera (3.0%) were positive and four sera (1.7%) were equivocal, giving a specificity for the Focus IgG ELISA of 97.0% for this population. The seven IgG positive sera may represent either IgG false-positive sera or prior exposure to a flavivirus, either by natural infection or vaccination. Therefore, the Focus IgG specificity in this population may actually be >97.0%. The final serum group evaluated for IgG reactivity was from individuals either naturally infected or vaccinated to non-WNV flaviviruses, including JE virus, SLE virus, dengue virus, and YF virus (Table 2). The cross-reactivity in the Focus WNV IgG ELISA for this study population ranged from a high of 95% cross-reactivity with sera from secondary dengue infection to a low of 15% for sera from YF vaccinees. The overall IgG cross-reactivity for the 131 sera in this flavivirus group was 35%. It is interesting that only 41% of the sera from naturally infected SLE patients were WNV IgG positive, whereas nearly all of the secondary dengue infection sera were WNV IgG positive. Also, only 22.5% of the JE vaccine recipients were WNV IgG positive, even though WNV is part of the JE serogroup.
|
View this table: [in a new window] |
TABLE 2. Focus WNV ELISA IgG results for non-WNV seraa
|
IgM ELISA performance. The lefthand columns of Table 3 present the Focus IgM results obtained for 583 sera tested at 16 different SPHLs during the 2002 WNV season. There was an overall concordance of 97.2% (566 of 582 samples) between the SPHL IgM results and the Focus IgM results. Although this sample set is heavily biased toward IgM-positive samples, it does allow for a good indication of the sensitivity of the Focus IgM assay (99.6% in this group). Three samples were SPHL positive, but Focus equivocal or negative; however, neither PRNT nor SPHL IgG results were available for these samples. One of the three samples was tested by Focus IgG ELISA and found to be negative. The specificity of the Focus IgM ELISA with this study group was only 68%; however, only 44 of the 583 samples were SPHL IgM negative, so very few samples were available to assess specificity in this panel. For the 14 sera that were Focus IgM positive and SPHL IgM negative, all remained Focus IgM positive after the background subtraction method described below. Neither SPHL IgG nor PRNT results were available for the 14 sera; only 2 of the 14 had Focus IgG results available, and both were IgG positive.
|
View this table: [in a new window] |
TABLE 3. Concordance of IgM ELISA results for the WNV (SPHL) panel and for the encephalitis/meningitis panela
|
|
View this table: [in a new window] |
TABLE 4. Focus WNV ELISA IgM results for non-WNV seraa
|
Preliminary studies using the background subtraction method were performed on known WNV IgM-positive and -negative sera. First, 35 consensus WNV IgM-negative sera, for which both the SPHL and the Focus results were IgM negative, were run in the Focus WNV IgM assay with the background subtraction method, and they remained IgM negative. Next, 36 sera positive for WNV IgM in the Focus assay and confirmed positive for WNV by PRNT results supplied by SPHLs were tested by the background subtraction method. All 36 sera remained Focus WNV IgM positive after background subtraction. Finally, 131 sera determined to be IgM positive by the CDC IgM ELISA were evaluated with or without the background subtraction modification of the Focus assay. On initial screening with the Focus kit, 130 sera were found to be positive for IgM, and one sample was equivocal with an index of 1.1. After the background subtraction method, 130 sera remained IgM positive and 1 sample was equivocal with the Focus IgM ELISA. The sample originally equivocal with an index of 1.1 became positive with an index of 1.2. Also, one sample originally positive with an index of 1.2 became equivocal with a post-background subtraction index of 1.1.
The background subtraction method was then applied to the following serum panels: a encephalitis/meningitis panel, a blood donor panel, a flavivirus panel, and the autoimmune panel. The IgM results for each serum panel when the background subtraction method was used are shown in Tables 3 and 4. For the encephalitis/meningitis panel, the two sera determined to be positive and two sera determined to be equivocal by Focus but negative by the CDC method were interpreted as negative after background subtraction. The 45 sera found to be positive by the CDC ELISA method retained the result obtained in the original Focus assay; i.e., 44 sera were IgM positive, and 1 serum remained equivocal after the background subtraction method was applied (Table 3). After the background subtraction method was applied, the Focus IgM ELISA still exhibited 97.8% sensitivity for the entire encephalitis/meningitis panel, and the specificity increased to 100%. The two sera in the blood donor panel that were IgM positive in the Focus IgM assay became negative after background subtraction; thus, all 236 donor sera were considered WNV IgM negative (Table 4).
When the background subtraction method was applied to the flavivirus serum panel (Table 4), the IgM results for the JE sera remained unchanged. For the dengue sera, three of the four WNV IgM-positive sera remained positive, four of the five originally WNV IgM-equivocal sera remained equivocal, and the remaining two sera were negative after background subtraction. Only three WNV IgM-positive sera from the SLE group were available for background subtraction studies, and all three remained WNV IgM positive. Finally, the five WNV IgM-positive sera in the autoimmune serum panel, as well as three sera from the IgM panel, were all interpreted as negative when the background subtraction method was applied. Table 5 provides the details of the pre- and post-background subtraction index values for the sera tested in Tables 3 and 4. Two sera (PRNT identification [ID] numbers M-1 and M-2) are included in Table 5 as examples of the absorbance and index values obtained by the background subtraction method with consensus IgM ELISA- and PRNT WNV-positive samples. The pre- and post-background subtraction index values for the 44 consensus IgM-positive sera from Table 3 are not included in Table 5.
|
View this table: [in a new window] |
TABLE 5. Absorbance and index values with or without background subtraction for IgM ELISA
|
|
|
|---|
Three serum panels were used to determine the performance characteristics of the Focus IgG and IgM ELISAs: the SPHL panel composed primarily of WNV-positive samples, the meningitis/encephalitis panel composed primarily of WNV-negative samples, and a blood donor panel from an area where WNV is not endemic. The first two panels included sera that were accurate indicators of the sensitivity of both the IgG and the IgM assays. The Focus IgG kit had a sensitivity of 97.6% with these two panels (100 and 95.3%, respectively), whereas the IgM kit had a combined sensitivity of 99.3% (99.6% in the SPHL panel and 95.6% in the meningitis/encephalitis panel). All three panels were used to determine the specificity of both the Focus IgG and IgM kits; the values ranged from 41 to 97.4% for the Focus IgG kit and from 68 to 99.2% for the Focus IgM kit. Both of the low specificity values for the IgG and IgM kits were obtained with the SPHL serum panel. This panel proved problematic for determining the specificity of the Focus kits for two reasons. First, only 44 of the 583 samples in the SPHL panel were negative for WNV IgM by the CDC protocol and, second, only 103 of the 583 samples had IgG results generated by the CDC protocol. For WNV IgG, the Focus IgG kits found 76 WNV IgG-positive samples, whereas only 42 of these 76 samples were determined to be positive by the CDC IgG protocol. However, 29 of the 34 CDC IgG-negative, Focus IgG-positive samples were positive for IgM by both the CDC IgM protocol and the Focus IgM assay, and 10 of 10 samples tested by PRNT were positive for WNV antibody. It may also be possible that the Focus IgG ELISA may be giving false-positive results for some or all of 24 sera where no PRNT results were available. The concordance of positive IgM results for these 29 sera may indicate a lack of sensitivity for determining IgG by the CDC IgG protocol rather than a lack of specificity by the Focus IgG assay. The SPHL panel had only 44 IgM negatives of the 583 tested by using the CDC protocol; however, 14 of the 44 CDC IgM protocol negative samples were positive by the Focus IgM assay. As with the IgG samples in the SPHL serum panel, the apparent lack of IgM specificity by the Focus assay may be due to lack of sensitivity of the CDC protocol performed at 16 different SPHLs. All 14 IgM discordant sera in this panel remained positive after the background subtraction method was used in the Focus IgM assay, indicating that these samples were not false positive by the Focus IgM assay. It should be noted that the SPHL serum panel represents WNV results generated from 16 different public health laboratories; discrepancies would be expected when such a large number of laboratories are involved. The impact of such a large number of laboratories submitting results and the consistency of the results could not be determined in the present study. In fact the relatively low number of discordant results obtained in this panel of nearly 600 sera demonstrates good concordance among the 16 public health laboratories. For the reasons outlined above, the encephalitis/meningitis panel and the blood donor panel were the most appropriate sera panels to determine the specificity of the Focus assays. The former panel included sera from meningitis/encephalitis symptomatic patients, most of who do not have WNV infection, and the latter panel includes sera from an WNV-naive healthy population. The specificity for the two combined panels was 98.4% (100 and 97.0%, respectively) for the Focus IgG assay, and 99.3% for the Focus IgM assay (99.3 and 99.2%, respectively). As noted previously, nearly 16% of the samples in the meningitis/encephalitis panel did not have the IgG results reported with the CDC IgG ELISA due to high background signal. These samples were primarily negative in the Focus IgG ELISA.
Heterophile antibodies in human sera react with proteins from other mammalian species; examples include human anti-mouse immunoglobulin, human anti-horse immunoglobulin, and human anti-bovine albumin (5, 11, 25). The presence of heterophile antibodies is known to interfere in numerous serologic assays, in particular IgM assays (5, 11). Although the Focus IgM assay gave a very high specificity with the panels investigated, when sera known to be problematic for IgM assays were tested, an increased number of false positives were encountered. This observation was especially true for rheumatoid factor-positive sera. Further, although the Focus IgM assay was >99% specific, one false-positive IgM result would be expected for approximately every 200 sera screened for the presence of WNV IgM. Internal studies have indicated the occasional false-positive IgM samples encountered with the Focus IgM assay are typically due to heterophile antibodies; however, these heterophile antibodies are heterogeneous and have reactivity to multiple mammalian proteins. To reduce the level of false-positive IgM results caused by heterophile antibodies to zero, a background subtraction method was used. The method did not adversely affect the sensitivity of the Focus IgM assay and produced a specificity of 100%. All false-positive results in the meningitis/encephalitis panel, the blood donor panel, the autoimmune panel, and the IgM panel were eliminated. Since the background subtraction step is only performed on samples initially IgM positive and the number of WNV IgM-positive samples routinely encountered is relatively low, only a small proportion of samples typically require testing by the background subtraction method.
The serologic detection of WNV or other flavivirus antibodies is confounded by the variable cross-reactivity detected by using the typical antibody screening methods such as ELISA and immunofluorescence assay (1, 12). The cross-reactivity is highest among viruses within a serocomplex, but cross-reactivity also occurs between flaviviruses from different serocomplex groups (4, 15, 26). Typically, the PRNT is required to determine the infecting flavivirus, since virus-specific neutralizing antibodies are detected in the PRNT. The Focus IgG and IgM assays demonstrated cross-reactivity with other flaviviruses, especially for the IgG assay, which had an overall 35% cross-reactivity. For the Focus IgM assay, 26% of patients with either secondary dengue or SLE infection were WNV IgM positive, whereas only 3% of JE and YE vaccinees were IgM positive.
Overall, the Focus Technologies IgG and IgM assays gave highly concordant results with the CDC WNV IgG and IgM protocols, and sensitivity and specificity values of 96 to >99%. The Focus IgM assay will yield specificity values of 100% when the background subtraction method is used.
|
|
|---|