Journal of Clinical Microbiology, November 2004, p. 5392-5393, Vol. 42, No. 11
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.11.5392-5393.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
Department of Pathology and Laboratory Medicine, School of Medicine,1 Biostatistics, Rollins School of Public Health, Emory University, Atlanta, Georgia2
Received 26 March 2004/ Returned for modification 14 June 2004/ Accepted 11 August 2004
|
|
|---|
|
|
|---|
For laboratories using the AMPLICOR test, version 1.5 has replaced version 1.0 in clinical practice. A recent study comparing versions 1.0 and 1.5 of the test showed that there was close agreement among the values obtained with the different test versions, and it was suggested that it was not necessary to reestablish a baseline viral load when changing versions of the test (2). However, since implementing the microwell plate version 1.5 test we have noted some patients with viral load values of <50 copies/ml of plasma with the version 1.0 test that were >50 copies/ml by the version 1.5 test. This observation led us to compare the sensitivity and viral load values between the MWP v1.0, the MWP v1.5, and the COBAS AMPLICOR HIV-1 MONITOR (COBAS v1.5) tests.
HIV-1 RNA (subtype B; AcroMetrix, Benicia, Calif.) was obtained at a concentration of 5,000 copies/ml and was diluted in human plasma to a concentration of 100 copies/ml of plasma. Quantification of the standard was confirmed by testing 10 replicates of the material in the MWP v1.5 test. After confirming the concentration, the control material was further diluted to 25 and 50 copies/ml in human plasma. Aliquots of the control material were frozen at 70°C until viral load testing was performed. Viral load testing was done using the ultrasensitive MWP v1.0 test, the MWP v1.5 test, and the COBAS v1.5 test. Testing was performed according to the manufacturer's protocols, using 500 µl of specimen. According to the manufacturer, the lower limit of quantification of the ultrasensitive tests is 50 copies/ml of plasma. One lot of reagents was used for each of the three tests. Forty replicates of the 25-copy/ml sample, 40 replicates of the 50-copy/ml sample, and 52 replicates of the 100-copy/ml sample were tested in each of the three assays. Two samples, a 25-copy/ml replicate in both the MWP v1.0 and COBAS v1.5 tests, gave invalid results due to failure to amplify the internal control and were eliminated from the data analysis. For this study, any sample with an optical density (OD) of >0.2 in the sample well was considered positive, even if the calculated viral load was <50 copies/ml of plasma. Data were analyzed using the
2 or Fisher's exact test based on the percentage of positive samples for each concentration and test and combined across the three concentrations. Probit regression analysis of percent positive versus log10 copies/ml was performed with SAS Proc Probit software (version 8) to compare slope and intercept estimates between the three tests and to determine for each test the viral load value and the 95% confidence interval (CI) at which 95% of the results are expected to be positive (1). For replicates of the 100-copy/ml sample that had a detectable viral load, the geometric mean and 95% confidence interval were calculated for each of the three tests (4). The mean viral load was compared between the three tests with a one-way analysis of variance, and pairwise comparisons were made with t tests. Reported P values are two-sided. A Bonferroni adjustment (P < 0.0167; that is, 0.05/3) was used for the pairwise comparisons between the three tests.
The percentage of samples that were positive for each of the three tests is shown in Table 1. When the 25-copy/ml sample was tested, 30, 72, and 64% of the replicates were positive with the MWP v1.0, MWP v1.5, and COBAS v1.5 tests, respectively. The differences in sensitivity were statistically significant when both the COBAS v1.5 and MWP v1.0 tests (P = 0.003) and the MWP v1.5 and MWP v1.0 tests (P = 0.0002) were compared. The percentage of positive replicates increased for both the 50- and 100-copy/ml samples for all three tests and reached a high of 98% for the MWP v1.5 and COBAS v1.5 tests with the 100-copy/ml sample. The differences in the percentage of positive replicates for the 50- and 100-copy/ml samples were not statistically significant for the three tests. When the results for the 25-, 50-, and 100-copy/ml samples were combined, the MWP v1.5 and COBAS v1.5 tests were both more sensitive than the MWP v1.0 test (for COBAS v1.5 versus MWP v1.0, P = 0.0016; for MWP v1.5 versus MWP v1.0, P = 0.0002). A limitation of the study is that all testing was done using a single lot of reagents for each test, so the impact of lot-to-lot variation was not assessed.
|
View this table: [in a new window] |
TABLE 1. Percentage of replicates with a detectable viral load and probits for the three tests
|
Probit analysis was used to estimate the limit of detection for the three tests (Fig. 1). The slopes of the three lines are not different (P = 0.74), but there are significant differences between the intercepts for the version 1.5 tests compared to the MWP v1.0 test (for COBAS v1.5 versus MWP v1.0, P < 0.001; for MWP v1.5 versus MWP v1.0, P < 0.001; for COBAS v1.5 versus MWP v1.5, P = 0.57). Viral load values that could be detected in 95% of the replicate tests were calculated to be 129 (95% CI, 94 to 238), 79 (95% CI, 56 to 220), and 77 (95% CI, 57 to 158) copies/ml for the MWP v1.0, MWP v1.5, and COBAS v1.5 tests, respectively.
![]() View larger version (9K): [in a new window] |
FIG. 1. Plot of probit (estimated probability) versus log10 copies/ml for the MWP v1.0 ( ), MWP v1.5 ( ), and COBAS v1.5 ( ) tests.
|
|
|
|---|
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||