Journal of Clinical Microbiology, December 2004, p. 5921-5922, Vol. 42, No. 12
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.12.5921-5922.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
Laboratoire de Biologie Clinique, HIA Val-de-Gràce, Paris,1 Laboratoire de Biologie Clinique, HIA Percy, Clamart,2 Laboratoire de Biologie Clinique, HIA Desgenettes, Lyon, France,4 Centre Paul Faure, Djibouti, Republic of Djibouti3
Received 17 October 2003/ Returned for modification 17 November 2003/ Accepted 20 July 2004
|
|
|---|
|
|
|---|
From January to April 1999, patients attending the Paul Faure clinic (a center specializing in tuberculosis diagnosis and treatment) in Djibouti, Republic of Djibouti, and presenting an LN with a diameter of >10 mm were included in the study. A standardized questionnaire was filled out for each patient. Specimens were collected by fine-needle (18-gauge) aspiration of the LN and then collected in a sterile tube. If the volume of the specimen was <0.1 ml, the needle and syringe were rinsed with 1 ml of sterile water. All specimens were shipped at room temperature (using air carrier) to Paris, France, and generally arrived within 48 h of collection. All analyses of specimens were performed in Paris, France. After decontamination using the N-acetyl-L-cysteine-sodium hydroxide procedure, specimens were separated into two equal parts. The first part was used for microscopy and culture (solid media, Lowenstein-Jensen and Coletsos; liquid medium, Mycobacterial Growth Indicator Tube [MGIT]). The second part was used to perform MTD. The remainder of the specimen was stored.
Isolates of M. tuberculosis complex strains were identified by routine biochemical methods and DNA probes (Accuprobe; Gen-Probe, Inc., San Diego, Calif.). The cutoff value for a positive MTD result was set at 1,000,000 relative light units (RLU), and the equivocal zone was defined as values between 30,000 and 1,000,000 RLU (1, 5). Each run included positive and negative amplification controls. The positive amplification control was prepared from a 104 to 105 dilution of a suspension of M. tuberculosis with a McFarland standard of 1. Sterile water was used as the negative amplification control. The run was validated when the negative-control value was <20,000 RLU and the positive-control value was >500,000 RLU.
All specimens were analyzed for the presence of amplification inhibitors by adding an aliquot of the positive control to the sample. A 1/5 dilution in sterile water was used to remove inhibitors. If there was a discrepancy between culture and MTD results, the final diagnosis was made by considering the patient's clinical, biological, and radiological findings (8); the factors taken into account were a positive culture for another sample from the same patient, clinical signs with highly suggestive lesions on chest X-ray, a known history of tuberculosis, or a favorable outcome with antituberculosis treatment. Assay performance was evaluated by comparing the assay results to the final diagnosis.
From January 1999 to April 1999, 197 LN aspiration specimens from various sites (cervical, 98; submaxillar, 42; axillar, 21; supraclavicular, 19; inguinal, 10; others, 7) were obtained from 153 patients. Of the 197 specimens, 30 were smear positive and culture positive, 89 were smear negative and culture negative, and 78 were smear negative and culture positive. A total of 108 strains belonging to the M. tuberculosis complex were isolated (M. tuberculosis [89.9%], Mycobacterium africanum [4.6%], Mycobacterium canetti [4.6%], and Mycobacterium bovis [0.9%]). No atypical Mycobacterium species was isolated.
Of the 115 samples testing positive with MTD, 30 were smear and culture positive, 70 were smear negative and culture positive, and 15 were smear and culture negative. Of the 82 MTD-negative samples, none were smear and culture positive, 8 were smear negative and culture positive, and 74 were smear and culture negative. Four samples (2%) exhibited amplification inhibition (two culture-negative samples and two culture-positive samples; for the latter, inhibitors were successfully removed after sample dilution).
The eight false-negative specimens (MTD negative and culture positive) represent 7.4% (8 of 108) of the culture-positive samples; they did not contain amplification inhibitors. Some researchers have noted that, in contrast to pulmonary specimens, the lack of sensitivity of molecular diagnosis performed on extrapulmonary samples might result from the use of very small sample volumes and an irregular dispersion of the aggregated bacteria in paucibacillary specimens (4, 8). Our data are in accord with this hypothesis: seven of the eight discrepant samples had a very low volume (<0.1 ml), four samples had discrepancies between the culture medium used (three samples gave a negative result in MGIT and a positive result on solid medium, and one sample gave a positive result in MGIT and a negative result on solid medium), the time to culture positivity was high, and the number of colonies on slants was low (between one and three colonies per two tubes for seven samples). The 15 apparent false-positive specimens with MTD (MTD positive, culture negative) came from patients with confirmed tuberculosis, according to clinical, historical, and other biological findings (respiratory tuberculosis, 2 patients; positive response to treatment, 3 patients; history of tuberculosis, 3 patients; other LN with positive culture, 4 patients).
After the discrepant results were resolved, of the 153 total patients assessed, 95 (62%) had a conclusive diagnosis of tuberculosis. The sensitivity, specificity, positive predictive value, and negative predictive value for the 197 specimens tested were 87, 100, 100, and 83% for culture, respectively, and 93, 100, 100, and 90% for MTD, respectively (Table 1).
|
View this table: [in a new window] |
TABLE 1. Performance of culture and MTD on LN specimens after discrepant results were resolved
|
Given the high performance of MTD with fine-needle LN aspiration specimens, we recommend this technique for a rapid diagnosis of LN tuberculosis. The use of MTD and fine-needle LN aspiration could reduce the necessity for performing open LN biopsy when testing for LN tuberculosis; it could also be used in developing countries to perform epidemiological surveys.
|
|
|---|