JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chapin, K. C.
Right arrow Articles by Musgnug, M. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chapin, K. C.
Right arrow Articles by Musgnug, M. C.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, February 2004, p. 909-911, Vol. 42, No. 2
0095-1137/04/$08.00+0     DOI: 10.1128/JCM.42.2.909-911.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.

Evaluation of Sensititre Automated Reading and Incubation System for Automated Reading of Sensititre Broth Microdilution Susceptibility Plates

Kimberle C. Chapin1* and Michael C. Musgnug2

Department of Pathology, Lifespan Academic Medical Centers, Rhode Island and Miriam Hospitals, Providence, Rhode Island,1 Binax, Portland, Maine2

Received 11 July 2003/ Returned for modification 10 September 2003/ Accepted 13 November 2003


    ABSTRACT
 Top
 Abstract
 Introduction
 References
 
The Sensititre Automated Reading and Incubation System (ARIS) performs automated reading of Sensititre identification and susceptibility plates. ARIS susceptibility results were compared to manually read results. A total of 212 isolates (3,352 organism-antimicrobial agent combinations) yielded 99.4% essential agreement and categorical error rates of 0.2% minor, 0% major, and 1.0% very major. ARIS yields accurate susceptibility results compared to plates read manually.


    INTRODUCTION
 Top
 Abstract
 Introduction
 References
 
Microdilution susceptibility tests are manually interpreted by visual inspection or using an automated instrument. The Sensititre Automated Reading and Incubation System (ARIS) (TREK Diagnostic Systems, Inc., Cleveland, Ohio) is an automated system that detects fluorescence liberated by bacterial enzymes that cleave fluorogenic substrates in the broth. A computer algorithm converts this fluorescent signal to a MIC. Two different instruments are available to perform automated reading of the fluorescent plates: the semiautomated Autoreader, which requires off-line incubation of plates and reads one plate at a time, and the fully automated, robotics-driven ARIS, which incubates and reads plates with minimal user intervention.

Sensititre susceptibility plates have been available for more than 20 years. Plates produced in the early 1980s were interpreted by nonfluorescent methods and interpreted manually by visualizing macroscopic growth with a light box. Early studies yielded results equivalent to the National Committee for Clinical Laboratory Standards (NCCLS) frozen reference microdilution method (2-4). Subsequently, susceptibility plates that used a fluorogenic substrate strip and allowed automated interpretation by the Autoreader, a semiautomated system, as well as manual interpretation were developed. Studies in the 1980s showed that the plates read by the Autoreader yielded results that were equivalent to the results from plates that were visually inspected using a light box (8, 9).

At this time, there are two Sensititre plate modifications. While all Sensititre susceptibility plates can still be read manually, the two modifications use fluorogenic substrates to allow for automated interpretation. Both modifications of Sensititre plates can be used in conjunction with both the semiautomated Autoreader and the fully automated ARIS instruments. One plate format has dried fluorogenic substrate in the wells. This substrate in well (SIW) modification simplifies plate setup by eliminating the addition of a separate substrate strip. Recently, ARIS compared favorably to a popular commercial automated method, MicroScan WalkAway (Dade Behring, West Sacramento, Calif.) (1) using Sensititre SIW plates. SIW plates are preferred for testing sites that have custom-formatted plates and/or have higher testing volume. The second plate format requires the addition of a fluorogenic substrate strip to the broth prior to inoculation. Plates requiring a substrate strip may be preferable for laboratories with smaller testing volume where manual or semiautomated interpretation is appropriate. Both plate formats, those requiring a substrate strip and SIW plates, produce the same results (Roger Grist [TREK Diagnostic], personal communication).

The objective of this investigation was to evaluate the performance of the ARIS for automated reading of Sensititre plates requiring the addition of a substrate strip compared to visual inspection of the same plates. In addition, the background of increasing resistance to antimicrobial agents should provide a greater challenge for this system compared to earlier studies.

A total of 212 clinical isolates were evaluated (Table 1). Appropriate quality control strains recommended by NCCLS (6, 7) were tested each day of testing. Isolated colonies that had been incubated overnight on solid medium were added to demineralized water to prepare a 0.5 McFarland suspension as verified by a Sensititre nephelometer. A fluorogenic substrate strip was added to a 10-ml sample of Sensititre cation-adjusted Mueller-Hinton broth. Thirty minutes later, a 10-µl aliquot of the organism suspension was transferred to the broth containing the substrate strip. Sensititre MG and MH susceptibility plates (for gram-positive and gram-negative organisms, respectively) were inoculated with this suspension. A Sensititre AutoInoculator, a robotics-driven dosing platform, delivered 50 µl of this suspension to each well of the susceptibility plates. Adhesive plate seals were added, and plates were placed in the ARIS at 35°C for 18 to 24 h. Automated plate reading was performed by the ARIS, and manual plate reading was done using the SensiTouch computer-driven light box.


View this table:
[in this window]
[in a new window]
 
TABLE 1. Clinical isolates tested in this study

 
Only those antimicrobial-organism combinations suggested by NCCLS as appropriate for routine use and reporting were evaluated (7). The ranges of antimicrobial concentrations tested are shown in Table 2. Results from the automated read and manual read were compared for essential agreement (results within 1 twofold dilution). Results were analyzed for the following categorical errors: minor (intermediate versus susceptible or resistant result), major (ARIS, resistant result; manual, susceptible result [falsely resistant]), and very major (ARIS, susceptible result; manual, resistant result [falsely susceptible]). Any minor error that was in essential agreement was not tallied as an error. For example, a result of 8 µg/ml (susceptible) compared to 16 µg/ml (intermediate) would be in essential agreement and therefore not be tallied as a minor error.


View this table:
[in this window]
[in a new window]
 
TABLE 2. Sensititre antimicrobial agents and concentrations tested

 
Essential agreement was 99.5% for 122 gram-negative isolates (2,092 antimicrobial-organism combinations). Categorical errors were as follows: 3 (0.1%) minor, 0 (0%) major, and 5 (1.0%) very major errors. Ninety gram-positive isolates (1,260 antimicrobial-organism combinations) yielded 99.4% essential agreement and the following categorical errors: 2 (0.2%) minor, 0 (0%) major, and 4 (1.1%) very major errors (Table 3).


View this table:
[in this window]
[in a new window]
 
TABLE 3. Accuracy of ARIS for reading Sensititre susceptibility plates

 
Very major errors occurred with Pseudomonas aeruginosa and ciprofloxacin and trimethoprim-sulfamethoxazole and with Acinetobacter spp. and ciprofloxacin, gentamicin, and trimethoprim-sulfamethoxazole. Enterococcus faecalis and rifampin yielded three very major errors, while Enterococcus faecium and rifampin yielded one very major error (Table 4).


View this table:
[in this window]
[in a new window]
 
TABLE 4. Details of very major errors

 
Although ciprofloxacin and trimethoprim-sulfamethoxazole each yielded two very major errors, this represents only 2% of gram-negative isolates. It is unclear why 12% of Enterococcus isolates yielded errors with rifampin.

The ARIS performs well for automated reading of Sensititre susceptibility plates requiring the addition of a substrate strip compared to results determined manually by visual inspection. Performance exceeds 90% essential agreement, which is generally accepted as satisfactory (5). Categorical error rates are also well within accepted limits. The Sensititre ARIS/Autoreader automated system appears to work well in this era of new antimicrobials and increasing resistance and proves to be a viable alternative in automated antimicrobial susceptibility testing.


    ACKNOWLEDGMENTS
 
This study was funded by TREK Diagnostic Systems and the Robert E. Wise, MD, Research and Education Institute, Lahey Clinic.


    FOOTNOTES
 
* Corresponding author. Mailing address: Department of Pathology, Lifespan Academic Medical Centers, Rhode Island and Miriam Hospitals, 593 Eddy St., Providence, RI 02903. Phone: (401) 444-2526. Fax: (401) 444-8514. E-mail: kchapin{at}lifespan.org. Back


    REFERENCES
 Top
 Abstract
 Introduction
 References
 

  1. Chapin, K. C., and M. C. Musgnug. 2003. Validation of the Automated Reading and Incubation System with Sensititre plates for antimicrobial susceptibility testing. J. Clin. Microbiol. 41:1951-1956.[Abstract/Free Full Text]
  2. Gavan, T. L., R. N. Jones, and A. L. Barry. 1980. Evaluation of the Sensititre system for quantitative antimicrobial drug susceptibility testing: a collaborative study. Antimicrob. Agents Chemother. 17:464-469.[Abstract/Free Full Text]
  3. Hansen, S. L., and P. K. Freedy. 1983. Concurrent comparability of automated systems and commercially prepared microdilution trays for susceptibility testing. J. Clin. Microbiol. 17:878-886.[Abstract/Free Full Text]
  4. Jones, R. N., T. L. Gavan, and A. L. Barry. 1980. Evaluation of the Sensititre microdilution antibiotic susceptibility system against recent clinical isolates: three-laboratory collaborative study. J. Clin. Microbiol. 11:426-429.[Abstract/Free Full Text]
  5. Jorgensen, J. H. 1993. Selection criteria for an antimicrobial susceptibility testing system. J. Clin. Microbiol. 31:2841-2844.[Free Full Text]
  6. National Committee for Clinical Laboratory Standards. 2000. Methods for dilution antimicrobial susceptibility test for bacteria that grow aerobically. Approved standard M7-A5. National Committee for Clinical Laboratory Standards, Wayne, Pa.
  7. National Committee for Clinical Laboratory Standards. 2002. Performance standards for antimicrobial susceptibility testing. 12th informational supplement. M100-S12. National Committee for Clinical Laboratory Standards, Wayne, Pa.
  8. Nolte, F. S., K. K. Krisher, L. A. Beltran, N. P. Christianson, and G. E. Sheridan. 1988. Rapid and overnight microdilution antibiotic susceptibility testing with the Sensititre breakpoint Autoreader system. J. Clin. Microbiol. 26:1079-1084.[Abstract/Free Full Text]
  9. Staneck, J. L., S. D. Allen, E. E. Harris, and R. C. Tilton. 1985. Automated reading of MIC microdilution trays containing fluorogenic enzyme substrates with the Sensititre Autoreader. J. Clin. Microbiol. 22:187-191.[Abstract/Free Full Text]


Journal of Clinical Microbiology, February 2004, p. 909-911, Vol. 42, No. 2
0095-1137/04/$08.00+0     DOI: 10.1128/JCM.42.2.909-911.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.





This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chapin, K. C.
Right arrow Articles by Musgnug, M. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chapin, K. C.
Right arrow Articles by Musgnug, M. C.


Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
Antimicrob. Agents Chemother. Clin. Microbiol. Rev.
Clin. Vaccine Immunol. ALL ASM JOURNALS