Leptospirosis usually results from contact with the urine of infected animals (13). The diagnosis of leptospirosis is mainly based on serological tests, with the microscopic agglutination test (MAT) considered the standard methodology (8, 12). The serological test for Leptospira is based on the formation of bacterial aggregates resulting from the addition of serum samples to the Leptospira suspension. The agglutination leads to a significant change in the analyzed particles, as observed by dark-field microscopy by the MAT procedure.
A variety of serological tests other than MAT have been developed for the diagnosis of leptospirosis. Among them are the complement fixation test (33), several enzyme-linked immunosorbent assay formats (1, 29), the macroscopic slide agglutination test (14), the microcapsule agglutination test (9), the indirect hemagglutination assay (20), the dipstick assay (27), and other methods (3, 15, 22, 30). Each assay has its own advantages, drawbacks, and limitations (4, 18). Despite its widespread use, MAT has several limitations. The test is difficult to perform and control, the results are difficult to interpret, and it is time-consuming and labor-intensive (31). The interpretation of MAT results is subjective and may cause quality assurance difficulties. One of the disadvantages of serologic testing by MAT compared to that by other techniques is its low sensitivity, particularly with early acute-phase specimens (3, 5, 10). In this work, we describe the use of the flow cytometry (FCM) technique for the serological diagnosis of leptospirosis. It is shown that the diagnosis of leptospirosis and the definition of the serogroup involved are feasible, based on the changes in the light scatter parameters forward scatter (FSC) and side scatter (SSC). By the FCM technique, the sizes and the shapes of the cells can be determined by measurement of FSC and SSC (17, 26, 32). Whereas FSC is related to the cell size and the optical refraction index of the outer membrane of the cell, SSC is related to the cell's granularity. Analysis is possible due to the highly developed new generation of flow cytometric analyzers with the capability of observing particles with diameters of 0.5 µm, which is as small as a variety of bacterial species (2, 7, 11, 23-25, 28, 35).
FCM analysis was found to be objective, sensitive, and rapid. The duration of the whole process, i.e., the times for incubation of the sera, analysis, and interpretation of the results, was less than 1.5 h.
|
|
|---|
1:200, according to the case definition of the Centers for Disease Control and Prevention (6). In the present study patients positive for leptospirosis were considered those suspected of having leptospirosis by the clinicians because they exhibited clinical symptoms typical of the disease. Their sera were sent to the Israel Institute for Biological Research and were found to be positive by MAT. Also, nine of these patients were known to work in areas of leptospirosis outbreaks from which L. interrogans serovar hardjo was isolated.
Sera from 26 patients whose paired sera were found to be negative by MAT served as negative controls: 4 serum samples from patients with murine typhus, 4 serum samples from patients with Mediterranean spotted fever, 4 serum samples from patients with Q fever, 5 serum samples from patients with syphilis, and 9 serum samples from patients with other clinical symptoms resembling leptospirosis.
Bacterial culture. Twenty-one reference serovars of living leptospiral spirochetes were used; among these were 19 pathogenic serovars (L. interrogans) and 2 nonpathogenic (L. biflexa) serovars. Details about the serovars are listed in Table 1.
|
View this table: [in a new window] |
TABLE 1. Reference serovars used in this study
|
= 488/10 nm), a photomultiplier for SSC (
= 488/10 nm), and a photomultiplier for green fluorescence (FL1; emission
= 530/30 nm). The instrument settings included logarithmic amplifiers on all detectors. All experiments were performed for a fixed time of analysis (30 s). Acquisition and analysis were performed with CellQuest software. Preparation of samples for FCM analysis. Human sera were diluted (1:100) in saline-formaldehyde (0.14%) and filtered through a low-protein-binding 0.45-µm-pore-size syringe filter. Leptospiral organisms were grown for a week in EMJH medium (nos. 279410 and 279510; Becton Dickinson), counted with a Petroff-Hausser counting chamber under a dark-field microscope to confirm the presence of 1 x 108 to 2 x 108 bacteria/ml, and then harvested. Each serovar was diluted 1/10 in saline-formaldehyde (0.14%), incubated in an equal volume of serum at room temperature for 60 min, and analyzed by FCM.
Quantification of agglutination. Quantification of the agglutination process based on the light scatter parameters was achieved by comparison of the agglutinated to the nonagglutinated Leptospira dot plots. Nonagglutinated leptospires could be observed as a subpopulation under region R2 on a dot plot of the light scatter parameters (FSC and SSC) (Fig. 1). Agglutinated leptospires could be observed as a subpopulation under region R3 on the same dot plot (Fig. 1).
![]() View larger version (56K): [in a new window] |
FIG. 1. Dot plots of L. interrogans serovar icterohaemorrhagiae obtained by FCM. (A) Leptospira in phosphate-buffered saline; B(I) and B(II) light scatter (SSC-FSC) and fluorescence (SSC-FL1) dot plots, respectively, of L. interrogans serovar icterohaemorrhagiae in the presence of rabbit anti-icterohaemorrhagiae serum (30 min, 37°C) and the secondary antibody goat anti-rabbit immunoglobulin G-FITC (20 min, 37°C); (C) light scatter dot plots of Leptospira in the presence of rabbit anti-icterohaemorrhagiae serum (1:2,000) after 1 min (I), 5 min (II), and 10 min (III) of incubation at 37°C; (D) same as panel C after 35 min of incubation at 37°C with the serum dilution levels of 1:8,000 (I), 1:4,000 (II), and 1:2,000 (III). R1 (green), the region of FITC-stained agglutinated Leptospira; R2 (red), nonagglutinated Leptospira; R3 (pink), agglutinated Leptospira.
|
![]() | (1) |
|
|
|---|
Figure 1A shows the light scatter parameters of nonagglutinated Leptospira. Even though the leptospiral size is in the micron range, and hence their signals were at the limit of sensitivity of FCM, the Leptospira population could be observed by using logarithmic amplification of the FSC and SSC parameters. The Leptospira population was defined as region R2, in which a concentration dependency of the bacteria was observed (data not shown).
Upon addition of the specific rabbit antiserum and anti-rabbit immunoglobulin G-FITC conjugate, the majority of the Leptospira population shifted from region R2 to region R3 [Fig. 1B(I)], which reflected stronger FSC and SSC signals. The same events that reached the new SSC and FSC region (region R3) [Fig. 1B(I)] were also the events with higher fluorescence signals and were observed in region R1 on the SSC-FL1 dot plot [Fig. 1B(II)]. This correlation indicates that the larger particles observed by the light scatter parameters (FSC and SSC) in region R3 are the agglutinated leptospires; hence, they light up by specific immune staining, as shown in the SSC-FL1 dot plot (region R1). No agglutination was observed by performing the same experiment with the same experimental setup but with nonimmune rabbit serum (data not shown). The correlation between the events located in region R3 and those located in region R1 indicates that light scatter parameters are sufficient for the analysis of agglutinated Leptospira.
The shifts of the Leptospira signals from region R2 to region R3 were found to be dependent on the serum concentration and the incubation time (Fig. 1C and D, respectively). By the FCM technique, it was possible to observe agglutination of Leptospira in an incubation time shorter than 5 min with a serum dilution of 1:2,000. The titer in the same serum sample by end-point titration by MAT was 1:3,200 after incubation for 1 h, while by the FCM technique agglutination could be observed after 35 min of incubation, even when a dilution of 1:8,000 was used. Hence, compared to MAT, FCM analysis is more sensitive and rapid.
FCM analysis of sera from suspected human leptospirosis cases.
Twenty-seven serum samples from 18 patients were examined by MAT and FCM analysis. Since a single titer
1:200 by MAT is considered positivity for leptospirosis, FCM analysis was performed with a serum dilution of 1:200. The sera were incubated for 60 min with each Leptospira serovar and then subjected to FCM analysis.
Quantification of the agglutination process, based on the light scatter parameters, was achieved by calculating Af, as described in the Materials and Methods section. Calculation of Af values for negative control sera enabled us to set a threshold that distinguished between negative and positive sera. In order to set the threshold, we measured the Af values for each of the 26 negative serum samples (described in Materials and Methods) against 13 serovars, resulting in a total of 338 tests. The average Af value was calculated to be 1.8 ± 1.5, and serum samples with Af values
9.3 (average plus five times the standard deviation) were considered negative for leptospirosis.
Figure 2A presents the results of a typical FCM analysis of a serum sample (from patient 2) positive by MAT and analyzed for 13 different Leptospira serovars. It was concluded from the FSC and SSC results that the serum sample was positive for serovars icterohaemorrhagiae copenhageni (serovar 1 in Fig. 2A), sejroe sejroe M84 (serovar 5), szwajizak szwajizak (serovar 6), and icterohaemorrhagiae RGA (ATCC 43642) (serovar 29). All of the same serovars except sejroe sejroe M84 were also found to be positive by MAT, with titers of 1:400, 1:400, and 1:200, respectively.
![]() View larger version (45K): [in a new window] |
FIG. 2. (A) Light scatter dot plots of 13 different Leptospira serovars (as presented in the Materials and Methods section and indicated in parentheses in each panel) in the presence of human serum sample 2/II. R2, nonagglutinated Leptospira; R3, agglutinated Leptospira. (B) Af for serum sample 2/II in comparison to that for a negative serum sample.
|
Table 2 summarizes the predominant serogroups obtained by MAT and FCM analyses of the 27 serum samples from 18 leptospirosis patients. In addition to exhibiting typical clinical symptoms of the disease, the sera of these patients were found to be positive for leptospires by MAT, which was performed against 21 serovars. Moreover, nine of these patients (patients 3, 5, 7, 8, 9, 15, 16, 17, and 18) were known to work in areas where leptospirosis outbreaks had occurred and from which serovar hardjo had been isolated.
|
View this table: [in a new window] |
TABLE 2. FCM and MAT results for leptospirosis patients and negative controls
|
In some cases it was possible to identify leptospirosis in patients by FCM significantly before it was detected according to the appropriate titers by MAT. For example, the indication of positivity for leptospirosis by FCM appeared in the acute-phase sera of patients 2, 6, and 16, whereas by MAT it appeared only in the sera obtained later. These results demonstrate that the FCM technique can reliably be used as a tool for the diagnosis of leptospirosis at an early stage of the disease and for the identification of the infecting serogroup.
From Table 2 one can see that the Af value did not always correspond to the MAT titers. For example, the titer obtained by MAT was 1:400 for both patients 2 and 3, while the Af values were 330 and 40, respectively. Despite these differences, the predominant serogroup appeared to be the same by both methods. In three other patients (patients 4, 11, and 14), although the patients were found to be Leptospira positive by both the MAT and the FCM techniques, the predominant serogroups identified and defined by MAT and FCM were different. The predominant serogroups found by MAT were icterohaemorrhagiae, icterohaemorrhagiae, and canicola, respectively, whereas the predominant serogroups found by FCM were szwajizak, szwajizak, and ballum, respectively. However, the serogroups that were predominant by MAT also gave positive Af values. These differences can be explained by the differences in the natures of the two different methods and by the cross-reactivities of the serovars.
|
|
|---|
By implementation of this methodology, FCM analysis enabled the detection of the serogroup in all patients, and in three patients (patients 2, 6, and 16) serogroup detection occurred in the acute phase, when the MAT result was still negative.
However, the Af value did not always correspond to the MAT titers, and in 3 of 18 patients (patients 4, 11, and 14), the predominant serogroup determined by MAT differed from that determined by FCM. This can be explained by the differences in the natures of the two methods: microscopy takes into account only the presence or the absence of aggregates, whereas FCM measures light scatter parameters and the analysis is more precise, as it considers size, shape, and number of the aggregates. Moreover, FCM analysis can detect very small aggregates not visible by light microscopy, as can be found in the early stages of agglutination. This may lead to positive Af values and negative MAT results in the early stages of the disease. Moreover, it was recently shown (19) that serological analysis by MAT could not always predict and identify the infecting serovar in individual patients. The reasons for the poor predictive ability of MAT could emanate from the cross-reactivity between serogroups and from the paradoxical reaction of an acute-phase or an early-convalescent-phase serum sample. It is possible that analysis by FCM overcomes some of these problems as a result of its capability to analyze multiple parameters (FSC, SSC, and fluorescence) and will be able to improve the specificities and sensitivities of serological tests.
In addition, an intrinsic limitation of MAT is the subjective interpretation of the results and the difficulties in ensuring standardization between laboratories. The FCM methodology eliminates these drawbacks due to the accuracy and objectivity stemming from the nature of the FCM analysis. Another advantage of FCM is its rapidity, as the entire procedure, including incubation time and analysis, is completed in 1.5 h. Furthermore, the analysis can be automated and used to perform large numbers of tests.
In conclusion, FCM uses standard equipment, is available in many hospitals, and is used mainly for blood counts and other immunological purposes. This method can easily be used for the diagnosis of leptospirosis because of its sensitivity and objectivity and because automated procedures can be applied to FCM.
|
|
|---|