Journal of Clinical Microbiology, May 2004, p. 2094-2100, Vol. 42, No. 5
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.5.2094-2100.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
Institute of Transfusion Medicine and Immunohematology, German Red Cross, Johann Wolfgang Goethe University, Frankfurt, Germany
Received 31 October 2003/ Returned for modification 28 November 2003/ Accepted 25 December 2003
|
|
|---|
|
|
|---|
Before the PCR assay for the SARS CoV was available, diagnosis of this atypical pneumonia was based solely on clinical and epidemiological findings: acute febrile illness with respiratory symptoms not attributed to another cause and a history of exposure to an individual with a suspect or probable case of SARS or the individual's respiratory secretions or other bodily fluids (15).
The introduction of specific PCR testing led to the faster and more reliable confirmation of infections with the new strain of coronavirus, thus helping to gain control over the epidemic (20). Even though no new infections have been reported for several months, a renewed outbreak in colder seasons seems likely because the reservoir of the virus has not been definitely identified (1, 18).
In the present study we investigated two reverse transcription (RT)-PCR assays using the Roche LightCycler apparatus: the new Roche SARS CoV Quantification kit (referred to here as the Roche assay), codeveloped by Roche Diagnostics, Mannheim, Germany, and the Genome Institute of Singapore, was compared with the first commercially available assay, the Artus assay. The sensitivities and the ranges of validity of both assays were assessed.
|
|
|---|
RNA was prepared from the viral standard with a Qiagen viral RNA kit according to the instructions of the manufacturer. Purified RNA (external standard RNA) was diluted to final concentrations of 101, 102, 103, 104, 105, and 106 copies/ml.
Artus assay. The Artus assay is a ready-to-use system for the detection of SARS CoV RNA. The assay contains reagents and enzymes for the specific amplification of an 80-bp region of the SARS CoV genome, and additionally, the assay contains a second heterologous amplification system to identify possible PCR inhibition. Primer and probe sequences were described by Drosten et al. (9). Internal SARS CoV standards, which allow the determination of the pathogen load, are supplied with the assay kit.
Real-time quantitative amplification of SARS CoV RNA was performed according to the instructions of the manufacturer. A total of 5 µl of RNA extract was transferred into reaction tubes containing 15 µl of PCR reagents. RT was performed at 50°C for 10 min; and amplification was performed for 1 cycle of 95°C for 10 s and 50 cycles of 95°C for 2 s, 55°C for 12 s, and 72°C for 10 s. Finally, cooling was performed at 40°C for 30 s.
Roche assay. The Roche assay contains a reaction mixture that amplifies a 180-bp target sequence of the replicase 1AB/polymerase gene of SARS CoV. Specific probes emit fluorescent light after hybridization to the target sequence. An internal control sequence (180 bp) is amplified by the same pair of primers that amplify the target sequence, but the internal control sequence is detected with different hybridization probes.
Real-time quantitative amplification of SARS CoV RNA was performed according to the instructions of the manufacturer. The company does not publish the primer sequences. A total of 5 µl of RNA extract was transferred into reaction tubes containing 15 µl of PCR reagents. RT was performed at 61°C for 20 min; and amplification was performed for 1 cycle of 95°C for 30 s and 45 cycles of 95°C for 5 s, 55°C for 15 s, and 72°C for 10 s. Finally, cooling was performed at 40°C for 30 s.
Statistical analysis. The standard deviation (SD) and coefficient of variation (CV) of the real-time PCR test were calculated by using Excel 2000 software. Sensitivity was estimated by probit analysis with SPSS (version 11.5) software. Student's unpaired t test was performed with the data from the probit analysis. Statistical significance was considered if P was <0.05.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Precisions of the Roche and Artus assays for detection of SARS CoVa
|
![]() ![]() View larger version (33K): [in a new window] |
FIG. 1. Linear ranges of SARS CoV assays. (A) Correlation between nominal SARS CoV RNA concentrations and measured SARS RNA concentrations analyzed with the external standard RNA. , Roche assay; , Artus assay; , line of equality. (B) Correlation between nominal and measured SARS CoV RNA concentrations analyzed with kit-specific internal standards. The correlation factors (R2 values) showed no significant differences.
|
|
View this table: [in a new window] |
TABLE 2. Sensitivities of Roche and Artus assays for detection of SARS CoVa
|
![]() View larger version (26K): [in a new window] |
FIG. 2. Real-time PCR of SARS CoV by two real-time PCRs: the Roche assay (I) and the Artus assay (II). Real-time PCR runs for SARS CoV with six external standard RNA concentrations (A, 106 copies/ml; B, 105 copies/ml; C, 104 copies/ml; D, 103 copies/ml; E, 102 copies/ml; F, 101 copies/ml; NTC, no-template control) are shown. The Roche assay demonstrates positive results only for the first five concentrations (A to E), whereas the Artus assay shows positive results for all six concentrations (A to F).
|
Agreement between both assays. To assess the level of agreement between the two assays, we plotted the data on a logarithmic scale (Fig. 3) and drew a line of equality on which all points would lie if both assays gave exactly the same values at the same concentrations. At higher external standard RNA concentrations (>104 copies/ml) the level of agreement between the two both methods approached an asymptote along the line of equality. However, measurements below 104 copies/ml disagreed considerably. Therefore, the best way to estimate the intermethod differences would be to take the mean values obtained by both methods and plot those values against the differences in the means. According to Bland and Altman (4), limits of agreement are defined as the mean of differences ± two times the SD. It is assumed that 95% of the data lie between these limits if the differences are normally distributed. As demonstrated in Fig. 4A, there was good agreement between the two assays with high external standard RNA concentrations (106 copies/ml). The mean of the differences was close to zero (0.17), as expected for good agreement, and 95% of the data were between 0.36 and 0.02 after logarithmic transformation. However, for low external standard RNA concentrations (102 copies/ml; Fig. 4C), the mean of the differences was 1.45 and the limits of agreement (mean ± two times the SD) were 1.72 and 1.18, respectively.
![]() View larger version (20K): [in a new window] |
FIG. 3. Agreement between Roche assay and Artus assay. The results for external standard RNAs (106 to 101 copies/ml) were plotted against each other. The line of equality is represented by a dotted line. , measured values.
|
![]() View larger version (20K): [in a new window] |
FIG. 4. Limits of agreement according to Bland and Altman (4). The means of both methods (x axis) were plotted against the differences between the means of the two methods (y axis). (A) External standard RNA concentration of 106 copies/ml; (B) external standard RNA concentration of 104 copies/ml; (C) external standard RNA concentration of 102 copies/ml. The mean is indicated by a solid line; the means ± two times the SD (2SD) are indicated by dotted lines.
|
|
|
|---|
Our results demonstrate that the Roche assay shows a linear range from 1 x 104 to 2 x 108 copies/ml, whereas the Artus assay is linear from 1 x 102 to 1 x 107 copies/ml.
The 95% detection limits were shown to be 3,982 copies/ml for the Roche assay and 38 copies/ml for the Artus assay.
According to the instructions of the manufacturers, 5 µl of extract is used for each PCR. Therefore, in the case of the Artus assay, 0.19 copies per PCR mixture could be detected with 95% probability. Nevertheless, no false-positive measurements were obtained.
As a 95% detection limit below 2 to 5 copies is regarded as unrealistic, even for ultrasensitive assays, we assume that the external quantification standard is underestimated by at least one decimal unit. Thus, we calculate 95% probability limits of >40,000 copies/ml for the Roche assay and >380 copies/ml for the Artus assay.
Published concentrations of SARS CoV in different clinical specimens range from 108 copies/ml in sputum to 190 copies/ml in plasma (9). Since these concentrations were determined with the same external standard RNA preparations (9), revision is strongly recommended. As testing of sputum (which contains approximately 108 copies/ml) represents the main application, the use of both kits is suitable for verification of the disease. Due to high viral levels in sputum, the poor performance of the Roche assay with lower virus concentrations is of minor importance. Nevertheless, one can imagine that other clinical specimens will be used to test for SARS CoV. As reported previously, PCR testing of donated blood for SARS CoV may be used in an epidemic. We previously showed that the Artus assay has sufficient sensitivity to detect the low virus burdens in plasma samples, even in pooled material (17). Our data confirm the previously published high sensitivity of the Artus assay, whereas the Roche assay may not be reliable when viremia levels are below approximately 40,000 copies per ml.
Two possible explanations for the poor performance of the Roche assay with lower virus concentrations can be considered. In addition to the different lengths of the amplified sequences used in the Roche and Artus assays, the enzymes used for RT-PCR may be the reason. The Artus assay uses a combination of Moloney murine leukemia virus reverse transcriptase and Taq DNA polymerase, whereas the Roche assay applies the one-enzyme Tth (Thermus thermophilus) DNA polymerase assay, as described by Myers and Gelfand (14). Tth polymerase in combination with manganese ions for RT-PCR has been shown to be less susceptible to inhibitors and GC-rich genomes (2, 5, 12, 14, 16). However, the lack of sensitivity at low virus concentrations, presumably due to the insufficient RT activity of Tth polymerase, has been reported previously (5, 7, 8, 10, 11, 13).
In conclusion, we show that both the Roche and the Artus assays may be suitable for the verification of SARS by examination of sputum samples. Additionally, the Artus assay could even be used to detect SARS CoV in clinical specimens with low virus loads. Thus, the Artus assay provides a wider range of applications. Furthermore, we believe that the amount of the external standard RNA which was previously used to quantify the virus loads in different clinical specimens has been underestimated and is higher than has been reported previously. Therefore, in our opinion a repeat quantification of the virus load is necessary. Irrespective of the high virus level in sputum, an ultrasensitive PCR test is needed for blood donor services. Each donor receives a brief examination by medical professionals prior to blood donation. However, it is conceivable that a SARS CoV-infected blood donor may not be suspected if, for example, he or she has received antipyretic treatment. An improved PCR-based test would reduce the risk of transmission of SARS CoV by infected blood products.
We declare that we have no competing financial interests.
|
|
|---|