Journal of Clinical Microbiology, June 2004, p. 2636-2643, Vol. 42, No. 6
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.6.2636-2643.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
Divisions of Clinical Microbiology,1 Infectious Diseases,3 Department of Health Sciences Research, Mayo Clinic and Foundation, Rochester, Minnesota 559052
Received 8 September 2003/ Returned for modification 5 December 2003/ Accepted 17 March 2004
|
|
|---|
3.5 versus
72 h). The use of this test could have important implications for the effective control and prevention of nosocomial outbreaks of VRE. |
|
|---|
8 µl/ml, the clinical significance of VanC resistance is unknown (14). VanA and VanB phenotypes can occur in either Enterococcus faecalis or Enterococcus faecium strains. We and others have previously determined that based on DNA sequence information at least two additional vanB alleles exist, vanB2 and vanB3 (7, 15). The U.S. Centers for Disease Control and Prevention and the Society of Healthcare Epidemiology of America have provided recommendations to prevent the spread of VRE in institutional settings (2b,10). These recommendations include screening of patients by perianal or rectal swab or fecal surveillance cultures to identify carriers of VRE and subsequent isolation or cohorting of VRE carriers. As well, patients experiencing active VRE infection are isolated. Such a policy, when activated at 32 health care facilities in a region of the Midwest United States, significantly reduced or eliminated the transmission of VRE in these facilities (12).
Screening feces or perianal or rectal swab specimens for VRE by culture may be challenging. Various selective culture media, both agar and broth, have been developed that contain various amounts of vancomycin (8 to 64 µl/ml) (2, 8). Limitations of these methods include the following: (i) other vancomycin-resistant organisms, including Leuconostoc spp., Pediococcus spp., and Lactobacillus spp., may grow with these media; (ii) some Enterococcus spp. with vanB phenotypes may not grow due to MICs in the 6 to 8 µl/ml range; (iii) the time requirements for VRE confirmation frequently exceed 48 h; and (iv) Enterococcus spp. may exist in a viable but nonculturable state (3b). Relevant to points ii and iv, it is not surprising that a high percentage of false-negative results (42%) has been reported when culture-based detection methods are used to detect VRE colonization in stool specimens (3).
Recently, a real-time PCR assay, the Roche LightCycler vanA/vanB detection assay (Roche Molecular Diagnostics, Indianapolis, Ind.) became commercially available. This test method uses fluorescent energy transfer (FRET) probes and the LightCycler instrument (Roche Diagnostics Corporation) and can detect VRE (vanA and vanB genes) directly from fecal or perianal swab samples. Extraction of specimens for this assay can also be conveniently performed using a specialized buffer (Stool Transport and Recovery [S.T.A.R.] buffer; Roche Diagnostics Corporation, Indianapolis, Ind.) and the automated MagNa Pure instrument (Roche Applied Science) as well as commercially available manual extraction methods. The objective of the present study was to compare the Roche LightCycler vanA/vanB detection assay to culture for detection of VRE from perianal swabs. Performance characteristics, as well as turn around time for results, were compared for each test method.
|
|
|---|
Archived bacterial isolates. The one hundred archived clinical isolates of Enterococcus spp. were previously identified by the Mayo Clinic Microbiology Laboratory over the time period 1992 to 1995 and stored at 70°C. These enterococcal isolates were classified by demonstrating 6.5% NaCl tolerance and growth on bile-esculin agar with esculin hydrolysis. Identification to the species levels of Enterococci was based on fermentation of sugars, arginine hydrolysis, motility, pigmentation and growth on tellurite agar. Susceptibility testing was performed by an agar dilution method following guidelines of the National Committee for Clinical Laboratory Standards (11). Mueller-Hinton agar with vancomycin concentrations of 2 through 256 µg/ml and teicoplanin concentrations of 8 and 16 µg/ml were used, and the MICs were determined after 24 h of incubation at 35°C.
(i) Culture and DNA extraction. Each Enterococcus sp. isolate was inoculated to a Trypticase soy agar plate (Becton Dickinson Microbiology Systems, Sparks, Md.) containing 5% sheep blood and incubated at 35°C for 24 h. Three colonies were transferred into 100 µl of sterile water and boiled in a dry heat block at 100°C for 10 min. The suspension was centrifuged at 20,800 x g for 1 min.
(ii) Multiplex PCR-RFLP. A single multiplex PCR was performed using colonies growing on blood agar plates as previously described (14). This assay detects the presence of vanA, vanB, van-C1, or vanC-2 genes. The PCR products generated were digested with the enzyme MspI at 37°C overnight. The digested PCR products were electrophoresed on a 3% Nu Sieve agarose gel containing ethidium bromide to discriminate among the van genes.
Clinical study. (i) Collection of specimens. All specimens were collected using a swab collection and transport system (Culture Swab; Becton Dickinson Microbiology Systems, Cockeysville, Md.). The perianal specimen was obtained by swabbing the anal verge area without rectal insertion.
(ii) Perianal swab culture.
All perianal swabs were inoculated to an Enterococcosel culture plate (containing vancomycin at either 6 or 8 µg/ml) prior to processing for DNA testing. The plates were examined after incubation for 24 and 48 h for the presence of black colonies (bile esculin positive) which were Gram stained. Gram-positive cocci were plated to a blood agar plate for isolation and after 24 h were Gram stained and checked for catalase and pyrrolinodyl peptidase activity. The catalase-negative and pyrrolinodyl peptidase-positive isolates were reported as Enterococcus spp., and susceptibility testing was performed. Those isolates for which the MIC of vancomycin was
32 µg/ml were reported as VRE. For quality control purposes, each new lot of Enterococcosel agar was tested with vancomycin-resistant and -susceptible Enterococcus faecalis strains as recommended by the National Committee for Clinical Laboratory Standards (11).
(iii) Swab DNA extraction. After culture was performed the swab was swirled in a 1.5-ml screw-cap tube containing approximately 250 µl of 0.1-mm-diameter zirconia-silica beads (Biospec Products, Inc., Bartlesville, Okla.) and 300 µl of sterile distilled water. The capped tube was processed on a FastPrep Instrument (Qbiogene, Inc., Carlsbad, Calif.) for 30 s at 6.5. The tube was centrifuged at maximum speed for 1 min, and 100 µl of the supernatant was added to 100 µl of commercially available S.T.A.R. buffer (Roche Diagnostics Corporation) in a MagNA Pure sample cartridge. The DNA was extracted with the automated MagNA Pure LC instrument using the Total Nucleic Acid isolation kit (Roche Applied Science). Quality control for DNA extraction was assessed as follows. A positive control plasmid at a concentration of 100 copies/µl in S.T.A.R. buffer was stored at room temperature. This control, along with a negative control of 200 µl of S.T.A.R. buffer, was included in all MagNA Pure extraction runs.
LightCycler vanA/vanB PCR for archived bacterial isolates and clinical study. For each reaction mixture, 5 µl of the boiled lysate (from colonies) or extracted DNA (from swabs) was added to 15 µl of the PCR reagent mix. The PCR reagent mix contains the following: 2 µl of LightCycler FastStart DNA Master Hybridization Probe mix with enzyme, 7 µl of sterile water, and 2 µl of MgCl2 supplied with the FastStart reagents (Roche Diagnostics Corporation catalog no. 3 003 248), 2 µl of vanA/vanB Primer/Hybridization probe (Roche catalog no. 3 334 961), and 2 µl of the LightCycler vanA/vanB Recovery Template (Roche catalog no. 3 334 970). The LightCycler instrument in which color compensation was installed was programmed as described in Table 1. LightCycler software versions 3.1 and 3.5 were used for all the experiments. Sterile water was used as a negative control. A positive control, LightCycler vanA/vanB, Template DNA (Roche catalog no. 3 334 988) and a negative control were included in each run.
|
View this table: [in a new window] |
TABLE 1. Programming of the LightCycler instrument
|
The LightCycler instrument amplifies and monitors the fluorescent development of the target nucleic acid after each cycle. A melting curve analysis was used to differentiate the vanA, vanB, and vanB-2/3 gene targets. Representative results for analyses of real-time data are shown in Fig. 1 and 2. A representative melting curve analysis, which confirms the identification of the amplicon as belonging to one of the VRE genotypes, is shown in Fig. 3. The analytical sensitivity of the assay is less than 50 organisms per reaction. The LightCycler vanA/vanB detection assay is designed to detect and differentiate vanA, vanB, and vanB-2/3 in a single tube using two primer and probe sets.
![]() View larger version (30K): [in a new window] |
FIG. 1. Melting curve analysis for LightCycler vanA/vanB detection assay. The melting temperatures (means ± standard deviations) corresponding to the vancomycin-resistant genotypes were determined by melting curve analysis and are as follows: vanA, 67 ± 2.5°C; vanB, 60 ± 2.0°C; and vanB-2,3, 56 ± 2.0°C.
|
![]() View larger version (22K): [in a new window] |
FIG. 2. Detection of VRE DNA with the LightCycler instrument and the LightCycler vanA/vanB Primer/Hybridization probes. Positive results are indicated by an upward-deflecting curve as seen here in both the positive control and the positive sample. F2 refers to the fluorescence emission for the LC-Red640.
|
![]() View larger version (22K): [in a new window] |
FIG. 3. Quantitative representation or cycling curve analysis of recovery template (internal control) for the LightCycler PCR assay. The recovery template FRET probe has a reporter dye (LC-Red640) different from that of the FRET probe used to detect target DNA in the sample (LC-Red640) and is detected in channel F3 of the LightCycler instrument. This quality control step indicates whether inhibition of the PCR occurred in any of the patient samples or the positive and negative controls. Amplification of the recovery template should occur with each of these analyses except when the amount of target DNA in the patient's sample significantly exceeds that of the recovery template DNA.
|
Additional testing included (i) culture of the original perianal swab into heart infusion (HI) broth (Becton Dickinson Microbiology Systems) with subculture of HI broth to both an Entercoccosel plate and CNA plate (Becton Dickinson Microbiology Systems) (the same concentration of vancomycin [either 6 or 8 µg/ml] was used for this Enterococcosel subculture plate as that of the original Enterococcosel agar plate used for the specimen); (ii) PCR using the LightCycler vanA/vanB detection assay of a swab of undifferentiated bacterial growth, if present, from the original Enterococcosel plate; (iii) PCR using the LightCycler vanA/vanB detection assay of a swab of undifferentiated bacterial growth, if present, from the Enterococcosel and/or CNA subculture plates inoculated from the HI broth; (iv) Multiplex PCR-RFLP testing of DNA extracted from the original perianal swab (see method for multiplex PCR-RFLP above); and/or (v) standard Enterococcosel plate culture of a new (subsequent) perianal swab specimen from the same patient.
For the discordant patient specimens, all the results for microbiology tests that were ordered 120 days prior to the date the VRE study culture was ordered were reviewed. Other cultures from which VRE were isolated were noted.
Specificity panel evaluation for LightCycler PCR assay. The specificity of the LightCycler vanA/vanB detection assay was determined by evaluation of DNA extracted from pure cultures of a variety of gram-positive and gram-negative bacteria (see list of organisms below in Results). These bacteria included many gram-positive enterococci as well as other bacteria and parasites that are considered normal flora, colonizers, or cause of infection in the gastrointestinal tract.
Analytical sensitivity for LightCycler PCR assay. To determine analytical sensitivity, dilutions of the plasmid control were tested in triplicate.
Analysis of data for clinical evaluation. For each of the evaluations using the two different concentrations of vancomycin in the Enterococcosel plate, the results for the LightCycler PCR were compared to the results for Enterococcosel plate culture to determine sensitivities, specificities, and positive and negative predictive values. Therefore, the Enterococcosel plate culture was considered the gold standard.
Statistical analysis. Confidence intervals for sensitivity, specificity, and positive and negative predictive values were based on exact binomial probabilities.
Assessment of assay time requirements. The approximate time required to complete each test, including specimen processing and assay time, was recorded.
|
|
|---|
The following Enterococcus spp. were evaluated for the presence of either the vanA or vanB genes and found to be negative by the LightCycler PCR assay: five isolates of Enterococcus gallinarum containing the vanC-1 gene, four isolates of Enterococcus casseliflavus containing the vanC-2,3 gene, one isolate of Enterococcus faecium, and one isolate of Enterococcus faecalis with the vanC-1 gene. Enterococcal isolates tested which did not contain the van genes were as follows: Enterococcus faecium (2 isolates), Enterococcus faecalis (11 isolates), Enterococcus raffinosus, Enterococcus avium, and E. casseliflavus. As expected strains of enterococci containing the vanC gene did not produce a signal by melt curve analysis with the VRE detection assay.
Analytical sensitivity for LightCycler PCR assay. The analytical sensitivity was determined to be less than 10 targets/µl (50 copies/reaction tube).
Clinical evaluation. (i) LightCycler assay versus Enterococcosel plate containing vancomycin at 6 µg/ml. Among 421 perianal swabs, 11 (3%) were identified as positive by culture and 25 (6%) were identified as positive by LightCycler PCR (vanA [n = 9], vanB [n = 2], and vanB-2/3 [n = 14]). PCR inhibition was determined by lack of recovery template (internal control) amplification for 5 of 421 (1.2%) specimens evaluated using the LightCycler method. The sensitivity, specificity, and positive and negative predictive values for the LightCycler versus culture comparison are shown in Table 2. The mean and median for LightCycler crossing points (the cycle number at which the assay was positive) for concordant results (n = 11) were 28.7 and 26, respectively, and for discordant results (n = 13) were 30 and 30, respectively.
|
View this table: [in a new window] |
TABLE 2. Sensitivities, specificities, and predictive values for LightCycler PCR assay compared to Enterococcosel screening plates for detection of VRE from perianal swabsa
|
The mean and median for LightCycler crossing points for concordant results (n = 9) were 25.3 and 24, respectively, and those for discordant results (n = 16) were 28.7 and 29, respectively.
(iii) Additional testing for discordant results: cultures, conventional PCR (multiplex PCR-RFLP), and LightCycler PCR. Fourteen discordant results (LightCycler positive, culture negative) were noted for specimens screened using the Enterococcosel plate containing vancomycin at 6 µg/ml. Original specimen and/or original Enterococcosel plates were available for all of these cases. Additional testing demonstrated positive results for 7 of 14 (50%) of these cases.
Twenty-one discordant results (LightCycler positive, culture negative) were observed for specimens screened using the Enterococcosel plate containing vancomycin at 8 µg/ml. Original specimens or original Enterococcosel culture plates were available for additional testing for 16 of these 21 cases; additional testing demonstrated positive results for 11 of 16 of these cases. For one additional case for which the original specimen and original Enterococcosel plate were not available, a new specimen was obtained and was found to be positive by Enterococcosel plate culture. In summary, positive results were noted for 12 of 17 (71%) of available discordant cases for which an Enterococcosel plate containing vancomycin at 8 µg/ml was used.
Assessment of assay time requirements. Approximately 2.5 h was required to complete specimen processing (extraction of nucleic acid from the sample), and 1 h was required for analysis and reporting of results (total time requirement = 3.5 h). Due to the automation of extraction (MagNA Pure) and analysis (LightCycler), the actual hands-on time was approximately 20 to 25 min for a full run of 32 samples.
|
|
|---|
Our protocol for evaluating discordant positive results for the LightCycler compared with culture showed that many of the samples that were LightCycler PCR positive but culture negative represented true-positive samples. Several possibilities exist which could explain cases where the LightCycler PCR method was positive but the culture negative. Nonviable enterococci may have been present in specimens which were shed from the small intestine or upper region of the colon. Alternatively, viable enterococci may have been present but in a nonculturable state as recently described by del Mar Lleò and colleagues (3b). Another possibility is that some vanB-containing enterococci may have been inhibited by vancomycin at 8 or 6 µg/ml. Discordant results occurred more frequently with vanB gene detection (with vancomycin at 8 µg/ml, 17 of 20 [58%]; with vancomycin at 6 µg/ml, 12 of 14 [86%]). Finally, recovery of VRE may relate to the amount or type of specimen cultured. Relevant to this last point, D'Agata and colleagues showed that a single screening culture for VRE from rectal swabs, was only 58% sensitive compared to serial quantitative stool cultures and skin cultures (3). Therefore, sampling of enough specimen or different specimen types for our patients may have enhanced the recovery of VRE and further reconciled the number of discordant results.
For some of our discordant samples, additional testing of the original DNA extract using PCR-RFLP was negative. This is expected, as PCR-RFLP, which requires visual inspection of bands on gels, is less sensitive than real-time PCR, which automatically measures fluorescence of hybridized probes. Also, most of the samples tested were from DNA extracts of the original specimen stored for up to 18 months. Degradation of target DNA likely occurred in these samples.
Unfortunately there is no clinical gold standard for assessing the accuracy of tests for detecting VRE in perianal specimen. Because the presence of VRE in perianal swabs indicates colonization and not disease, arbitration of discordant PCR positive versus culture-negative results by medical history review is not possible.
The LightCycler VRE assay can be performed in considerably less time than culture. In our practice, VRE culture results for perianal swabs require a minimum of 72 h. In contrast, LightCycler PCR, including extraction of nucleic acid from the specimen requires 3.5 h. Considering both the enhanced sensitivity and turnaround time for results for LightCycler PCR, use of this assay in place of culture should have significant implications for infection control. Because of the speed and ease of performance of the LightCycler PCR test, it is conceivable that all patients could be screened for VRE before admission to healthcare institutions, including hospitals and nursing homes. This strategy would be similar to the "search and destroy" strategy recently reported as an effective and cost-saving method for preventing nosocomial outbreaks of methicillin-resistant Staphylococcus aureus (MRSA) infections in a Dutch hospital (20). At that healthcare facility, patients who are considered at high risk for MRSA are isolated until nasal cultures are demonstrated to be negative for MRSA. If the same strategy were used for VRE, it is possible that similar results for VRE control and cost savings would be realized. However, if one screens patients before admission with a real-time PCR assay, one should also eliminate the additional costs incurred for quaranting patients (non-VRE carriers) until culture-based susceptibility results are available. Indeed, in a "search and destroy" approach some patients may require isolation for 3 days or longer before results could be confirmed for culture-based methods.
Other studies have shown that the more rapid provision of bacteria identification and antibiotic susceptibility results can be cost saving. Doern and colleagues demonstrated that same-day versus overnight provision of results for bacterial identification and antimicrobial susceptibility to physicians at their institution resulted in statistically significantly fewer laboratory studies ordered per patient and a statistically significant savings per patient hospitalization of
$4,000. Over a year's time this represented a total cost savings of $2,403,162 (4). Barenfanger and colleagues showed in a similarly designed study that provision of more rapid results for bacterial identification and antimicrobial susceptibility decreased length of hospital stay for patients an average of 2.0 days, decreased the mortality rate from 9.6 to 7.9% and resulted in an annual cost savings of $4,189,500 (1).
Other investigators have developed PCR assays for the direct detection of van genes from rectal or perirectal swabs. Satake and colleagues (18) developed a multiplex PCR assay to detect vanA, vanB, vanC-1, and vanC-2 genes. Samples were extracted using two commercially available column technologies in a sequential fashion (QIAamp tissue kit column; QIAGEN Inc., Chatsworth, Calif.; Centerisep gel filtration column, Princeton Separation, Inc., Adelphia, N.J.). PCR and detection were accomplished using a conventional thermocycler and gel electrophoresis, respectively. No vanB-containing enterococci were isolated by culture. Compared to culture the sensitivity and specificity of the vanA assay were 88.5 and 99.6%, respectively. The vanA gene was detected in one sample from which no enterococci were isolated. No internal control (recovery template) was used to assess PCR inhibition. Paule and colleagues (16) showed more impressive results using a multiplex PCR vanA and vanB assay. DNA was extracted from rectal or perianal swabs using a MasterPure DNA purification kit (Epicentre Technologies, Madison, Wis.), amplified using conventional thermocycling and amplified product was detected by direct visualization of 1.5% agarose gels stained with SYBR green I (Molecular Probes, Eugene, Oreg.). This manual PCR assay detected statistically significantly more VRE from either rectal or perianal swabs than the standard Enterococcosel plate containing vancomycin at 6 µg/ml.
Recently, Pallidino and colleagues (13) used a real-time LightCycler PCR assay to detect vanA and vanB genes directly from rectal swabs. DNA extraction was performed using the QIAmp DNA Stool Mini kit (QIAGEN, Inc.) and the assay was developed using hybridization probes. This real-time PCR assay was positive for 45 of 100 specimens; in contrast a positive result was obtained for 43 of 100 specimens using a standard agar plate culture method. This assay was also used to test enrichment broth cultures for these same specimens. For this determination, 88 of 100 specimens were positive by PCR, representing a 95% increase in sensitivity. The PCR inhibition rate for DNA extracts from negative rectal swabs spiked with a vanB containing Enterococcus faecium strain was 55%. It is likely that this amount of inhibition contributed to the relative lack of sensitivity for the direct specimen PCR method versus the enrichment broth PCR method. If one assumes for this study that due to inhibition of PCR
50% of true positives were not detected, then if the inhibition rate is decreased to
1%, twice as many specimens (90 versus 45) specimen would test PCR-positive. This improvement represents a 100% increase in sensitivity which is similar to the results we found for our real-time PCR assay versus the standard culture method using a vancomycin concentration of 6 µg/ml in the Entercoccosel agar plate. We detected slightly over twice as many positives by PCR versus culture using either the culture plate containing vancomycin at 6 or 8 µl/ml. This represented an increase of 109 and 141%, respectively. The inhibitory effects of stool on PCR are well known. In our experience the use of S.T.A.R. buffer and the MagNA Pure instrument significantly decreases inhibition. In fact, in the present study, the inhibition rate was
1%. The use of the MagNA Pure also automates the extraction process. The time required for specimen extraction of 32 samples is approximately 2.5 h, and the assay run time is
1 h. Because both of those processes are automated, actual hands-on time is
20 to 25 min.
The LightCycler VRE detection assay has the potential to be used to identify VRE cultivated on conventional agar plate media or in blood culture bottles. With this approach the identification process for VRE could be shortened by 24 h or more. Elsayed and colleagues (6) recently showed that by using a similar real-time PCR method that MRSA could be detected 24 to 36 h sooner directly from blood culture bottles that were smear-positive for gram-positive cocci; appropriate therapies could therefore be offered 24 to 36 h sooner. We are currently evaluating the ability of this assay for identifying VRE in blood culture bottles that are positive for gram-positive cocci by Gram staining.
In summary, the results of the present study suggest that the Lightcycler VRE detection assay is considerably more sensitive than the standard Enterococcosel culture method for detecting VRE from perianal swabs. The LightCycler method also provides results much faster than culture (
3.5 versus
72 h). The use of this test method could have important implications for the effective control and prevention of nosocomial outbreaks of VRE.
|
|
|---|