Journal of Clinical Microbiology, August 2004, p. 3581-3588, Vol. 42, No. 8
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.8.3581-3588.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
The Johns Hopkins Medical Institutions, Division of Microbiology, Baltimore, Maryland
Received 25 March 2004/ Returned for modification 4 May 2004/ Accepted 11 May 2004
|
|
|---|
13% for both extraction methods. Clinical specimens tested by Monitor were quantified using the CTM48 HCV ASR assay. Characteristics of the regression line included a slope of 0.98 and y intercept of 0.23. Quantification by the two methods was correlated (r = 0.97). CTM48 HCV ASR assay values were on average twofold lower than those for Monitor HCV v. 2.0. These data suggest that our assay combines the characteristics of qualitative and quantitative PCR platforms into a single test. |
|
|---|
Most laboratorians have become familiar with qualitative and quantitative molecular assays for HCV. However, health care practitioners often request these tests inappropriately and fail to interpret results correctly, in part due to an incomplete understanding of the performance characteristics and the utility of different assay formats. The potential for mismanagement of patients exists in this setting.
In response, many laboratories have implemented diagnostic and therapeutic algorithms. In a typical diagnostic algorithm, a qualitative assay is performed to ascertain a diagnosis of chronic HCV infection. If the result is positive, the specimen is then tested by a quantitative assay to determine baseline viral load. In therapeutic algorithms, a quantitative test is performed to assess viral load during treatment. If no HCV RNA is detected, a qualitative assay is performed to assess whether viral clearance has occurred.
Algorithms have been effective for clinicians, but they are cumbersome for laboratories that must also offer individual tests. Laboratories that offer algorithms must implement a system for storage and tracking of specimens and results to ensure that the correct tests are performed in a timely manner. However, the potential for errors exists, even with an optimally designed system.
The development of a single test platform that combines the performance characteristics of qualitative and quantitative assays would greatly simplify HCV diagnostics for health care practitioners and clinical laboratories. The advent of quantitative PCRs in which amplification and detection occurs simultaneously (real-time PCR) offers a potential solution. The broad measurable range and increased sensitivity of real-time PCR make it an ideal platform for the consolidation of HCV molecular assay platforms.
We have investigated the performance of the Roche COBAS TaqMan48 (CTM48) HCV analyte-specific reagent (ASR) for the detection and quantification of HCV RNA in plasma. These reagents utilize TaqMan chemistry to amplify and detect sequences in the 5' untranslated region (UTR) of HCV. The aim of this study was (i) to determine the measurable range, LOD, and precision of quantification of HCV Gt1 after extraction by manual column and automated instrumentation, (ii) to determine whether genotype bias occurred by comparing the LOD and measurable range of quantification of Gt1, -2, and -3 viruses after RNA extraction by automated instrumentation, and (iii) to determine the correlation of quantification by Monitor HCV v. 2.0 and our CTM48 HCV ASR assay using patient specimens.
|
|
|---|
Accuracy of HCV quantification can vary widely due to a number of reasons, including variability in testing proficiency by individual laboratories and the lack of a "gold standard" method for quantification. Therefore, viruses were requantified using the method employed by the manufacturer to verify the stated concentrations (Monitor HCV v. 2.0 for quantifiable calibrators and Teragenix Gt3 virus; COBAS Taqman HCV ASR for Teragenix Gt1 and Gt2 viruses). New values were assigned if the concentration observed after requantification differed from the expected value by a factor of 1.5-fold or greater. Viruses were diluted in defibrinated human plasma (Boston Biomedica, Inc., West Bridgewater, Mass.) for measurable range and LOD experiments. For LOD determinations, viruses were diluted to 100 IU/ml (in 10-fold increments) and then further diluted to 50, 37.5 (for Gt1 automated extraction only), 25, 10, and 5 IU/ml. A sufficient volume of each dilution was prepared to allow extraction of 20 replicates per extraction method.
Plasma samples that had been previously tested by Monitor HCV v. 2.0 were extracted and tested with the CTM48 HCV ASR assay. Specimens with concentrations greater than Monitor's upper limit of quantification (5.7 log10 IU/ml) were diluted 100-fold and retested by Monitor HCV v. 2.0. Samples were not tested simultaneously on both platforms but were frozen at 70°C until CTM48 HCV ASR testing. These studies were performed following guidelines for human subject experimentation. Approval was received from the Johns Hopkins Medicine Institutional Review Board to conduct these experiments.
RNA extraction methods. HCV RNA was extracted by an automated method (MagAttract viral RNA M48 reagents on a BioRobot M48 instrument; QIAGEN, Valencia, Calif.) and a manual method (QIAamp MinElute virus vacuum kit; QIAGEN). Plasma input and elution volumes for the automated method were 300 and 65 µl, respectively, as per the manufacturer's recommendations.
For manual extraction, the manufacturer's protocol was followed with two exceptions: the protease was rehydrated in distilled water instead of the supplied protease resuspension buffer, and carrier RNA was reconstituted in AVE buffer to which quantitative standard reagent (QS) was added. Plasma input and elution volumes were 500 and 60 µl, respectively. For each extraction method, QS was added to each sample such that the final concentration after elution was approximately 1,100 copies in 50 µl of eluate.
QS contains a synthetic virus ("armored RNA") comprised of MS2 bacteriophage bearing RNA sequences complementary to the primers used to amplify HCV and an intervening unique sequence that is used to detect QS during the real-time PCR. QS is used to correct for the presence of reaction inhibitors and other variables that can affect the efficiency of real-time reverse transcription-PCRs. QS was added to the lysis buffer to prevent degradation of the armored RNA.
HCV RNA amplification and detection. The Monitor HCV v. 2.0 test was performed according to the manufacturer's instructions. For the CTM48 HCV assay, working Master Mix was prepared by combining manganese (170 µl) and ASR Master Mix reagents (1.4 ml). Master Mix contains deoxynucleoside triphosphates, oligonucleotide primers (upstream and downstream) containing sequences from the HCV 5' UTR, fluorescent-labeled probes (complementary to 5' UTR and QS-specific sequences), Z05 polymerase (thermostable polymerase with reverse transcriptase activity), and uracil-N-glycosylase (to prevent amplification of contaminants from previous reactions). The working Master Mix (50 µl) was added to each K-tube in the K-Carrier, followed by 50 µl of test sample. The K-Carrier was placed in the CTM48 instrument for reverse transcription, amplification, and detection. The HCV PCR profile used for amplification is shown in Table 1. Results from the six virus preparations used for calibration (1.7 to 6.7 log10 IU/ml in 10-fold increments, four to six replicates per concentration) were imported into ASR external calibration software v. 2.1 to derive the three coefficients that defined the calibration curve for each ASR lot. Data from a minimum of three replicates at each concentration were used to derive the three calibration coefficients. Calibrator data that caused the standard deviation of the QS crossing threshold to exceed 0.5 were excluded from the coefficient calculation. Quantification was performed in subsequent runs by loading the calibration coefficients into the CTM48 instrument software (AmpliLink v. 3.0.1). The calibration coefficients are specific to each ASR lot, therefore the calibration protocol was repeated and new coefficients were calculated for each new lot of ASR reagents.
|
View this table: [in a new window] |
TABLE 1. Instrument settings for CTM48 ASR assaya
|
0.4) were used to define the upper and lower limits of quantification. (ii) LOD. To determine the LOD, viruses were serially diluted to concentrations ranging from 50 to 5 IU/ml. Twenty replicates were tested at each concentration, except for Gt1 at 5 IU/ml extracted by the automated method (n = 10) and Gt3 at 50 IU/ml (n = 5). The LOD was defined as the concentration at which 95% of replicates were detected (7).
(iii) Precision. Precision of the CTM48 ASR assay after automated and manual extraction was assessed using a Gt1 virus at 2.0, 4.0, and 6.0 log10 IU/ml. For intra-assay precision, 20 replicates were tested at each concentration. To calculate interassay precision, 10 additional replicates were tested on two additional runs performed on two separate days, and the data were combined with the initial run of 20 replicates.
(iv) Correlation between quantification using COBAS Monitor v. 2.0 and CTM48 HCV ASR. To assess whether quantification by the real-time PCR ASR and quantitative conventional PCR methods correlated, 106 specimens that had been quantified by COBAS Monitor HCV v. 2.0 were quantified using the CTM48 HCV ASR assay. Specimens throughout the range of quantification were chosen for analysis. Data were logarithmically transformed prior to analysis. The distribution of genotypes was as follows: Gt1, n = 76; Gt2, n = 8; Gt3, n = 6; Gt4, n = 1. Genotypes were determined by direct sequencing of core-E1 sequences (1) or Versant HCV genotyping assay (LiPA; Bayer, Tarrytown, N.Y.). Genotype could not be determined for 15 specimens.
Statistical Methods. Regression lines and their characteristics were determined using Microsoft Excel (Microsoft Office 2001; Microsoft Corp., Redmond, Wash.).
|
|
|---|
6.4 for Gt1 (Fig. 1), 1.8 to
6.8 for Gt2 (Fig. 2), and 1.6 to
6.8 for Gt3 (Fig. 3). Some replicates were not detected at the lowest concentrations of Gt1 (0.8 and 1.1 log10 IU/ml), Gt2 (0.9 and 1.2 log10 IU/ml), and Gt3 (1.3 log10 IU/ml). As a result, these concentrations failed to meet the criterion of acceptable variability used to define the limits of quantification. The lower limit of quantification after manual extraction of Gt1 (1.0 log10 IU/ml) was less than that observed after automated extraction (1.5 log10 IU/ml), and quantification was linear up to the highest concentration of virus tested (6.1 log10 IU/ml; Fig. 4).
![]() View larger version (12K): [in a new window] |
FIG. 1. Measurable range (IU/ml) of Gt1 quantification after automated RNA extraction. The dashed line indicates theoretical trend line of complete agreement between expected and observed measurements.
|
![]() View larger version (11K): [in a new window] |
FIG. 2. Measurable range (IU/ml) of Gt2 quantification after automated RNA extraction. The dashed line indicates theoretical trend line of complete agreement between expected and observed measurements.
|
![]() View larger version (12K): [in a new window] |
FIG. 3. Measurable range (IU/ml) of Gt3 quantification after automated RNA extraction. The dashed line indicates theoretical trend line of complete agreement between expected and observed measurements.
|
![]() View larger version (11K): [in a new window] |
FIG. 4. Measurable range (IU/ml) of Gt1 quantification after manual extraction. The dashed line indicates theoretical trend line of complete agreement between expected and observed measurements.
|
|
View this table: [in a new window] |
TABLE 2. LODs of CTM48 HCV ASR
|
|
View this table: [in a new window] |
TABLE 3. Precision of CTM48 HCV ASR
|
![]() View larger version (17K): [in a new window] |
FIG. 5. Correlation between quantification with CTM48 HCV ASR assay and Monitor HCV v. 2.0. Solid arrow, lower limit of quantification for Monitor HCV v. 2.0; dashed arrow, lower limit of quantification for CTM48 HCV ASR. The dashed line indicates the theoretical trend line of complete agreement in quantification between the two assays.
|
![]() View larger version (11K): [in a new window] |
FIG. 6. Difference in quantification between CTM48 HCV ASR and Monitor HCV v. 2.0 (CTM48 HCV ASR log10 IU/ml minus Monitor HCV v. 2.0 log10 IU/ml) throughout the range of quantification. Mean, average of CTM48 HCV ASR and Monitor HCV v. 2.0 log10 values; SD, standard deviation (0.37).
|
|
|
|---|
User-defined calibrators are an alternative to commercial reagents. These are typically residual plasma (typically 2 to 3 ml) or a large-volume specimen from a patient with a high viral load. Each of these two approaches has disadvantages. Namely, new calibrators must be extensively recharacterized when residual plasma volumes have been depleted and consent from an institutional human subject review board may be required in order to perform large-volume phlebotomy.
The use of external calibrators raises additional issues. One question is whether it can be assumed that assays performed after calibration meet the same performance standard achieved during calibration. This may not be true, and laboratories should consider quality assurance measures such as tracking crossing thresholds (or elbow values, as designated for the CTM48 HCV ASR assay) of controls and analyzing individual controls within a run.
A second question is whether quantification is accurate at concentrations greater and less than the assay calibrators. At the low concentrations, our data suggest that quantification is accurate at concentrations two- to fivefold less than the lowest calibrator (50 IU/ml). We could not assess whether the upper limit of quantification exceeded the concentration of the highest calibrator because the concentrations of virus reagents were equivalent to calibrators. We observed high viral loads in patient specimens (see Fig. 5, with the caveat that these may not be accurate measurements), suggesting the need for accurate measurement in this range.
Our data demonstrate that our CTM48 ASR assay has a measurable range of at least 5 orders of magnitude, at least 10-fold greater than that of Monitor HCV v. 2.0 (4). An accurate determination of the upper limit of quantification will require the use of calibrators with viral loads greater than 7.0 log10 IU/ml. Users will need to define such reagents since they are not currently available commercially.
The potential for genotype-dependent bias in quantification exists, as demonstrated by the decreased efficiency of quantification of non-1 genotypes observed with the first version of HCV Monitor (6). This bias was corrected in Monitor v. 2.0 (4), and we did not observe any evidence of its occurrence in our assay with the CTM48 ASR reagents. The measurable range also appeared to be independent of extraction method, since linearity observed with manual and automated RNA preparation was largely similar.
QS is provided with the CTM48 ASR. In this assay, the QS is used to assess amplification inhibition in individual specimens. We found that addition of QS to the lysis buffer during extraction was critical for assay performance. Although the QS is an armored RNA product, it may be subject to nuclease attack in native specimens.
The LODs of our CTM48 HCV ASR assay were lower than those for Monitor HCV v. 2.0 (600 IU/ml) and comparable to those of the conventional qualitative PCR platform COBAS Amplicor HCV v. 2.0 (4). No significant genotype bias was observed after automated HCV RNA extraction. The LOD after manual extraction with the MinElute vacuum manifold protocol was lower than after automated extraction with the BioRobot M48. One explanation for this might be the greater effective concentration of plasma after manual versus automated extraction (a 1.8-fold increase in concentration after manual compared to automated extraction). It is also possible that the inclusion of protease K in the lysis buffer of the MinElute reagents (not included in MagAttract viral RNA reagent) improves the efficiency of HCV RNA extraction.
Precision throughout the range of quantification was within an acceptable range for a real-time PCR method. Variability in quantification was greatest at low concentrations as expected for a PCR-based amplification method. We expected that precision using an automated extraction instrument would be better than a manual method, but our data demonstrate that reproducibility was comparable between the two methods.
The major benefit of manual extraction using the MinElute reagents and vacuum manifold setup was an apparent increase in sensitivity and a decrease in the lower limit of quantification of our CTM48 HCV ASR assay. Automated extraction could be implemented for routine testing, given the comparable performance characteristics and obvious labor savings. However, extraction with MinElute could be incorporated selectively, to provide an ultrasensitive assay for HCV RNA detection.
We chose to determine the correlation between Monitor HCV v. 2.0 and CTM48 HCV ASR using the automated RNA extraction method because we planned to use the latter assay to replace Monitor in routine diagnostic testing. We characterized the correlation throughout the range of quantification because many clinicians are now requesting viral load measurement (rather than testing with a qualitative HCV RNA detection assay) for patients receiving therapy. Consequently, viral loads in the range of 2.0 to 4.0 log10 IU/ml are now commonly observed. Our data demonstrated excellent correlation throughout the range of concentrations tested.
Interestingly, CTM48 HCV ASR yielded values (
0.2 log10 IU/ml) for Gt1 lower than those with Monitor HCV v. 2.0, as demonstrated by regression analysis results of Gt1 measurable range and quantification correlation of clinical specimens (primarily Gt1). Similar bias for Gt2 and -3 was not observed in measurable range experiments. The existence of bias in quantification of Gt2 and -3 in clinical samples could not be determined because of the small number of non-1 genotype specimens.
The potential sources of variation that could account for quantification differences between the two assays include extraction method and assay format (real-time PCR with external calibrator versus conventional PCR with internal QS). It was recently noted that HCV Monitor v. 2.0 overestimated an international (Gt1) standard concentration by approximately 0.5 log10 IU/ml (8). The authors hypothesized that the discrepancy was due to a potential flaw in QS conversion from copies to international units per milliliter, lending support to the possibility that assay format might explain quantification differences between the two tests. If Monitor v. 2.0 truly overestimates HCV Gt1 viral load, our data suggest that quantification by CTM48 HCV ASR may be more accurate.
The advantages of the CTM48 HCV ASR assay include increased throughput (48 positions per CTM versus 24 positions per COBAS instrument) and decreased turnaround time (
3 versus 6 h) compared to COBAS Monitor HCV v. 2.0. In addition, the incorporation of automated RNA extraction should result in substantial labor savings due to decreased hands-on time.
The disadvantages of our CTM48 HCV ASR assay include calibrators, the need to provide our own assay controls (manufacturers cannot provide positive and negative controls for ASRs, in accordance with United States Food and Drug Administration regulation 21 CFR 809.10[e][1][x]), and the need to perform extensive in-house verification. Another potential disadvantage is the need to perform additional quality assurance steps (on each run, as described above). One comparatively smaller concern is that the volume of extracted material is sufficient for only a single round of testing.
Our data suggest that our CTM48 HCV ASR assay has the combined performance characteristics of qualitative and quantitative platforms. Laboratories that wish to consolidate testing into a single platform must weigh the benefits of streamlining with the added burden of providing quality control and rigorously performing quality assurance.
ASR materials were provided by Roche Diagnostics Corporation.
|
|
|---|
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»