Journal of Clinical Microbiology, September 2004, p. 4307-4309, Vol. 42, No. 9
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.9.4307-4309.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
Laboratoire de Bactériologie-Hygiène, Centre Hospitalier Régional Universitaire de Lille, Lille,1 Laboratoire de Biologie Polyvalente, Hôpital de Roubaix, Roubaix, France2
Received 24 February 2004/ Returned for modification 18 March 2004/ Accepted 4 May 2004
|
|
|---|
|
|
|---|
After decontamination with N-acetyl-L-cysteine-2% NaOH (7) and centrifugation (3,000 x g for 20 min at 4°C), the sediments of the pulmonary specimens were stained with auramine-rhodamine fluorochrome for fluorescent microscopy and inoculated into both a Löwenstein-Jensen medium and an antibiotic-supplemented BacT/Alert MP culture bottle (bioMérieux, Marcy l'Etoile, France). Nonpulmonary specimens from normally sterile sites were cultured without prior decontamination. After inoculation, the remainder of the specimen was aliquoted into two portions (one for AMTDII and the other for the real-time PCR) and frozen until the amplification. NAA tests were applied in parallel, on the frozen aliquots from the same specimen, the same day, in a blind manner by one laboratory technician, independently of the smear and the culture results.
The AMTDII was carried out according to the manufacturer's instructions. Briefly, 450 µl of specimens was placed in lysing solution, and then 25 µl of lysate was used for amplification. All specimens were tested in duplicate. One of them was spiked with a known quantity of M. tuberculosis to detect inhibiting substances. All AMTDII series included one negative and one positive amplification control. The run was validated when the negative and positive controls were <20,000 and >1,000,000 relative light units (RLUs), respectively. For real-time PCR assay, whenever possible, approximately 500 µl of sample (minimal volume of 100 µl) was used for DNA extraction with the MagNA Pure LC DNA isolation kit III (Roche Molecular Biochemicals, Mannheim, Germany) as recommended by the manufacturer. The primers and the fluorogenic probe for the IS6110 sequence (GenBank accession no. X52471) were designed with the Primer Express software, version 2.0 (Applied Biosystems), and were obtained from Applied Biosystems (Warrington, United Kingdom). The nucleotide sequences of the forward and the reverse primers were 5'-CCGAGGCAGGCATCCA-3' (positions 1062 to 1077) and 5'-GATCGTCTCGGCTAGTGCATT-3' (positions 1112 to 1132). The sequence of the minor groove binder probe was 6-carboxyfluorescein-5'-TCGGAAGCTCCTATGAC-3' (positions 1095 to 1111). PCR amplifications were performed in 25-µl reaction volumes including 1x TaqMan Universal Master Mix (Applied Biosystems) containing dUTP and uracil-N-glycosilase, each primer at a final concentration of 100 nM, the fluorescent-labeled probe at a final concentration of 300 nM, and 5 µl of extracted DNA. Each run included positive and negative controls, the former being prepared from 102, 104, and 106 dilutions of 45-µg/ml DNA from M. tuberculosis strain H37Rv, and the latter consisted of sterile water. Quality control was acceptable when the negative control had an undetectable cycle threshold (CT) and the 102, 104, and 106 dilutions of H37Rv DNA had CT values between 17 and 19, 25 and 27, and 33 and 35, respectively. During the test period, the 102, 104, and 106 dilutions of H37Rv DNA had CT values of 18.6 ± 0.6, 25.7 ± 0.25, and 34.3 ± 0.21, respectively. Each sample was tested in duplicate, and an exogenous internal positive control (Applied Biosystems) was added to the reaction chamber in each duplicate and coamplified for detecting endogenous PCR inhibitors. Standard amplification parameters were as follows: 50°C for 2 min and 95°C for 10 min, followed by 45 cycles of 95°C for 15 s and 60°C for 1 min. Real-time data were analyzed with SDS software, version 1.0.
Clinical and radiological data were collected from 93 patients by review of their clinical records. A case of TB was defined as a patient with either specimens positive for MTB in culture or specimens negative for MTB in culture if the following criteria were present: (i) the specimen was isolated from a patient treated by antituberculous drugs, (ii) the specimen was isolated from a patient whose other specimens were culture positive, and (iii) the patient's clinical history provided evidence of TB sufficient to initiate antituberculous therapy.
Among the 100 specimens analyzed, 45 and 55 were isolated from 40 TB and 53 non-TB patients, respectively. Out of the 45 clinical specimens isolated from TB patients, 21 and 24 were pulmonary and nonpulmonary specimens, respectively. The pulmonary specimens consisted of sputum (n = 17), bronchial aspirate (n = 3), and bronchoalveolar lavage fluid (n = 1). The nonpulmonary specimens included lymph node (n = 8), tissues (n = 4), urine (n = 3), synovial fluid (n = 3), abscess (n = 3), and pleural fluid (n = 3). Amplification results with smears and cultures data are summarized in Table 1. Among the 21 pulmonary specimens isolated from TB patients, 16 (76%) and 19 (90%) were AMTDII and real-time PCR positive, respectively. The number of positive results was lower among the 24 nonrespiratory specimens with both methods. Indeed, 14 (58%) were found positive with AMTDII and 19 (79%) with real-time PCR. Nine (9%) specimens (four lymph nodes, three sterile fluids, one abscess, and one bronchial aspirate), of which five were from four different TB patients, could not be interpreted with AMTDII due to the presence of inhibitors of amplification. Complete inhibition of the AMTDII with inhibiting substances was previously reported, particularly in nonpulmonary specimens (3). Therefore, it was suggested that all specimens be decontaminated, even those from sterile sites, to reduce proteins and enzymes which likely inhibit the amplification reactions. However, the sample manipulations during decontamination may cause sampling errors such as (i) inappropriate specimen dilution, (ii) accidental aspiration of the pellet when removing supernatant, and (iii) cross-contamination. In addition, decontamination methods have been reported to cause a large reduction (80%) of CFU recovered in cultures (13). This may increase the likelihood of culture-negative results among paucibacillary specimens. In contrast, no inhibitor was detected with the real-time PCR method. Likely, the method of DNA extraction used for real-time PCR as well as the smaller volume amplified (5 versus 25 µl) may reduce inhibitors better than AMTDII. Finally, out the 40 TB patients, 29 (73%) and 37 (93%) were conclusively detected as positive with AMTDII and real-time PCR, respectively. Among the 55 non-MTB specimens, two cerebrospinal fluids and one bronchial aspirate yielded equivocal results (RLUs ranging from 200,000 to 500,000) with AMTDII. Some authors have stated that these samples should be retested (6, 12). Unfortunately, no residual specimens were available for retesting, and AMTDII could not be done on additional specimens because patients were discharged from the hospital with favorable clinical outcome in the absence of antituberculous treatment. In contrast, no equivocal results were observed with the real-time PCR method. Although the specimen processing for smear decontamination and culture was done in a separated area with distinct equipment for NAA testing, it is possible that the equivocal results were due to cross-contamination from a positive specimen in an adjacent test position. Indeed, the AMTDII required multiple user-dependent steps for manipulations that have the potential for cross-contamination. In contrast, the sample preparation for real-time PCR is automatically performed on the MagNA Pure and the SDS 7000 systems, thus limiting the carryover contamination.
|
View this table: [in a new window] |
TABLE 1. Comparison of AMTDII and real-time PCR for detection of TB in clinical specimens
|
2 test). As expected, a reduced sensitivity was seen among smear-negative specimens with both methods. However, the reduced sensitivity appeared to be less marked with real-time PCR (67% with real-time PCR versus 50% with AMTDII), but it was also not statistically significant. A low sensitivity of NAA testing on the smear-negative specimens is common, especially because few bacilli are present in these type of specimens (4, 5, 8, 9). |
View this table: [in a new window] |
TABLE 2. Sensitivity, specificity, and negative and positive predictive values of AMTDII and real-time PCRa
|
|
|
|---|