Journal of Clinical Microbiology, January 2005, p. 356-362, Vol. 43, No. 1
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.1.356-362.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Department of Virology, Danish Institute for Food and Veterinary Research, Lindholm, Kalvehave,1 The Royal Veterinary and Agricultural University, Laboratory of Virology, Frederiksberg C,2 Novo Nordisk A/S, Applied Trinomics, Måløv, Denmark,4 Centro de Investigación en Sanidad Animal (CISA-INIA), Valdeolmos, Madrid, Spain3
Received 22 April 2004/ Returned for modification 17 June 2004/ Accepted 3 September 2004
|
|
|---|
|
|
|---|
The appearance of typical clinical signs of vesicular disease in domestic livestock should always cause a suspicion of FMD. When clinical samples are negative for FMD virus (FMDV), rapid laboratory tests that reliably identify VSV are of urgent need. The diagnostic tools presently available for the diagnosis of VSV infection include virus isolation (by inoculation of cell cultures) and viral antigen detection (enzyme-linked immunosorbent assay and complement fixation and neutralization tests) (9). Detection of VSV by amplification of viral RNA by reverse transcription-PCR (RT-PCR) has been reported previously (4, 7, 8, 14). These assays are exclusively based on gel-based identification of PCR products and are designed to detect NJ (14), Ind-1 and/or NJ by a heminested assay (4), and Ind-1 and NJ by a multiplex assay (7, 8).
In this study, we developed a quantitative RT-PCR assay for the detection and differentiation of the two main VSV serotypes based on the novel primer-probe energy transfer (PriProET) detection system (5, 12). This fluorescence resonance energy transfer detection system combines probe-based real-time monitoring of PCR amplification with confirmation of probe hybridization from the melting temperature (Tm) curve. The PriProET system uses two primers, one of which is labeled with a donor fluorophore and a fluorescent probe labeled with a reporter fluorophore. The primers are extended during each amplification cycle. The fluorescent probe anneals to the extended reverse primer, which enables energy transfer from the donor to the reporter fluorophore. This allows quantification of the specific amplicon, as the fluorescence emitted from the reporter depends directly on the amount of amplicon formed. A region of the gene encoding the RNA-dependent RNA polymerase (the L gene) was amplified in the presence of two serotype-specific fluorescent probes. Amplification was primed with VSV-specific primers, one of which was labeled with the donor fluorophore 6-carboxyfluorescein (FAM). The two serotype-specific fluorescent probes were labeled with a reporter fluorophore, Texas Red (a serotype Ind-1-specific probe) or cyanine 5 (Cy5; a serotype NJ-specific probe). Generation of amplicons specific for either Ind-1 or NJ caused a fluorescence energy transfer between the fluorophores, and the fluorescent signal was monitored. The specificity of the reaction was further confirmed by the Tm profiles of the fluorescent probes. This quantitative RT-PCR assay was used to determine the amount of VSV RNA in clinical samples from experimentally infected pigs.
|
|
|---|
Design of primers and probes. VSV-specific primers were designed to target the most conserved regions of the L gene of the VSV genome. Conserved regions suitable as primer targets were selected from alignments of all available VSV sequences downloaded from the GenBank database. Additionally, the areas of interest of the VSV strains used in this study were sequenced. Totally conserved regions were absent, so regions with the fewest sequence differences were selected. In order to achieve equivalent amplification efficiencies, degenerate purine and pyrimidine bases were incorporated into the primers specific for the cDNA synthesized from the very divergent serotypes (3). The nucleotide analogues inosine, N6-methoxy-2,6-diaminopurine (dK), and 6H,8H-3,4-dihydropyrimido[4,5-c] (1,2)oxazin-7-one (dP) were used as alternatives to traditional degenerated positions; inosine was incorporated in the primer at a position where the sequence could be one of the four nucleotides, dK was incorporated at a position where the sequence could be either A or G, and dP was incorporated at a position where the sequence could be either C or T. For use in the PriProET PCR, the forward primer was covalently modified with a donor fluorophore (FAM) at the 5' end.
Two differently labeled fluorogenic probes that would hybridize to the VSV sequence between the sequences of the VSV-specific primers were designed. The specific fluorogenic probes were included in the assay to allow real-time detection of the 120-bp specific amplicon and simultaneous differentiation between serotypes Ind-1 and NJ. One specific probe (the Ind-1-specific probe), which has a covalently linked fluorescent reporter dye, Texas Red, attached to the 3'end, would hybridize to amplicons derived from serotype Ind-1 with a specific Tm of 67.1°C. Another specific probe (the NJ-specific probe) with a covalently linked fluorescent reporter dye, Cy5, attached to the 3' end would hybridize to the amplicon derived from serotype NJ with a specific Tm of 67.7°C. The probes would hybridize to the amplicons in a sequence-specific manner, depending on the template present in the sample, and generate a fluorescent signal due to the close proximity of the donor to the reporter fluorophore. These signals could be distinguished on the basis of the corresponding emission spectra (Texas Red, 610 nm; Cy5, 660 nm). The sequences of the primers and probes are listed in Table 1.
|
View this table: [in a new window] |
TABLE 1. Oligonucleotide sequences of primers and probes
|
|
|
|---|
![]() View larger version (25K): [in a new window] |
FIG. 1. Nucleotide sequence alignment of a part of the VSV L gene corresponding to nucleotides 7017 to 7131, numbered relative to the sequence with GenBank accession number J02428. The alignment represents all the different L-gene sequences from GenBank as of February 2004. GenBank accession numbers are as indicted for the following Ind-1 strains: Mudd Summers, K0237 and J02428 (16); 98COE, AF483864 (15); 85CLB, AF473865 (15); and 94GUB, AF473866 (15). GenBank accession numbers are as indicted for the following NJ strains: Hazelhurst, AY074803 and AY102918; Hazelhurst_a, M20166 (2); and Ogden, M29788 (1). Additionally, the four different VSV strains used in this study were sequenced, and their sequences were added to the alignment. The nucleotide sequence alignment was used to design the primers and probes. The framed sequences correspond to the target regions for the VSV-specific forward primer (5' labeled with FAM), the VSV-specific reverse primer, and the two serotype-specific probes (the Ind-1-specific probe 3' labeled with Texas Red and the NJ-specific probe 3' labeled with Cy5).
|
Multiplex detection of VSV. The ability of the PriProET assay to detect VSV in a multiplex format was assessed with virus grown in primary swine kidney cell cultures. Serotype Ind-1 or serotype NJ in Eagle's medium was subjected to RT-PCR with the VSV-specific primers that were designed, with the serotype-specific probes, and by the optimized PriProET RT-PCR protocol. The sensitivity, detection limit, and intra-assay variability of the PriProET assay were determined for each target by testing three replicates of 10-fold dilutions of cell culture-grown VSV serotypes Ind-1 and NJ. The standard curves obtained (Fig. 2E and F) displayed a linear relationship between the log amount of viral nucleic acids in the reaction mixtures and the observed CT, with calculated E values of 1.82 for serotype Ind-1 and 1.88 for serotype NJ. The average between-run correlation coefficients were 0.99 for both standard curves. The limits of detection were repeatedly found to be below 10 TCID50s/ml for both serotypes. The intra-assay variability was ±1 CT value, which corresponds to less than 5% variation between replicates. The interassay variability was evaluated by comparison of the observed CT values for the standards included in each experiment. The interassay variability observed was consistently at the same level as the intra-assay variability (data not shown).
![]() View larger version (25K): [in a new window] |
FIG. 2. (A to D) Detection of VSV RNA in cell culture material. Tenfold dilutions of cell culture-grown VSV serotype Ind-1, serotype NJ, and mixtures with each serotype were subjected to RNA extraction, cDNA synthesis, and, subsequently, PriProET PCR. In the mixed dilutions the concentration of Ind-1 was 10 times lower than the concentration of NJ. Each panel (A to D) displays the readout from the multiplex PriProET PCR, with the fluorescent signals from the Ind-specific probe (diamonds; Texas Red; 610 nm) and the NJ-specific probe (crosses; Cy5; 660 nm) indicated. For simplification, the results for only one sample are displayed in each amplification plot. (A) Sample positive for Ind-1 with a CT of 36; (B) sample positive for NJ with a CT of 33; (C) sample positive for Ind-1 with a CT of 36 and positive for NJ with a CT of 33; (D) sample negative for both serotypes. (E and F) Standard curves generated by analysis of 10-fold dilutions of VSV Ind-1 (E) and VSV NJ (F) from cell culture with known numbers of TCID50s per milliliter. Each dilution was analyzed in triplicate. The CT values plotted were obtained from amplification plots similar to those shown in panels A to D. Real-time PCR amplification efficiencies were calculated by the equation E = 101/slope (11). The intra-assay variability was ±1 CT, which corresponds to less than 5% variation between replicates.
|
Confirmation of assay specificity and robustness. The specificities of the reactions shown in Fig. 2 were confirmed by the fluorescent probe Tm profiles (Fig. 3A to D). The serotype Ind-1-positive reactions showed a specific Tm of 67.1°C (Fig. 3A), and the serotype NJ-positive reactions showed a specific Tm of 67.7°C (Fig. 3B). Double-positive reactions showed specific Tm peaks in both the Texas Red and the Cy5 channels (Fig. 3C). Specific Tms were absent from all negative reactions (Fig. 3D). The specificity of the assay was further evaluated by testing the cross-reactivity to viruses causing diseases with clinical symptoms similar to those caused by VSV. Purified RNA obtained from cell culture-propagated high-titer viral preparations of FMDV and swine vesicular disease virus was reverse transcribed and used as the template for the VSV PriProET assay. Positive amplifications and specific Tms were absent from these reactions (data not shown).
![]() View larger version (19K): [in a new window] |
FIG. 3. (A to D) Confirmation of specificity of amplification of VSV RNA. Each panel displays the readout (dF/dT) from the fluorescent probe Tm profile analysis, with the fluorescent signals from the serotype Ind-1-specific probe (solid lines) and the serotype NJ-specific probe (dotted lines) indicated. (A) Sample positive for Ind-1; (B) sample positive for NJ; (C) sample positive for both Ind-1 and NJ; (D) sample negative for both serotypes. (E) Cross-detection of VSV serotypes. The diagram displays the readout (dF/dT) from the fluorescent probe Tm profiles when only one of the two serotype-specific probes was included in the PriProET assay. Observe the larger temperature range in panel E compared to those in panels A to D; the Tm is shifted approximately 25 to 26°C. Serotype Ind-1 was cross-detected in the Tm profile with a Tm of 41.1°C (dotted line) when only the serotype NJ-specific probe was included in the assay. Likewise, serotype NJ was identified in reaction mixtures in which only the serotype Ind-specific probe was included (solid line; Tm = 42.9°C).
|
Detection of acute infections. The ability of the PriProET assay to detect VSV during acute infections was analyzed with experimentally infected pigs. Two groups of three pigs each were inoculated with either VSV serotype Ind-1 or VSV serotype NJ. All pigs had increased rectal temperatures (Fig. 4A and B), and four animals had vesicular lesions. Vesicles were observed in pig 1 at PIDs 7 and 11 and in pig 6 at PID 7. Local vesicular foot lesions at the injection site were observed in pigs 3 and 5 at PID 4. Nasal swab specimens were obtained on PIDs 1 to 4, 7, 9, 11, and 14 and examined by PriProET RT-PCR. VSV nucleic acids were detectable in the swabs at PID 2 for serotype NJ-infected pigs (Fig. 4C) and PID 3 for serotype Ind-1-infected pigs (Fig. 4D). The detection of viral RNA in nasal swab specimens preceded the detection of vesicular lesions. The largest amount of VSV in the nasal swab specimens was observed in pig 1 between PIDs 7 and 11. Similarly, vesicular materials obtained from the diseased pigs were analyzed by the PriProET assay; and selected blood samples and tissues obtained at necropsy, including tonsil and lymph node tissues, were likewise tested for the presence of viral RNA. All the samples obtained from the vesicular lesions contained large amounts of viral RNA. Vesicular materials obtained on PID 4 had CT values of 32 for pig 3 (infected with serotype NJ) and 29 for pig 5 (infected with serotype Ind-1), whereas the CT values for blood obtained from the same pigs were 43 and 47, respectively. The difference in CT values of 11 and 18 corresponds to approximately 103- to 105-fold higher virus contents in vesicles than in blood. Viral RNA was also detected in tonsil and lymph node tissues, but to a lesser extent compared to the amounts detected in vesicular materials.
![]() View larger version (18K): [in a new window] |
FIG. 4. (A and B) Rectal temperatures of the group of pigs infected with VSV serotype NJ (A) and VSV serotype Ind-1 (B). Asterisks mark the occurrences of vesicles. (C and D) Detection of viral RNA from serotypes NJ (C) and Ind-1 (D) in nasal swab samples by the multiplex PriProET assay. The viral loads in nasal swabs are displayed as arbitrary units, defined as 50 minus the observed CT value. All nasal swab specimens from NJ-infected animals were negative for Ind-1 by the multiplex PriProET assay, whereas all nasal swabs from Ind-1-infected animals were negative for NJ.
|
|
|
|---|
This report introduces a new real-time multiplex assay that allows the simultaneous detection and differentiation of viral RNA from the two major VSV serotypes, Ind-1 and NJ. We have previously described the PriProET system (12) and have provided details about the real-time detection system with pan-FMDV-specific primers and a pan-FMDV-specific probe. The cycling conditions for the pan-FMDV assay and the multiplex VSV assay are identical, which allows combined side-by-side detection of FMVD and VSV in the same real-time thermocycler. The PriProET assay for VSV described in this paper is a further refinement of the method for multiplex detection and differentiation. So far only a few diagnostic RT-PCR assays for VSV have been reported (4, 7, 8, 14). These are traditional gel-based assays, and none of these are designed for the quantitative detection of VSV. Two of the assays (7, 8) are multiplex assays that detect both VSV serotypes with a combination of serotype-specific primers. To our knowledge a VSV-specific primer set has not been described before, and the high degree of genetic variability in the VSV genomes complicates the design of such primers. Nevertheless, pan-virus-specific primers with sequences conserved between all known serotypes of a virus or even a family of viruses are very attractive for diagnostic purposes, as they result in very robust assays. Pan-virus-specific primers will, for example, allow the detection of genetic recombinants formed between different viral serotypes, and pan-virus-specific primers will likewise be assumed to detect future emerging serotypes of a given virus (5, 10, 12). Despite the extraordinarily high degree of sequence divergence between the two serotypes of VSV, even when the sequence divergence is compared to those of other RNA viruses, we were able to design a VSV-specific primer set in the present study. The primer set covers the very diverse serotypes Ind-1 and NJ, as demonstrated by the experiments performed in this study. Since these serotypes represent the majority of the viruses causing clinical cases of disease (13), it is likely that the primer set will amplify the majority of unknown VSV strains. Whether the primers also amplify serotype Ind-2 or Ind-3 has not been demonstrated due to a lack of nucleotide sequences for these strains. The uniform amplification of serotypes Ind-1 and NJ with equal sensitivities and efficiencies was made possible by the incorporation of nucleotide analogues (inosine, dK, and dP) in the primers. This is consistent with the findings of Hill et al. (3), who found that oligonucleotides containing up to six residues replaced by a mixture of dK and dP primed DNA synthesis efficiently. From a theoretical point of view, the use of multiple dK, dP, and especially inosine residues in the primer sequence results in a less specific hybridization and, as a result, a less specific diagnostic assay. However, this problem was circumvented in the assay described here, in which the specificity of the PCR amplicon was confirmed not only by hybridization of a specific probe but also by precise determination of the Tm for this hybridization. The end result is a unique combination of high specificity and robustness to sequence divergence. The robustness of the VSV assay developed is, however, not restricted to the use of VSV-specific primers. Use of the unique primer-to-probe energy transfer gives a unique robustness to the detection of sequence diversity in the probe region compared to the robustness of other probe-based real-time technologies, such as the TaqMan and molecular beacon detection systems. This robustness of the system was evaluated by testing the ability of the system to detect VSV types with multiple mutations in the probe sequence-specific region, as shown in Fig. 3E. VSV strains with up to 6 nucleotide differences in the probe-specific region could be detected by the Tm profile, although it could not be quantified. If a specific hybridization with a shifted Tm is observed in the Tm profile, the DNA of the amplicon can rapidly be sequenced for precise typing. In cases in which this DNA sequencing reveals completely new variants, cDNA will still be available for rapid PCR amplification and DNA sequencing of other regions of the viral genome. As such, the PriProET system is a powerful tool not only for the detection of viruses that have high mutation rates or that cover very distant genotypes but also as a tool for surveillance for newly emerging viral variants. In the future, we foresee that the unique sequence flexibility of the PriProET system can be extended to cover even whole virus families.
The sensitivity of the novel quantitative multiplex RT-PCR method was shown to be less than 10 TCID50s/ml and to be similar for both principal VSV serotypes when serial dilutions of serotype Ind-1 and NJ viral isolates were used. Nevertheless, to further validate the assay as a diagnostic tool, a collection of clinical samples from pigs experimentally infected with serotype Ind-1 or NJ was tested and the results were compared with those obtained by a previously described gel-based RT-PCR method (8). The results obtained in the present study reflect the qualitative results obtained with the same clinical materials by the gel-based RT-PCR. The multiplex PriProET method allowed relative quantification of the viral load in the clinical material and permitted the detection of VSV in nasal swab specimens at PIDs 2 and 3 for NJ- and Ind-1-infected pigs, respectively, even before the typical vesicular lesions appeared in the affected animals. Furthermore, large amounts of viral RNA were detected in the samples obtained from vesicular lesions. Therefore, whenever vesicles are present, they should be sampled for early diagnosis. A Tm analysis confirmed the specificity of the fluorescence yield for each of the experimental samples. In comparison to the gel-based method, the use of the ABI 7700 system allowed the simultaneous detection and differentiation of both serotypes in a closed-tube format, which represents a less laborious and much faster method. In addition, as the real-time detection format eliminates the need for postamplification handling of the PCR products, the risk of contamination is minimized, and thus, the risk of false-positive reactions is also minimized.
In summary, the results presented here show that the novel PriProET assay can be a powerful diagnostic tool that gives a definitive result in hours from the time of sample receipt and that it can be applied to different clinical materials. We have demonstrated that the use of the present quantitative RT-PCR method provides a sensitive tool for the simultaneous detection and differentiation of the two major VSV serotypes, Ind-1 and NJ. The merging of a fast and robust means of detection, identification, and quantification makes this method a very useful tool for diagnostic laboratories.
This study was partly funded by the European Union (Project QLK2-CT-2000-00486).
|
|
|---|