Journal of Clinical Microbiology, October 2005, p. 5288-5290, Vol. 43, No. 10
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.10.5288-5290.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Clinical Microbiology Service, Department of Pathology, New York Presbyterian Hospital, Columbia University Medical Center, New York, New York,1 Bronx Community College of the City University of New York, Bronx, New York2
Received 23 February 2005/ Returned for modification 12 April 2005/ Accepted 9 June 2005
|
|
|---|
|
|
|---|
This study evaluates the Cellenium 160US system (TREK Diagnostic Systems, Cleveland, Ohio), a urine screening system that applies selective molecular fluorescence staining to robot-based computerized microscopic video image analysis to obtain rapid results within 1 to 2 h. The threshold for detection of microorganisms is
104 organisms/ml urine, and it predicts the Gram reaction and morphology. We evaluated the performance of the Cellenium system in comparison to that of semiquantitative plate culture for the detection of clinically significant bacteriuria.
A prospective study of 1,199 randomly collected urine specimens (voided and catheterized from adult patients) received by the microbiology laboratory of Columbia University Medical Center, was conducted. The specimens were collected in a transport device (BD Vacutainer, Franklin Lakes, NJ) and processed within 1 h of receipt. The urine specimens were inoculated with a 1-µl calibrated loop onto Columbia agar-based agar (5% sheep blood) and MacConkey agar biplates (Becton Dickinson Microbiology Systems, Sparks, MD). The plates were incubated (24 h, 35°C) and examined for significant bacteriuria (
104 CFU/ml of one or two potentially pathogenic microorganisms). When three or more colony morphologies were detected, mixed commensal flora (MCF), defined as viridans group streptococci, coagulase-negative staphylococci other than Staphylococcus saprophyticus, lactobacilli, corynebacteria, and Gardnerella vaginalis, was reported. Bacterial identification was performed by using the MicroScan system (Dade Behring Inc., Deerfield, IL).
Urine specimens were placed into the Cellenium system in batches of 24, according to the manufacturer's instructions. Using robotics, the instrument dispensed aliquots of the samples into a cassette and mixed them with highly specific fluorescent probes that bind to bacteria, yeast, and other cells. The stained cells were collected through a filtering membrane to produce a monolayer which was illuminated with light of selected wavelengths to excite the fluorescent dyes. The computerized image analysis system employed homogeneous fluorescent beads (1-µm latex spheres) as calibration particles and, using proprietary bacterial recognition algorithms, processed the digital image. A threshold colony count of
104 organisms/ml urine was reported as a positive result by the instrument within 1 to 2 h.
When the culture and Cellenium system data were discordant, both analysis methods were repeated within 24 h by using the original urine specimens, which had been stored at 5°C. In addition, patient charts were reviewed for urine microscopy and urinalysis results.
Of the 1,199 urine specimens studied, 479 (39.9%) were culture positive and 720 (60.1%) were culture negative. A single bacterial pathogen was recovered from 172 (14.3%) specimens, two pathogens were recovered from 18 (1.5%) specimens, and 289 (24%) specimens had mixed pathogens or grew nonpathogens. Of the culture-positive specimens, 154 (32.2%) were determined to be clinically significant (Table 1).
|
View this table: [in a new window] |
TABLE 1. Clinically significant urine culture results
|
Among 44 clinically significant gram-positive bacteria, 14 (8 E. faecalis isolates and 6 group B streptococci) were not detected by the Cellenium system, resulting in a false-negative rate of 2.2% for gram-positive cocci (GPC). Chart review showed that 11 of 14 specimens were positive by urine microscopy or urinalysis. To further investigate the analytical sensitivity for GPC, pooled filtered urine specimens were seeded with serial 10-fold dilutions at 103 to 106 CFU/ml each of E. faecalis and group B streptococci. Cellenium system screens and cultures were performed. E. faecalis and group B streptococci were detected only at
105 CFU/ml, confirming the limitation of the Cellenium system for the detection of these gram-positive bacteria at 104 CFU/ml. Upon a repeat of the Cellenium system screen, 5 of the 14 samples with false-negative results tested positive, suggesting sampling error.
The Gram and morphology classifications (as defined by the Cellenium system) of the 53 culture-negative and Cellenium system-positive urine specimens demonstrated the following: cells (n = 13); gram-negative rods (GNRs; n = 16); GPC (n = 7); GNRs and gram-positive rods (GPRs; n = 4); GPC and GPRs (n = 3); GNRs and yeasts (n = 3); GNRs and GPC (n = 3); GPC, GNRs, and GPRs (n = 2); GPRs and yeasts (n = 1); and GPRs, yeasts, and GNRs (n = 1). Urinalysis and urine microscopy results were consistent with infection for 36 (68%) specimens reported as "false positive" by the Cellenium system. These specimens may have contained nonviable organisms due to prior antimicrobial therapy, anaerobes, fastidious organisms (e.g., Chlamydia or Ureaplasma), or preservative-sensitive species. False-negative culture results may also occur if the urine specimen is not mixed thoroughly prior to plating.
By using culture as the reference, the overall sensitivity (SN), specificity (SP), negative predictive value (NPV), and positive predictive value (PPV) for Cellenium system results were 89.5%, 94.4%, 97.1%, and 81.0%, respectively (Table 2). The Cellenium system detected gram-negative rods at 96.9% SN, 96.4% SP, 99.4% NPV, and 84.4% PPV (Table 3) and gram-positive cocci at 68.1% SN, 97.9% SP, 97.8% NPV, and 69.8% PPV (Table 4).
|
View this table: [in a new window] |
TABLE 2. Routine culture versus Cellenium system results (all clinically significant organisms)
|
|
View this table: [in a new window] |
TABLE 3. Routine culture versus Cellenium system results (gram-negative organisms)
|
|
View this table: [in a new window] |
TABLE 4. Routine culture versus Cellenium system results (gram-positive organisms)
|
In summary, the results of this study indicate that the Cellenium urine screen offers many features favorable for incorporation into a routine clinical microbiology laboratory, i.e., its high-throughput capacity, its ability to distinguish MCF, and its high negative predictive value. Overall, the Cellenium system offers an NPV of 97.1% and is comparable to the previously described Sysmex UF-100 analyzer (NPV, 95%) (7) and the Coral UTI screen system (NPV, 95.9%) (8). Although the sensitivity of detection of gram-negative uropathogens was reliable, the Cellenium system failed to detect some clinically significant gram-positive uropathogens. The setting of a lower threshold for detection at 103 CFU/ml may increase the sensitivity of detection for GPC. Additionally, detection of low-count bacteriuria (102 to 104 CFU/ml) has been reported to be of clinical significance in sexually active women (2, 3).
While automated systems aim at improving uropathogen detection (1, 5) and incorporating algorithms that allow the reporting of results within 1 to 2 h following receipt of the specimen in the laboratory and to minimize the unnecessary use of antibiotics, these cannot be universally recommended. The Cellenium 160US system has recently been voluntarily withdrawn from the market by the manufacturer. To increase marketing success, the system should have a smaller footprint and have the capacity to distinguish between viable and nonviable bacteria to avoid false-positive results, among the other caveats discussed here. Until a reliable system becomes available, culture still remains a "gold standard" for confirmation of a urinary tract infection.
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||