Previous Article | Next Article ![]()
Journal of Clinical Microbiology, October 2005, p. 5366-5368, Vol. 43, No. 10
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.10.5366-5368.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Centre for Infectious Diseases and Microbiology & NHMRC Clinical Centre of Research Excellence in Infectious Diseases, Hematology, and Health Ethics, University of Sydney Western Clinical School,1 Westmead Hospital,2 Centre for Infectious Diseases and Microbiology Laboratory Services, Westmead Hospital, Westmead, New South Wales 2145, Australia3
Received 17 May 2005/ Returned for modification 2 June 2005/ Accepted 6 July 2005
| ABSTRACT |
|---|
|
|
|---|
| TEXT |
|---|
|
|
|---|
In this study, we developed a nested Aspergillus-specific PCR assay into a nested qualitative real-time PCR assay for use with a LightCycler (LC) system (Roche Diagnostics, Australia). The first-round PCR is performed on a block-based PCR machine, while the second-round amplification and the product detection are carried out simultaneously using an Aspergillus genus-specific TaqMan probe and an LC instrument. We compared the performance of the assay with that of our routine nested PCR assay by using the same clinical specimens from a random population of immunosuppressed hematology patients at high risk for IA.
Probe design.
The Aspergillus-specific probe was designed using Primer Express software, version 1.0 (Applied Biosystems, Foster City, CA), and Oligo Explorer software, version 1.1.0 (T. Kuulasma, University of Kuopio, Kuopio, Finland [http://www.uku.fi/
kuulasma/OligoSoftware]). The probe was designed within a conserved region of the 18S rRNA genes (GenBank database) of A. fumigatus (AB008401), A. flavus (AF18SRR), A. nidulans (AB008403), A. niger (AN18SRR), A. ochraceus (AB008405), A. terreus (AB008409), A. ustus (AB002072), and A. versicolor (AB008411). This region (5'-TTG TTA AAC CCT GTC GTG CTG GGG ATA GAG-3') was not homologous to the 18S rRNA genes of other sequenced pathogenic fungi or mammalian DNA. The fluorogenic probe was labeled at the 5' end with FAM (6-carboxyfluorescein) and at the 3' end with TAMRA (6-carboxy-tetramethyl-rhodamine) and synthesized by Applied Biosystems, Foster City, CA.
DNA extraction. A GenElute Mammalian DNA kit (Sigma-Aldrich Co., St. Louis, MO) was used for all DNA extractions. Tissue specimens were extracted according to the manufacturer's instructions. Blood samples (500 µl) were lysed with 3 volumes of erythrocyte lysis buffer (0.155 M NH4Cl, 0.01 M NH4HCO3, and 0.1 mM EDTA [pH 7.4]) (10) for 10 min at 20°C and centrifuged at 6,600 rpm for 10 min. Cerebrospinal fluid, ascitic fluid, and bronchoalveolar lavage (BAL) specimens were centrifuged at 13,000 rpm for 10 min. Supernatants were discarded and the pellets resuspended in 200 µl of sorbitol buffer (1 M sorbitol, 100 mM EDTA, and 0.1% 2-mercaptoethanol) (13) and 200 U of lyticase (Sigma-Aldrich Co.). After incubation at 37°C for 60 min, spheroplasts were precipitated by centrifugation at 7,600 rpm for 5 min, resuspended in 180 µl of lysis solution T (GenElute Mammalian DNA kit) and 20 µl of proteinase K (Sigma-Aldrich Co.), and incubated at 55°C for 60 min. DNA was extracted according to the manufacturer's instructions with a final elution volume of 120 µl.
First-round PCR. Oligonucleotide primer pairs (AFU 7S, AFU 7AS; AFU 5S, AFU 5AS) (10) were used in the first- and second-round PCR amplification reactions, respectively. Reactions were performed in a final volume of 25 µl. The first-round PCR mix contained 1x PCR buffer (10 mM Tris-HCl [pH 8.3], 50 mM KCl, 1.5 mM MgCl2, 0.001% gelatin) (Applied Biosystems), 1 mM MgAc, 200 µM of the deoxynucleoside triphosphates (dNTPs) (Roche Diagnostics), 1.2 µM concentrations of primers AFU 7S (5'-CGGCCCTTAAATAGCCCG) and AFU 7AS (5'-GACCGGGTTTGACCAACTTT), 1.25 U Taq DNA polymerase (Applied Biosystems), and 10 µl of DNA. DNA amplification was performed on an Eppendorf Mastercycler gradient thermocycler (Eppendorf AG, Germany) by incubation at 94°C for 2 min, followed by 30 amplification cycles of 94°C for 20 s, 65°C for 30 s, and 72°C for 30 s, and a final extension at 72°C for 5 min.
Second-round PCRs. Conventional assay. The reaction mix contained 1x TTH buffer (10 mM Tris-HCl [pH 8.3], 50 mM KCl, 1.5 mM MgCl2, 0.01% Tween 20, 0.01% gelatin, 0.01% NP-40), 3 mM MgAc, 200 µM of dNTPs, 0.4 µM of primers AFU 5S (5'-AGGGCCAGCGGTACATCACCTTG) and AFU 5AS (5'-GGGRGTCGTTGCCAACYCYCCTGA), 0.5 U DyNAzyme DNA polymerase (Finnzymes, Finland), and 2 µl of the first-round PCR product. The amplification conditions were 96°C for 2 min, followed by 30 cycles of 96°C for 20 s, 65°C for 30 s, and 72°C for 30 s, and a final extension at 72°C for 5 min. The PCR products were separated by 1.5% agarose gel electrophoresis, stained with ethidium bromide, and visualized by UV transillumination. A negative control of molecular biology-grade water and a control to exclude the presence of inhibitory substances were included for each specimen. The PCR assay and/or DNA extraction was repeated if any controls failed. To minimize contamination, extractions and first- and second-round PCRs were performed in three separate rooms. Gels were run in a fourth room, with a unidirectional work flow between all rooms.
Real-time assay.
The second-round real-time PCRs were performed in glass capillaries (Roche Diagnostics) in a 20-µl final volume and contained 1x LC FastStart reaction mix (containing FastStart Taq DNA polymerase, reaction buffer, dNTPs, and 1 mM MgCl2) (Roche Diagnostics), 3.5 mM MgCl2, 0.5 µM of each primer, 0.1 µM Aspergillus-specific probe, and 2 µl of the first-round PCR product. Cycling parameters on the LC machine were 95°C for 10 min, followed by 35 cycles of 95°C for 0 s, 65°C for 5 s, and 72°C for 15 s, and a final cooling at 40°C for 30 s. Fluorescence data calculated by Light Cycler were collected at the end of each 65°C annealing step with a single fluorescence acquisition for each capillary. A sample was designated positive for Aspergillus species DNA when there was an exponential increase in fluorescence during the first 30 cycles of PCR amplification (crossing point,
30). Samples with crossing points of >30 cycles were repeated to verify the results. The sensitivity of the nested real-time assay was determined by extracting DNA from blood seeded with serial dilutions of either A. fumigatus, A. flavus, A. nidulans, or A. terreus conidia (104 to 100 CFU/ml).
The Aspergillus-specific probe hybridized to DNA extracted from all four Aspergillus spp. (Fig. 1). The sensitivities of both the real-time and conventional PCR assays were equivalent to 10 CFU of conidia/ml of blood for all Aspergillus spp. (shown for the nested real-time assay in Fig. 2).
|
|
The conventional nested PCR assay required up to 5 h 30 min for the first- and second-round PCR amplifications, agarose gel electrophoresis, ethidium bromide staining, and result analysis. The nested real-time PCR assay was less labor-intensive. Furthermore, the first- and second-round PCR amplifications were completed and results were available in less than 3 h.
The aim of this study was to transfer a nested Aspergillus-specific PCR assay onto the LC platform to realize the potential benefits of a real-time PCR platform while retaining the sensitivity of the nested method. To date, the majority of published real-time PCR assays are specific for A. fumigatus, yet in Australia, 10 to 15% of IA cases are due to other Aspergillus species (Australian Mycoses Interest Group, unpublished data). These species will not be detected by A. fumigatus-specific assays, as was demonstrated recently in a case of IA due to A. terreus (8). Real-time assays are less sensitive than nested PCR assays, including in our hands (see above and references 1, 4, 11, and 12). However, their rapidity and reduced potential for contamination make them an attractive option for a diagnostic laboratory.
To transfer the conventional nested PCR assay across to the LC platform, the second-round PCR conditions were modified, and a fluorescent-labeled Aspergillus-specific probe was incorporated into the amplification mix. This significantly reduced labor costs and the duration of the assay without compromising sensitivity or specificity. The use of sealed capillary tubes for the second-round reaction and the absence of post-PCR product manipulation also reduced the risk of contamination due to amplicon carryover.
We conclude that the adaptation of a nested PCR assay developed for the rapid diagnosis of IA from blood, BAL, and tissue specimens to a real-time PCR platform is a significant advance. This platform should be applicable to the diagnosis of other invasive infections.
| ACKNOWLEDGMENTS |
|---|
This work was funded by the National Health and Medical Research Council of Australia through a Clinical Centre of Excellence grant (264625).
| FOOTNOTES |
|---|
| REFERENCES |
|---|
|
|
|---|
3)-ß-D-glucan test in weekly screening for invasive aspergillosis in patients with hematological disorders. J. Clin. Microbiol. 42:2733-2741.
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. |
|---|---|
| Clin. Vaccine Immunol. | ALL ASM JOURNALS |
|---|