Journal of Clinical Microbiology, December 2005, p. 5950-5956, Vol. 43, No. 12
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.12.5950-5956.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Rush Medical College, Chicago, Illinois,1 University of North Carolina at Chapel Hill, Chapel Hill, North Carolina,2 The Macfarlane Burnet Institute for Medical Research and Public Health, Melbourne, Australia,3 New England Research Institute, Watertown, Massachusetts,4 University of Zurich, Zurich, Switzerland,5 University of Witwatersrand and National Health Laboratory Services, Parktown, South Africa,6 Center for Disease Control and Prevention, Atlanta, Georgia,7 Institute of Microbiology, Medical School, University of Genoa, Genoa, Italy,8 MTC/SMI, Karolinska Institute, Stockholm, Sweden,9 Uppsala University, Uppsala, Sweden,10 National Institute of Allergy and Infectious Diseases, National Institutes of Health, Bethesda, Maryland,11
Received 3 February 2005/ Returned for modification 20 July 2005/ Accepted 21 September 2005
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
For the Ultrasensitive p24 assay, a standard ELISA format is used for the capture and detection of HIV p24 coupled with a specific signal amplification to increase the assay sensitivity. Heat denaturation of the plasma prior to binding in the ELISA step helps to dissociate immune complexes and denature the antibodies (24) so that they no longer compete for binding to the p24 antigen, a phenomenon that has plagued prior versions of this assay (10, 19, 25). The assay also adds a kinetic readout using the Quanti-Kin Detection System Software that was developed and validated in O. E. Varnier's laboratory (8) to increase the linear range of the assay. HIV p24 concentrations are reported as femtograms of HIV-1 p24/ml of plasma (R.I.L.A.B., S.r.l., Genoa, Italy).
For the RT assay (Cavidi), the RT enzyme is separated from the virus particle using a solid-phase extraction manifold, and the amount of RT enzyme is quantified using a functional assay whereby the RT incorporates bromodeoxyuridine (BrdU) monophosphate into DNA using a poly(A) template bound to a 96-well plate (3, 16). BrdU is then quantified spectrophotometrically using anti-BrdU conjugated to alkaline phosphatase, followed by the addition of its substrate. The RT activity in the unknown sample is compared to that of a recombinant RT enzyme standard with a known concentration. The extrapolated result is reported as fg of RT/ml of plasma or as HIV-1 RNA equivalents/ml based on a conversion factor supplied by the manufacturer. The assay has undergone revisions (17) to improve sensitivity and results for both versions of the kit are reported here.
We sought here to demonstrate the potential utility of measuring HIV p24 or RT activity in plasma by comparing detection rates and sensitivity and correlating the surrogate marker with HIV RNA detection.
(These data were presented in part as a poster on 14 to 17 September 2003 at the 43rd Annual Interscience Conference on Antimicrobial Agents and Chemotherapy Meeting, Chicago, Ill. [control #3803].)
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
|
The third panel consisted of spiked samples (subtype B HIV). This panel was designed to evaluate the improved sensitivity in the newly enhanced RT assay (Cavidi, v2.0). The range in viral load for this panel was 0 to 10,000 HIV RNA copies/ml with an emphasis on replicates at the low end (100 to 6,000 HIV RNA copies/ml). Testing was performed by the VQA only using the v2.0 RT assay (Cavidi).
Roche Amplicor HIV-1 Monitor Test, v1.5. HIV RNA testing was performed as defined in the package insert. For the standard extraction method, virus from 0.2 ml of plasma was lysed in the kit lysis buffer and the HIV RNA was precipitated using isopropanol and pelleted by centrifugation. After being washed with ethanol, the RNA was resuspended in the kit dilution buffer. Extracted RNA was amplified and detected according to the manufacturer's instructions, and results were reported as HIV RNA copies/ml (or log10 converted results). The equipment that is necessary to perform the Roche Amplicor Monitor Test is listed in Table 3. The estimated cost for this equipment is $13,200 (U.S. dollars).
|
Version 2.0 of the RT assay (Cavidi) was modified to improve assay sensitivity (17). Briefly, the virus particles from 1 ml of plasma were coated, captured, and lysed in 330 µl of lysis buffer instead of 600 µl of buffer to concentrate the RT in the eluate; additionally, more RT (from the samples and standards) was added to the 96-well plate (30 and 150 µl of eluate instead of 15 and 75 µl, respectively). Finally, in an attempt to streamline and improve linearity of the assay, day 2 was a day of extended incubation rather than washing and labeling, and day 3 was used for the detection of the assay. Even though the assay still took 3 days, day 2 did not require any hands-on activity. The equipment that is necessary to perform the Cavidi Exavir Load Assay is listed in Table 3. The estimated cost for this equipment is $8,300 (U.S. dollars).
Perkin-Elmer HIV-1 p24 ELISA plus the ELAST ELISA Amplification System. The Ultrasensitive p24 assay (Perkin-Elmer) was performed according to the manufacturer's instructions using the kit lysis buffer. Briefly, the sample (50 µl of plasma) was lysed with 250 µl of prediluted kit lysis buffer, heated at 100°C for 5 min, and cooled to room temperature. A volume of 250 µl of specimen, control, or standard was then added to the respective wells of a 96-well plate. The plate was incubated at 4°C overnight and then washed in an automated plate washer. The captured p24 antigen was labeled with a biotinylated anti-p24 antibody, followed by a streptavidin-horseradish peroxidase treatment step. The biotinyl-tyramide reagent (provided in the ELAST kit) was used to "amplify" the streptavidin-coated bound antigen and was followed by a second streptavidin-horseradish peroxidase step that was then developed colorimetrically and read kinetically for 10 min, followed by a final reading after 30 min when the reaction was stopped with concentrated acid. Quantitative results were reported based on the algorithms incorporated into the p24 assay Quanti-Kin software.
The modified version (Perkin-Elmer plus EB) of this assay was performed according to the manufacturer's instructions, with one deviation. The specimen (50 µl) was first incubated with 25 µl of the external buffer (30 mM Tris-HCl [pH 7.2], 450 mM NaCl, 1.5% Triton X-100, 1.5% deoxycholic acid [sodium salt], 0.3% sodium dodecyl sulfate, 10 mM EDTA) at room temperature for 10 min (26). This sample was then treated with 225 µl of prediluted kit lysis buffer and heat denatured as described above. Quantitative results were also reported based on the algorithms incorporated into the p24 assay Quanti-Kin software. The equipment that is necessary to perform the Perkin-Elmer Ultrasensitive p24 test is listed in Table 3. The estimated cost for this equipment is $6,800 (U.S. dollars).
Statistics. Detection rates were based on the kit-defined limits. For the RT assay (Cavidi) and the Ultrasensitive p24 assay (Perkin-Elmer), no limit of detection is set, but instead each assay's limit of detection is defined based on the combined performance of negative or background specimens and the standard curve. The RT-PCR assay (Roche) limit of detection is 400 copies HIV RNA/ml.
Comparisons of reproducibility were based on coefficients of variation (CVs) because the assays measure different analytes. The mean and standard deviation was estimated from a model that allowed for left censoring where some of the results were below the limit of detection of an assay. A CV was calculated only if at least two positive results were obtained at a given concentration because unbiased estimates of the mean and standard deviation could not be obtained with fewer than two. The CVs for panel 1 were based on the intra-assay standard deviation.
Rank correlations were calculated for each panel in an attempt to provide an estimate of the relationship between the median log10 results obtained with the RT (Cavidi v1.0 and v2.0) assays or the Ultrasensitive p24 (Perkin-Elmer and Perkin-Elmer + EB) assays and the Roche assay or nominal concentrations. A separate correlation was calculated for each run of each assay in each laboratory. Negative samples were excluded from all analyses. The results that were below the limit of detection of an assay were assigned a value of 1.0. On panels 1 and 3, results from the various assays were correlated with nominal concentrations, but for panel 2 the results were correlated with the median RT-PCR (Roche) result because it included clinical samples that did not have a corresponding nominal concentration. On panel 1, correlations with nominal concentrations were calculated for the Ultrasensitive p24 (Perkin-Elmer plus EB) for nominal concentrations greater than 10,000 copies/ml and for the RT assay (Cavidi v1.0 and v2.0) and the RT-PCR assay (Roche) at all positive concentrations. The results from the Ultrasensitive p24 (Perkin-Elmer) were excluded because positive results were only obtained at the two highest concentrations. On panel 2, correlations were calculated for the Ultrasensitive p24 (Perkin-Elmer and Perkin-Elmer plus EB) and the RT assays (Cavidi v1.0 and v2.0). The table includes minimum, median, and maximum correlations for each assay, as well as the number of correlations calculated (i.e., the number of runs available).
| RESULTS |
|---|
|
|
|---|
|
For panel 1, the Ultrasensitive p24 assay (Perkin-Elmer) detected 100% of the specimens with virus loads of 250,000 copies/ml or greater; the addition of the external buffer increased the detection rate to 100% for samples with virus loads of 50,000 copies/ml or greater. Detection rates for the Ultrasensitive p24 (Perkin-Elmer) assay were also improved in panel 2, when the external buffer was included. Detection of samples from panel 2 ranged from 0 to 86% for the Ultrasensitive p24 assay, but improved to 66 to 99% when the external buffer was used in the procedure. Better sensitivity was observed in clinical specimens than in spiked samples, regardless of the buffer used.
For panel 2, the combination of intra- and interassay variation was used because the absence of replication within the panel prevented estimation of the intra-assay standard deviation. For panel 1, median CVs were 31, 31, 27, 27, and 13% for the RT-PCR (Roche), RT (Cavidi v1.0), RT (Cavidi v2.0), Ultrasensitive p24 (Perkin-Elmer), and modified p24 (Perkin-Elmer plus EB) assays, respectively. For panel 2, the median CVs were 26, 26, 17, 18, and 57%.
Correlation values between RT-PCR (Roche) and nominal concentrations were very good, indicating that the expected virus load of the panel members was very close to that which was obtained by RNA testing (Table 5). The Ultrasensitive p24 assay (Perkin-Elmer) gave reasonable correlation values with HIV RNA concentration, but this correlation was improved when an external buffer was used (Perkin-Elmer plus EB). Furthermore, better correlations between RT-PCR and Ultrasensitive p24 values were observed when panel members included only spiked samples versus spiked and clinical samples. The RT assays (Cavidi v1.0 and v2.0) correlated very well with virus load and did not seem to vary if the panel included spiked or clinical samples.
|
| DISCUSSION |
|---|
|
|
|---|
The search for a simplified assay to measure HIV has identified two potential candidates. The first, the Ultrasensitive p24 (Perkin-Elmer) assay, measures the HIV-specific core protein, which may be present as virion-associated or free p24 antigen, although free p24 is mostly bound in immune complexes (25, 27). The correlation between p24 detection and HIV RNA or CD4 cell counts has been evaluated in a number of studies. Schupbach and others have demonstrated that the p24 is predictive of CD4 depletion (14, 28) and disease progression (14, 35), that it decreases when ART is initiated (2, 23, 29), and that it correlates inversely with CD4 changes even when HIV RNA is so low that it cannot be quantitated (30, 31). However, disappointing results have also been reported using this kit (1, 21). Bonard et al. tested plasma samples from 14 patients treated with highly active antiretroviral therapy, found weaker changes in Ultrasensitive p24 antigen testing compared to HIV RNA results, and questioned the reliability of this assay as an RNA surrogate for clinical management of HIV-infected patients. Prado et al. demonstrated a lack of correlation between HIV RNA and the Ultrasensitive p24 assay (Perkin-Elmer) for monitoring patients undergoing a structured treatment interruption. Prado et al. found that discordance between HIV RNA detection and p24 detection after viral rebound was 76% of the treated patients and 49% of the naive patients enrolled in the study. The discordance was attributed to the presence of high-affinity immunoglobulins that interfered with the p24 assay detection. More recently, Schupbach et al. reported that p24 antigen was detectable and remained relatively unchanged during short STI (31). These contradictory findings demonstrate the importance of using standardize methods in attempts to compare assays and demonstrate clinical utility. The two earlier studies utilized the Ultrasensitive p24 kits in absence of the modified external buffer, while the third study included it. Our data demonstrate that the modified buffer does impact the sensitivity of the assay; fully optimized assays should be used for future studies trying to demonstrate utility of this assay for monitoring HIV disease progression.
The p24 data obtained in the current study need to be interpreted carefully. The panels utilized contained two distinct types of samples, those that were spiked with well-characterized virus and those that were obtained from HIV-infected individuals. The reality is that clinical samples will typically contain p24 that is both not virion associated and virion associated. However, in attempting to evaluate the performance of a particular assay there are benefits to using a well-characterized specimen, including the ability to decipher the contribution of immune-complexed versus virion-associated p24 antigen in the detection rates of this assay.
Limits of detection were quite high in samples where only virion-associated p24 was present (i.e., HIV spiked samples). It is estimated that 2,000 to 4,000 p24 Gag proteins are present in each virion, so it would take 6,150 to 12,500 virus particles or 12,500 to 25,000 copies of viral RNA to detect 1 pg of p24 protein (5). Since the limit of detection of the modified p24 assay is approximately 6 pg/ml, it is not surprising that the observed detection rate in spiked samples is approximately 100,000 copies/ml. When clinical samples were tested, the detection rate appeared to be greatly improved, and this is presumed be due to the presence of non-virion-associated p24. Since the contribution of this component may vary between patients (4, 25) and change with disease progression (13, 14, 35), the impact on HIV monitoring is not fully understood and explains the lower correlation with HIV RNA. HIV p24 detection rates were improved with the use of the external buffer (kindly provided by Jorg Schupbach), especially in HIV-spiked samples, where only virion-associated proteins were present. The improved sensitivity with spiked samples suggests that the buffer may help to facilitate dissociation of the protein from the virus particle. Although the added external buffer also appeared to improve the detection of non-B subtypes, a more likely explanation would be that improved detection of the non-B subtype samples was due to improved assay sensitivity.
CV data demonstrate that comparable results were obtained between laboratories with the p24 assay. Correlations between the p24 assay and HIV RNA were lower when the kit buffer was used (r = 0.58 to 0.67) than when the external buffer was used (r = 0.59 to 0.95). Furthermore, correlations were higher when only spiked samples were included. The correlations with HIV RNA reported in this study were not different than those reported (r = 0.48 to 0.69) in other studies which did not utilize the external buffer (20, 22). Respess et al. did document a difference in correlation between untreated (r = 0.69) and treated (r = 0.48) patients, similar to the difference noted in this study between spiked and clinical specimens. This may be attributed to the fact that treatment alters the presence of virion-associated and non-virion-associated p24 antigen that is available for detection. Although more data are need to demonstrate the utility of the p24 assay for monitoring patients on ART, preliminary data suggest that it may prove useful for pediatric HIV diagnosis (15, 22, 37, 33) or diagnosis of acute infection (20). The fact that a small sample volume is required is especially attractive for pediatric diagnosis, and ongoing studies are under way to evaluate the potential for using this assay for detection of p24 in blood spots.
The second candidate to be considered as a surrogate for HIV RNA testing is the RT (Cavidi) assay. A number of studies in Africa, Australia, and France have provided excellent correlations between the RT and HIV RNA assays (3, 9, 34, 36). The fact that the assay only monitors virion-associated RT contributes to this excellent correlation. In addition, the assay could be useful in the detection of other lentiviruses (18), including HIV-2, and the extracted RT may be used in the Cavidi HIV Phenotype RT kit for evaluating resistance in patients on anti-RT regimens (32, 38). The simplified equipment, which includes a standard spectrophotometer, a vacuum driven extraction manifold, and wash buckets for plate washing, make this assay especially attractive for resource-limited urban and rural settings. The assay result may be reported as fg of RT/ml of plasma, but the use of a conversion factor may be used to convert the result into a more familiar readout of HIV RNA equivalents/ml. While the utility of this assay for monitoring patients on ART has yet to be fully evaluated, the roll-out of the v2.0 assay, with its improved sensitivity, should have a major impact. A limiting factor of this assay is sample volume. While a 1-ml plasma input is not problematic for monitoring adult patients, it certainly may be problematic for monitoring pediatric patients. Smaller sample volumes (down to 0.25 ml) may be utilized; however, this reduces assay sensitivity by the dilution factor involved (data from Cavidi).
The RT assay (Cavidi v1.0) showed good detection rates for samples with viral loads of >10,000 copies/ml. The v2.0 assay improved this detection limit to 400 copies/ml, which is comparable to that observed with the HIV RT-PCR assay (Roche) routinely used in many industrialized countries. CV analyses demonstrated comparable reproducibility between the RT and RNA assays. Comparable results were also obtained between the laboratories performing the testing, suggesting the assay is quite robust. Higher CVs were noted in the v2.0 assay than in the v1.0 assay. This was mainly observed in samples with lower viral loads, a phenomenon that has also been noted in HIV RNA assays for samples with viral loads near the limit of detection (7). Correlation between HIV RNA (Roche) and the RT (Cavidi) results were excellent for both versions of the kit (r = 0.83 to 0.99) and did not vary per panel. These results are quite similar to those reported by other investigators (9, 34, 36) using both versions of the RT assay (Cavidi). Preliminary studies involving non-B subtypes (34, 36), suggest that clades CRF02_AG, C, and B are detected well with this assay, but more studies are needed to substantiate these observations and better define detection rates of other clades. Factors that could affect detection rates include the presence of RT inhibitors or resistance mutations in the RT gene. The data presented by Greengrass et al. (9) suggest that the presence of non-nucleoside RT inhibitor-associated resistance mutations may reduce the RT activity by 0.20 log10, but the presence or absence of nucleoside RT inhibitor resistance mutations was not associated with RT activity. More data are needed to confirm this observed effect on RT activity and its impact on monitoring HIV-infected patients.
The benefits of using HIV RNA surrogates are assay simplicity and cost reduction. While kit and reagent costs will vary by country (RT-PCR = $17 to $80, RT assay =
$20, Ultrasensitive p24 = $5 to $10), equipment requirements for the HIV RNA surrogate assays are simpler and ca. 40 to 50% less expensive. In addition, although HIV RNA assays require well-developed infrastructures to ensure sterility and containment of nucleic acids and amplified products, the HIV RNA surrogate assays have more simplistic requirements with respect to facilities since nucleic acid extractions and amplification are not part of the assay. Therefore, the use of an HIV RNA surrogate assay to screen patients for drug efficacy and compliance, followed by less-frequent confirmation testing with molecular assays, may provide a useful strategy in resource-poor settings. The two assays evaluated in this study offer feasible RNA surrogates and should be considered for use in this type of strategy.
| ACKNOWLEDGMENTS |
|---|
We thank Joseph Fitzgibbon of the National Institutes of Health for helpful comments and careful review of the manuscript.