Comparison of the Binax NOW Flu A Enzyme Immunochromatographic Assay and R-Mix Shell Vial Culture for the 2003-2004 Influenza Season
Robert C. Fader*
Scott and White Memorial Hospital and the Texas A&M University Health Science Center, Temple, Texas 76508
Received 1 July 2005/
Returned for modification 17 August 2005/
Accepted 22 September 2005

ABSTRACT
The Binax NOW Flu A enzyme immunochromatographic assay was compared
to viral culture with R-Mix shell vials for 455 nasal-wash or
nasal-aspirate specimens. The overall sensitivity, specificity,
positive predictive value, and negative predictive value of
the assay were 64.9%, 98.4%, 89.3%, and 93.2%, respectively.
However, the assay sensitivity decreased significantly with
increasing patient age.

TEXT
With the advent of antiviral agents that target the neuraminidase
enzyme, the ability to diagnose and discriminate between influenza
A and B viruses has become more important. Numerous enzyme immunoassays
(EIA) are available to the laboratorian but often suffer from
a lack of sensitivity, although specificity is often high (
1-
3,
6-
15). Fluorescent-antibody stains have demonstrated increased
sensitivity and can assess specimen quality but are not conducive
to rapid testing in a physician's office or small hospital laboratory
(
3,
6,
9,
10). Viral culture with shell vials provides excellent
recovery (
4,
5) but usually not in the time frame that can impact
treatment with antiviral agents. Therefore, the EIA are useful
to the frontline clinician who is faced with the decision of
whether to administer antiviral agents or to admit the patient
for inpatient care.
The purpose of this study was to compare the performance of the Binax NOW Flu A and Flu B immunochromatographic assay to that of viral culture utilizing R-Mix shell vials. The Binax NOW Flu A and Flu B assay consists of separate test strips for detection of a nucleoprotein of influenza A and B viruses in a lateral-flow format and is approved for nasal-wash, nasal-aspirate, and nasopharyngeal-swab specimens. Nasal-wash and nasal-aspirate specimens can be tested directly on the test strips, whereas nasopharyngeal-swab specimens must be treated with an extraction reagent prior to testing. Results are available within 15 min of sample delivery to the test strips. The test kits can be stored at room temperature. The study was performed during the 2003-2004 influenza season when influenza A virus (H3N2/Fujian) was the predominant circulating strain. Consequently, the influenza B virus component could not be evaluated.
All specimens were nasal-wash or nasal-aspirate specimens submitted to the virology laboratory at Scott and White Memorial Hospital. The majority of the specimens were collected in the emergency department, in the pediatric outpatient clinic, or from inpatient settings and were most often transported immediately to the laboratory through a pneumatic tube system. Specimens were tested within 30 min of receipt in the laboratory by use of the Binax NOW Flu A and Flu B test (Binax, Inc., Portland, ME), following the manufacturer's directions.
Once EIA testing was completed, the nasal-wash or nasal-aspirate specimens were placed into M-4 transport medium (Remel, Lenexa, KS) and were refrigerated at 2 to 8°C until processed for cell culture. Specimens in M-4 transport medium were passed through a 0.45-µm Acrodisc syringe filter (Pall Corp., Ann Arbor, MI) and were inoculated into two R-Mix shell vials (Diagnostic Hybrids, Inc., Columbus, OH) as previously described (6), except that the shell vials were centrifuged at 700 x g for 50 min. Any remaining specimen was stored at 80°C for future use. After an overnight incubation at 37°C, the cell monolayer was rinsed with phosphate-buffered saline, fixed in chilled acetone, rinsed again in phosphate-buffered saline, and finally stained with SimulFluor Flu A/Flu B direct fluorescent-antibody (DFA) stain (Light Diagnostics, Temecula, CA), following the manufacturer's instructions. Coverslips were removed to a glass microscope slide and examined with a fluorescence microscope with dual filters (fluorescein isothiocyanate and rhodamine). With this stain, influenza A virus-infected cells demonstrate a green fluorescence, whereas influenza B virus-infected cells stain with yellow-gold fluorescence. For cultures with initial negative results, the second shell vial was processed after an additional overnight incubation as described above but was stained with a SimulFluor respiratory screen kit (Light Diagnostics, Temecula, CA), following the manufacturer's instructions. This stain will cause respiratory syncytial virus-infected cells to fluoresce yellow-gold, while cells infected with influenza A or B virus, adenovirus, or parainfluenza virus will fluoresce green. When virus identification could not be established by the SimulFluor respiratory screen kit, the specimen was retrieved from 80°C storage, thawed, and reinoculated into two additional R-Mix shell vials. The shell vials were incubated for 48 h, and each shell vial was stained with either the SimulFluor Flu A/Flu B DFA stain or the SimulFluor Para 1,2,3/Adeno DFA stain (Light Diagnostics, Temecula, CA), according to the manufacturer's instructions. The parainfluenza virus component of this stain causes infected cells to fluoresce green, whereas the adenovirus component causes infected cells to fluoresce yellow-gold.
The Binax NOW Flu A component of the EIA had an overall sensitivity, specificity, positive predictive value, and negative predictive value of 64.9%, 98.4%, 89.3% and 93.2%, respectively (Table 1). However, when analyzed based on the subject age group, the assay sensitivity decreased significantly. The sensitivity for the 0- to 5-year-old group was 84.6% but fell to 33.3% for those patients over the age of 50 (Table 1). The specificity of the assay for all age groups remained above 97%. The positive predictive value ranged from 83.3% for the over-50-year-old age group to 100% for the 21- to 50-year-old age group. The negative predictive values ranged from 86.3% for the 6- to 20-year-old age group to 96.9% for the 0- to 5-year-old age group. Of the six false-positive EIA results, one culture was positive for respiratory syncytial virus, one culture grew a parainfluenza virus, and the remaining four were negative for virus isolation.
This is the largest reported study of the Binax NOW Flu A and
Flu B immunochromatographic assay. During the 2003-2004 respiratory
virus season, Scott and White Memorial Hospital served as a
study site for the FluMist nasal vaccine (Aviron, Mountain View,
CA). Consequently, a large number of specimens were processed
for viral culture (2,159 total specimens with 860 total specimens
positive for respiratory viruses). The Binax NOW Flu A and Flu
B test was used primarily to assess patients prior to admission
for grouping purposes or to determine whether antiviral therapy
should be initiated. Viral culture was performed with R-Mix
shell vials because previous influenza surveillance studies
in our laboratory indicated that the R-Mix FreshCells recovered
more respiratory viruses than 14-day rhesus monkey kidney cell
cultures with less hands-on time (unpublished data). The results
of this study indicate that the Binax NOW Flu A test is a rapid,
user-friendly test for the presence of the influenza A virus.
The test is easy to perform with minimal hands-on time and is
suitable for rapid testing within or outside of the clinical
virology laboratory. However, even when testing was limited
to nasal-wash specimens or nasopharyngeal aspirates, which are
considered the most optimal specimen types (
3,
8,
11), the Binax
NOW Flu A EIA provided mixed results with regard to test sensitivity.
The assay performed well with specimens obtained from children
less than 5 years of age but provided unacceptable sensitivity
when specimens were obtained from older children or adults.
This outcome is similar to the results of Landry et al. (
11)
in their study comparing the Binax NOW Flu A and Flu B test
and the Directigen Flu A+B test (BD Microbiology Systems, Cockeysville,
MD) to viral culture and spin-enhanced fluorescent-antibody
stain. Landry et al. also noted increased assay sensitivity
in young children (<2 years of age) and decreased sensitivity
with specimens obtained from adult patients. Similar results
have been reported by Weinberg and Walker (
15), who evaluated
the Binax NOW Flu A and B assay in two age groups, younger than
and older than 9 years of age. This decrease in test sensitivity
in adult patients has also been noted with other enzyme immunoassay-based
tests for influenza (
9,
10,
13,
14) and has been suggested to
be a result of less viral shedding in adult patients (
9). Nevertheless,
a positive Binax NOW Flu A test result in an adult patient does
indicate a high likelihood that the patient is infected with
influenza virus.
The one obvious weakness of this study is the fact that the 2003-2004 influenza season in Texas was predominantly one of influenza A virus. Of the 773 positive influenza virus cultures recovered during the 2003-2004 respiratory season, only one was positive for influenza B virus. Consequently, the performance of the influenza B virus component of the assay could not be determined. The assay format, as tested in this study, consisted of individual test strips for influenza A and B viruses. This format provides the laboratory with the option of testing for either influenza A or B virus, depending on the influenza viruses circulating during the season. However, the company has now begun to manufacture an assay in which both influenza A and B viruses can be identified on a single test strip.
In summary, the results of the study indicate that while the Binax NOW Flu A EIA is relatively sensitive in specimens collected from children under the age of 5, the assay provides unacceptable sensitivity in older children and adults and must be used with caution in these populations. Depending on the clinical situation, further testing by culture, fluorescent-antibody stain, or PCR should be considered in the event of an initial negative result.

FOOTNOTES
* Mailing address: Microbiology, Scott and White Memorial Hospital, 2401 South 31st Street, Temple, TX 76508. Phone: (254) 724-2714. Fax: (254) 724-8776. E-mail:
rfader{at}swmail.sw.org.


REFERENCES
1 - Cazacu, A. C., G. J. Demmler, M. A. Neuman, B. A. Forbes, S. Chung, J. Greer, A. E. Alvarez, R. Williams, and N. Y. Bartholoma. 2004. Comparison of a new lateral-flow chromatographic membrane immunoassay to viral culture for rapid detection and differentiation of influenza A and B viruses in respiratory specimens. J. Clin. Microbiol. 42:3661-3664.[Abstract/Free Full Text]
2 - Cazacu, A. C., S. E. Chung, J. Greer, and G. L. Demmler. 2004. Comparison of the Directigen Flu A+B membrane enzyme immunoassay with viral culture for rapid detection of influenza A and B viruses in respiratory specimens. J. Clin. Microbiol. 42:3707-3710.[Abstract/Free Full Text]
3 - Covalciuc, K. A., K. H. Webb, and C. A. Carlson. 1999. Comparison of four clinical specimen types for detection of influenza A and B viruses by optical immunoassay (FLU OIA test) and cell culture methods. J. Clin. Microbiol. 37:3971-3974.[Abstract/Free Full Text]
4 - Dunn, J. J., C. Gordon, C. Kelley, and K. C. Carroll. 2003. Comparison of the Denka-Seiken INFLU A · B-Quick and BD Directigen Flu A+B kits with direct fluorescent-antibody stain and shell vial culture methods for rapid detection of influenza viruses. J. Clin. Microbiol. 41:2180-2183.[Abstract/Free Full Text]
5 - Dunn, J. J., R. D. Woolstenhulme, J. Langer, and K. C. Carroll. 2004. Sensitivity of respiratory virus culture when screening with R-Mix Fresh Cells. J. Clin. Microbiol. 42:79-82.[Abstract/Free Full Text]
6 - Fong, C. K. Y., M. K. Lee, and B. P. Griffith. 2000. Evaluation of R-Mix FreshCells in shell vials for detection of respiratory viruses. J. Clin. Microbiol. 38:4660-4662.[Abstract/Free Full Text]
7 - Hindiyeh, M., C. Goulding, H. Morgan, B. Kenyon, J. Langer, L. Fox, G. Dean, D. Woolstenhulme, A. Turnbow, E. Billetdeaux, S. Shakib, C. Gordon, A. Powers, G. Vardeny, M. Johnson, L. Skodack-Jones, and K. Carroll. 2000. Evaluation of BioStar FLU OIA assay for rapid detection of influenza A and B viruses in respiratory specimens. J. Clin. Virol. 17:119-126.[CrossRef][Medline]
8 - Kaiser, L., M. S. Briones, and F. G. Hayden. 1999. Performance of virus isolation and Directigen Flu A to detect influenza A virus in experimental human infection. J. Clin. Virol. 14:191-197.[CrossRef][Medline]
9 - Landry, M. L., S. Cohen, and D. Ferguson. 2000. Impact of sample type on rapid detection of influenza virus A by cytospin-enhanced immunofluorescence and membrane enzyme-linked immunosorbent assay. J. Clin. Microbiol. 38:429-430.[Abstract/Free Full Text]
10 - Landry, M. L., and D. Ferguson. 2003. Suboptimal detection of influenza virus in adults by the Directigen Flu A+B enzyme immunoassay and correlation of results with the number of antigen-positive cells detected by cytospin immunofluorescence. J. Clin. Microbiol. 41:3407-3409.[Abstract/Free Full Text]
11 - Landry, M. L., S. Cohen, and D. Ferguson. 2004. Comparison of Binax NOW and Directigen for rapid detection of influenza A and B. J. Clin. Virol. 31:113-115.[CrossRef][Medline]
12 - Rodriguez, W. J., R. H. Schwartz, and M. M. Thorne. 2002. Evaluation of diagnostic tests for influenza in a pediatric practice. Pediatr. Infect. Dis. 21:193-196.
13 - Ruest, A., S. Michaud, S. Deslandes, and E. H. Frost. 2003. Comparison of the Directigen Flu A+B test, the QuickVue influenza test, and clinical case definition to viral culture and reverse transcription-PCR for rapid diagnosis of influenza virus infection. J. Clin. Microbiol. 41:3487-3493.[Abstract/Free Full Text]
14 - Shultze, D., Y. Thomas, and W. Wunderli. 2001. Evaluation of an optical immunoassay for the rapid detection of influenza A and B viral antigens. Eur. J. Clin. Microbiol. Infect. Dis. 20:280-283.[CrossRef][Medline]
15 - Weinberg, A., and M. L. Walker. 2005. Evaluation of three immunoassay kits for rapid detection of influenza virus A and B. Clin. Diagn. Lab. Immunol. 12:367-370.[Abstract/Free Full Text]
Journal of Clinical Microbiology, December 2005, p. 6133-6135, Vol. 43, No. 12
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.12.6133-6135.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Roh, K. H., Kim, J., Nam, M.-H., Yoon, S., Lee, C. K., Lee, K., Yoo, Y., Kim, M. J., Cho, Y.
(2008). Comparison of the Seeplex Reverse Transcription PCR Assay with the R-mix Viral Culture and Immunofluorescence Techniques for Detection of Eight Respiratory Viruses. Annals of Clinical & Laboratory Science
38: 41-46
[Abstract]
[Full Text]
-
Mehlmann, M., Bonner, A. B., Williams, J. V., Dankbar, D. M., Moore, C. L., Kuchta, R. D., Podsiad, A. B., Tamerius, J. D., Dawson, E. D., Rowlen, K. L.
(2007). Comparison of the MChip to Viral Culture, Reverse Transcription-PCR, and the QuickVue Influenza A+B Test for Rapid Diagnosis of Influenza. J. Clin. Microbiol.
45: 1234-1237
[Abstract]
[Full Text]
-
Leland, D. S., Ginocchio, C. C.
(2007). Role of Cell Culture for Virus Detection in the Age of Technology. Clin. Microbiol. Rev.
20: 49-78
[Abstract]
[Full Text]
-
Zitterkopf, N. L., Leekha, S., Espy, M. J., Wood, C. M., Sampathkumar, P., Smith, T. F.
(2006). Relevance of influenza a virus detection by PCR, shell vial assay, and tube cell culture to rapid reporting procedures.. J. Clin. Microbiol.
44: 3366-3367
[Abstract]
[Full Text]
-
Agoritsas, K., Mack, K., Bonsu, B. K., Goodman, D., Salamon, D., Marcon, M. J.
(2006). Evaluation of the Quidel QuickVue Test for Detection of Influenza A and B Viruses in the Pediatric Emergency Medicine Setting by Use of Three Specimen Collection Methods.. J. Clin. Microbiol.
44: 2638-2641
[Abstract]
[Full Text]