Previous Article | Next Article ![]()
Journal of Clinical Microbiology, March 2005, p. 1059-1063, Vol. 43, No. 3
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.3.1059-1063.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Division of Clinical Microbiology,1 Department of Health Sciences Research,2 Department of Dermatology, Mayo Clinic College of Medicine, Rochester, Minnesota3
Received 19 July 2004/ Returned for modification 14 September 2004/ Accepted 9 November 2004
|
|
|---|
|
|
|---|
Since May 2000, our laboratory has implemented the use of LightCycler PCR (Roche Molecular Biochemicals, Indianapolis, Ind.) for the routine diagnosis of dermal and genital HSV infections. Because of the large number of specimens processed in our laboratory, we also implemented the routine use of the MagNA Pure LC automated extraction system (Roche Molecular Biochemicals) for the detection of HSV DNA from dermal and genital tract swabs. The MagNA Pure LC system provided our laboratory with a standardized, efficient, and cost-effective method for extraction of nucleic acids from specimens compared to the manual IsoQuick system (Orca Research, Inc., Bothell, Wash.) and another automated extraction system (BioRobot 9604; QIAGEN, Inc., Chatsworth, Calif.) (4).
In an effort to optimize the performance characteristics of the LightCycler PCR assay, decrease cost, improve turnaround time, and increase the availability of our automated extraction system (MagNA Pure LC) to process other types of specimens, we evaluated the swab extraction tube system (SETS) (catalog no. 3 315 568; Roche Applied Science) as a potential effective method for extraction of HSV DNA from dermal and genital swabs that could replace the MagNA Pure LC system. In a study published earlier, a group in our laboratory has demonstrated the suitability of this extraction method for the detection of group A streptococcus target DNA from throat swabs using LightCycler PCR (14).
(This work was presented in part at the 20th Annual Clinical Virology Symposium, Clearwater Beach, Fla., 25 to 28 April 2004.)
|
|
|---|
Consecutive dermal and genital swabs from 642 adult and pediatric patients were collected from December 2003 through January 2004. Seventy-nine of 642 (12.3%) specimens were excluded because patients declined to provide permission to use their specimens (Minnesota Statute 144.335) or because the specimens were taken from sources other than skin or genital tract. A total of 563 dermal and genital swabs sufficient to detect a 2.1% difference between the two methods were available for analysis.
Collection of specimens. Swabs from dermal specimens (n = 234) and genital swabs (n = 329) from patients suspected of having HSV infections were initially extracted by laboratory technologists, placed in 2 ml of serum-free medium, processed by the MagNA Pure LC system, and assayed by LightCycler PCR per our standard (laboratory) practice. The original swabs were discarded. However, the collection and transport tubes containing a gel or sponge with remaining fluid were then used as a specimen source for our study. For the (study) comparison, a double-swab (CultureSwab; Becton Dickinson Microbiology Systems, Cockeysville, Md.) was inserted into each collection tube (Fig. 1). The laboratory technologist made sure that both swabs were soaked with material; one swab was picked randomly and processed using the MagNA Pure LC system, and the other swab was processed by SETS. The time period between receiving and processing the specimens varied between 24 and 72 h.
![]() View larger version (33K): [in a new window] |
FIG. 1. Specimen collection and processing. Noted centrifuge speed is as on an Eppendorf 5417 (Brinkman Instruments).
|
(ii) SETS. Two hundred microliters of swab neutralization buffer (2 mM EDTA and 30 mM Tris [pH 8.4]) was first placed in the outer tube of the SETS. The remaining swab from the clinical specimen was inserted into the inner tube of the SETS. The exposed shaft of the CultureSwab was then broken above the swab portion contained in the inner tube using a gloved hand and Bio-Screen biohazard wipe (Fisher Scientific, Pittsburgh, Pa.). The SETS with the swab was capped and centrifuged at 14,000 x g for 1 min using a microcentrifuge with an aerosol-tight fixed-angle rotor (Eppendorf 5417C; Brinkman Instruments, Westburg, N.Y.) to extract the fluid from the swab. The inner tube containing the dry swab was then discarded, and the outer tube containing fluid was covered with a screw cap, labeled, and placed in a 100°C heating block for 2 min.
(iii) LightCycler PCR assay. Five microliters of extracted specimen processed by the MagNA Pure LC system and SETS was added, each separately, to 15 µl of LightCycler FastStart Master Hybridization probe mix containing the following: 6.6 µl of water, 2.4 µl of 25 mM magnesium chloride, 2.0 µl of LightCycler-herpes simplex virus 1/2 primer/hyb probes (catalog no. 03355632001; Roche Diagnostics), 2.0 µl of LightCycler FastStart DNA Master hybridization probes (catalog no. 3003248; Roche Diagnostics), and 2.0 µl of a 1:8 dilution of LightCycler-herpes simplex virus 1/2 recovery template (catalog no. 03355624001; Roche Diagnostics) per reaction mixture. The LightCycler PCR was performed using the following protocol: (i) an initial step of 10 min at 95°C; and (ii) 45 cycles, with 1 cycle consisting of 10 s of denaturation at 95°C, 15 s of annealing at 55°C, and 15 s at 72°C with fluorescence measurement. Melting curve analysis was performed after PCR as follows: 95°C for 0 s, 40°C for 1 min, and 85°C at 0.2°C/s with continuous measurement of the fluorescence. LightCycler software was used to determine the genotype and cycle threshold (CT) for each specimen processed by either method.
Statistical analysis. The positivity rates for the MagNA Pure LC system and SETS were estimated, and 95% exact binomial confidence intervals were calculated for each HSV genotype and specimen source (dermal and genital swabs) separately. The positivity rates of the two different methods (MagNA Pure LC system and SETS) were compared in pairs by the sign test. The numbers of cycles before the detection of HSV DNA were compared by the paired t test or Wilcoxon signed-rank test as appropriate.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Detection of HSV DNA by LightCycler PCR after extraction of DNA from dermal and genital swabs by the MagNA Pure LC system and by SETSa
|
|
View this table: [in a new window] |
TABLE 2. Detection of HSV DNA by LightCycler PCR after extraction of DNA from dermal and genital swabs by the laboratory (original swabs) and by SETSa
|
|
View this table: [in a new window] |
TABLE 3. Detection of HSV DNA by LightCycler PCR after extraction of DNA from dermal and genital swabs by the laboratory (original swabs) and by the MagNA Pure LC system (dual swab)a
|
![]() View larger version (12K): [in a new window] |
FIG. 2. Comparison of the detection of HSV DNA extracted by SETS and MagNA Pure LC system.
|
|
View this table: [in a new window] |
TABLE 4. Positive results of specimens or swabs processed by the MagNA Pure LC system and SETS by HSV subtype and specimen source
|
|
View this table: [in a new window] |
TABLE 5. Discrepant results between specimens processed by the MagNA Pure LC system, SETS, and laboratory result (original specimens)
|
|
|
|---|
Rapid real-time PCR cycling, and most importantly the amplification and detection of amplicons in a closed vessel in automated platforms, has allowed integration of this seminal technology into the mainstream clinical laboratory for the diagnosis of many viral and other microbial infections. Since May 2000, our laboratory implemented real-time PCR as a replacement test for cell culture technology for the detection of HSV DNA from dermal and genital swabs (3).
Using real-time PCR, we and others have achieved higher sensitivities ranging from 28 to 108% in the clinical laboratory compared to the sensitivities of conventional cell culture methods (2, 3, 9). Although test costs and reimbursement of real-time PCR are similar or better than cell culture methods, the associated capital equipment expenses, especially for nucleic acid extraction technology, are a potential drawback for laboratories wanting to incorporate this new level of viral diagnosis into the clinical laboratory.
We previously evaluated the technical performance of three nucleic acid extraction methods (IsoQuick, MagNA Pure LC system, and BioRobot 9604) for recovery of nucleic acid prior to amplification of target DNA of HSV by real-time LightCycler PCR (4). In the present study, we compared a manual nucleic acid extraction method, SETS, to the MagNA Pure LC system in order to improve work flow, increase turnaround time, and potentially decrease the costs associated with automated technology particularly for laboratories which process fewer of these specimens daily than larger reference centers.
Interestingly, in our evaluation, the SETS extraction method was more sensitive than the MagNA Pure LC system in that HSV DNA was detected in more specimens (179 versus 157, respectively; P < 0.0001) than the automated method when the two methods were compared head to head under exactly the same conditions. The higher positivity rate detected by the laboratory (original swabs) compared to that of SETS (although not statistically significant [P = 0.82]) results from the fact that the laboratory had the advantage of processing the original specimen. The high CT for the positive specimens detected by the laboratory indicates a low initial HSV DNA copy number in the original swab, and the copy number of HSV DNA could have been even lower in the remaining fluid that was used to test the SETS method.
The technical processing for comparing SETS with the MagNA Pure LC system was experimentally designed to reflect existing methods for nucleic acid extraction. Thus, for MagNA Pure LC extraction, one of two dual swabs was initially swirled in 2 ml of serum-free medium prior to real-time PCR amplification. Alternatively, the comparison swab for extraction of nucleic acid by SETS was centrifuged in 0.2 ml of swab neutralization buffer; the resultant extract was then heated at 100°C for 2 min before real-time PCR amplification. Therefore, the volume of the SETS extraction was concentrated 10-fold relative to the MagNA Pure LC method. The MagNA Pure LC system can process a variable number of samples (1 to 32) in one run. To process 30 samples, it requires approximately 30 min of hands-on time to aliquot the samples, prepare the machine, and load the reagents and 1.5 h to complete the nucleic acid extraction. The cost of MagNA Pure LC extraction in our laboratory is $5.69 per sample (excluding the cost of acquiring the MagNA Pure LC instrument).
The SETS method requires less than 1 min per sample to insert the swab, break the shaft, and close the inner tube. The centrifugation time is 1 min, and the heating time is 2 min. Therefore, it takes between 25 and 35 min to complete the DNA extraction of 30 specimens with a cost of $0.5 per sample.
Overall, our study shows that HSV DNA extraction using SETS is faster, less expensive, and more sensitive than the MagNA Pure LC system and that SETS could replace the MagNA Pure LC system in the laboratory diagnosis of HSV infections using LightCycler PCR.
|
|
|---|
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»