Journal of Clinical Microbiology, July 2005, p. 3438-3442, Vol. 43, No. 7
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.7.3438-3442.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
David A. Bruckner,2 and
Andrew E. Simor3
Gen-Probe Incorporated, San Diego, California,1 UCLA Medical Center, Department of Pathology and Laboratory Medicine, Los Angeles, California,2 Sunnybrook and Women's College Health Sciences Centre, Department of Microbiology, Toronto, Canada3
Received 29 November 2004/ Returned for modification 24 January 2005/ Accepted 21 March 2005
|
|
|---|
|
|
|---|
Common methods for susceptibility testing include broth microdilution, disk diffusion, and screening agars, which require timely incubations after isolate identification. Alternatively, the Oxoid PBP2' latex agglutination (OLA) assay (Oxoid, United Kingdom) is a rapid test for differentiating MRSA from methicillin-sensitive Staphylococcus aureus (MSSA) on agar plates (2, 11, 14, 23, 24, 25, 26, 27). Unfortunately, reliable testing of CoNS with the OLA assay requires induction with oxacillin (Ox), and a heavier inoculum size and longer lysis are critical for accurate results (11, 14, 26).
Recently, molecular assays for methicillin resistance and Staphylococcus identification using mecA and Staphylococcus-specific genes have yielded fast and reliable results (4, 6, 9, 12, 13, 16, 20). Molecular assays have an advantage because they do not rely on time-consuming incubations or phenotypic expression (7). However, because of the technical expertise required, many formats are not practical for routine use in most clinical laboratories.
The objective of this study was to evaluate the performance of a Gen-Probe (San Diego, Calif.) prototype DNA probe hybridization protection assay (HPA). This assay, currently under development, can simultaneously identify a Staphylococcus isolate and the mecA gene for rapid detection of MRSA and methicillin-resistant CoNS (MRCoNS) (Fig. 1). This is the first report describing the use of DNA probe HPA for detecting single-stranded DNA. A total of 589 clinical isolates and 24 American Type Culture Collection (ATCC) isolates were tested using the oxacillin MIC, the DNA probe HPA, and the OLA assay. Specificities for 368 clinical isolates representing 288 S. aureus and 80 CoNS strains were tested on subculture plates. To test assay sensitivity and specificity in a clinical setting, 221 isolates representing 110 S. aureus and 111 CoNS strains were tested from primary culture plates. Each clinical isolate represented only one strain per patient. Ox MIC was the reference method, since it is used in the laboratory for routine susceptibility testing. PCR of the mecA gene was used to resolve any discrepant results.
![]() View larger version (20K): [in a new window] |
FIG. 1. Simultaneous identification of Staphylococcus and determination of mecA status. Positive probe results are indicated in red.
|
Probe analysis was performed using HPA (1, 18), which employs probes with chemiluminescent acridinium ester labels complementary to a highly conserved bacterial rRNA gene region or antibiotic resistance gene. These highly conserved bacterial rRNA gene probes have been extensively characterized for identification specificity (15). Gen-Probe Incorporated supplied all proprietary reagents. The assay was run in white 96-well microtiter plates (Whatman, Clifton, N.J.). Each microtiter plate contained three probe mixtures in individual wells to make up the reaction set (Fig. 1). A 1-µl loop of cells was lysed at 100°C in 150 µl of a succinate-buffered detergent reagent. After the addition of 100 µl of base solution and a 10-min 60°C incubation, the reaction mixture was cooled to room temperature and 100 µl of a hybridization buffer was added. Of this mixture, 50 µl was aliquoted to each of the three wells and a 50-µl oil overlay was added to prevent evaporation throughout the 60°C assay. HPA was automated using a BMG luminometer (Lumistar 500; BMG Technologies, Offenburg, Germany), and the results were available in under 60 min.
The probe results for detecting MRSA and MRCoNS from both primary and subculture plates are summarized in Tables 1 and 2. Of the 398 S. aureus clinical isolates, 196 were Ox resistant (MICs
4 µg/ml) and 202 were Ox susceptible (MICs
2 µg/ml) (Table 1). The mecA probe results demonstrated that 184 of the 196 Ox-resistant isolates produced positive probe results and that 199 of the 202 Ox-susceptible isolates produced negative probe results. The mecA probe results showed 184/196 MRSA and 199/202 MSSA. The mecA probe, OLA, and PCR were negative for 12/196 MRSA isolates formally defined as borderline resistant (MICs
4 to 8 µg/ml). Regarding the three OLA-, mecA probe-, and PCR-positive MSSA samples, the initial OLA results for two of the three were negative, but upon using the modified protocol, both samples converted to OLA positive. Converted OLA-positive isolates may reflect low expression levels of the mecA protein due to regulatory mutations.
|
View this table: [in a new window] |
TABLE 1. Results for 411 S. aureus isolates in Ox MIC, GP probe, and OLA assays
|
|
View this table: [in a new window] |
TABLE 2. Results for 202 CoNS clinical and ATCC isolates in Ox MIC, GP Probe, and OLA assays
|
0.5 µg/ml) and 62 were Ox susceptible (MICs
0.25 µg/ml) (Table 2). Of the 129 MRCoNS isolates, 116 were positive by mecA probe, PCR, and OLA. Eleven of the 13 discrepant MRCoNS isolates (MIC = 0.5 to 1.0 µg/ml) were S. saprophyticus, S. lugdunensis, or S. warneri (Table 2). Mechanisms of resistance other than those mediated by mecA may be responsible for these results. Other investigators using similar molecular methods with CoNS have observed this phenomenon (8, 11). Based on these MIC cutoffs, CLSI suggests further testing for clarification of resistance status using molecularly based techniques (17). The other two discrepant MRCoNS isolates were OLA positive and PCR negative and were identified as S. intermedius (MIC
0.5 µg/ml) and S. warneri (MIC = 16 µg/ml). The two OLA-positive MRCoNS isolates that disagreed with the mecA probe and PCR results might reflect mutations. However, with three different DNA probes widely dispersed along the mecA gene, three different mutations yielding a false-negative result are unlikely. A possible explanation is that the material tested by the mecA probe and PCR was different from the colonies used for MIC and OLA. This hypothesis is supported by OLA-negative results when glycerol retention samples of these colonies were grown on plates containing an oxacillin disk. These isolates were originally identified only as CoNS using a coagulase test, so a correlation at the species level could not be performed between the first and second rounds of test isolates.
Of 62 MSCoNS isolates, 56 were negative by mecA probe, PCR, and OLA. PCR confirmed the six mecA probe-positive MSCoNS isolates, with one displaying a positive OLA result identified as S. warneri. Of the five that disagreed with OLA, four were S. epidermidis and one was S. sciuri. The five OLA-negative isolates may represent an unexpressed mecA gene due to mutations in the regulatory regions. The single OLA-positive isolate may be an example of down regulation as opposed to nonexpression of the mecA gene.
Testing of 24 ATCC Staphylococcus isolates showed 100% agreement for S. aureus with the mecA probe and PCR. Of the 11 ATCC CoNS isolates, 1 of 3 MRCoNS isolates was mecA probe and PCR negative and was identified as S. saprophyticus. OLA results for the ATCC isolates agreed with MIC, mecA probe, and PCR except for one false positive for a methicillin-sensitive S. warneri isolate and the false negative for S. saprophyticus (Table 2). The OLA manufacturer's insert states that S. warneri may give a false positive, and in such cases, the Ox MIC should be used for determining resistance.
Rapid identification of Staphylococcus, more specifically S. aureus, along with the methicillin resistance status has been the focus of many recent studies (4, 6, 9, 12, 13, 20). Many direct and amplified formats have problems with false negatives due to lack of an internal control to verify that sufficient target was present and accessible. The DNA probe system described here addresses this problem by including an internal control consisting of rRNA gene-targeted Staphylococcus identification probes (Fig. 1). If there is sufficient DNA target to give a valid identification signal, there should be sufficient target for determination of the mecA status, thus eliminating false-negative results.
This study demonstrates that the identification of an isolate as a member of the Staphylococcus genus, and more specifically as S. aureus, coupled with the determination of mecA status in a 1-hour timeframe is feasible. The DNA probe identification results for S. aureus displayed 100% sensitivity and specificity. The mecA DNA probe results for S. aureus demonstrated 93.9% sensitivity and 98.6% specificity, and CoNS had 89.4% sensitivity and 91.4% specificity compared to MICs. When PCR was used as the means to resolve the mecA discrepant results, the mecA probe test demonstrated 100% sensitivity and specificity with all S. aureus and CoNS isolates tested.
We have shown that this Staphylococcus-S. aureus-mecA probe system is an accurate, easy, and rapid assay. The DNA probe sample-handling methodology requires less than 5 min of hands-on time, is amendable to batch processing, and is performed in under 15 min. The probe hybridization and detection steps are fully automated and can be completed in 45 min using the automated microtiter instrument. The data are automatically exported into a spreadsheet at the end of the run, which makes analysis simple. Coupled with effective communication to the physician, this 1-hour DNA probe assay could improve management and infection control practices for methicillin-resistant Staphylococcus infections.