Department of Microbiology, Laboratory Group Heidelberg, D-69126 Heidelberg, Germany
Received 24 November 2004/ Returned for modification 4 January 2005/ Accepted 2 May 2005
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
Fast results provided by these instruments are not only due to short incubation times for identification (ID) and antimicrobial susceptibility testing (AST) of bacterial isolates but also to hands-on time for setting up the different devices. We compared the different manipulation steps required by the VITEK 2 and Phoenix systems for setting up bacterial isolates for ID and AST. Additionally, the time to result was recorded for each system. The workflow analysis was the main focus of this evaluation; however, efficient workflow must generate accurate results. Therefore, performance characteristics with regard to ID and AST were also investigated. ID results were compared to API systems (bioMérieux), and AST results were compared to standard broth microdilution as recommended by the National Committee for Clinical Laboratory Standards (NCCLS) (14).
(The findings of this study were partly presented at the 103rd Gen. Meet. Am. Soc. Microbiol., abstr. C-250, 2003.)
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
Phoenix. For the Phoenix system, the combined ID and AST NMIC/ID 14 panel for gram-negative bacilli and the PMIC/ID 13 panel for gram-positive cocci were used. The setup of the panels was performed according to the manufacturer's instructions. The Phoenix ID broth was inoculated with bacterial colonies from Columbia blood agar and adjusted to a 0.5 to 0.6 McFarland standard using the Crystal Spec Nephelometer (BD Diagnostic Systems). After supplementing the AST broth with one drop of indicator dye, 25 µl of the ID suspension was transferred to the AST broth to achieve a final inoculum density of 1.5 x 108 CFU/ml. The ID and the AST broths were poured into the respective side of the panel placed on the Phoenix inoculation station. The inoculated panels were closed and placed into the transport caddy, and, after entering the accession number, the panels were placed into the Phoenix instrument.
VITEK 2. For ID testing with VITEK 2, the ID-GPC card was used for gram-positive cocci and the ID-GNB card was used for gram-negative rods. For AST, the AST-P 523 panel was used for gram-positive cocci and the AST-N021 card was used for gram-negative bacilli. A sufficient number of colonies was suspended in sterile saline (0.45%) and adjusted to a 0.5 McFarland turbidity standard using the DensiChek Densitometer (bioMérieux). The inoculated tube was placed in a cassette on the VITEK 2 Smart Carrier Station. The sample number was entered and associated with an ID and AST card. The sample accession numbers and card identification bar code numbers were scanned, and the information was stored on the cassette memory chip. The SCS cassette with the cards and the test tubes was placed on the VITEK 2 instrument where the inoculation of the AST cards was automatically performed by the instrument.
Antibiotics. The following antimicrobial agents were tested: for Enterobacteriaceae and NFGNB, ampicillin, piperacillin, piperacillin-tazobactam, meropenem, cephazolin, cefuroxime, ceftazidime, cefepime, and levofloxacin; for staphylococci, penicillin, oxacillin, erythromycin, clindamycin, teicoplanin, vancomycin, rifampin, ofloxacin, gentamicin, tetracycline, and cotrimoxazol; for enterococci, teicoplanin, vancomycin, and tetracycline.
Reference system. For ID, the API 20 E, ID 32 GN, API 32 Staph, and API 32 Strep systems (bioMérieux) were used as comparator systems. Frozen microdilution panels (Biotest AG, Dreieich, Germany) were used as the reference method for AST. These panels were set up according to NCCLS guidelines (14).
Detection of special resistance mechanisms. MRSA were detected by Oxascreen Agar (BD Diagnostic Systems). Additionally, the presence of the mecA gene was determined by an in-house PCR test using the following primers: 5'-ACA TCT ATT AGG TTA TGT TGG-3' and 5'TAT ATT CTT CGT TAC TCA TGC-3'. Gel electrophoresis was performed for the detection of the 492-bp amplification products. The ESBL resistance mechanism was detected using the combined disk methods recommended by NCCLS (14).
Quality control. The following reference strains were used for AST according to NCCLS guidelines (13): Escherichia coli ATCC 25922, E. coli ATCC 35218, Pseudomonas aeruginosa ATCC 27853, S. aureus ATCC 29213, Enterococcus faecalis ATCC 29212, S. aureus ATCC 43300, and Klebsiella pneumoniae ATCC 700603.
Workflow. The average manipulation time required to set up each system was determined for 39 batches of seven organisms each. Two skilled technicians were involved in the practical procedure. Using a stopwatch, all steps were timed by an observer not involved in the testing process. The following steps associated with the workflow were included: preparation of bench and devices, adjustment of McFarland standard of 0.5 to 0.6 with bacterial inoculum, setup of VITEK 2 cards and Phoenix panels, loading of the system, entry of accession numbers, and the follow-up work. The details associated with the manipulation steps are provided in Table 2. Considering that the procedures of both systems are not exactly identical, similar workflow steps were directly compared, whereas differing steps are mentioned separately. Additionally, the time for maintenance of both instruments was measured twice during the study of approximately two month.
|
| RESULTS |
|---|
|
|
|---|
|
|
AST. The overall category agreement for all isolates was 97.0% for both instruments. The minor error rates, major error rates, and very major error rates for all bacterial isolates tested were 3.0, 0.3, and 0.6 and 2.8, 0.2, and 1.7 for Phoenix and VITEK 2, respectively (P values of 0.76, 0.75, and 0.09). The AST results of all single antimicrobials and organisms included in this study are shown in Table 5. The 11 ESBL-producing Enterobacteriaceae isolates were correctly detected by Phoenix; VITEK 2 missed one K. oxytoca isolate. The oxacillin resistance mechanism was correctly detected by both systems.
|
| DISCUSSION |
|---|
|
|
|---|
Workflow. An important issue when conducting workflow comparative studies is the experience of the technicians with each system. Our technicians were well trained and experienced in the operation of both systems, as several previous studies have been performed with both of these systems (7 and U. Eigner, A. Caganic, D. Schuhmacher, R. Englert, and A. M. Fahr, Abstr. 99th Gen. Meet. Am. Soc. Microbiol. 1999, abstr. C-85, 1999, and A. M. Fahr, U. Wild, and U. Eigner, Abstr. 13th Eur. Cong. Clin. Microbiol. Infect. Dis. 2003, abstr. P-755, 2003). In this study, the mean time to set up one bacterial isolate was 3.0 min for the Phoenix system and 1.5 min for the VITEK 2 system (P < 0.001). The longer hands-on time for the Phoenix instrument was mainly due to the additional manual steps of transferring 25 µl of ID broth to AST broth, dropping of the indicator dye into the AST broth, inoculating the panels with the respective broth, closing the panels, placing the panels in the transport caddy, and entering the panels into the instrument. Additionally, setting up the McFarland standard with CrystalSpec was more time-consuming compared to DensiChek. According to BD Diagnostic Systems this will be improved with a new version in the near future.
Meyer et al. performed a timed workflow study comparing the Phoenix instrument with the Microscan Walkaway-96 (W. A. Meyer III, D. Lockwood, W. Brasso, and J. Reuben, Abstr 100th Gen. Meet. Am. Soc. Microbiol. 2000, abstr. C-309, 2000). The following workflow steps were measured to process 50 isolates: logging of isolates into ID and AST system (3.3 min), manual labeling of the panels (7.7 min), inoculation of panels (14.6 min), and the loading of the panels into the instrument (3.8 min). These selected values were similar to our respective manipulation steps. Unfortunately the inoculum preparation time was not evaluated in this study due to the significant differences between the processes normally associated with overnight ID and AST analysis (Microscan) and "rapid" ID and AST analysis (Phoenix). Therefore, the final results are not comparable to our evaluation, where the differences between both instruments were mainly due to specimen inoculum preparation procedures.
Our result of 1.5 min for setting up one isolate for the VITEK 2 instrument is in accordance with other investigators, considering the fact that each study design is slightly different. Larone et al. reported 1.3 min technical hands-on time to set up one isolate with VITEK 2 compared to 1.7 min for VITEK classic and 1.9 min for Microscan (D. H. Larone, L. J. Tucci, and D. O. Samide, Abstr. 100th Gen. Meet. Am. Soc. Microbiol. 2000, abstr. C-279, 2000). Goessens et al. calculated a mean operating time of 0.9 min per specimen for VITEK 2 and 2.3 min for Microscan Walkaway (W. H. F. Goessens, A. H. J. A. Van Vliet, and H. A. Verbrugh, Abstr. 9th Eur. Cong. Clin. Microbiol. Infect. Dis. 1999, abstr. P-0822, 1999). Abele-Horn et al. reported a time frame of 15.0 to 20.0 min for inoculum preparation and setting up 20 species with the VITEK 2 instrument (0.8 to 1.0 min per isolate), in comparison to 30.0 to 40.0 min with API systems and agar diffusion tests (1.5 to 2.0 min per isolate) (1).
Comparing both instruments, W. S. Seto reported 2.4 min and 2.0 min from initiation of inoculum preparation to loading into the machine with Phoenix and VITEK 2, respectively (personal communication). Silver et al. compared the Phoenix with the VITEK 2 instrument for 325 isolates of staphylococci and also stated that Phoenix required more preparation time than VITEK 2 without showing any detailed data (D. M. Silver, L. Louie, and A. E. Simor, Abstr. 102nd Gen. Meet. Am. Soc. Microbiol. 2002, abstr. C-132, 2002).
TTR. In this study, TTR was determined when the results of all antibiotics on one card or panel were final. The mean TTR evaluated for the different bacterial groups were significantly longer for Phoenix in comparison to VITEK 2. To our knowledge no other direct studies had been published on comparative TTR results between both instruments. The results available are from separate studies for each instrument, therefore a comparison with this study is not valid.
For complete ID and AST results with the Phoenix system, Donay et al. required 7.5 h, 16.0 h, 13.0 h, 10.5 h, and 11.0 h for E. coli, all NFGNB isolates, P. aeruginosa (exclusively), S. aureus, and E. faecalis, respectively (6). Endimiani et al. reported an average TTR of 14.8 h ± 1.6 h for NFGNB and a lower TTR of 13.0 h ± 1.8 h for Acinetobacter baumannii (7).
With regard to VITEK 2, the results reported in this study were almost in accordance with other evaluations. In a previous study we collected AST TTR for 3,143 fresh clinical isolates reported by VITEK 2 in our routine laboratory (A. M. Fahr, U. Eigner, D. Bertsch, M. Kirstahler, M. Holfelder, and A. Turnwald-Maschler, Abstr. 13th Eur. Cong. Clin. Microbiol. Infect. Dis. 2003, abstr. P-760, 2003). The mean TTR for all bacteria tested was 8.1 h, with 7.2 h for Enterobacteriaceae, 12.1 h for P. aeruginosa, 6.9 h for S. aureus, and 9.0 h for enterococci. Similar TTR results for VITEK 2 were reported by other working groups (12, 15).
For ID and AST, several studies have compared either the Phoenix or VITEK 2 system with reference methods (1, 6, 7, 8, 9, 10, 11, 12, 13, 15, 16, 17).
ID. Both systems were equally good at identifying all four bacterial groups examined (96% to 100%). Similar results for the Phoenix instrument were reported by Stefaniuk et al. and Fahr et al., whereas Donay et al. found results of 89.4% and 86.2% for NFGNB and enterococci (6, 8, 16). For VITEK 2, correct ID results of 95% and 97.8% were reported for routine strains and 88.5% for a challenge set of gram-negative organisms (9, 13, 15). Comparing both systems Gross et al. found a concordance of 96% for gram-negative and gram-positive isolates (R. Gross, U. Hörling, and G. Peters et al, Abstr. 12th Eur. Cong. Clin. Microbiol. Infect. Dis. 2002, abstr. P-703, 2002).
AST. For all bacterial groups the CA was >95% for both instruments, only for NFGNB we observed a CA of 84.8% for Phoenix and 86.9% for VITEK 2. For both instruments similar results were obtained for mE and ME, whereas VITEK 2 had a larger but not significant VME rate compared to Phoenix for Enterobacteriaceae, NFGNB, and staphylococci. For the Phoenix system similar CA of >95% are reported for Enterobacteriaceae, staphylococci, and enterococci with a low VME rate of <1.5% (6, 8, 16). For P. aeruginosa, Donay et al. reported a CA of 75.8% with a VME rate of 0.3% for Phoenix (6). For VITEK 2 comparable results with our study were reported for gram-negative isolates (12, 13). For 138 Enterobacteriaceae tested Pérez-Vázquez et al. observed an essential agreement of 97.8% with a VME rate of 6.1%, mainly due to piperacillin (15). Good performance in detecting vanA, vanB, and vanC1 resistance mechanisms in enterococci was stated by van den Braak et al. for VITEK 2 (17). Comparing both instruments, Gross et al. showed comparability between Phoenix and VITEK 2 of 94% for all bacterial groups tested (R. Gross et al., 12th Eur. Cong. Clin. Microbiol. Infect. Dis. Abstr. 2002, abstr. P-704, 2002).
In conclusion, the mean setup time for one bacterial isolate was significantly longer with Phoenix when compared to VITEK 2, whereas VITEK 2 required more maintenance time than Phoenix. TTR for all bacterial groups was shorter with VITEK 2 when compared to the Phoenix system, so more results will be available with the VITEK 2 instrument on the same day. VITEK 2 had a larger VME rate than the Phoenix system, the difference being not significant. CA, mE, ME, and correct ID rates were equally good for both systems.
| ACKNOWLEDGMENTS |
|---|
This work was partly sponsored by bioMérieux, Marcy L'Ètoile, France.