Journal of Clinical Microbiology, August 2005, p. 3901-3905, Vol. 43, No. 8
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.8.3901-3905.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Departments of Pediatrics,1 Laboratory Medicine,4 Medicine, University of Washington,5 Department of Biostatistics, Children's Hospital and Regional Medical Center, Seattle, Washington,3 Departments of Pediatrics, Chang-Gang Children's Hospital, Kaohsiung, Taiwan2
Received 18 March 2005/ Accepted 6 May 2005
|
|
|---|
= 0.74 [P < 0.0001] and
= 0.56 [P = 0.0001], respectively). The sensitivity of the HD p24 Ag assay was similar when plasma and whole blood on filter paper were contrasted to the real-time RT-PCR assay (80% versus 82.5% and 78.6% versus 83.3%, respectively). However, while the specificity of the HD p24 Ag assay of plasma on filter paper was 100%, the specificity was diminished in whole-blood specimens. The storage of specimens on filter paper for 2 weeks at 37°C, 24°C, or 0°C did not alter the detection or quantification of HD p24 Ag. These results suggest that transport and storage of plasma on filter paper and quantification of HD p24 Ag may be a reliable method for HIV-1 load monitoring. |
|
|---|
The quantification of p24 antigen (Ag) has been considered for viral load monitoring (3, 4, 13, 15, 19, 21-23, 25-28). The p24 Ag assay measures the virus directly and has been shown to correlate with levels of plasma HIV-1 RNA in untreated subjects (21, 23, 30) and persons receiving antiretroviral therapy (3, 23, 27, 28). In addition, detection of p24 Ag has been explored for early diagnosis of pediatric HIV-1 infection (19, 20, 24). Assays that quantify p24 Ag are less expensive, time consuming, labor intensive, and costly than the RNA assay (21, 23, 27-29) and may be a practical alternative for viral load monitoring.
Dried specimens collected on filter paper have the potential to further reduce the expense of viral load testing by simplifying sample collection, storage, and shipment (14, 17, 18). Transport of specimens on filter paper has proved effective for detection of HIV-1 DNA (1, 7, 8, 9), quantification of HIV-1 RNA (4, 6, 12, 18), and monitoring the emergence of drug resistance mutations (8). Here we report the evaluation of filter paper for transport and storage of human whole blood and plasma for subsequent quantification of viral core Ag (p24 Ag). Our goal was to examine a simple, inexpensive, practical method for specimen processing and storage that could be integrated with an existing inexpensive validated kit for HIV-1 load quantification.
|
|
|---|
30% humidity, or at 0°C for 2 weeks prior to storage at 80°C. HD p24 Ag assay. Specimens were eluted from filter paper by incubation in 0.5% Triton X-100 buffer for 60 min in a 1.5-ml polypropylene tube (USA Scientific, Ocala, FL). The eluted whole blood or plasma in 0.5% Triton X-100 was separated from the filter paper by centrifugation at 13,600 x g for 3 min in an Ultrafree-MC tube (Millipore, Billerica, MA). The eluate was then transferred to a 2-ml screw-cap polypropylene tube and heated at 100°C for 5 min in a dry heat block and assayed using the HIV-1 p24 enzyme-linked immunosorbent assay (ELISA) kit (Perkin-Elmer Life Sciences, Wellesley, MA) boosted by the ELAST ELISA amplification system (Perkin-Elmer Life Sciences) (26). Briefly, treated whole blood or plasma (0.25 ml) was transferred to wells of the HIV-1 p24 Ag ELISA kit, and the wells were covered and incubated for 2 h at room temperature on a microtiter plate shaker. Wells were washed with 1x wash buffer, and then 0.1 ml of biotinylated detector antibody was added and the mixture was incubated for 1 h at 37°C. After a wash, 0.1 ml of streptavidin-horseradish peroxidase solution was added to the wells and the plate was incubated for 15 min at 37°C. After a wash, 0.1 ml of biotinyl-tyramide solution was added to the wells and the plate was incubated for 15 min at room temperature. Following a wash, 0.1 ml of streptavidin-horseradish peroxidase diluted in buffer solution was added to the wells and the plate was incubated for 15 min at room temperature. O-Phenylenediamine substrate solution (0.1 ml) was added to the wells after a wash. At the end of 30 min, the colorimetric reaction was stopped by the addition of 0.1 ml of stop solution and the optical density (OD) read at 490 nm using an ELISA reader. Kit standards were included in each assay run. The assay lower limit of detection was established as the mean of the negative controls plus 3 standard deviations. All specimens were tested in duplicate, and the mean of the two values was used in statistical analysis.
Quantification of plasma HIV-1 RNA by real-time PCR. Viral particles were pelleted from blood plasma (500 µl) by centrifugation at 23,000 x g for 1 h at 4°C, lysed (5.75 M guanidinium isothiocyanate, 190 mM dithiothreitol, 25 mM Tris-HCl, 30 µl of glycogen [10 mg/ml]), and then extracted with isopropanol (600 µl) and ethanol (1 ml of a 70%, vol/vol, solution). The overall specimen preparation schema followed the manufacturer's procedure for the Roche Monitor HIV-1 RNA US-RT-PCR assay, with the exception that the pellet was resuspended in 50 µl rather than 100 µl of lysis buffer. Real-time PCR was done on each specimen using a master mixture (40 µl) composed of manganese acetate at 25 mM (6 µl; Applied Biosystems, Foster City, CA), 5x EZ buffer with ROX (10 µl; SYNTHEGEN, Houston, TX), deoxynucleoside triphosphates at 10 mM (6 µl; Amersham Pharmacia Biotech, Piscataway, NJ), forward primer HXB2-Gag-F (CAA GCA GCC ATG CAA ATG TT) at 20 µM in AE buffer (1 µl), reverse primer SK431-B (TGC TAT GTC ACT TCC CCT TGG TTC TCT) at 20 µM in AE buffer (1 µl), probe HXB2-gag (6-carboxyfluorescein-AAA GAG ACC ATC AAT GAG GAA GCT GCA GAA-6-carboxytetramethylrho-damine) at 10 µM (0.5 µl), 12.75 µl diethyl pyrocarbonate H2O, RNase inhibitor at 26.4 U/µl (0.75 µl; Amersham Pharmacia Biotech), and rTth at 2.5 U/µl (2 µl; Applied Biosystems), and sample (10 µl) was added to each reaction tube. The tube was sealed and placed in the reaction plate and then into the TaqMan PCR machine (ABI Prism 7700). The PCR program thermocycler conditions were set at 60°C for 30 min for reverse transcription, 95°C for 5 min for denaturation, and 42 cycles of amplification at 95°C for 20 s, 52°C for 20 s, and 60°C for 1 min. All primers were high-performance liquid chromatography purified and purchased from Invitrogen (Frederick, MD), and all probes were high-performance liquid chromatography purified and purchased from Applied Biosystems. The standard curve for the assay was prepared from Armored RNA HIV-1 of subtype B (Ambion, Inc., Austin, TX) in TSM buffer (10 mM Tris, pH 7.0; 100 mM NaCl; 1 mM MgCl2; 0.1% gelatin; diethyl pyrocarbonate H2O; carrier Escherichia coli tRNA at 10 µg/µl). The working stock contained 30,000 RNA copies/µl, as verified by measuring the A260 and confirmed using the Roche Monitor HIV-1 RNA assay.
HIV-1 genotyping. HIV-1 subtypes of pol were assessed as described previously (10) using the Stanford database HIVseq Sequence Analysis Program (www.hivdb.stanford.edu).
Statistical analysis.
HD p24 Ag and HIV-1 RNA concentrations were log10 transformed for all analyses. Sensitivity statistics were calculated for the samples from HIV-1-infected individuals. Specificity statistics were calculated using samples from 20 uninfected controls. Spearman rank correlation coefficients were used to evaluate the strength of the relationship between HIV-1 RNA concentrations and HD p24 Ag levels. Spearman's rho (
) statistic was calculated from the rankings of observations within the population sampled. Two features of the Spearman statistic support its use for this analysis: (i) that it is not limited to detecting linear relationships and (ii) that it requires no arbitrary assumptions to allow inclusion of measurements that fall below the detection limit. The Spearman correlation coefficients were calculated for both plasma and whole-blood samples.
Linear regression was chosen to analyze the effects of different short-term storage temperatures on measurements of HD p24 Ag, as the fit of the linear model and the degree by which the slope differs from 1.0 is critical to determining whether temperature affects the assay results. Slope estimates and r2 statistics (i.e., the proportion of the total variance explained by the linear model) are reported. Similarly, for HD p24 Ag measurements in samples on filter paper and samples in polypropylene tubes, linear regression slopes and r2 values are reported. For the linear regression modeling, observations that fell in the undetectable range were assigned a value equal to the detection limit of the assay, 102 fg/ml.
|
|
|---|
Concordance of HIV-1 detected by HD p24 Ag and by real-time PCR. The concordance of qualitative detection of HIV-1 in plasma specimens stored on filter paper and tested in the HDp24 Ag assay to plasma specimens stored in tubes and evaluated by the real-time PCR of HIV-1 RNA was 87.5% (35 of 40), with 75% (30 of 40) of the specimens being positive in both assays and 12.5% (5 of 40) being negative in both assays. The nonconcordant specimens had low viral loads (<3.5 log10) (Fig. 1). The sensitivities of the two assays were similar, with a virus detected in 33/40 (82.5%) specimens by the HIV-1 RNA assay and in 32/40 (80%) by the HD p24 Ag assay.
![]() View larger version (24K): [in a new window] |
FIG. 1. Plasma HIV-1 loads determined by detection of HD p24 Ag from filter paper and HIV-1 RNA concentrations determined by real-time PCR in 40 specimens from HIV-1-infected patients. The subtype CRF01_AE specimens are circled, and the single subtype C specimen tested negative for both RNA and p24 Ag (however, a previous specimen from the subject with subtype C had been sequenced).
|
HD p24 Ag was not detected in any of the 20 plasma specimens from laboratory workers without HIV-1 infection when stored on filter paper or in tubes (100% specificity). Testing of whole blood on filter paper yielded a higher mean OD compared to plasma in the specimens from both HIV-1-infected and uninfected individuals. The mean difference in OD (± standard deviation) between whole blood and plasma in non-HIV-1-infected persons was 0.12 ± 0.05, and in HIV-1-seropositive persons it was 0.08 ± 0.02. In fact, the OD for HD p24 Ag in the whole-blood specimens from all 20 non-HIV-1-infected individuals exceeded the negative cutoff and thus tested falsely positive.
Quantification of HIV-1 by HD p24 Ag compared to real-time PCR of plasma viral RNA.
The quantification of HD p24 Ag in plasma on filter paper correlated with the plasma HIV-1 RNA level. The Spearman rank correlation of these two assays was
= 0.74 (P <0.0001) (Fig. 1). The HD p24 Ag assay detected virus in 30% (3 of 10) of specimens with viral loads of <1,000 RNA copies/ml, 88.9% (8 of 9) of specimens with viral loads between 1,000 and 10,000 copies/ml, and 100% (21 of 21) of specimens with viral loads of >10,000 copies/ml. The quantification of plasma HD p24 Ag stored in polypropylene tubes yielded very similar results to that of plasma stored on filter paper. The concordance was 97.5% (39 of 40 specimens), with 77.5% (31 of 40) positive in both assays and 20% (8 of 40) negative in both assays. The slope of the linear regression on the log10-transformed values was 1.037, with 95% confidence bounds of 0.96 and 1.11. The proportion of total variation explained by the linear model, r2, was 0.95.
The relationship between quantity of HD p24 Ag measured in whole blood on filter paper and plasma HIV-1 RNA levels was statistically significant but not as strong as for plasma. The Spearman rank correlation for testing of whole blood on filter paper compared to plasma in tubes was
= 0.56 (P = 0.0001) (Fig. 2). Detection of p24 Ag increased with the plasma HIV-1 RNA. The HD p24 Ag assay detected viral Ag in 37.5% (3 of 8) of the specimens with viral loads of <1,000 RNA copies/ml, 57.1% (4 of 7) of the specimens with viral loads between 1,000 and 10,000 copies/ml, 93.8% (15 of 16) of the specimens with viral loads between 10,000 and 100,000 copies/ml, and 100% (11 of 11) of the specimens with viral loads of >100,000 copies/ml.
![]() View larger version (23K): [in a new window] |
FIG. 2. HIV-1 loads determined by detection of HD p24 Ag in whole blood on filter paper and HIV-1 RNA concentrations determined by real-time PCR in plasma in specimens from 42 HIV-1-infected patients. The subtype CRF01_AE specimens are circled, and the single subtype C specimen tested negative for both RNA and p24 Ag (however, a previous specimen from the subject with subtype C had been sequenced).
|
![]() View larger version (11K): [in a new window] |
FIG. 3. Stability of p24 Ag in whole blood (A) and plasma (B) on filter paper at different temperatures for 2 weeks prior to freezing at 80°C is shown compared to an aliquot of plasma from the same specimen frozen in a polypropylene tube at 80°C within 6 h of collection. The samples stored at 37°C are represented by plus signs, the samples stored at 24°C by circles, and the samples stored at 0°C by triangles.
|
|
|
|---|
The correlation of HD p24 Ag in whole blood stored on filter paper to the plasma HIV-1 RNA load was relatively weak, for reasons not fully elucidated in our study. The OD readings from whole blood were increased compared to the concordant plasma specimens. Opsonization of pathogens by complement and antibodies can lead to their binding to complement receptor 1 (CD35), also called the immune adherence receptor, on primate erythrocytes (16). Erythrocyte-associated HIV-1 RNA has been detected in persons with or without detectable virus in plasma, providing evidence that HIV-1 immune complexes bind to erythrocytes (13). Since the p24 Ag is largely bound in immune complexes (11), quantification of this Ag in whole-blood samples would be expected to detect a higher level of this viral protein compared to testing of plasma. However, the higher OD readings were also evident in the whole blood of non-HIV-1-infected laboratory workers (20 of 20 of whom tested in the low positive range for HD p24 Ag), suggesting that a component of whole blood increases the OD. One possibility is that hemoglobin, released from erythrocytes lysed at 100°C, may coat the bottom of the microtiter well and remain there even after several washes. The hemoglobin could increase the optical absorbance and the OD. Importantly, if the relatively high OD of whole blood was derived from a nonspecific component, this could lessen the quantitative correlation between HD p24 Ag and HIV-1 RNA, especially at low viral loads. Such a nonspecific effect could explain the flattening of the curve we observed with testing of HD p24 Ag in whole blood (Fig. 2). These results stand in contrast to studies showing comparable quantification of HIV-1 RNA in dried whole-blood and plasma spots (5, 18) and to plasma stored in tubes (18). Importantly, the false-positive reactions of whole blood on filter paper indicate that this method should not be used for early diagnosis of infants without further evaluation of negative specimens to determine the cutoff for a positive reaction, as others have done (19).
A significant correlation between HD p24 Ag and HIV-1 RNA has been reported (3, 19, 23, 28), primarily in studies evaluating HIV-1 subtype B. However, in several other studies the correlation was weak or absent (2, 5, 22). Bonard and colleagues showed a weak correlation (r = 0.33) for quantification except for HIV-1 RNA levels of >5 log10 copies/ml (r = 0.62; P < 0.001) in African adults infected primarily with HIV-1 CRF02_AG strains (2). Bürgisser and colleagues found a poor correlation (r = 0.39) in a study of mostly non-B subtypes (5), and plasma p24 Ag levels did not parallel HIV-1 RNA levels following structured treatment interruptions (22). However, Pascual and colleagues reported a correlation between plasma p24 Ag and HIV-1 RNA when assessing a panel of subtypes A to F and 18 specimens from Malawi (all subtype C) (21). Our study evaluated mostly subtype B variants. The paucity of data on non-B subtype virus suggests that further validation of the assay, especially with non-B subtypes, is warranted.
An important finding of our study was that whole blood and plasma on filter paper could be stored for 2 weeks at ambient temperatures without adversely affecting the measurement of p24 Ag. The conditions evaluated, 2 weeks at 37°C or 24°C, should encompass the room temperatures of most laboratories. Importantly, the use of desiccant and moisture-tight bags is advisable in environments with high humidity to minimize the risk of fungal growth on the blood spots.
The procedures used to perform the HD p24 Ag assay are relatively simple and do not require costly equipment. Also, the assay is also generally less expensive than commercial HIV-1 RNA assays. The cost of the HD p24 Ag assay has been estimated to amount to 5 to 20% of a HIV-1 RNA assay (21, 23, 29).
The reliable preservation and transport of plasma p24 Ag on filter paper have far-reaching epidemiological and public health implications. The close correlation we observed between HD p24 Ag in plasma stored on filter paper and HIV-1 RNA levels in plasma stored in tubes and the stability of HD p24 Ag over a range of environmental temperatures suggest that the assay could provide a convenient option for viral load monitoring in regions where transport of specimens to laboratories for testing is problematic. Testing of whole blood on filter paper would provide even greater convenience; however, in our evaluation, quantification of HD p24 Ag in whole blood was inferior to testing of plasma because a component of whole blood appeared to nonspecifically increase the OD readout of the assay. In conclusion, the attributes of filter paper for specimen transport and laboratory safety, combined with the close correlation of HD p24 Ag with plasma viral RNA and the stability of p24 Ag over a large range of temperatures, provides impetus for continued refinement of the assay as a practical and economical tool for monitoring viral loads in HIV-1-infected patients.
|
|
|---|