Discordance between Viral Loads Determined by Roche COBAS AMPLICOR Human Immunodeficiency Virus Type 1 Monitor (Version 1.5) Standard and Ultrasensitive Assays Caused by Freezing Patient Plasma in Centrifuged Becton-Dickinson Vacutainer Brand Plasma Preparation Tubes
Hossein Salimnia,1*
Ellen C. Moore,2
Lawrence R. Crane,3
Rodger D. MacArthur,3 and
Marilynn R. Fairfax1
Department of Pathology and DMC University Laboratories,1
Division of Allergy, Immunology and Rheumatology, Department of Pediatrics,2
Division of Infectious Diseases, Department of Medicine, Wayne State University School of Medicine, 540 E. Canfield, Detroit, Michigan 482013
Received 11 March 2005/
Returned for modification 25 April 2005/
Accepted 25 May 2005

ABSTRACT
The Roche COBAS AMPLICOR human immunodeficiency virus type 1
(HIV-1) Monitor (version 1.5) standard and ultrasensitive viral
load assays often gave discordant results, with viral loads
from the standard assay exceeding those from the ultrasensitive
assay by more than 0.5 log
10 for approximately 20% of specimens
received. We began studies to determine the extent, magnitude,
and reproducibility of the discordance between the assays and
to discover and eliminate the cause of this discordance. Until
then, we revised our standard operating procedure to include
both standard and ultrasensitive testing on all specimens submitted
for viral load determinations. Discordant results usually recurred
on retesting. They were most prevalent for specimens with ultrasensitive
viral loads of <1,000 and rare for specimens with viral loads
of >10,000. Often, standard assay results exceeded those
of the ultrasensitive assay by 50- to 100-fold. At higher viral
loads, the difference between the standard and ultrasensitive
assays persisted, but the percent difference was smaller and
rarely caused discordance. The proportion of discordant results
was significantly higher in specimens from pediatric patients
than in specimens from adults. The ultrasensitive viral load
determinations generally agreed with the results of the B-DNA
(Bayer) viral load assays. If the plasma was transferred from
the centrifuged plasma preparation tubes before freezing, standard
and ultrasensitive results were concordant with each other and
with values determined on plasma from lavender-topped EDTA tubes.

INTRODUCTION
Human immunodeficiency virus (HIV) viral load determinations
represent the standard of care in the initial assessment of
HIV-infected persons and in the subsequent management of the
disease (
4,
11,
13). Spurious high viral loads may cause patients
to be started unnecessarily on antiretroviral therapy. For the
patient on therapy, such elevated results could prompt unnecessary
repeat testing, HIV genotyping, or medication changes. Pregnant
women could receive unnecessary caesarean sections. Providers
may wrongly conclude that there is nonadherence to a prescribed
antiretroviral regimen. Erroneous conclusions may be drawn if
the patient is participating in a clinical trial. Thus, accuracy
and reproducibility of the viral load results are important
considerations. The Roche standard and ultrasensitive assays,
available in manual and automated (COBAS) versions, are the
most common tests for viral load quantification, and the Bayer
Versant human immunodeficiency virus type 1 (HIV-1) RNA 3.0
(B-DNA) assay (Tarrytown, New York) is also widely used. The
standard and ultrasensitive assays have been reported to correlate
well both within versions and between versions 1.0 and 1.5 (
2,
3,
5,
6), although at very low viral loads version 1.5 is the
more sensitive (
2,
9).
Our laboratory offered both (COBAS) standard and ultrasensitive assays (version 1.5) and performed the ordered test. When standard and ultrasensitive assays were performed with the same patient sample, the ultrasensitive result was less than 50 RNA copies/ml, while the standard result was 3,000. This discordance recurred on retesting the specimen. Determination of the HIV-1 viral loads for 20 pediatric patients by both standard and ultrasensitive methods revealed similar discordances in six of the results, with the standard results being higher. From July to December 2004, we revised our standard operating procedure to perform both standard and ultrasensitive assays on all plasma specimens submitted for viral load determination and began investigations to determine which result was correct and how to eliminate the problem.
(A preliminary version of these data was presented informally by E. C. Moore at the Pediatric AIDS Clinical Trials Group Winter Meeting, Baltimore, Md., 1 to 4 December 2004. Some information was also provided to Becton-Dickinson to allow them to alert other clients to the discrepancy and the techniques we developed to avoid it.)

MATERIALS AND METHODS
Patients.
Patient specimens used in the study were submitted to Detroit
Medical Center (DMC) University Laboratories for viral load
testing between July and December of 2004. Eighty percent of
the patients came from the DMC adult HIV clinic (1,600 patients)
or the pediatric clinic (about 90 patients). The remainder of
the specimens were submitted by infectious-disease physicians
or came from patients in our component hospitals or the high-risk-pregnancy
clinic. This study was approved by the Human Subjects Committee
of Wayne State University.
Specimen collection and transport.
Except as described for the individual experiments, blood was collected in plasma preparation tubes (PPT) by standard venipuncture and maintained at room temperature until centrifuged at 1,100 x g for 20 min within 4 h of collection. The PPT was then frozen at 20°C until the day of testing, when it was thawed at room temperature. The PPT is an evacuated plastic tube for the collection of venous blood. It contains dried K2EDTA as an anticoagulant. It also contains a proprietary gel material that, during centrifugation, allows the cellular elements to pass through and forms a barrier between the plasma and the cells. This technology permits transportation of the PPT without decanting the cell-free plasma. The tube is intended to prepare plasma for molecular diagnostic techniques such as PCR. Publications from Roche and Becton-Dickinson (7, 8) support freezing the PPT for the transport and storage of specimens intended for viral load determination. These handling techniques were validated for both standard and ultrasensitive assays in our laboratory in 2000 (see Discussion.)
Viral load testing.
The Roche COBAS AMPLICOR HIV-1 Monitor (version 1.5) standard and ultrasensitive assays were carried out according to instructions on the package inserts. Standard and ultrasensitive assays are performed with the same kit but differ in sample preparation. The standard assay begins with untreated plasma, which is lysed, extracted, precipitated, and redissolved. A sample derived from 25 µl of plasma is used in the assay. For the ultrasensitive assay, plasma is ultracentrifuged to sediment the virus. The supernatant plasma is discarded, and the viral pellet is lysed, extracted, precipitated, and redissolved. A sample prepared from 250 µl of original plasma is assayed. The standard assay has a linear range from 400 to 750,000 RNA copies/ml, while the ultrasensitive assay is linear between 50 and 100,000 RNA copies/ml. The standard and ultrasensitive assays are capable of detecting virus below 400 and 50 genome copies per ml, respectively, but the results are nonlinear. The quantity present is reported as <400 or <50, with the designation (D) or (ND) appended to indicate whether viral nucleic acids were detected or not detected, respectively. The data presented in Fig. 1 were transformed to the base 10 logarithm (log10 HIV-1 RNA copies/ml).
Definition of discordant results.
Discordant results were arbitrarily defined to be those in which
the viral loads determined by the standard and ultrasensitive
assays differed by more than 0.5 log
10, or about 3.2-fold, although
the actual difference was generally larger. We chose this number
in part because it represents the difference that most clinicians
assume is clinically significant (
10). At high copy numbers,
the assay variation is approximately 0.2 log
10 (
1,
10). Other
authors have shown that the assay variation is larger at the
lower ends of the linear ranges (
6). The differences between
the lower limits of the quantification ranges of the standard
and ultrasensitive assays introduce unavoidable anomalies into
the evaluation of discordance. A specimen with a standard result
of 400 and an ultrasensitive result of <50 is discordant
(minimum difference of eightfold). For a sample with a standard
result of <400 and an ultrasensitive result of <50, concordance
cannot be excluded. The difference in the detection limits may
cause an unavoidable underestimate of discordance.
Statistical analysis.
Data were analyzed using SPSS for Windows v. 11.0 (SPSS, Inc., Chicago, Ill.). Fisher's exact tests were used to compare discordance rates in specimens from children versus rates in specimens from adults for each ultrasensitive range and for combined ranges.

RESULTS
Comparison of ultrasensitive and standard viral loads.
Viral load results from both standard and ultrasensitive assays
were available for 1,265 specimens (including the 814 presented
in Fig.
1). Of these, 230 (18.2%) were discordant. Two hundred
eighteen discordant specimens were sent to our reference laboratory
for repeat testing, and both standard and ultrasensitive results
were received for 179 specimens. One hundred eleven specimens
(62.0%) remained discordant, while 68 (38.0%) became concordant
due to a decrease in standard results. The number of discordant
specimens per run (22 specimens) varied between 0 and 13.
Figure 1 shows a plot of log10-transformed standard and ultrasensitive results from consecutive adult and pediatric patient specimens during the first 3 months of the study. These specimens are included in the 1,265 discussed above. One hundred forty-five specimens were included from 96 HIV-infected pediatric patients (only the first two specimens/patient are included), aged between 3 months and 17 years, including 38 females. The overall discordance rate was 49.0%; for the 135 specimens with ultrasensitive viral loads less than 10,000 RNA copies/ml, it was 51.9%. No pediatric specimen had an ultrasensitive viral load greater than 100,000 RNA copies/ml. Figure 1 also shows 669 results from 572 adult patients (five patients have three specimens each), ages 18 to 77, including 198 females. The overall discordance rate was 20.1%, and the rate among specimens with ultrasensitive viral loads of
10,000 RNA copies/ml was 25.3%. Fifty-seven adult specimens (9%) exceeded 100,000 RNA copies/ml. Even in concordant specimens, where the difference between the standard and ultrasensitive viral loads was <0.5 log, the viral load result determined by the standard assay was generally the higher.
The percentage of specimens with discordant results was higher from children than from adults. To determine whether the differences in percentage of discordant specimens were significant for children versus adults, the data from Fig. 1 were stratified by ultrasensitive viral load and are shown in Table 1. The difference in proportion of discordant specimens from children versus from adults is significant for all specimens and for those with viral loads of <1,000, both stratified and aggregated as shown in the table. For viral loads between 1,000 and 9,500, the proportion of discordant specimens is higher for children (22.2%) than for adults (11.1%); however, there are so few pediatric patients with viral loads in this range that we do not have adequate power to demonstrate that the pediatric and adult discordance rates differ significantly. Causes of the difference in discordance rates between children and adults are under investigation.
View this table:
[in this window]
[in a new window]
|
TABLE 1. Difference in percentage of discordant HIV viral load determinations from children versus from adults, stratified by ultrasensitive assay results
|
B-DNA results.
When the problem with discordant results was first detected,
we did not know whether the standard or ultrasensitive results
(or neither) were correct. Fifty-six discordant specimens with
adequate remaining plasma were sent to a reference laboratory
for viral load determinations by the B-DNA technique. Of these,
49 were concordant with the ultrasensitive results (using the
same 0.5 log
10 definition applied to comparisons between standard
and ultrasensitive results). Of seven (12.5%) specimens with
discordance between ultrasensitive and B-DNA results, only three
ultrasensitive results exceeded the B-DNA results by more than
fivefold. No B-DNA viral load was higher than its corresponding
ultrasensitive result. Thus, no B-DNA result could have been
concordant with the standard result on the same specimen. This
information suggests that the results of the ultrasensitive
assay are generally reliable (data not shown). Additional B-DNA
results are shown in Table
2.
Effect of handling of PPT on viral load results.
Several factors made us suspect that freezing PPT might cause
the discordant results. Among these were the nearly contemporaneous
reports that the PPT could not be used for certain hepatitis
assays, a suggestion from Cheryl Jennings of Rush Medical Center
and the Viral Quality Assurance Laboratory, and information
from Roche that they could reproduce our discrepant results
with our specimens but not with those from a collaborating laboratory.
To test our hypothesis, we drew two blood specimens each from
13 patients for viral load testing: one specimen for testing
with an EDTA tube and one for testing with a PPT. Within 2 h
of collection, the EDTA specimen was centrifuged for 20 min
at 1,100
x g. The plasma was frozen in two aliquots, one of
which was sent to a reference laboratory for B-DNA testing.
The PPT was centrifuged according to our usual protocol. Plasma
(850 µl) was removed and frozen in a transfer tube; the
remainder was frozen in situ in the PPT. On the day of testing,
each of the three specimens was thawed at room temperature and
tested by the standard and ultrasensitive assays. The results
are shown in Table
2. Columns 2 to 4 show the results obtained
with plasma from the lavender-topped EDTA tube, including the
B-DNA results. Columns 5 and 6 show the results from the plasma
aliquot frozen in a transfer tube after removal from the centrifuged
PPT. The results from the PPT plasma frozen in the aliquot tube
are concordant with each other, with the B-DNA results, and
with the lavender-topped EDTA results. Twelve of the 13 specimens
exhibited viral loads below the linear range of all assays.
Columns 7 and 8 show the standard and ultrasensitive results
obtained from the plasma frozen in situ in the PPT. Eight of
the 13 viral loads determined from plasma frozen in the PPT
were discordant (denoted by bold type). For seven of these results,
the standard viral load increased from <400 (none detected)
to between 450 and 3,500 RNA copies/ml. In addition, three specimens
with ultrasensitive results below the linear range in the first
two test conditions were elevated to within the linear range
(65, 80, and 600 RNA copies/ml) after being frozen in situ.
Sample no. 13 had results of 600 and 5,500 RNA copies/ml in
the ultrasensitive and standard assays, respectively, when assays
were performed on plasma frozen in situ; the other assays done
with this specimen detected no viral RNAs. Freezing patient
plasma in situ in PPT can cause elevations in both the standard
and ultrasensitive viral loads determined for that plasma. Increases
in viral loads caused by freezing plasma in PPT are more numerous
and larger in the standard assay, but the change in ultrasensitive
results may also be clinically relevant.

DISCUSSION
Freezing plasma samples in the PPT in which they are collected
elevates artifactually the results of the standard viral load
assay but has less effect on the ultrasensitive results. At
low viral loads, the standard results may exceed the ultrasensitive
results by 100-fold or more. Overall, about 18% of the specimens
received during this study were discordant, and specimens from
children had a significantly higher rate of discordant results
than those from adults. We do not know why only some specimens
were discordant or why pediatric patients differed significantly
from the adult patients in the proportion of discordant specimens.
These questions are under investigation.
Whatever unknown factors contribute to the discordant results, the discrepancies can be eliminated if plasma specimens collected in PPT are transferred to a second tube after centrifugation but prior to freezing. We assume that freezing the PPT releases nonparticulate HIV-1 genetic material from cells in or below the separator gel into the overlying plasma. Since the viral load assay employs reverse transcriptase PCR technology, it can amplify any virus-specific template, either DNA or RNA, that may be present. A recent study by the Viral Quality Assurance Laboratory (NO1-AI-85354) determined that a single copy of HIV-1-specific proviral DNA could increase the detected viral load by 100 to 200 RNA copies (Cheryl Jennings [Rush Medical Center], personal communication).
For the standard assay, nucleic acids are extracted directly from plasma samples, so any contaminating genetic material would be amplified. For the ultrasensitive assay, virions are sedimented from the plasma prior to nucleic acid extraction, apparently leaving the contaminating material in the supernatant fluid, which is discarded. If the amount of HIV genetic material released into the plasma were especially large, the viral load determined by the ultrasensitive assay could be falsely elevated by free HIV genetic material remaining in the small amount of plasma contaminating the viral pellet. This hypothesis may explain the higher ultrasensitive viral load results shown in Table 2, column 7 (frozen in the PPT) compared to column 5 (frozen in a transfer tube) and to the B-DNA results in column 4. It also could explain why 12.5% of the B-DNA results obtained for discordant specimens were also discordant with the ultrasensitive results. Thus, freezing plasma in the PPT is not acceptable, even if the ultrasensitive HIV-1 viral load assay is planned. Other results, not presented here, suggest that if specimens in PPT are markedly delayed in transit or otherwise mishandled, the standard assay results are similarly compromised relative to those of the ultrasensitive assay. With the agreement of our physicians, we have discontinued the standard assay and offer only the ultrasensitive assay.
One published paper (12), one recent poster (R. Murphy, B. Berzins, A. Leake, M. Till, V. Stosor, J. Stanton, and F. Palella, 12th Conference on Retroviruses and Opportunistic Infections, Boston, Mass., no. 738, 22 to 25 February 2005), and anecdotal reports indicate that others have noted the discrepancy between results obtained in EDTA and PPT but did not determine the cause. Our laboratory had some previous indications that viral load determinations were not accurate: failure to amplify viral RNA in the HIV genotyping assay sometimes occurred despite detected viral loads ranging from 1,000 to 5,000 copies/ml. Since we changed our specimen processing protocol for the viral load determinations, these genotyping failures no longer occur.
Previous papers showing that viral loads were not affected by freezing the PPT (7, 8) made PPT the collection method of choice for many laboratories. Since PPT became available prior to highly active antiretroviral therapy, most specimens used in the original validations (7, 8), and in our own as well, had viral loads greater than 10,000 genome copies per ml. The numeric difference between the discordant standard and ultrasensitive viral loads usually ranged from 350 to 10,000 RNA copies/ml. Only when highly effective antiretroviral therapy substantially reduced the viral load did a discrepancy of several hundred or even several thousand RNA copies/ml become meaningful.

ACKNOWLEDGMENTS
We thank the technologists of the DMC University Laboratories
Molecular Microbiology Laboratory for their patience, understanding,
and support during the months of duplicate testing required
until we eliminated the cause of the discrepant results. We
thank Michael Kruger for performing the statistical analysis.
Roche Molecular Diagnostics paid for all duplicate testing without restriction on test type or location.

FOOTNOTES
* Corresponding author. Mailing address: DMC University Laboratories, 4201 St. Antoine, Detroit, MI 48201. Phone: (313) 745-4609. Fax: (313) 993 8754. E-mail:
hsalimni{at}dmc.org.


REFERENCES
1 - Brambilla, D., P. S. Reichelderfer, J. W. Bremer, D. E. Shapiro, R. C. Hershow, D. A. Katzenstein, S. M. Hammer, B. Jackson, A. C. Collier, R. S. Sperling, M. G. Fowler, R. W. Coombs, et al. 1999. The contribution of assay variation and biological variation to the total variability of plasma HIV-1 RNA measurements. AIDS 13:2269-2279.[CrossRef][Medline]
2 - Brambilla, D. J., C. Jennings, R. Morack, S. Granger, and J. W. Bremer. 2004. Comparison of the sensitivities of the version 1.5 and version 1.0 ultrasensitive Roche AMPLICOR HIV-1 MONITOR kits at low concentrations of human immunodeficiency virus RNA. J. Clin. Microbiol. 42:2819-2820.[Abstract/Free Full Text]
3 - Brambilla, D. J., S. Granger, C. Jennings, and J. W. Bremer. 2001. Multisite comparison of reproducibility and recovery from the standard and ultrasensitive Roche AMPLICOR HIV-1 MONITOR assays. J. Clin. Microbiol. 39:1121-1123.[Abstract/Free Full Text]
4 - Coombs, R. W., S. L. Welles, C. Hooper, P. S. Reichelderfer, R. T. D'Aquila, A. J. Japour, V. A. Johnson, D. R. Kuritzkes, D. D. Richman, S. Kwok, J. Todd, J. B. Jackson, V. DeGruttola, C. S. Crumpacker, J. Kahn, et al. 1996. Association of plasma human immunodeficiency virus type 1 RNA level with risk of clinical progression in patients with advanced infection. J. Infect. Dis. 174:704-712.[Medline]
5 - Erali, M., and D. R. Hillyard. 1999. Evaluation of the ultrasensitive Roche Amplicor HIV-1 Monitor assay for quantitation of human immunodeficiency virus type 1 RNA. J. Clin. Microbiol. 37:792-795.[Abstract/Free Full Text]
6 - Hill, C. E., A. M. Green, J. Ingersoll, K. A. Easley, F. S. Nolte, and A. M. Caliendo. 2004. Assessment of agreement between the AMPLICOR HIV-1 MONITOR test versions 1.0 and 1.5. J. Clin. Microbiol. 42:286-289.[Abstract/Free Full Text]
7 - Holodniy, M., L. Mole, B. Yen-Lieberman, D. Margolis, C. Starkey, R. Carroll, T. Spahlinger, J. Todd, and J. B. Jackson. 1995. Comparative stabilities of quantitative human immunodeficiency virus RNA in plasma from samples collected in VACUTAINER CPT, VACUTAINER PPT, and standard VACUTAINER tubes. J. Clin. Microbiol. 33:1562-1566.[Abstract]
8 - Holodniy, M., L. Rainen, S. Herman, and B. Yen-Lieberman. 2000. Stability of plasma human immunodeficiency virus viral load in VACUTAINER PPT plasma preparation tubes during overnight shipment. J. Clin. Microbiol. 38:323-326.[Abstract/Free Full Text]
9 - Jackson, L. B., E. Piwowar-Manning, L. Johnson-Lewis, R. Bassett, L. M. Demeter, and D. Brambilla. 2004. Comparisons of versions 1.0 and 1.5 of the UltraSensitive AMPLICOR HIV-1 MONITOR test for subjects with low viral load. J. Clin. Microbiol. 42:2774-2776.[Abstract/Free Full Text]
10 - NCCLS. 2003. Quantitative molecular methods for infectious diseases, p. 19-20. Approved standard MM6-A (ISBN 1-56238-508.9). NCCLS, Wayne, Pa.
11 - O'Brien, W. A., P. M. Hartigan, D. Martin, J. Esinhart, A. Hill, S. Benoit, M. Rubin, M. S. Simberkoff, and J. D. Hamilton. 1996. Changes in plasma HIV-1 RNA and CD4+ lymphocyte counts and the risk of progression to AIDS. N. Engl. J. Med. 334:426-431.[Abstract/Free Full Text]
12 - Squires, K., A. Lazzarin, J. M. Gatell, W. G. Powderly, V. Pokrovskiy, J.-F. Delfaissy, J. Jemsek, A. Rivero, W. Rozenbaum, S. Schrader, M. Sension, A. Vibhagool, A. Thiry, and M. Giordano. 2004. Comparison of once-daily atazanavir with efavirenz, each in combination with fixed-dose zidovudine and lamivudine, as initial therapy for patients infected with HIV. AIDS 36:1011-1019.
13 - Welles, S. L., J. B. Jackson, B. Yen-Lieberman, L. Demeter, A. J. Japour, L. M. Smeaton, V. A. Johnson, D. R. Kuritzkes, R. T. D'Aquila, P. A. Reichelderfer, D. D. Richman, R. Reichman, M. Fischl, R. Dolin, R. W. Coombs, J. O. Kahn, C. McLaren, J. Todd, S. Kwok, C. S. Crumpacker, et al. 1996. Prognostic value of plasma HIV-1 RNA levels in patients with advanced HIV-1 disease and with little or no zidovudine therapy. J. Infect. Dis. 174:696-703.[Medline]
Journal of Clinical Microbiology, September 2005, p. 4635-4639, Vol. 43, No. 9
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.9.4635-4639.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Kran, A.-M. B., Jonassen, T. O., Sannes, M., Jakobsen, K., Lind, A., Maeland, A., Holberg-Petersen, M.
(2009). Overestimation of Human Immunodeficiency Virus Type 1 Load Caused by the Presence of Cells in Plasma from Plasma Preparation Tubes. J. Clin. Microbiol.
47: 2170-2174
[Abstract]
[Full Text]
-
Marinovich, A., Narlieva, M., Bashir, M., Zingman, B. S., Fox, A.
(2006). False-Positive Result from a Bayer Versant Human Immunodeficiency Virus Type 1 Branched-DNA Viral Load Assay, with a Possible Role for Light Leakage after Inadequate Maintenance of the Analyzer. J. Clin. Microbiol.
44: 4288-4289
[Full Text]