Previous Article | Next Article ![]()
Journal of Clinical Microbiology, November 2006, p. 4288-4289, Vol. 44, No. 11
0095-1137/06/$08.00+0 doi:10.1128/JCM.01298-06
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
| LETTER TO THE EDITOR |
|
|
|---|
The patient was a 28-year-old male who was referred for HIV testing after a needle stick injury from a needle contaminated by blood from a newly diagnosed HIV-positive patient. He was given postexposure prophylaxis consisting of zidovudine, lamivudine, and nelfinavir for 4 weeks. Baseline and 6-week HIV-1/2 antibody enzyme immunoassay (EIA) testing produced negative results. An HIV viral load assay was ordered at the same time as the 6-week EIA, for which the patient's blood was collected in a Greiner K2 EDTA tube. This tube was centrifuged and then frozen at 70°C within 65 min of collection. The following day, the Versant HIV-1 RNA 3.0 bDNA assay was performed according to the manufacturer's instructions, using the Bayer System 340 analyzer, and produced a result of 955 copies/ml. No error code was generated by the analyzer during this test.
Immediately after this test was performed, a Bayer technician examined the analyzer by prior arrangement (see below). It was discovered that the read head of the instrument was not adequately compressing the tops of the sample wells, potentially resulting in light leakage between wells. Maintenance was performed to correct this problem. A repeat viral load bDNA assay on the same serum 2 days later produced a result of <75 copies/ml. Additional EIAs performed 3 days and 45 days after the initial viral load assay were repeatedly negative (7 and 12 weeks following the needle stick exposure). An additional viral load assay using new serum collected 2 days after the first viral load also produced a result of <75 copies/ml. In the 4.5 months since the needle stick exposure, the patient continues to be asymptomatic, with negative HIV testing, and is presumed to be HIV negative.
For several weeks prior to this test, the analyzer generated intermittent error reports due to abnormally high luminescence intensities for certain standard specimens. The analyzer did not indicate problems with the read head during our regularly scheduled operator-performed maintenance. We notified the manufacturer, and a maintenance visit was arranged, as described above. Meanwhile, we sought to determine whether light leakage could be occurring by placing an empty strip of sample wells just below the strip of wells containing the standard specimens. Table 1 shows the luminescence results for the standard-specimen and empty wells in the first and second rows under the before-maintenance heading, respectively. The third row contains the luminescence results for routine-patient samples in the strip immediately below the empty strip. There was a strong correlation between the luminescence intensities of the empty wells and those of the adjacent wells immediately above, below, and diagonal to each empty well (P < 0.01). It is notable that the sample well immediately below our patient's initial (presumed false-positive) well had a very high luminescence intensity, corresponding to a viral load of >500,000 copies/ml (data not shown).
|
View this table: [in a new window] |
TABLE 1. Luminescence intensities
|
In summary, we present a case of a false-positive HIV result obtained using the Versant HIV-1 bDNA viral load assay, apparently caused by inadequate maintenance resulting in light leakage between sample wells during luminescence reading by a Bayer System 340 analyzer. While we present one way that a false-positive result may be generated by a viral load assay, it is important to note that other false positives previously reported with multiple assay types show that there can be diverse causes. Clinicians using HIV viral load assays for HIV diagnosis must be cognizant of the potential for false-positive results and prepare their patients for this possibility. In addition, the phenomenon we report here could conceivably cause known HIV-positive patients' viral loads to be erroneously reported above certain critical values, indicating treatment failure where none may be occurring and leading to unwarranted changes in clinical management. This case underscores the need for the proper maintenance of the apparatus used in a chemiluminescent assay to ensure that light leakage is not occurring between samples. Even isolated deviations of the luminescence intensities of standard specimens should prompt a thorough evaluation of the assay, and preventive maintenance of the analyzer, before further clinical use of the assay. Future design modifications of these assays may need to incorporate improved detection of light leakage.
Published ahead of print on 27 September 2006. |
|
|---|
|
Adrian Marinovich Momka Narlieva Mohammed Bashir Barry S. Zingman Amy Fox* Montefiore Medical Center Virology 111 East 210 Street Bronx, NY 10467
| ||||||
| * Phone: (718) 920-5567, Fax: (718) 920-2854, E-mail: afox{at}montefiore.org |
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2010 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»