Multicenter Performance Evaluation of a New TaqMan PCR Assay for Monitoring Human Immunodeficiency Virus RNA Load
L. Perrin,1
J. M. Pawlotsky,2
M. Bouvier-Alias,2
C. Sarrazin,3
S. Zeuzem,3 and
G. Colucci4*
Clinical Virology, University of Genève, Genève, Switzerland,1
Department of Bacteriology and Virology, Hôpital Henri Mondor, Créteil, France,2
Department of Internal Medicine, University Hospital, Homburg-Saar, Homburg, Germany,3
Roche Molecular Diagnostics, Rotkreuz, Switzerland4
Received 8 March 2006/
Returned for modification 30 August 2006/
Accepted 12 September 2006

ABSTRACT
A TaqMan real-time PCR assay, the COBAS TaqMan human immunodeficiency
virus (HIV) (HPS-CTMHIV) PCR assay, recently developed for the
quantification of HIV type 1 RNA in plasma, was evaluated in
comparison with the licensed COBAS AMPLICOR HIV-1 MONITOR (CAHIM)
assay. In this study, we have analyzed the tests' sensitivities,
precisions, and linearities using multiple replicates of a standard
panel of HIV RNA covering a 7-logarithm range of concentrations,
as well as serial threefold dilutions of high-titer clinical
samples. The subtype inclusivity was also investigated, using
a panel of subtypes A to H, while a collection of 160 clinical
samples was analyzed to assess the tests' specificities and
the systems' similarities. The results of these experiments
showed that the HPS-CTMHIV assay has a sensitivity of 53 copies/ml
(95% hit rate), 100% specificity, and good intra- and interassay
precision. The results of the HPS-CTMHIV assay were linear in
the 50- to 10
7-copies/ml range, with a correlation coefficient
(
R) for expected versus observed results of 0.98. Compared to
the CAHIM assay, the HPS-CTMHIV assay showed a high correlation
(
R = 0.99) across the dynamic range of RNA concentrations that,
for the CAHIM assay, requires two different sample preparations.
Equivalent performances were also observed for the two systems
in the detection and quantification of HIV subtypes A to H.
These data indicate that the HPS-CTMHIV assay may be one of
the tests of choice for monitoring viral load throughout the
course of HIV infection and during highly active antiretroviral
therapy.

INTRODUCTION
The quantification of human immunodeficiency virus type 1 (HIV-1)
RNA in plasma has revolutionized the management of infected
patients. This parameter is the main prognostic factor for evaluating
the progression of the disease and for timing the initiation
of treatment and monitoring its efficacy (
1,
5,
7,
10,
11,
14,
15,
16). Commercially available assays for the quantification
of HIV-1 viral RNA use a variety of techniques, from reverse
transcriptase PCR to the branched-DNA assay (
2,
12,
20,
21).
The COBAS AMPLICOR HIV-1 MONITOR (CAHIM) test v. 1.5, a quantitative reverse transcriptase PCR assay (3), has been widely used in research as well as in clinical practice. Quantification of HIV-1 RNA with this assay requires two sample preparation procedures: in the first, ultrasensitive method, 500 µl of plasma is centrifuged to increase the input of plasma RNA to achieve a quantitative range of 50 to 2,000 copies/ml of plasma, whereas in the second, standard protocol, with a starting plasma volume of 200 microliters, only an amount of RNA corresponding to 25 microliters of plasma is introduced in the PCR, resulting in a dynamic range of 500 to 500,000 copies/ml. A single-step technique with a wider dynamic range is therefore desirable for routine laboratory applications. The real-time PCR technology already developed for several noncommercial viral assays offers not only broader quantification ranges but also faster turnaround time and more effective prevention of contamination (due to a closed-tube configuration) (4, 13, 23).
In this investigation, we compared the linearity, reproducibility, and sensitivity of the COBAS AMPLICOR HIV-1 MONITOR test v. 1.5 with those of a newly developed real-time PCR assay, the COBAS TaqMan HIV (HPS-CTMHIV) assay, which has a semiautomated format and dedicated hardware and software.

MATERIALS AND METHODS
In this multicenter trial, we evaluated the performance characteristics
of the HPS-CTMHIV PCR assay, a semiautomated real-time PCR test
based on a dual-labeled hybridization probe targeting the
gag region. After manual HIV RNA extraction using the High Pure
kit (Roche Diagnostics, Mannheim, Germany), which consists of
subsequent steps of lysis and glass particle RNA capture and
purification, the TaqMan assay is performed using a dedicated
COBAS TaqMan 48 instrument and the results are expressed in
numbers of HIV RNA copies/ml. The prevention of carryover contamination
and the integrity of the sample are ensured by the use of AmpErase
and by an internal control (quantitation standard) (
8,
17).
All experiments were performed at sites 1, 2, and 3. The sensitivity, specificity, linearity, and subtype performances were analyzed in comparison with those of the CAHIM, a quantitative PCR assay licensed for diagnostic use (3). The HPS-CTMHIV assay and the CAHIM assay were used in accordance with the manufacturer's instructions included in the test package inserts.
Precision, sensitivity, and specificity.
Intra- and interassay reproducibilities were investigated using an eight-member standard HIV RNA panel with titers ranging from 30 to 5 x 106 copies/ml (AcroMetrix, Benicia, CA). Six replicates of each standard panel member were analyzed on three consecutive days. These experiments were also used to estimate the assay's limit of detection, calculated by probit analysis, while the specificity was assessed by testing 100 samples obtained from healthy donors or patients with HIV-unrelated diseases.
Linearity and dynamic range.
Linearity and dynamic range were investigated using 13 threefold serial dilutions and an undiluted sample of a 5 x 106-copies/ml concentration of the AcroMetrix HIV DNA panel member as well as of a high-titer clinical specimen selected at the trial sites. Each dilution was run in triplicate.
Subtype identification.
The HPS-CTMHIV assay's ability to detect and quantify different subtypes with the same efficiency was analyzed using a standard panel consisting of eight members, representing the group M clade, subtypes A to H, tested in triplicate on three consecutive days (HIV-1 RNA clade performance panel PRD201; Boston Biomedica Inc.).
Comparison of assays.
Parallel experiments were performed at one of the trial sites (site 1) with the CAHIM assay, a quantitative PCR assay with a 4-log dynamic range of concentrations of 50 to 750,000 copies/ml obtained by combining two different sample preparation procedures, the standard and the ultrasensitive (21). In addition, to compare the main performance characteristics of the two systems, we also analyzed their relative performances in quantifying 60 clinical samples taken from patients with HIV infection and different viral loads.
Both the HPS-CTMHIV assay and the CAHIM assay have been approved by the European Union for in vitro diagnostic applications (CE mark) and are being reviewed by the FDA.

RESULTS
In this study, we report the results of a three-site multicenter
evaluation of the HPS-CTMHIV assay, a real-time PCR system specifically
developed to improve the ease of use, turnaround time, and overall
efficacy of current quantitative PCR assays.
Using an HIV RNA standard panel, we investigated the HPS-CTMHIV assay's precision and detection limit across a range of concentrations from 10 to 107 copies/ml. In several experiments, run on three consecutive days with six replicates of each panel member, we observed a mean total precision (intra- and interassay reproducibility) of 56% (31% to 81% coefficient of variation) and a 95% detection limit of 53 copies/ml as estimated by probit analysis (Fig. 1 and Tables 1 and 2). The assay results were linear over a 5-log range from 10 to 107 copies/ml, and the correlation between the expected and observed results was very high, with an R of 0.98 (Fig. 2). Similar findings were made when data obtained with the HPS-CTMHIV assay were compared to those produced with the CAHIM assay using both the standard and the ultrasensitive procedure. Samples with titers higher than 105 copies/ml had to be further diluted, as they were above the upper limit of the CAHIM assay's reportable range (Fig. 3a). An excellent correlation between the two tests was also noted when we analyzed 72 clinical samples taken from patients with chronic HIV infection and a wide range of viremia levels (Fig. 3b).
We also used a collection of 100 samples obtained from healthy
donors or patients with HIV-unrelated diseases to assess the
HPS-CTMHIV assay's specificity. No HIV RNA was detected by either
the HPS-CTMHIV or the CAHIM assay, with 100% specificity (data
not shown). All major M clade subtypes, including subtypes A
to H, were accurately detected and quantified when analyzed
at two different concentrations, 300 and 3,000 copies/ml (Fig.
4a and b).

DISCUSSION
The HPS-CTMHIV test has been developed to standardize the real-time
technique based on TaqMan instruments and to improve the performance
of the existing PCR assays (
13). Based on this evaluation, the
HPS-CTMHIV assay appears to have achieved these goals, confirming
its expected wide dynamic range and increased sensitivity. Our
data indeed show that the assay can quantify concentrations
ranging from 10 to 10
7 copies/ml, thus including and extending
the range obtainable by the CAHIM test, the FDA- and CE-approved
test we used for system comparison. However, the CAHIM assay
makes use of two different sample preparation procedures, the
standard and the ultrasensitive. Specimens that are negative
by the former, at a detection limit of 400 copies/ml, have to
be reevaluated using the latter, which allows for a range of
50 to 5
x 10
4 copies/ml (
21). Although laboratory managers may
decide beforehand which preparation to use based on the expected
HIV titers of the sample, the possibility of avoiding repeat
testing by using the HPS-CTMHIV test instead of the CAHIM test
represents a significant advantage in terms of cost effectiveness
and workflow.
The HPS-CTMHIV assay's precision, as assessed by intra- and interassay reproducibility, is good and is comparable to that of the CAHIM assay. Further improvement is expected when the RNA extraction procedure, presently labor-intensive and lengthy, is automated (6). Medium-sized to large laboratories, with high sample turnover, may indeed benefit from fully automated sample preparation using dedicated instrumentation (6, 18). In this respect, the COBAS AmpliPrep instrument, a nucleic acid extractor already in use in conjunction with the CAHIM (COBAS AmpliPrep/COBAS AMPLICOR HIV MONITOR) reagents, is being validated as the front-end instrument for the HPS-CTMHIV test (19).
The issue of potential carryover has already been solved for the CAHIM assay through pretreatment with uracyl-N-glycosylase, the enzyme that specifically inactivates dUTP-containing amplicons. This step is also included in the HPS-CTMHIV assay, with additional safety derived from the closed-tube procedure typical of the hydrolysis probe technology used with TaqMan instruments.
This investigation was not designed to evaluate the efficacy of HIV-1 RNA quantification for the various non-B subtypes, since only two dilutions of a single strain corresponding to the most frequent HIV-1 subtypes were tested. However, in this study we show that both tests are able to detect the subtypes tested at low HIV RNA levels (300 copies/ml) and provide similar results at the higher titers tested (3,000 copies/ml). Additional studies of larger numbers of clinical samples are warranted to confirm these data and to assess the ability of the HPS-CTMHIV assay to detect newly emerging strains and recombinants (9, 22).
In conclusion, we have shown that the HPS-CTMHIV test provides performances similar to those of the licensed CAHIM test with regard to specificity, sensitivity, and precision and has additional advantages in terms of a larger dynamic range, high throughput, and absence of carryover during the PCR amplification procedure.

ACKNOWLEDGMENTS
COBAS, TaqMan, AmpliPrep, AMPLICOR, High Pure, and AmpErase
are trademarks of Roche.

FOOTNOTES
* Corresponding author. Mailing address: Roche Molecular Diagnostics, Roche Instrument Center AG, Lettenstrasse, 6343 Rotkreuz, Switzerland. Phone: 41 41 7992815. Fax: 41 41 7992845. E-mail:
giuseppe.colucci.gc1{at}roche.com.

Published ahead of print on 27 September 2006. 

REFERENCES
1 - Coffin, J. M. 1995. HIV population dynamics in vivo: implications for genetic variation, pathogenesis, and therapy. Science 267:483-489.[Abstract/Free Full Text]
2 - Collins, M. L., B. Irvine, D. Tyner, E. Fine, C. Zayati, C. Chang, T. Horn, D. Ahle, J. Detmer, L. P. Shen, J. Kolberg, S. Bushnell, M. S. Urdea, and D. D. Ho. 1997. A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml. Nucleic Acids Res. 25:2979-2984.[Abstract/Free Full Text]
3 - Hill, C. E., A. M. Green, J. Ingersoll, K. A. Easley, F. S. Nolte, and A. M. Caliendo. 2004. Assessment of agreement between the AMPLICOR HIV-1 MONITOR test versions 1.0 and 1.5. J. Clin. Microbiol. 42:286-289.[Abstract/Free Full Text]
4 - Holland, P. 1991. Detection of specific polymerase chain reaction product by utilizing the 5'-3' exonuclease activity of Thermus aquaticus DNA polymerase. Proc. Natl. Acad. Sci. USA 88:7276-7280.[Abstract/Free Full Text]
5 - Hughes, M. D., V. A. Johnson, M. S. Hirsch, J. W. Bremer, T. Elbeik, A. Erice, D. R. Kuritzkes, W. A. Scott, S. A. Spector, N. Basgoz, M. A. Fischl, R. T. D'Aquila, et al. 1997. Monitoring plasma HIV-1 RNA levels in addition to CD4+ lymphocyte count improves assessment of antiretroviral therapeutic response. Ann. Intern. Med. 126:929-938.[Abstract/Free Full Text]
6 - Jungkind, D. 2001. Automation of laboratory testing for infectious diseases using the polymerase chain reactionour past, our present, our future. J. Clin. Virol. 20:1-6.[CrossRef][Medline]
7 - Katzenstein, D. A., S. M. Hammer, M. D. Hughes, et al. 1996. The relation of virologic and immunologic markers to clinical outcomes after nucleoside therapy in HIV-infected adults with 200 to 500 CD4 cells per cubic millimeter. N. Engl. J. Med. 335:1091-1098.[Abstract/Free Full Text]
8 - Longo, M. C., M. S. Berninger, and H. L. Hartley. 1990. Use of uracil DNA glycosylase to control carry-over contamination in polymerase chain reactions. Gene 93:125-128.[CrossRef][Medline]
9 - McCutchan, F. E., J. L. Sankale, S. M'Boup, B. Kim, S. Tovanabutra, D. J. Hamel, S. K. Brodine, P. J. Kanki, and D. L. Birx. 2004. HIV type 1 circulating recombinant form CRF09_cpx from west Africa combines subtypes A, F, G, and may share ancestors with CRF02_AG and Z321. AIDS Res. Hum. Retrovir. 20:819-826.[CrossRef][Medline]
10 - Mellors, J. W., A. Munoz, J. V. Giorgi, J. B. Margolick, C. J. Tassoni, P. Gupta, L. A. Kingsley, J. A. Todd, A. J. Saah, R. Detels, J. P. Phair, and C. R. Rinaldo, Jr. 1997. Plasma viral load and CD4+ lymphocytes as prognostic markers of HIV-1 infection. Ann. Intern. Med. 126:946-954.[Abstract/Free Full Text]
11 - Mellors, J. W., C. R. Rinaldo, Jr., P. Gupta, R. M. White, J. A. Todd, and L. A. Kingsley. 1996. Prognosis in HIV-1 infection predicted by the quantity of virus in plasma. Science 272:1167-1170.[Abstract]
12 - Mulder, J., N. McKinney, C. Christopherson, J. Sninsky, L. Greenfield, and S. Kwok. 1994. Rapid and simple PCR assay for quantitation of human immunodeficiency virus type 1 RNA in plasma: application to acute retroviral infection. J. Clin. Microbiol. 32:292-300.[Abstract/Free Full Text]
13 - Niesters, H. B. 2004. Molecular and diagnostic clinical virology in real time. Clin. Microbiol. Infect. 10:5-11.[CrossRef][Medline]
14 - O'Brien, W. A., P. M. Hartigan, E. S. Daar, M. S. Simberkoff, J. D. Hamilton, et al. 1997. Changes in plasma HIV RNA levels and CD4+ lymphocyte counts predict both response to antiretroviral therapy and therapeutic failure. Ann. Intern. Med. 126:939-945.[Abstract/Free Full Text]
15 - O'Brien, W. A., P. M. Hartigan, D. Martin, et al. 1996. Changes in plasma HIV-1 RNA and CD4+ lymphocyte counts and the risk of progression to AIDS. N. Engl. J. Med. 334:426-431.[Abstract/Free Full Text]
16 - Riddler, S. A., and J. W. Mellors. 1997. HIV-1 viral dynamics and viral load measurement: implications for therapy. AIDS Clin. Rev. 1997-1998:47-65.
17 - Rosenstraus, M., Z. Wang, S.-Y. Chang, D. DeBonville, and J. P. Spadoro. 1998. An internal control for routine diagnostic PCR: design, properties, and effect on clinical performance. J. Clin. Microbiol. 36:191-197.[Abstract/Free Full Text]
18 - Schorling, S., G. Schalasta, G. Enders, and M. Zauke. 2004. Quantification of parvovirus B19 DNA using COBAS AmpliPrep automated sample preparation and LightCycler real-time PCR. J. Mol. Diagn. 6:37-41.[Abstract/Free Full Text]
19 - Stelzl, E., A. Kormann-Klement, J. Haas, E. Daghofer, B. I. Santner, E. Marth, and H. H. Kessler. 2002. Evaluation of an automated sample preparation protocol for quantitative detection of hepatitis C virus RNA. J. Clin. Microbiol. 40:1447-1450.[Abstract/Free Full Text]
20 - Stevens, W., T. Wiggill, P. Horsfield, L. Coetzee, and L. E. Scott. 2005. Evaluation of the NucliSens EasyQ assay in HIV-1-infected individuals in South Africa. J. Virol. Methods 124:105-110.[CrossRef][Medline]
21 - Sun, R., J. Ku, H. Jayakar, J.-C. Kuo, D. Brambilla, S. Herman, M. Rosenstraus, and J. Spadoro. 1998. Ultrasensitive reverse transcription-PCR assay for quantitation of human immunodeficiency virus type 1 RNA in plasma. J. Clin. Microbiol. 36:2964-2969.[Abstract/Free Full Text]
22 - Tong, C. Y. W., J. Mullen, R. Kulasegaram, A. De Ruiter, S. O'Shea, and I. L. Chrystie. 2005. Genotyping of B and non-B subtypes of human immunodeficiency virus type 1. J. Clin. Microbiol. 43:4623-4627.[Abstract/Free Full Text]
23 - Weiss, J., H. Wu, B. Farrenkopf, T. Schultz, G. Song, S. Shah, and J. Siegel. 2004. Real time TaqMan PCR detection and quantitation of HBV genotypes A-G with the use of an internal quantitation standard. J. Clin. Virol. 30:86-93.[CrossRef][Medline]
Journal of Clinical Microbiology, December 2006, p. 4371-4375, Vol. 44, No. 12
0095-1137/06/$08.00+0 doi:10.1128/JCM.00505-06
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Scott, L. E., Noble, L. D., Moloi, J., Erasmus, L., Venter, W. D. F., Stevens, W.
(2009). Evaluation of the Abbott m2000 RealTime Human Immunodeficiency Virus Type 1 (HIV-1) Assay for HIV Load Monitoring in South Africa Compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 Assays. J. Clin. Microbiol.
47: 2209-2217
[Abstract]
[Full Text]
-
Oliver, A. R., Pereira, S. F., Clark, D. A.
(2007). Comparative Evaluation of the Automated Roche TaqMan Real-Time Quantitative Human Immunodeficiency Virus Type 1 RNA PCR Assay and the Roche AMPLICOR Version 1.5 Conventional PCR Assay. J. Clin. Microbiol.
45: 3616-3619
[Abstract]
[Full Text]