Journal of Clinical Microbiology, December 2006, p. 4395-4399, Vol. 44, No. 12
0095-1137/06/$08.00+0 doi:10.1128/JCM.01411-06
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
National Center for Antimicrobials and Infection Control, Statens Serum Institut, Copenhagen, Denmark,1 Department of Clinical Microbiology, Central Hospital, Växjö, Sweden,2 AB BIODISK, Solna, Sweden3
Received 8 July 2006/ Returned for modification 29 August 2006/ Accepted 5 October 2006
|
|
|---|
4 mg/liter. Our findings suggest that the current CLSI zone diameter breakpoints should be adjusted from resistance (R)
19 mm to R
21 mm. In conclusion, cefoxitin disk diffusion testing and Etest MIC testing can accurately predict the presence of the mecA gene in S. aureus. Testing can be reliably performed using incubation temperatures of 35 to 36°C and incubation times of 18 to 22 h. We suggest MRSA interpretive criteria of susceptible (S)
4 mg/liter and R > 4 mg/liter, corresponding to S
22 mm and R
21 mm for the 30-µg disk and S
17 mm and R
16 mm for the 10-µg cefoxitin disk. These criteria resulted in only one mecA-positive isolate being misclassified as susceptible. |
|
|---|
Phenotypic detection of MRSA has been problematic ever since its discovery in the early 1960s. The emergence of low-level-resistant MRSA clones acquired in the community has only added to these difficulties.
Detection of the mecA gene or its product, penicillin binding protein (PBP2a), is considered the gold standard (5) for MRSA confirmation. Recent investigations suggest that disk diffusion using cefoxitin is superior to most previously recommended phenotypic methods, including oxacillin disk diffusion and oxacillin screen agar testing (8, 14, 19, 21). In 2005, the Clinical and Laboratory Standards Institute (CLSI) published zone diameter (6) breakpoint guidelines for cefoxitin. However, a number of technical issues remain regarding the use of cefoxitin as a predictor for methicillin resistance. The CLSI M100-S15 document stipulates an incubation time of 24 h unless the isolate has a zone diameter of
19 mm (i.e., resistant), in which case it can be reported after 18 h of incubation. However, in a recent publication, performance was equally good at 18 h of incubation (21).
For methicillin and oxacillin, incubation temperature is known to affect the test results (12, 17). For oxacillin, a maximum of 35°C for testing of staphylococci is specified by CLSI. For cefoxitin, high accuracy has been found using standard incubation temperatures, i.e., 35 to 37°C (8, 10, 19-21), but one study showed relatively low sensitivity at 37°C (2). In a recent investigation, zone diameters obtained on IsoSensitest agar with semiconfluent inoculum were dependent on whether the plates were incubated at 35°C or 36°C (R. Skov, unpublished results). Two studies have shown marginally improved accuracy when tests were performed at 30°C compared to 35°C and 37°C (4, 8). The 10-µg cefoxitin disk has been shown to be superior to the 30-µg disk with IsoSensitest agar and semiconfluent growth (20).
The CLSI M100-S15 document recommends oxacillin MIC testing for MRSA detection by the MIC method, and no criteria are as yet available for the use of cefoxitin as an alternative (6).
In this study, the influence of incubation time (18 h and 24 h) and temperatures (30°C, 35°C, 36°C, and 37°C) on the performance of 10- and 30-µg cefoxitin disks and cefoxitin Etest on Mueller-Hinton agar were evaluated for mecA-positive and mecA-negative S. aureus.
|
|
|---|
One mecA-positive isolate was later excluded from the investigation as population analysis showed it to be a susceptible phenotype (see the population profile analysis), leaving a total of 95 mecA-negative and 50 mecA-positive isolates.
Three reference strains were included for quality control: one mecA-positive strain (S. aureus ATCC 43300) and two mecA-negative strains (S. aureus ATCC 25923 and ATCC 29213).
All strains stored at 80°C were subcultured on two consecutive days using 5% Danish blood agar (SSI Diagnostika) prior to use.
spa typing and MLST. spa typing and multilocus sequence typing (MLST) were performed as previously described (7, 11). The spa types (t) and MLST sequence types (ST) were assigned through the Ridom (http://www.ridom.de) and MLST databases (http://www.mlst.net), respectively. Based on spa and/or sequence types, the isolates were given a predicted clonal complex annotation.
Susceptibility testing. In the first phase of the study (phase I), all 145 strains were tested by disk diffusion using cefoxitin and Etest (AB Biodisk, Solna, Sweden) using cefoxitin and oxacillin. All methodological variants were assessed using the same inoculum. The inoculum was standardized to 0.5 McFarland turbidity. Disk diffusion was done with 10-µg and 30-µg disks (Oxoid, Basingstoke, United Kingdom) using Mueller-Hinton BBL II agar (Becton Dickinson, Heidelberg, Germany). Agar plates were incubated overnight (18 to 19 h) in ambient air at 30°C, 35°C, and 36°C in stacks no higher than five plates.
Inhibition zone diameters were read from the back of the agar plate using reflected light and calipers to read to the nearest millimeter at the inner zone edge. For isolates with a zone size of >19 mm after 18 to 19 h for the 30-µg cefoxitin disk at 35°C, all plates were further incubated and read after 24 h as specified in M100-S15 (6).
Etest oxacillin MIC testing was performed according to the manufacturer's instructions using Mueller-Hinton BBL II agar supplemented with 2% NaCl (wt/vol) and incubation at 35°C for a full 24 h. Etest cefoxitin MIC testing was done using Mueller-Hinton BBL II agar without NaCl supplementation. Plates were incubated in ambient air at 30°C, 35°C, and 36°C. Cefoxitin MICs were read after 18 to 19 h of incubation. Isolates for which cefotixin MICs were
4 mg/liter were incubated further and read after a total of 24 h.
The influence of incubation at 37°C was investigated in phase II using a subset of 54 strains (35 mecA-negative and 19 mecA-positive strains) incubated simultaneously at 35°C and 37°C. For phase II we selected the most challenging strains, those with the largest cefoxitin zone diameters and those that during phase I had shown results influenced by variations in incubation temperature. For phase II, MICs and zone diameters were read after 18 h and 24 h of incubation.
In both phase I and phase II, S. aureus strains ATCC 29213, ATCC 25923, and ATCC 43300 were included for quality control on all test runs. The temperature in the incubators was monitored at 10-min intervals for the entire incubation period using TinytagPlus digital thermometers with accompanying software (Gemini Data Loggers Ltd., Chichester, West Sussex, United Kingdom). The temperature curves showed that with empty incubators the maximum variation during a 12-h test period was ±0.25°C (data not shown). With plates on the shelves, the time to reach the preset temperature on each shelf was dependent on the number of plates placed on the shelf. In phase I, when shelves were full, the temperature only reached the preset temperature at the end of the incubation time. This was true for both temperatures. Each strain was subjected to the same conditions for each of the two temperatures. The temperature at the end of incubation was 34.6 to 34.8°C and 35.6 to 36.4°C when incubators were set at 35°C and 36°C, respectively. In phase II the preset temperatures were reached on all shelves within 4 h, and the temperatures were stationary between 34.7 to 35.6°C and 36.8 to 37.1°C when incubators were set at 36°C and 37°C, respectively (detailed data not shown).
Intraassay variation was studied using S. aureus ATCC 25923, ATCC 29213, and ATCC 43300 tested for 10 days using two different batches of Mueller-Hinton II agar (BBL) and two different batches of 10-µg and 30-µg disks (Oxoid). Each agar plate was read independently by five technologists, giving a total of 50 measurements per Mueller-Hinton agar/disk combination.
Population analysis. Two mecA-positive isolates obtained from Norway (strains 9-8 and 10-22), previously undetectable by any phenotypic cefoxitin method (19, 20), were tested by population analysis profile using various concentrations of cefoxitin and oxacillin in agar. Colonies from a fresh overnight culture were inoculated into 5 ml tryptic soy broth (SSI Diagnostika) and incubated overnight at 35°C. Tenfold dilutions of the culture were prepared in 0.9% NaCl, and 20 µl from each dilution was spot inoculated in duplicate onto Mueller-Hinton agar plates containing twofold dilutions of oxacillin (0.25 to 128 mg/liter) and cefoxitin (0.25 to 128 mg/liter). The agar plates were incubated a full 24 h at 35°C, and colonies were counted to plot the population analysis profile. Strain 10-22 showed a highly heterogeneous resistance pattern against oxacillin and displayed colonies of up to 64 mg/liter (data not shown). Since strain 9-8 failed to show any colonies at 4 mg/liter in repeat experiments, even in the undiluted samples, it was concluded that this mecA-positive isolate was an oxacillin-susceptible phenotype, and it was omitted from further analysis. This is a well-described phenomenon that can be caused by a defect in one or more enzymes needed for an isolate to express oxacillin resistance to, e.g., one of the fem genes (1).
Statistics. Statistical differences in MICs obtained at different temperatures and/or incubation times were analyzed by Wilcoxon's matched pairs test, with P < 0.05 considered significant. Distributions were reported by medians and ranges. For statistical analyses, the Statistica software program (version 7.0; Stat Soft Inc., Tulsa, TX) was used.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Predicted clonal complexes (CC) based on spa type of the 50 mecA-positive isolates
|
4 mg/liter and resistant (R) > 4 mg/liter, only one mecA-positive isolate (isolate 10-22 from Norway) was incorrectly categorized as non-MRSA after both 18 h and 24 h of incubation at all temperatures (see also data from the population profile analysis for this isolate in Material and Methods). The oxacillin MIC for this isolate was 4 mg/liter. Another mecA-positive isolate (isolate 1748) was phenotypically sensitive to oxacillin (MIC, 1 mg/liter) but was resistant to cefoxitin (MIC, 16 mg/liter). |
View this table: [in a new window] |
TABLE 2. Oxacillin Etest MIC at 35°C after 24 h of incubation and cefoxitin Etest MIC at 30, 35, and 36°C after 18 h and 24 h of incubation
|
21 mm for the 30-µg disk and
17 mm for the 10-µg disk in both phase I and phase II.
![]() View larger version (21K): [in a new window] |
FIG. 1. Zone diameter distribution for 30-µg cefoxitin disk (A) and 10-µg cefoxitin disk (B) against 145 S. aureus isolates (phase I) and repeat testing with 54 S. aureus isolates (phase II) at 35°C and 18 h of incubation. The current CLSI breakpoint for the 30-µg disk is shown by a dashed line, and the suggested breakpoints are shown by black lines in both figures. , mecA negative, phase I; , mecA positive, phase I;
|
25 mm and
23 mm for phase I and II, respectively, and
17 mm for the 10-µg disk in both phases I and II. Increasing the incubation time from 18 h to 24 h or the temperature from 35°C to 36°C did not affect inhibition zone diameter distributions for either disk for mecA-negative or mecA-positive isolates (data not shown).
Increasing the temperature from 35°C to 37°C was slightly more problematic; for the 30-µg disk, a mecA-negative strain was classified as positive, and with the 10-µg disk a mecA-positive strain was classified as susceptible. Decreasing the temperature from 35°C to 30°C made one strain falsely susceptible by both disks (data not shown).
Repeated testing of quality control strains. The results for the 10-day repeated testing of S. aureus ATCC 29213 (i.e., CLSI quality control strain for routine MIC testing) and S. aureus ATCC 25923 (i.e., CLSI quality control strain for routine disk diffusion testing) are shown in Table 3 (all methodological variants were tested with the same inoculum suspension). S. aureus ATCC 25923, the quality control strain normally used by the CLSI, on one day (day 5) gave significantly different results from the rest of the nine days both for the 10-µg and the 30-µg disks. Omitting the outlier results from day 5, the following medians (ranges) were obtained for S. aureus ATCC 25923: 25 mm (22 to 28 mm) for the 30-µg disk and 20 mm (18 to 22 mm) for the 10-µg disk. For S. aureus ATCC 29213, tighter zone diameter ranges were obtained for both disks, giving medians (ranges) of 25 mm (23 to 27 mm) for the 30-µg disk and 18 mm (17 to 20 mm) for the 10-µg disk. There was no difference between batches of agar or disks (data not shown).
|
View this table: [in a new window] |
TABLE 3. Intraassay variation for 10-µg and 30-µg cefoxitin disks for S. aureus ATCC 29213 and ATCC 25923
|
|
|
|---|
Results obtained suggest that both cefoxitin inhibition zone diameters and MICs for mecA-positive strains are influenced by incubation temperature, although they are influenced to a lesser degree than for oxacillin. Most importantly, detection of MRSA by the cefoxitin-based method was not affected by temperature variations between 35°C and 36°C, while at 37°C disk diffusion gave one false phenotypic categorization. In the International Organization for Standardization (ISO) methodology standard for MIC determinations currently under development (22), the incubation temperature is specified as 34 to 37°C. For oxacillin testing with staphylococci, there is a note stating that the temperature should not exceed 35°C. Although cefoxitin is influenced to a lesser degree, it may be appropriate to include the same note for cefoxitin testing. No advantage was seen with the 30°C incubation temperature as proposed by others (4, 8). Importantly, there was no benefit in extending the incubation time from 18 h to a full 24 h. This is in agreement with the findings of Swenson and Tenover (21). It is a major advantage to clinical laboratories that standard methodology (medium, inoculum, incubation time, and temperature) can be used for the detection of MRSA.
For the 30-µg disk, our findings suggest MRSA interpretive breakpoints of S
22 mm and R
21 mm, i.e., 2 mm larger than the criteria published by the CLSI (6). This proposal is supported by the results of several previous investigations (Table 4) in which mecA-negative isolates all have had zone diameters of 22 mm or more, and occasionally mecA-positive isolates exhibiting zone diameters of 20 and 21 mm have been reported (4, 9, 10, 16, 21, 23, 24).
|
View this table: [in a new window] |
TABLE 4. Published zone diameter ranges (CLSI 30-µg cefoxitin disk) for mecA-positive and mecA-negative S. aureus isolates
|
17 mm and R
16 mm. The contents of both disks had comparable sensitivity and specificity for detection of MRSA. The lower disk content produces smaller zones and thereby reduces the interference with results for other antibiotic disks tested on the same agar plate. The higher disk content may have a slightly higher specificity, as reflected by the larger gap between mecA-positive and mecA-negative zone diameter results.
In our study, cefoxitin MIC testing with Etest was accurate for MRSA detection. The results suggest interpretive breakpoints of S
4 mg/liter and R
8 mg/liter using CLSI terminology and S
4 mg/liter and R > 4 mg/liter in EUCAST (the European Committee of Antimicrobial Susceptibility Testing) terminology. However, the cefoxitin MIC mode for mecA-negative strains is close to the suggested breakpoints. The proposed breakpoints are supported by data published by other investigators (Felten et al. [8], Fernandes et al. [9], Swenson and Tenover [21], and Votta et al. [M. Votta, D. Turner, B. Turng, T. Wiles, J. Reuben, Abstr. 44th Intersci. Conf. Antimicrob. Agents Chemother., abstr. D-47, 2004]). However, in an investigation with many borderline oxacillin-resistant S. aureus strains by Swenson et al., cefoxitin breakpoints of S
6 mg/liter, instead of 4 mg/liter, and R
8 mg/liter were shown to have better specificity but a slightly lower sensitivity (J. Swenson, D. Lonsway, S. McAllister, A. Thompson, L. Jevitt, J. Patel, Abstr. 45th Intersci. Conf. Antimicrob. Agents Chemother., abstr. D-1732, 2005). In this study, there was no difference in sensitivity or specificity using a breakpoint of S
6 mg/liter and R
8 mg/liter instead of S
4 mg/liter and R
8 mg/liter.
In conclusion, this study provides further evidence that cefoxitin is an accurate surrogate marker for the detection of MRSA in routine susceptibility testing for disk diffusion and MIC testing. Incubation temperature should not surpass 36°C. Incubating for a full 24 h did not improve results obtained after 18 h. Hence, standard conditions currently used by clinical laboratories for routine susceptibility testing as described by CLSI can be used, and the cefoxitin disk can be included among other disks relevant for susceptibility testing of Staphylococcus aureus. The current CLSI cefoxitin zone diameter breakpoints should be adjusted by a 2-mm increase.