Journal of Clinical Microbiology, March 2006, p. 729-737, Vol. 44, No. 3
0095-1137/06/$08.00+0 doi:10.1128/JCM.44.3.729-737.2006
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
Klinik für Innere Medizin II, Universitätsklinikum des Saarlandes, Kirrberger Str., 66421 Homburg/Saar, Germany,1 Abteilung für Virologie, Universitätsklinikum des Saarlandes, Kirrberger Str., 66421 Homburg/Saar, Germany,2 Roche Diagnostics, Molecular Diagnostics Development, Nonnenwald 2, 82372 Penzberg, Germany3
Received 16 August 2005/ Returned for modification 28 September 2005/ Accepted 1 December 2005
|
|
|---|
|
|
|---|
For measurements of HCV RNA, different qualitative and quantitative tests based on target (reverse transcription [RT]-PCR and transcription-mediated amplification) and signal amplification (branched DNA [bDNA]) techniques with different lower detection limits and linear ranges of amplification are commercially available. All quantitative HCV RNA assays are standardized to IU on the basis of the first WHO HCV international standard, 96/790 (2, 7, 11, 17, 23, 24, 27). However, continuing limitations are the lack of complete automation (22), the necessity for dilutions for quantification by standard PCR-based assays (6), the relatively low sensitivity of quantitative HCV RNA assays (31), and the need for different test systems for qualitative and quantitative HCV RNA measurements (31). Furthermore, standardization of results to IU are mainly based on HCV genotype 1 panels, and little is known about the variability of commercially available HCV RNA assays for quantification of different HCV genotypes.
Real-time PCR methods for the quantification of HCV RNA have the advantage of linear amplification over a broad dynamic range, together with an integrated, automated detection system. With efficient HCV RNA extraction, they have the potential to achieve lower detection limits of less than 10 IU/ml.
In the present study, we compared two real-time RT-PCR-based assays (HCV RNA extraction with the manual High Pure System, together with the COBAS TaqMan 48 Analyzer [HPS/CTM], and HCV RNA extraction with the automated COBAS Ampliprep instrument, together with the COBAS TaqMan 48 Analyzer [CAP/CTM]), one standard RT-PCR-based assay (COBAS Amplicor HCV Monitor 2.0 [CAM]), and one signal amplification assay (Versant HCV Quantitative 3.0 [bDNA]) for HCV RNA quantification of clinical samples of HCV genotype 1a-, 1b-, 2a/c-, 2b-, 3a-, 4-, and 5-infected patients in Europe. For genotype 1- and genotype 2- and 3-infected patients, in addition to baseline samples, week 12 and week 4 samples, respectively, were analyzed. For estimation of assay precision in the lower range and investigation of apparent underquantification of certain HCV genotypes by the HPS/CTM assay, HCV RNA measurements of samples after 1:20 dilution were performed.
|
|
|---|
The COBAS TaqMan HCV Test for use with the High Pure System, the COBAS AmpliPrep/COBAS TaqMan HCV Test, and the COBAS Amplicor HCV Monitor Test v2.0 are currently not available for in vitro diagnostic use in the United States.
Undiluted clinical serum or plasma samples from European patients with chronic hepatitis C infected with HCV genotype 1a/b (n = 40), 2a/c (n = 14), 2b (n = 11), 3a (n = 24), 4 (n = 9), or 5 (n = 10) were used for parallel testing with the four different assays in a single determination. To guarantee identical conditions, serum or plasma samples stored at 80°C were thawed to generate appropriate aliquots (50 to 850 µl) for the different assays. Prior to being tested with the four different assays, all aliquots were stored again at 80°C. The same procedure was carried out for testing of 1:20 dilutions in a subset of samples with sufficient volumes available. For testing at week 12 and week 4, 24 samples from genotype 1a/b- and 24 samples from genotype 2- and 3-infected patients were available. All HCV RNA measurements with the four different assays were performed in the virologic laboratories of the Saarland University Hospital.
All patients had been enrolled in prospective studies and had been treated in the hepatology outpatient clinics of the University Hospitals of Homburg and Frankfurt, Germany. Treatment was performed with polyethylene glycol-alpha interferon 2a at 180 µg per week subcutaneously, plus ribavirin orally (800 mg for genotype 2- and 3- and 1,000/1,200 mg for genotype 1- and genotype 4- and 5-infected patients). Genotype 2- and 3-infected patients were treated for 24 weeks, and genotype 1- and genotype 4- and 5-infected patients were treated for a maximum of 48 weeks.
Virologic response in the trials was assessed by a qualitative HCV RNA assay with a lower sensitivity of 50 IU/ml (COBAS Amplicor HCV 2.0; Roche Diagnostics). According to the qualitative HCV RNA results, patients were defined as (i) virologic nonresponders (NR) (HCV RNA positive at the end of treatment), (ii) end-of-treatment responders with relapse (REL) (HCV RNA negative at the end of treatment but positive thereafter), and (iii) virologic sustained responders (SR) (HCV RNA negative at the end of treatment and at the end of follow-up). Genotype 1-infected patients with a positive qualitative HCV RNA test at week 24 were defined as nonresponders, and treatment was discontinued according to the study protocol.
Written informed consent was obtained from each patient, and the studies were approved by the Ethics Committees of Medical Research in Homburg and Frankfurt in accordance with the 1975 Declaration of Helsinki. All specimens in this evaluation represented leftover samples from the studies described above; no bleeds were obtained from patients specifically for this HCV RNA quantification study of different HCV gentoypes.
Genotyping of HCV according to the classification of Simmonds et al. (29) was performed by reverse hybridization assay (INNO LiPA HCV-II; Innogenetics, Gent, Belgium).
Real-time RT-PCR-based assays (HPS/CTM and CAP/CTM). For testing with the HPS/CTM assay, HCV RNA was isolated from a 0.5-ml aliquot of controls and clinical specimens using the manual High Pure System Viral Nucleic Acid kit procedure. A known amount of HCV quantification standard RNA was introduced into each specimen, along with the lysis reagent. For adsorption of HCV RNA and quantification standard to a glass fiber surface, isopropanol was added to the lysis mixture before centrifugation through a column with a glass fiber filter insert. After removal of unbound substances, HCV RNA and HCV quantification standard RNA were eluted from the glass fiber particles.
For testing with the CAP/CTM assay, HCV RNA was isolated from 0.85-ml aliquots of controls and clinical specimens using the automated COBAS Ampliprep instrument. HCV quantification standard was added to the sample in order to achieve full process control. After a protease incubation step, the lysis reagent, together with the magnetic glass particles, was introduced into each specimen. HCV RNA and HCV quantification standard RNA were bound to the surfaces of magnetic glass particles. After completion of several washing steps, the adsorbed nucleic acids were eluted at elevated temperature with an aqueous solution.
After HPS- and CAP-based extraction of nucleic acids, samples and controls were processed for amplification and detection using the COBAS TaqMan 48 Analyzer according to the instructions of the manufacturer. While the compositions of the reagents for amplification and detection are not identical, both assays (HPS/CTM and CAP/CTM) rely on the same test principles. Reverse transcription and amplification were carried out using primers that bind within the highly conserved 5' nontranslated region of HCV; deoxynucleoside triphosphates, including deoxyuridine and Thermus species strain Z05 polymerase (a single-tube, single-enzyme, single-primer set process). In the presence of manganese (Mn2+), Z05 has both reverse transcription and DNA polymerase activities. The quantification standard represents a noninfectious RNA construct containing fragments of HCV sequences with primer binding regions identical to those of the HCV 5' nontranslated region, leading to an amplification product of the same length and base composition as the HCV target RNA. For destruction of potential contaminating DNA from previous amplifications, the AmpErase system (AmpErase) was used. For detection of amplification products, the assays utilize real-time PCR technology with two different dual-labeled fluorescent oligonucleotide probes, which are able to bind HCV target amplicon and quantification standard amplicon, respectively, within the regions spanned by the primers. The two different probes for the HCV target and the quantification standard are labeled with two different fluorescent reporter dyes. The reporter fluorescence is suppressed in the intact probe by the proximity of the quencher dye due to inductive-resonance-based energy transfer (Förster-type energy transfer). During elongation, the hybridized dual-labeled oligonucleotide probe is cleaved by the 5'-3' exonuclease activity of Z05 polymerase, leading to the separation of reporter and quencher dyes. Within each cycle during the annealing and elongation phase of PCR, the increasing emission of fluorescence light from such cleaved dual-labeled oligonucleotides is collected independently for the HCV target and quantification standard at different wavelengths. The larger the original HCV RNA amount of a specimen, the earlier the fluorescence of the reporter dye rises above certain assigned fluorescence levels (the critical-threshold value), whereas for the constant titer of quantification standard RNA, the fluorescence of the reporter dye should appear at the same cycle for all specimens. By comparison of critical-threshold values obtained for the target HCV RNA and the quantification standard RNA, the original HCV RNA concentration of the specimen is calculated.
Due to the large dynamic range of the HPS/CTM and CAP/CTM assays, none of the specimens investigated in the first part of this study, addressing the analysis of undiluted pretreatment samples, had to be diluted.
Standard RT-PCR-based assay. Extraction of nucleic acids for testing with the CAM assay was carried out from a 0.1-ml aliquot of controls and clinical specimens according to the manufacturer's instructions. Briefly, nucleic acids (HCV RNA, together with the quantification standard RNA) were isolated by lysis of virus particles with a chaotropic reagent, followed by a standard precipitation of RNA with alcohol. Subsequently, reverse transcription and PCR amplification by Thermus thermophilus polymerase and HCV RNA quantification with the CAM assay were performed as previously described in detail (9). Predilutions were performed for HCV RNA concentrations above 500,000 IU/ml. The lower detection limit of the CAM version 2.0 assay, according to the manufacturer, is 600 IU/ml.
The CAM, HPS/CTM, and CAP/CTM assays are standardized against the first WHO HCV international standard (96/790), and titer results are automatically reported in international units (IU/ml).
Signal amplification-, bDNA-based assay. HCV RNA quantification with a signal amplification-based assay was performed with the third generation of the branched DNA nucleic acid probe test (bDNA). Several modifications to enhance signal amplification and to reduce nonspecific binding in comparison to version 2.0 were introduced; the bDNA 3.0 test procedure has been described in detail elsewhere (27).
The bDNA assay version 3.0 is standardized for IU, and the assay has been reported to be linear over its entire dynamic range from the lower detection limit of 615 IU/ml up to 8 million IU/ml (27).
Dilutions. In samples of different HCV genotypes (1a/b, n = 5; 2a/c, n = 3; 2b, n = 4; 3a, n = 3; 4, n = 3; and 5, n = 3) with sufficient volume remaining after initial testing, 1:20 dilutions were performed, and six replicates of each sample were tested by each of the four quantitative HCV RNA assays. The results for diluted samples were compared with the HCV RNA concentrations derived from a single determination of the original undiluted specimens divided by 20.
Data analysis. Results are expressed as mean, median, and standard deviation (SD) as appropriate. Correlation coefficients (R) were calculated for the linearity of the assays.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. HCV RNA concentrations of genotype 1- to 5-infected patients by four different assays
|
![]() View larger version (22K): [in a new window] |
FIG. 1. Correlation of HCV RNA concentrations of clinical samples between the CAM and the CAP/CTM assays. Results are shown separately for HCV genotype 1a/b, 2a/c, 2b, 3a, 4, and 5 samples. In addition to the single HCV RNA concentrations, the identity line and the regression line (boldface) are shown.
|
![]() View larger version (22K): [in a new window] |
FIG. 2. Correlation of HCV RNA concentrations of clinical samples between the CAM and the HPS/CTM assays. Results are shown separately for HCV genotype 1a/b, 2a/c, 2b, 3a, 4, and 5 samples. In addition to the single HCV RNA concentrations, the identity line and the regression line (boldface) are shown.
|
![]() View larger version (21K): [in a new window] |
FIG. 3. Correlation of HCV RNA concentrations of clinical samples between the CAM and the bDNA assays. Results are shown separately for HCV genotype 1a/b, 2a/c, 2b, 3a, 4, and 5 samples. In addition to the single HCV RNA concentrations, the identity line and the regression line (boldface) are shown.
|
|
View this table: [in a new window] |
TABLE 2. HCV RNA concentrations of genotype 1- to 5-infected patients by four different assays after 1:20 dilution
|
Analyses of week 4 and 12 samples.
For genotype 1-infected patients, HCV RNA concentrations at baseline and week 12 were analyzed with the four different assays. A mean HCV RNA decline of
2 log10 IU/ml was observed for all SR and REL patients by the CAP/CTM and HPS/CTM assays due to the lower detection limit of approximately 10 IU/ml for these assays (28). However, also for the CAM and bDNA assays, results at week 12 were either
2 log units below baseline or below the detection limit of 600 to 615 IU/ml for all SR and REL patients (Table 3). Prediction of a virologic relapse after the end of treatment was not possible on the basis of the HCV RNA decline between baseline and week 12. The mean differences of HCV RNA concentrations between baseline and week 12 were even higher in REL patients than in SR patients (Table 3).
|
View this table: [in a new window] |
TABLE 3. HCV RNA concentrations of genotype 1-infected patients: baseline compared with week 12 during therapy
|
2 log IU/ml at week 12 but were HCV RNA positive at week 24. Furthermore, measurement of HCV RNA concentrations at baseline and week 12 with different assays (e.g., CAP/CTM at baseline and bDNA at week 12) would have led to an aberrant continuation of therapy in up to three NR patients.
For all HCV genotype 2- and 3-infected patients, a sharp HCV RNA decline was observed from baseline to week 4, and only one patient exhibited virologic nonresponse due to breakthrough during further antiviral therapy (Table 4). As for genotype 1-infected patients, prediction of virologic relapse was not possible by analyses of HCV RNA decline between baseline and week 4 in patients with genotype 2 and 3 infections.
|
View this table: [in a new window] |
TABLE 4. HCV RNA concentrations of genotype 2- and 3-infected patients: baseline compared with week 4 during therapy
|
For genotype 1 and genotypes 2 and 3, mean HCV RNA concentrations at baseline were higher in REL than in SR patients (7.0 x 106 versus 3.8 x 106 IU/ml and 6.4 x 106 versus 4.1 x 106 IU/ml, respectively). However, the differences did not reach statistical significance (Mann-Whitney U test).
|
|
|---|
A dual-labeled probe-based real-time RT-PCR assay (TaqMan) has been automated and standardized for detection of different RNA and DNA targets, including HCV RNA (CTM). For measurement of HCV RNA concentrations, the COBAS TaqMan test was combined with a manual (HPS) sample preparation assay or an automated (CAP) sample preparation step. In the present study, these new real-time RT-PCR-based assays (HPS/CTM and CAP/CTM) were compared with the widely distributed quantitative HCV RNA assays CAM and bDNA.
For comparison of the CAP/CTM assay with CAM as a reference test, a high correlation for HCV RNA concentrations of genotype 1, 2a/c, 2b, 3a, and 5 samples was observed. The mean differences were always below ±0.3 log10 IU/ml (i.e., twofold), and thus, a linear correlation between the two assays is present. Only in genotype 4 samples was a slight underquantification (mean HCV RNA concentrations, 0.52 log IU/ml) detected, which requires further investigation on the basis of a larger number of samples.
Comparison of the HPS/CTM assays with the CAM test showed a good correlation for genotype 1 and 2b samples, with a mean difference of 0.13 and 0.18 log10 IU/ml. For genotype 2a/c, 3a, 4, and 5 samples, a significant underquantification of 0.78 to 1.51 log10 IU/ml was observed. Due to this underquantification, version 1 of the HPS/CTM assay is restricted by the manufacturer to use in patients with genotype 1 and 6 infections only. Unfortunately for the present study, no specimens of genotype 6-infected patients, which are very rare in Europe, were available. The reasons for underquantification of certain HCV genotypes by the HPS/CTM assay are unknown. Because real-time RT-PCR-based assays represent a complex optimization of the different steps of nucleic acid preparation, amplification, and detection for the different HCV isolates, subtypes, and genotypes, multiple reasons for the underquantification of specific HCV genotypes have to be considered. General mismatch of primers or the TaqMan probe located within a highly conserved part of the 5' nontranslated region of the HCV genome, which was already used for amplification by the CAM assay, was excluded by the manufacturer. However, suboptimal binding of oligonucleotides due to the secondary structure of the internal ribosome entry site located within the 5' nontranslated region may be possible. In the present study, we demonstrate that the underquantification in the HPS/CTM assay can be partially compensated for by 1:20 predilution of the samples. Thus, transfer of an inhibitor for PCR amplification or detection from nucleic acid extraction appears to be a conceivable explanation for the underquantification by the HPS/CTM assay. Restriction to certain HCV genotypes (2a/c, 3a, 4, and 5) may be explained by differences in the amplification efficiencies between HCV genotypes.
For the comparison of the bDNA assay with CAM, relatively stable lower HCV RNA concentrations were observed in genotype 1, 2a/c, 2b, and 3a samples (0.48 to 0.64 log10 IU/ml), despite the standardization of both assays to the first international WHO HCV standard (96/790). For genotype 4 and 5 samples, a good correlation of HCV RNA concentrations between bDNA and CAM was detected. It is known that linearity of quantification is lost for CAM for HCV RNA concentrations above 1 x 105 to 5 x 105 IU/ml (2, 11). However, in previous studies, predilution for the CAM assay was performed for samples of >850,000 IU/ml only (2, 11). In the present study, HCV RNA titers above 500,000 IU/ml were considered to be outside the linear range of the CAM assay and were retested after dilution. This may explain differences from previous CAM/bDNA comparative studies. Interestingly, after predilution (1:20) for samples of all genotypes, the correlation between CAM and bDNA improved to a mean difference of approximately 0.3 log10 IU/ml. This is in accordance with recent findings, with a recommendation of predilution for all samples before measurement with CAM due to saturation effects (21). However, also in the prediluted samples, a principal difference of a significantly lower estimation of HCV RNA concentrations in genotype 1 to 3 samples by the bDNA assay compared to the CAM and CAP/CTM assays was detectable throughout the present study, which may reflect differences in the process for calibration to the WHO standard of the assays.
In patients with chronic hepatitis C genotype 1 infection treated with (pegylated) interferon plus ribavirin, decisions about early treatment discontinuation are currently based on quantitative HCV RNA measurements at baseline and week 12 of therapy and a qualitative HCV RNA measurement at week 24 (3, 5, 8, 22). In the present study, reliable treatment decisions at week 12 could be made on the basis of all HCV RNA assays tested (HPS/CTM, CAP/CTM, CAM, and bDNA). However, it is important to note that a change between different quantitative HCV RNA assays from baseline to week 12 may lead to incorrect treatment decisions and therefore should be avoided. Furthermore, the importance of the week 24 discontinuation rule was demonstrated by the fact that a significant number of NR patients achieved a >2-log-unit decline at week 12 but were HCV RNA positive at week 24.
Taken together, comparison of the standard PCR-based CAM assay with the real-time PCR-based, automated CAP/CTM assay showed a good correlation of HCV RNA quantification for HCV genotypes 1 to 5. By comparison of the CAM assay with the HPS/CTM test, a significant underquantification of HCV genotypes 2a/c, 3a, 4, and 5 was observed, while for genotype 1 and 2b samples, a high concordance of HCV RNA concentrations was detected. For the bDNA assay, constant
0.5-log-unit-lower HCV RNA concentrations were observed in comparison with the CAM assay for genotype 1 to 3 samples.
Because of the high significance of HCV RNA concentrations in the management of the current (pegylated) interferon-ribavirin combination therapies and for future antiviral treatment options based on direct antiviral drugs (3, 12, 14, 26, 32), HCV RNA assays must be carefully analyzed for reliable quantification of all of the different HCV genotypes.
|
|
|---|