Our aim was to develop a simple sensor to signal an early warning of impending catheter encrustation and blockage. The idea was to produce sensors by impregnating polymeric materials with pH indicators. The sensors, located in the catheter drainage system, continuously monitor the pH at the surface of the catheter over a 7-day period (the interval after which the urine drainage bags are usually replaced). Bromothymol blue (BTB) was chosen as the indicator because of its transition from yellow to dark blue over the pH range 6 to 8. A change from an acid to an alkaline reaction at the sensor surface signals infection and biofilm formation by P. mirabilis. The appearance of the signal indicates that action should be taken to avoid an acute clinical episode. Our objectives were to (i) develop polymers suitable for the manufacture of sensors, (ii) test their ability to signal catheter encrustation in laboratory models, and (iii) identify the optimum location for the sensor in the drainage system.
The sensors were prepared by dissolving cellulose acetate (Acordis Ltd., Coventry, United Kingdom) in acetone and adding a mixture of BTB (Sigma-Aldrich, St. Louis, Mo.) and sulfuric acid. Under these conditions, the BTB becomes covalently bound to the cellulose acetate. Polyethylene glycol was added as a plasticizer and to control the rate of movement of ions through the matrix, thus controlling the response rate of the material to changes in pH. The mixture was spread over a glass plate to allow the polymer mixture to set. Strips (2 cm long by 1 cm wide by 1 mm thick) of the calcium acetate-BTB polymer were then prepared, washed in water to remove residual acid, and stored at 4°C.
The bladder model has been described previously (14). It consists of a glass chamber maintained at 37°C by a water jacket. Each model was sterilized by autoclaving, and then a 4.7-mm all-silicone catheter (Bard Ltd., Crawley, United Kingdom) was inserted into the chamber through its base. The catheter retention balloons were inflated with water and the catheters connected to drainage tubes and bags in the normal way. Sterile urine was pumped into the chambers at a rate of 0.5 ml/min, so that residual volumes collected below the catheter eyeholes before flowing through the drainage tube to the collection bags. The artificial urine used in the experimental work was based on that devised by Griffith and colleagues (4). Its composition and method of sterilization have been described previously (14). Sets of models were primed with urine (30 ml), which was then inoculated with 100 µl of a 4-h culture of 2 x 108 CFU/ml of P. mirabilis B2, a strain that had been isolated from a patient's encrusted catheter. After an hour to allow the test organism to establish itself in the bladder urine, the urine supply was switched back on, and the models operated until the catheters became blocked. The times required for catheters to become blocked were recorded. Under these conditions, the pH of the urine in the bladder rises from 6.1 to >8.0 in under 12 h.
Initial experiments were performed using noninfected models in which sensors placed at the catheter-drainage tubing junction were exposed to flow regimens of phosphate buffers at pH 6.2 and 8.0. Spectrophotometric analysis (at 450 nm) showed that in all cases, <10% of the indicator eluted from the polymer over 8 days. To test the ability of the sensors to signal P. mirabilis infection, models were supplied with urine and inoculated with P. mirabilis. Sensors were inserted into the drainage system at the catheter-drainage tube junction and in the urine bag. The times required for sensors to signal and catheters to become blocked in four replicated experiments are presented in Table 1. The appearance of the sensors at various times is illustrated in Fig. 1. It is clear that sensors located in the bags gave stronger and earlier signals. The bag also has the advantage of being in a more conspicuous site.
|
View this table: [in a new window] |
TABLE 1. Comparison of the times required for sensors placed in a urine bag and at the catheter-drainage tube junction to produce the signal indicating the presence of P. mirabilis in catheterized bladders
|
![]() View larger version (149K): [in a new window] |
FIG. 1. Appearance of sensors located at the catheter-drainage bag junction (A) and in the drainage bag (B) at 0 h, at the catheter-drainage bag junction at 26 h (C), and in the drainage bag at 16 h (D) after inoculation with P. mirabilis. The catheter became blocked at 51 h. The sensor registers intermediate colors between yellow and blue, but these are transitory. The positive reactions are indicated by the dark-blue colors shown in panels C and D. Other colors are considered to indicate negativity.
|
![]() View larger version (57K): [in a new window] |
FIG. 2. Scanning electron micrographs showing the early stages of encrustation on a catheter (A and B) removed at 12 h from a model colonized by P. mirabilis when a signal was recorded from the sensor. Encrustation was also present on the sensor at this time (C).
|
In conclusion, the results show that the cellulose acetate-bromothymol blue sensor is capable of giving a clear, simple signal of infection of the catheterized urinary tract by P. mirabilis. The device could be inserted into all types of drainage systems. The optimum location is the urine bag. It signals the early stages of catheter encrustation in time to allow action to avoid the clinical crises induced by catheter blockage. The sensor would have greater value if we had an effective strategy to inhibit the encrustation process that could be implemented in those cases where the sensor gives its signal. Such a strategy has been developed (12). It involves inflating the catheter retention balloon with a solution of triclosan, rather than water. We are now planning a clinical study to test the efficacy of the sensor-modulator system in patients.
|
|
|---|