Journal of Clinical Microbiology, May 2006, p. 1831-1833, Vol. 44, No. 5
0095-1137/06/$08.00+0 doi:10.1128/JCM.44.5.1831-1833.2006
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
Wilford Hall USAF Medical Center, Lackland AFB, Texas 78236,1 Walter Reed Army Institute of Research, Rockville, Maryland 20850,2 Henry M. Jackson Foundation for the Advancement of Military Medicine, Rockville, Maryland 20852,3 Triservice AIDS Clinical Consortium, Rockville, Maryland 208524
Received 20 October 2005/ Returned for modification 16 February 2006/ Accepted 21 February 2006
|
|
|---|
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Identification of microparticle types used in the Multispot devicea
|
These data are promising but remain incomplete, since they do not include an assessment of a patient population such as the Triservice AIDS Clinical Consortium (TACC) Natural History Study (NHS) (http://www.hivforum.org/cohorts/TACC%200305.pdf). This prospective continuous enrollment cohort of current and former U.S. military beneficiaries contains a small minority of patients whose serum anti-gp41 titers waned significantly over time, as measured by quantitative enzyme immunoassay (EIA) as well as by Western blotting (WB). In a previous study, antibody levels became sufficiently low to yield false-negative OQ results in 4% of subjects analyzed (7). This phenomenon appears to be restricted to patients undergoing effective highly active antiretroviral therapy (HAART). Others have previously reported decreases in anti-gp41 antibody levels (4) and IgG-secreting cells (6) among HAART-treated patients, possibly due to a decline in HIV-1-specific CD4+ T lymphocytes (10) from a lack of persistent antigen exposure during suppressive therapy. Since patients may undergo HIV testing under circumstances where they are unwilling or unable to disclose their serostatus, testing devices should perform well even on samples from HAART-experienced HIV-positive individuals. This study examined whether the Multispot gp41 antibody detection approach would render the test less sensitive in such a population, where rapid testing was previously less sensitive (7).
This study was approved by institutional review boards at each institution. All Multispot testing was done according to the package insert (1) in the Immunology Research Laboratory at Wilford Hall USAF Medical Center by a single technician. Multispot results were scored using an intensity scale of 0 to 4, where each spot was graded from 0 (no reaction) to 4 (strongest reaction), and are expressed according to the spot order given in Table 1.
Study samples were derived from repository sera maintained at 70°C. Specimens were identified by starting on the study protocol approval date and retrospectively selecting the 248 most recently collected unique-subject specimens associated with a TACC NHS visit with sufficient volume for testing. An additional serum repository sample was tested from each of the four subjects with false-negative results identified in a previous TACC NHS OQ study (7) with specimens obtained on the same date as previously studied. An aliquot from each sample subjected to Multispot testing was frozen at 70°C until shipment on dry ice to the Department of Molecular Diagnostics and Pathogenesis, Walter Reed Army Institute of Research, where HIV-1 Bio-Rad rLAV EIA (Bio-Rad Laboratories, Redmond, WA) and HIV-1 Bio-Rad WB were performed and reported according to the package inserts.
The study was powered (248 samples) to have an 80% likelihood of determining a significant difference between 96% sensitivity and the lower 95% confidence interval (CI) of the Multispot sensitivity stated in the package insert (99.94%) based on the 96% sensitivity of OQ in this population (7). Multispot sensitivity was expressed as a percentage with 95% binomial confidence intervals, descriptive statistics and Mann-Whitney U comparisons were performed using SPSS 13.0 (SPSS, Inc., Chicago, IL), and proportion comparison was accomplished by the Fisher exact test (EpiInfo 2000 version 1.1.2).
Samples included in the main study had been collected at the Wilford Hall USAF Medical Center clinical site from June 2004 through March 2005. Study subjects exhibited a mean time since HIV diagnosis of 8.7 years and a mean CD4 count of 637 cells/µl, with only 13 (5.2%) being below 200 cells/µl. HAART was being given to 153 (61.7%) subjects for a mean of 4.5 years (Table 2). All 248 specimens were reactive for HIV-1 using Multispot as well as the EIA/WB algorithm, yielding a Multispot sensitivity of 100% (95% CI, 98.80 to 100.00%) (Table 3). Spots 1, 3, and 4 were all universally reactive, whereas spot 2 was universally nonreactive. Two subjects failed to produce gp41 WB bands but were nevertheless reactive to both Multispot HIV-1 spots. By chance, 65 subjects contributed samples to both OQ (7) and Multispot evaluations, including three of four subjects that had false-negative results by OQ.
|
View this table: [in a new window] |
TABLE 2. Study subject characteristicsa
|
|
View this table: [in a new window] |
TABLE 3. Comparative reactivities between the current Multispot study and OQ HIV-1 false-negative specimens
|
This study supports the use of Multispot in settings where test subjects might be HAART exposed with attendant low or undetectable anti-gp41 antibody titers. Low titers were suggested by weak semiquantitative reactions in some individuals, but reactivity remained above the level of Multispot detection. Because a single specialized immunology technologist performed the Multispot testing, the performance characteristics of Multispot in this evaluation may not be generalizable to settings where testing is performed by laboratory generalists.
Multispot sensitivity (100%; 95% CI, 98.80 to 100.00) was superior to OQ sensitivity (7) in this population (96%; 95% CI, 90.17 to 98.91; P = 0.00672). The explanation for this difference may involve the flowthrough immunoconcentration design of Multispot, which affords the detection of lower antibody titers than the lateral-flow approach employed by OQ. Differences in antigen structure and preparation may also explain these findings, but the exact composition of antigens is not available for either device (1, 8). It is unlikely that our observations are due to subject selection bias, as similar subjects were drawn from the same population using the same approach in both studies and because Multispot correctly identified samples from individuals who had a false-negative result by OQ. While we did not perform live, side-by-side comparisons between OQ and Multispot for this study, having done so would have been unlikely to yield different results, since the two studies drew from the same patient population, used serum samples from the same repository, included a substantial number of the same individuals, and identified similar decrements in WB gp41 bands in the OQ false-negative specimens.
The views and opinions expressed herein do not necessarily reflect those of the U.S. Air Force, the U.S. Army, or the Department of Defense. Use of trade names is for identification only and does not imply endorsement by the U.S. government.
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||