Journal of Clinical Microbiology, June 2006, p. 2141-2146, Vol. 44, No. 6
0095-1137/06/$08.00+0 doi:10.1128/JCM.00049-06
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
Laboratory Medicine Unit, European Institute of Oncology, Milan, Italy,1 Division of Pathology and Laboratory Medicine, European Institute of Oncology, Milan, Italy,2 Analytical Cytology and Biomolecular Unit, CSPO, Scientific Institute of Tuscany, Florence, Italy,3 Division of Epidemiology and Biostatistics, European Institute of Oncology, Milan, Italy,4 Preventive Gynecology Unit, European Institute of Oncology, Milan, Italy5
Received 10 January 2006/ Returned for modification 24 February 2006/ Accepted 29 March 2006
|
|
|---|
|
|
|---|
The detection of HR HPV in cervical samples has been proposed to improve the efficacy of cervical carcinoma screening programs and to triage women with ambiguous or borderline cervical smears (1). Women with persistent HR HPV positivity have a clearly enhanced risk of developing a premalignant lesion and, hence, may be more closely monitored (3, 15). Moreover, HR HPV testing may be associated with Pap smear to monitor women who have been treated for high-grade cervical intraepithelial neoplasia (CIN) (16, 30). Several studies have revealed that HPV testing yields a high negative predictive value, approaching 100%, for high-grade CIN lesions and cervical carcinomas (lesions
CIN3) (21, 22, 31).
Testing for HPV relies on the detection of viral DNA. The only test currently approved by the U.S. Food and Drug Administration for the detection of HPV DNA is the Hybrid Capture 2 (HC2) system (Digene Corporation, Gaithersburg, Md.). The HC2 assay is a ready-to-use test for routine diagnostics and uses a liquid hybridization format followed by signal amplification to detect 13 HR HPV types (i.e., genotypes 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, and 68) by means of an RNA cocktail probe; the test does not distinguish individual HPV types. The sensitivity is approximately 4,700 viral copies/ml of cervical sample suspension. The evaluation of its laboratory performance has shown that HC2 is a reliable and reproducible test (4, 6, 8); both characteristics are fundamental for a test with potential widespread use. Different laboratories have also used many PCR-based methods. PCR methods are considered the "gold standard" for analytical sensitivity to detect infectious organisms, including HPV. However, PCRs for HPV detection are currently performed as "home brew" methods, lack standardization, show different sensitivity and specificity, are time-consuming, and require a demanding job for the laboratory. In the last years, some studies have compared the performance of HC2 and PCR, showing a good level of agreement between the two methods (19, 28).
Recently, a standardized PCR-based technique (AMPLICOR HPV test; Roche Molecular Systems) for the detection of the 13 HR HPV genotypes has been commercialized. This new test uses amplification of target DNA by PCR and nucleic acid hybridization for the detection of HR HPV genotypes in cervical cells collected into a transport medium. This test could be used in clinical diagnostic laboratories, but few data are currently available on its performance in a clinical setting (17, 27) and, in particular, no data exist on its comparison with the HC2 assay.
The purpose of this study was to examine the performance of two commercially available assays, the HC2 assay and the AMPLICOR HPV test, to detect the presence of HR HPV in cervical samples of women who attended the Colposcopy Clinic for an evaluation of an abnormal Pap smear or for a follow-up examination after a CIN treatment. This analysis will be useful to better identify the assay more suitable for a particular application as well as to inform the clinicians about the correct interpretation of the results of any assay.
|
|
|---|
Sample collection. Cervical samples were taken using a Cervex brush (Rovers Medical Devices B.V., Holland). The collected specimen was first smeared on a glass for conventional cytological evaluation, and then the brush was washed in a vial containing PreservCyt solution (Cytyc Corporation, Boxborough, MA) and transferred to the laboratory for HPV analyses (H. A. Cubie, K. S. Cuscheri, and C. Moore, 21st Int. Papillomavir. Conf., abstr. 121, 2004). The HC2 assay was performed weekly. A 500-µl aliquot of each sample was removed and frozen at 80°C for DNA extraction and subsequent testing for HPV DNA by PCR. In case of discrepant results, the specimen was genotyped using the LINEAR ARRAY HPV genotyping test (Roche Molecular Systems).
HC2 HR HPV assay. Sample material collected in PreservCyt medium was made suitable for HC2 assay using a sample conversion kit (Digene Corporation). HPV DNA testing by the HC2 assay method was performed with the automated HC2 assay system according to the manufacturer's protocol. HC2 is a sandwich capture molecular hybridization assay that utilizes chemiluminescent detection to provide a semiquantitative result. The assay is calibrated to detect approximately 4,700 genome equivalents (or 1 pg/ml) of target HPV, represented by an RLU (relative light unit) measurement greater than or equal to the cutoff value calculated in each run by a series of standards. A measurement less than the cutoff was scored as negative. The samples were analyzed for the presence of HR HPV types 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, and 68. Positive and negative controls (provided by the manufacturer) were included in each run.
AMPLICOR HPV test. (i) DNA extraction. DNA, both HPV and cellular, was released by lysing cervical specimens under denaturing conditions at elevated temperatures in the presence of proteinase K. DNA purification was obtained in columns with a silica-based membrane using vacuum processing.
(ii) PCR amplification. The AMPLICOR HPV test uses biotinylated primers to define a sequence of approximately 165 bp in length within the polymorphic L1 region of the HPV genome. The primers, pooled in the same PCR master mix, are designed to amplify viral DNA from the same 13 types included in the HC2 assay. Also with the same master mix, the ß-globin gene was amplified (268 bp amplicon) to test whether the extracted DNA would be suitable for use as a template for PCR.
(iii) Hybridization reaction. Capture probes, representing regions internal to the amplified sequences, were used to identify the viral or human DNA. Following PCR amplification, the amplicons were chemically denatured to form single-stranded DNA and added to separated wells of microwell plates (MWP) coated with either HR HPV probes or ß-globin-specific oligonucleotide probe. The hybridization was done at 37°C for 1 h.
(iv) Detection reaction. Following the hybridization reaction, the MWP was washed to remove unbound material, and an avidin-horseradish peroxidase conjugate was added to each well to link the biotin. The MWP was washed again to remove unbound conjugate, and a substrate solution containing hydrogen peroxide and 3,3',5,5'-tetramethyl benzidine (TMB) as a chromogen was added. The colorimetric reaction was done for 10 min at room temperature in the dark and stopped by the addition of a weak acid. The absorbance at 450 nm was measured immediately using an automated microwell plate reader.
According to the manufacturer's specifications, the AMPLICOR HPV test could detect genotypes 16, 18, 33, 35, 39, 45, 51, 56, and 68 at 100 copies/ml and genotypes 31, 52, 58, and 59 at 240 copies/ml, with a positivity rate greater than 95%. All genotypes are detected with a 100% positivity rate at 480 copies/ml.
LINEAR ARRAY HPV genotyping test. (i) DNA extraction. The specimen preparation was the same as in the case of the AMPLICOR HPV test.
(ii) PCR amplification. The LINEAR ARRAY HPV genotyping test uses biotinylated primers to define a sequence of nucleotides within the polymorphic L1 region of the HPV genome that is approximately 450 bp long. A pool of HPV primers is designed to amplify HPV DNA from 37 genotypes including 13 HR HPV genotypes (6, 11, 16, 18, 26, 31, 33, 35, 39, 40, 42, 45, 51, 52, 53, 54, 55, 56, 58, 59, 61, 52, 64, 66, 67, 68, 69, 70, 71, 72, 73, 81, 82, 83, 84, CP6108, and IS39; HR HPV genotypes are indicated in boldface type). Capture probe sequences are located in polymorphic regions of L1 bound by these primers. An additional primer pair targets the human ß-globin gene. AmpliTaq Gold DNA polymerase is utilized for amplification of the HPV target DNA and the ß-globin control.
(iii) Hybridization reaction. Following PCR amplification, the HPV and the ß-globin amplicon were chemically denatured to form single-stranded DNA by the addition of denaturation solution. Aliquots of denatured amplicon were then transferred to the appropriate well of the typing tray that contained hybridization buffer and a single LINEAR ARRAY HPV genotyping strip that was coated with HPV and ß-globin probe lines. The biotin-labeled amplicon hybridized to the oligonucleotide probes only if the amplicon contained the matching sequence of the complementary probe.
(iv) Detection reaction. Following the hybridization reaction, the LINEAR ARRAY HPV genotyping strip was washed to remove any unbound material. Streptavidin-horseradish peroxidase conjugate was then added to the strip. The streptavidin-horseradish peroxidase conjugate bound to the biotin-labeled amplicon hybridized to the oligonucleotide probes on the strip. The strip was washed to remove any unbound conjugate, and a substrate solution containing hydrogen peroxide and TMB was added to each strip. In the presence of hydrogen peroxide, the bound streptavidin-horseradish peroxidase catalyzed the oxidation of TMB to form a blue-colored complex, which precipitated at the probe positions where hybridization occurred. The LINEAR ARRAY HPV genotyping strip was then read visually by comparing the pattern of blue lines to the reference guide. The limit of detection is reported by the manufacturer for HPV genotypes 16, 18, 31, and 45. Genotypes 16, 18, and 45 were detected with a 100% positivity rate at 300, 3,000, and 900 copies/ml, respectively. Genotype 31 was detected with a 92% positivity rate at 3,000 copies/ml.
Statistical analysis. Sensitivity, specificity, and accuracy of the AMPLICOR HPV test and of the HC2 assay were determined against the presumed HPV status based on the combination of both tests and on genotyping results in the case of discordant findings. The 95% confidence intervals were calculated using the exact method. Agreement between the two tests was assessed by Cohen's kappa statistic, with values of 0.00 to 0.20 indicating poor agreement, 0.21 to 0.40 indicating fair agreement, 0.41 to 0.60 indicating moderate agreement, 0.61 to 0.80 indicating good agreement, and 0.81 to 1.00 indicating excellent agreement. Marginal homogeneity of the two tests was assessed by McNemar's test. All tests were two-sided, and a P value of <0.05 was considered significant.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Concordance between the results of the 168 samples tested by the HC2 assay and AMPLICOR HPV testa
|
|
View this table: [in a new window] |
TABLE 2. Overview of discordant cases
|
|
View this table: [in a new window] |
TABLE 3. HC2 assay and AMPLICOR HPV test performance
|
|
View this table: [in a new window] |
TABLE 4. Results of cervical cytology related to HPV DNA detection with HC2 assay and AMPLICOR HPV test
|
![]() View larger version (17K): [in a new window] |
FIG. 1. Association between histological results and HPV testing. Both the HC2 assay and the AMPLICOR HPV test identified the patient with SCC, the AMPLICOR HPV test missed 1/5 CIN2 patients, and the HC2 assay missed 2/7 CIN3 subjects. On the figure, the percentages of HPV test positivity are reported according to histological result. Neg., negative.
|
|
|
|---|
Two recent studies evaluated the performance of the newly introduced AMPLICOR HPV test. The study by Monsonego et al. (17) assessed its performance both in the detection of cervical pathology in women with an abnormal Pap smear and in addition to the Pap smear in the screening setting. In this study, no other method for HPV detection was used, and the authors found that, together with colposcopy and high-grade squamous intraepithelial lesion cytology, the AMPLICOR HPV test is a powerful independent predictor of high-grade CIN2 and 3. The other study compared the performance of the AMPLICOR HPV test with that of the INNO-LIPA HPV detection/genotyping assay (27): the authors found the two tests to be fairly well correlated, with an absolute agreement between the two tests of 97.5%.
Instead, we found discordant results in 17% of the samples. The genotyping showed that, in all cases of HC2 assay-positive and AMPLICOR HPV test-negative samples, no HR HPV could be detected, but in 7 of 8 cases, one or more LR HPV type has been isolated. Recent studies have demonstrated that the Digene HC2 HR assay may have cross-reactivity with LR HPV types (5, 20), known to cause some cytological abnormalities, which never progress to cancer. All of the LR HPV types present in these 7 specimens had already been described as cross-reactive. In particular, in 4 samples, HPV genotype 66 has been isolated. Although this could be considered an advantageous cross-reactivity, due to the fact that HPV genotype 66 has recently been added to the HR HPV group (18), these cross-reactions can nevertheless lead to false-positive results, affecting patient management.
Among the 20 samples which were AMPLICOR HPV test positive and HC2 assay negative, an HR HPV genotype was detected in 8 samples. In 7 of the remaining specimens, no HPV type was found: this could be due to either a low viral load (the AMPLICOR HPV test has a higher sensitivity than LINEAR ARRAY HPV genotyping) or a false-positive result due to the presence of contaminants in the sample. Assuming the worst situation, in which all untyped specimens could have been false positives by the AMPLICOR HPV test, the specificity of the AMPLICOR HPV test decreased slightly but was still comparable to that of the HC2 assay. In 5 cases, LR HPV genotypes were detected. To date, there are no reports evaluating the possibility of cross-reactivity with genotypes other than the 13 included in the panel. Larger studies are probably needed to clarify this issue.
The two tests gave substantially concordant results. However, in women with normal smears there was a difference in the positivity rate between the two assays. The AMPLICOR HPV test gave a significantly higher number of positive results in this group of women, probably reflecting the higher sensitivity. This finding has to be taken into account when the AMPLICOR HPV test is used in the screening setting. As elegantly pointed out by Snijders et al. (24), it is necessary to distinguish between the clinical sensitivity and the analytical sensitivity. Whether a method with a higher analytical sensitivity would result in a better performance in terms of clinical sensitivity and specificity is still a matter of investigation. It could be suggested that, for epidemiological purposes, the highest sensitivity is better, while for clinical purposes, there might be a viral load threshold below which the HPV infection could be not relevant (10, 25, 26).
Another important measure of a diagnostic test is the rate of false-negative results. In our study, we had only one SCC patient who was correctly identified by both tests. The AMPLICOR HPV test detected all of the CIN3 patients (7 patients), while the HC2 assay failed to detect 2 of 7 patients. In these two CIN3 HC2 assay-negative patients, the genotyping demonstrated the presence of HPV genotype 16 alone in one sample and HPV genotype 39 plus HPV genotype 45 in the other. We might speculate that these samples contained a low copy number of the HPV genome or that a deficient sample quality may have played a role. Unfortunately, it was not possible to determine the viral load, but similar experiences with false-negative results for HR HPV detection with HC2 have been reported previously (9, 13). In any case, the low number of samples does not allow us to draw any definite conclusion.
The slight differences observed between the two tests by the present investigation may have some clinical implications. At present, clinical use of HPV testing is directed to screening, triage of borderline smear results, and follow-up after conservative treatment of CIN. In the screening setting, the HPV test is generally used in combination with a Pap smear. The higher rate of positive results in patients with a normal smear observed for the AMPLICOR HPV test in comparison with the HC2 assay suggests that the latter is more suitable for the screening setting, where the management of cytology-negative and HR HPV-positive patients is still a clinical dilemma. Conversely, in the present study, the AMPLICOR test showed a slightly better accuracy in detecting CIN3 in comparison with the HC2 assay. While this is a minor advantage in the screening setting, where CIN3 is a rare occurrence and the HPV test is always coupled with a Pap smear, it may represent a true improvement in the triage of borderline smear results, where the rate of high-grade lesions is higher and the correct identification of the disease is of utmost importance for the subsequent management of these patients. In addition, it has recently been emphasized that, besides detecting the presence of one of the HR HPV genotypes using pooled probes, it could be of value to distinguish between the individual types and, in particular, to detect the presence of HPV 16 and/or HPV 18. These two HPV genotypes have been shown to be far more aggressive and dangerous than the other HR HPV genotypes, with 10-year cumulative incidences of
CIN3 among HPV genotype 16-positive and HPV genotype 18-positive women of 20% and 15%, respectively (10). Castle et al. reported that women with atypical squamous cells of undetermined significance or low-grade squamous intraepithelial lesions who were HPV genotype 16 positive had the highest 2-year risk for
CIN3 compared to women who were HPV negative; in particular, this risk was calculated to be fivefold greater than the increased risk in women who were positive for other oncogenic HPV types. The authors suggest that distinguishing the higher absolute risk for cervical precancer in HPV genotype 16-positive women from the lower risk posed by other oncogenic HPV types might have clinical implications (7). Their conclusions corroborate the fact that, besides the persistence of the infection, specific HPV types bear different oncogenic potentials. Therefore, if the future management of patients with preneoplastic lesions will include not only HR HPV detection but also HPV genotyping, the combination of any HPV assay with the LINEAR ARRAY will give adequate and appropriate information.
In conclusion, the present study shows that the two commercially available kits for HR HPV detection are easy to implement in the clinical laboratory, are reproducible, and give quite comparable results. The present data also evidence some slight differences in test performance, suggesting a different optimal clinical use, which should be further evaluated in further (prospective) clinical studies.
We thank Alessio Tricca, laboratory technologist, for help in performing tests.
|
|
|---|