Alphabio Laboratory, Marseille, France,1 Department of Hepatology and Gastroenterology, Saint-Joseph Hospital, Marseille, France,2 Department of Biostatistics and Epidemiology, CDL Pharma, Marseille, France,3 Department of Hepatology and Gastroenterology, Arnault Tzanck Institute, Saint Laurent du Var, France4
Received 25 January 2006/ Returned for modification 3 April 2006/ Accepted 13 April 2006
|
|
|---|
|
|
|---|
Despite the introduction of IU, discrepancies may occur when patients are monitored using different types of assays (19). In a previous study, false-positive and false-negative results, as well as variations in the HCV RNA level of >1 to 2 log IU, were observed. These results may have an impact on the management of patients receiving interferon therapy, for example, when treatment decisions are made using a single HCV RNA determination (11). Two standardized HCV RNA quantitative assays based on real-time PCR were recently developed (Abbott RealTime HCV and Roche Cobas TaqMan HCV). Both have a high sensitivity (based on the lower limit for detecting HCV RNA) and a great dynamic range and were standardized using the second WHO panel. The principle of real-time PCR techniques is to detect amplicon synthesis and to deduce the number of viral genomes in the starting clinical sample during the PCR rather than at the end (15). The main advantage of these tests is that each has been standardized, which makes results by different laboratories comparable through reporting in IU. Because of the use of viral kinetics during polyethylene glycol (PEG)-interferon-ribavirin therapy and the development of specific new anti-HCV drugs, assessment of the efficacy of anti-HCV drugs needs to be based not on end-point PCR assays but on real-time PCR.
To determine if the new real-time PCR-based assays can become the standards for HCV RNA detection and quantitation, we investigated the prognostic relevance of HCV RNA viral loads at baseline, week 4, and week 12 to the virological response to antiviral therapy. Also, we compared the performances of the two commercially available real-time PCR assays.
|
|
|---|
All tested positive for HCV RNA and negative for human immunodeficiency virus. Other causes of chronic hepatitis were excluded by appropriate serological testing and liver histology. The HCV genotype was determined by sequence analysis using the HCV genotyping Trugene assay (Bayer, Puteaux, France) (12). There were 41 genotype 1, 5 genotype 2, 10 genotype 3, and 3 genotype 4 patients. All patients were treated with 800 mg ribavirin per day orally plus PEG-interferon alfa-2b (1.5-µg/kg subcutaneous injection once a week) for 48 weeks.
The virological response was assessed by a qualitative HCV RNA assay with a lower limit of sensitivity of 50 IU/ml (HCV Amplicor 2.0; Roche Diagnostics, Meylan, France). According to the qualitative HCV RNA results, patients were defined as virologic sustained responders (HCV RNA negative 6 months after the end of therapy) or nonresponders.
Sera for HCV RNA analysis. A total of 177 serum samples were obtained from the 59 patients. Each of the specimens was divided into four aliquots and frozen to 80°C within 2 h of collection (10). Each aliquot was used for HCV RNA quantitation by the following assays: Cobas TaqMan HCV (Roche Diagnostics, Meylan, France), with a detection limit of 15 IU/ml (7), and the Abbott RealTime HCV quantitative assay (Abbott Laboratories, Rungis, France), with a detection limit of 10 IU/ml.
The HCV RNA viral load was measured at the following three time points: baseline and 4 and 12 weeks after the beginning of therapy. SVR was determined 6 months after the end of therapy with the Amplicor qualitative assay (Roche Diagnostics, Meylan, France). Values for HCV RNA are reported in IU/ml. All assays were done at Alphabio Laboratory.
Roche Taqman HCV assay.
The Cobas TaqMan HCV test (Roche Molecular Systems, Inc., Branchburg, NJ) is a real-time nucleic acid amplification assay for quantitative detection of HCV RNA in human serum or plasma. Like the TaqMan HCV analyte-specific reagent (TaqMan HCV ASR; Roche Molecular Systems), this assay was developed for use with the recently introduced Cobas TaqMan 48 analyzer (CTM 48; Roche Molecular Systems). Amplification and detection were performed according to the manufacturer's instructions for TaqMan HCV with a CTM 48 analyzer and Amlilink software, v. 3.0.1 (Roche Diagnostics, Meylan, France). The lower limit of detection is 15 IU/ml, with
95% probability, using 1 ml of serum (7).
Abbott RealTime HCV assay. The RealTime HCV assay (Abbott Molecular Inc., Des Plaines, IL) was performed at Abbott Laboratories according to the manufacturer's specifications. Total nucleic acids were extracted from 0.5 ml serum or plasma by magnetic microparticle technology, using an Abbott m1000 automated sample preparation system. The internal control (IC), derived from the hydroxypyruvate reductase gene of the pumpkin plant, Cucurbita pepo, was introduced as armored RNA into the sample lysis buffer. The IC was processed simultaneously with each sample. RNA was captured by magnetic microparticles, washed to remove unbound sample components, and eluted. A second step consisted of adding master mix (HCV oligonucleotide reagent including primers and probes, thermostable rTth polymerase enzyme, and activation reagent) to extract nucleic acid samples into 96-well optical reaction plates. Plates were sealed and placed on an Abbott m2000rt instrument (m2000rt; Abbott Molecular Inc.) for reverse transcription, PCR amplification, and detection/quantitation.
The HCV probe is a short linear oligonucleotide labeled with a fluorophore at the 5' end and a quencher at the 3' end. The PCR primers and probe target conserved regions of the 5' untranslated region of the HCV genome. In the absence of an HCV target, the fluorescence is quenched. In the presence of the HCV target sequence, the HCV probe preferentially hybridizes to the target sequence, allowing fluorescence detection. A second set of primers and a probe labeled with a different fluorophore are used for amplification and detection of the internal control. The IC probe is a single-stranded DNA oligonucleotide with a fluorophore at the 5' end and a quencher at the 3' end. IC probe fluorescence is quenched in the absence of IC target sequences. In the presence of IC target sequences, IC probe hybridization to complementary sequences separates the fluorophore and quencher and allows fluorescence emission and detection. The noncompetitive internal control is detected at all HCV levels.
The HCV- and IC-specific probes are each labeled with a different fluorophore, thus allowing for simultaneous detection of both amplified products at each cycle. The amplification cycle at which a reactive level of fluorescent signal is detected by the m2000rt instrument is proportional to the log of the HCV RNA concentration present in the original sample. The homogeneous format and sealed PCR tray eliminate contamination by amplified products.
The assay meets the Second WHO International Standard for HCV RNA. The lower limit of detection is 12 IU/ml, with
95% probability. The dynamic range of the assay extends from 12 to 100,000,000 IU/ml. The Abbott RealTime HCV assay detects and quantitates genotypes 1 to 6.
Statistical analysis. Values are expressed as medians or means (with 95% confidence intervals). Correlations between assays were evaluated by linear regression. The chi-square test or Wilcoxon test was used for data comparison. P values of <0.05 were considered significant.
|
|
|---|
![]() View larger version (18K): [in a new window] |
FIG. 1. Correlation between the results for 59 HCV-infected patients tested by Abbott RealTime HCV and Roche Cobas TaqMan HCV, irrespective of the genotype. Results shows a good correlation between the two assays (R2 = 0.94).
|
![]() View larger version (17K): [in a new window] |
FIG. 2. Correlation between the results for genotype 1 patients tested by the Abbott and Roche assays (R2 = 0.94).
|
![]() View larger version (16K): [in a new window] |
FIG. 3. Correlation between the results for non-genotype 1 patients tested by the Abbott and Roche assays (R2 = 0.94).
|
![]() View larger version (23K): [in a new window] |
FIG. 4. Histograms of median viral loads measured by the Abbott and Roche assays for all genotypes (A), genotype 1 patients (B), and non-genotype 1 patients (C). Bars with dots stand for the Abbott assay, and bars with horizontal lines stand for the Roche assay.
|
|
View this table: [in a new window] |
TABLE 1. Mean interassay differences between Abbott and Roche RT-PCR assays
|
Figure 5 shows that there were no differences between the Abbott and Roche assays in viral load detection and viral load decline after 4 weeks and 12 weeks of treatment. The percentages of patients who became HCV RNA negative within 4 weeks and remained negative after 12 weeks were similar for both assays (P = 1.00) (Fig. 5). With the Abbott assay, all patients (14/14) who had RVR at week 4 were HCV RNA negative at week 12, and with the Roche assay, 1/15 patients who had RVR at week 4 was HCV RNA positive at week 12 (not significant).
![]() View larger version (24K): [in a new window] |
FIG. 5. Histograms of differences between the Abbott and Roche assays in viral load detection and viral load decline after 4 weeks and 12 weeks of treatment.
|
For genotype 1 patients, we found no EVR difference in PPV and NPV values between the Roche and the Abbott assays: the NPV was 100% for both, and PPVs were 24% and 29%, respectively. All genotype 1 patients with viral load drops of <2 log IU/ml at week 12 (EVR) were nonresponders (NPV = 100%) (eight patients for the Abbott assay and seven for the Roche assay). At treatment week 12, 83% (20/24) and 96% (23/24) of patients with SVR tested negative for HCV RNA by the Abbott and Roche assays, respectively (not significant).
(iii) Qualitative and quantitative RT-PCR comparison. For a subsequent population of patients with low viral loads, we compared the absolute viral loads at week 12 measured by both the qualitative HCV Amplicor 2.0 (Roche) assay and the two real-time RT-PCR assays. The quantitative assay diagnostic values were slightly higher than the qualitative assay values but were not significantly different (Table 2).
|
View this table: [in a new window] |
TABLE 2. Diagnostic values for quantitative and qualitative assays after 12 weeks of treatment
|
|
|
|---|
Regarding the absolute values at baseline, week 4, and week 12, the two assays differed significantly, without any consequence on the virological response. Considering the linearity of quantification of the different genotypes using the two assays, we observed an excellent correlation for all genotypes (R2 = 0.94), regardless of the genotype. Previous data have reported that the Cobas Taqman assay cannot provide a linear correlation with genotypes 2 and 3 compared with the Cobas Amplicor Monitor 2.0 assay (18). A recent study reported that this phenomenon can be corrected by a modified extraction procedure with linear quantification of genotypes 1 to 6 (7).
End-point-PCR-based assays (Cobas Monitor and LCx) and the signal amplification assay (Versant; Bayer) should now be called "old-generation HCV RNA quantitation," as opposed to "new-generation real-time PCR" assays, for at least the following three reasons: (i) the absence of complete automation, (ii) the need to dilute for quantification above the upper range (8), and (iii) the fact that the former assays were standardized with the first WHO panel, and due to the different secondary standards used for calibration, the IUs measured by these assays are different. In a recent comparative study (11), we demonstrated that the absolute values obtained with these assays are different and that the HCV RNA viral load decline, assessed by different assays during antiviral therapy, can give different results regardless of the use of international units.
Novel therapeutic strategies against HCV are mainly focused on the development of agents that inhibit the specific steps of the life cycle of the virus. Protease and polymerase inhibitors require the highest HCV RNA sensitivity and dynamic range, because results are mainly based on their ability to reduce the viral load and to reach the largest percentage of patients with SVR (1, 17, 20).
In conclusion, the high sensitivity and large dynamic range offered by the new real-time PCR-based assays must definitively close the discussion about the use of a qualitative or quantitative assay for HCV clinical and therapeutic management. This was confirmed by our findings related to the slightly higher NPVs and PPVs observed with real-time PCR assays than with the qualitative HCV Amplicor 2.0 assay (Table 2). These observations emphasize the need for commercial HCV RNA quantification assays with a broader range of linear quantification, such as real-time PCR-based assays. One other topic is the cost of real-time PCR technology. National health departments might review their respective reimbursement policies, as real-time PCR assay costs are equivalent to those of the older generation of assays.
The 2002 NIH consensus conference (14) gave recommendations for the use of molecular tests, in addition to enzyme-linked immunosorbent assays, for diagnosis. These recommendations concern the decision to treat, optimal treatment schedules, and assessment of the viral response to antiviral therapy. For all of these indications, we propose that a real-time quantitative PCR assay is fully adequate for clinical and therapeutic management of chronic HCV infection without additional costs.
|
|
|---|