Multicenter Comparison of Nucleic Acid Extraction Methods for Detection of Severe Acute Respiratory Syndrome Coronavirus RNA in Stool Specimens
A. Petrich,1*
J. Mahony,1
S. Chong,1
G. Broukhanski,2
F. Gharabaghi,3
G. Johnson,4
L. Louie,5
K. Luinstra,1
B. Willey,6
P. Akhaven,6
L. Chui,7
F. Jamieson,2
M. Louie,8
T. Mazzulli,6
R. Tellier,3
M. Smieja,1
W. Cai,1
M. Chernesky,1
S. E. Richardson,3 for the Ontario Laboratory Working Group for the Rapid Diagnosis of Emerging Infections
St. Joseph's Hospital, Hamilton, Ontario, Canada,1
Ontario Ministry of Health, Etobicoke, Ontario, Canada,2
Hospital for Sick Children,3
St. Michael's Hospital,4
SunnyBrook & WHC HSC,5
Mt. Sinai Hospital, Toronto, Ontario, Canada,6
Edmonton Public Health Laboratory, Edmonton, Alberta, Canada,7
Calgary Public Health Laboratory, Calgary, Alberta, McMaster University, Hamilton, Ontario, and The University of Toronto, Toronto, Ontario, Canada8
Received 28 November 2005/
Returned for modification 6 February 2006/
Accepted 24 April 2006

ABSTRACT
The emergence of a novel coronavirus (CoV) as the cause of severe
acute respiratory syndrome (SARS) catalyzed the development
of rapid diagnostic tests. Stool samples have been shown to
be appropriate for diagnostic testing for SARS CoV, although
it has been recognized to be a heterogeneous and difficult sample
that contains amplification inhibitors. Limited information
on the efficiency of extraction methods for the purification
and concentration of SARS CoV RNA from stool samples is available.
Our study objectives were to determine the optimal extraction
method for SARS CoV RNA detection and to examine the effect
of increased specimen volume for the detection of SARS CoV RNA
in stool specimens. We conducted a multicenter evaluation of
four automated and four manual extraction methods using dilutions
of viral lysate in replicate mock stool samples, followed by
quantitation of SARS CoV RNA using real-time reverse transcriptase
PCR. The sensitivities of the manual methods ranged from 50%
to 100%, with the Cortex Biochem Magazorb method, a magnetic
bead isolation method, allowing detection of all 12 positive
samples. The sensitivities of the automated methods ranged from
75% to 100%. The bioMérieux NucliSens automated extractor
and miniMag extraction methods each had a sensitivity of 100%.
Examination of the copy numbers detected and the generation
of 10-fold dilutions of the extracted material indicated that
a number of extraction methods retained inhibitory substances
that prevented optimal amplification. Increasing the volume
of sample input did improve detection. This information could
be useful for the extraction of other RNA viruses from stool
samples and demonstrates the need to evaluate extraction methods
for different specimen types.

INTRODUCTION
The emergence of a novel coronavirus (CoV) as the cause of severe
acute respiratory syndrome (SARS) and its spread throughout
the world catalyzed the development of rapid diagnostic tests.
SARS CoV has been shown to replicate in the gastrointestinal
tract (
4), and consistent with this, stool samples were shown
to be appropriate for diagnostic testing for SARS CoV. Peiris
et al. (
7) found a positivity rate of 97% (65/67 samples) for
detection of SARS CoV nucleic acid in stool samples at 14 days
after the onset of symptoms. By contrast, Chan et al. (
1) found
a lower overall stool positivity rate of 26.2% (70/267), with
a 42.9% (9/21) positivity rate within 1 week of the onset of
symptoms, a 68.0% (17/25) positivity rate between 1 and 2 weeks
of onset, and a 70.8% (34/48) positivity rate between 2 and
4 weeks of onset. Preliminary studies performed in our laboratories
indicated that variations in RNA extraction methods could explain
the differences seen in these studies. We anticipated that the
optimization of extraction methods for stool samples could potentially
increase the sensitivity of amplification detection of SARS
CoV, especially early in infection. Stool samples have been
recognized to be heterogeneous and difficult samples for use
for molecular analysis. Bile salts, hemoglobin, and polysaccharides
have been identified as factors that inhibit amplification assays
(
11,
6); and reverse transcriptase (RT) has been described to
be particularly sensitive to inhibition (
12). Although these
amplification inhibitors are well known, limited information
is available on the efficiency of various extraction methods
for the purification and concentration of SARS CoV RNA from
stool samples.
Our primary study objective was to determine the optimal extraction method for the detection of SARS CoV RNA in stool samples, and our secondary objective was to examine the effect of an increased volume of stool specimen on the ability to detect SARS CoV RNA. We conducted an international multicenter evaluation of automated and manual extraction methods using dilutions of inactivated viral lysate in replicate mock stool samples, followed by single-site quantitation of SARS CoV RNA by real-time RT-PCR.

MATERIALS AND METHODS
Virus.
A heat-inactivated cell culture lysate of SARS CoV strain Tor-2
was supplied by Martin Petric, British Columbia Centre for Disease
Control, Vancouver, British Columbia, Canada. The lysate, which
contained 10
8 50% tissue culture infective doses per ml of SARS
CoV, was used to spike the stool samples.
Generation of mock stool suspensions.
Five stool samples submitted to the Regional Virology Laboratory in Hamilton, Ontario, Canada, for routine testing were pooled and diluted with RNase-free water to make a 10% (vol/vol) suspension. An aliquot was extracted and tested by RT-PCR (5) to ensure that it was negative for SARS CoV nucleic acid. SARS CoV lysate was serially diluted in the 10% stool suspension to generate 10-fold dilutions ranging from 102 to 107. An approximate detection limit was determined by extracting a 140-µl aliquot of each dilution by use of a QIAamp viral RNA kit (QIAGEN Inc., Mississauga, Ontario, Canada) and performing a quantitative RT-PCR targeting the nucleocapsid gene with primers APNF and APNR and a TaqMan probe, as described previously (5), in replicates of three.
Panel composition.
A panel of 20 samples was assembled and sent to seven laboratories. The panel contained 12 potentially positive samples and 8 negative samples. The four dilutions of the SARS CoV viral lysate in stool suspension contained in the panel comprised samples with 1 dilution lower than the endpoint dilution, the endpoint dilution, and 2 dilutions higher than the endpoint dilution (104 and 107). It was anticipated that the 107 dilution would not be positive by all extraction methods. Three replicates for each SARS CoV dilution were included for a total of 12 possible positive results. The negative samples comprised two stool suspensions (10%; vol/vol) to which nothing was added and six stool suspensions spiked with other viral lysates (parainfluenza virus 3, influenza A virus, human coronaviruses 229E and OC43, respiratory syncytial virus, and adenovirus). The order of the samples was randomized by a computer-based random numbers sequence, and test panels were sent to two laboratories to ensure that the samples performed as expected prior to wider distribution to all seven laboratories. The participating laboratories included St. Joseph's Healthcare, Hamilton, Ontario, Canada (site 1); Mount Sinai Hospital, Toronto, Ontario, Canada (site 2); Hospital for Sick Children, Toronto, Ontario, Canada (site 3); British Columbia Center for Disease Control, Vancouver, British Columbia, Canada (site 4); Edmonton Public Health Laboratory, Edmonton, Alberta, Canada (site 5); Central Public Health Laboratory, Etobicoke, Ontario, Canada (site 6); the Centers for Disease Control and Prevention, Atlanta, Ga. (site 7); Sunnybrook and WHC HSC, Toronto, Ontario, Canada (site 8).
Nucleic acid extraction.
The panels were extracted by eight different methods (seven commercial methods and one in-house method) at seven different laboratories. Two extraction methods were each performed at two different laboratories: QIAamp Viral RNA kit with 140 µl at two sites (sites 1 and 6) and the NucliSens automated extractor method (bioMérieux, St. Laurent, Quebec, Canada) with 100 µl and 200 µl at sites 4 and 7, respectively. Four methods were performed with two different sample input volumes: the QIAamp Viral RNA method with 140 µl and 560 µl, the miniMag method (bioMérieux) with 140 µl and 200 µl, the NucliSens automated extractor method (bioMérieux) with 100 µl and 200 µl, and the QIAamp UltraSens method (QIAGEN Inc.) with 140 µl and 500 µl. The participating laboratories were asked to elute RNA into a final volume of 50 µl, dispense the volume into two aliquots, and send the extracted nucleic acid back to a central site for testing by RT-PCR. Two laboratories (site 7 and site 4) did not extract into 50 µl, as they followed protocols that they had optimized in-house: the QIAamp MDx method (QIAGEN Inc.) with 80 µl and the bioMérieux NucliSens automated extractor method with 30 µl. All methods were performed according to the manufacturers' protocols. The extraction methods included four manual methods (QIAamp UltraSens, QIAamp Viral RNA, Magazorb [Cortex Biochem, San Leandro, Calif.], and in-house guanidinium thiocyanate methods), three automated methods (the MagNA Pure [Roche Diagnostics, Laval, Quebec, Canada], QIAamp MDx [QIAGEN Inc.], and bioMérieux NucliSens automated extractor methods), and one semiautomated method (the bioMérieux MiniMag method), with sample input volumes ranging from 100 µl to 560 µl for the different methods (Table 1). One manual in-house method used guanidinium thiocyanate buffer, followed by isopropanol precipitation, as described by Johnson et al. (3). The MagNA Pure LC TNA isolation kit, which uses a lysis buffer, followed by binding of nucleic acid to magnetic particles coated with silica, was used with the MagNA Pure method. The miniMag and Cortex Biochem methods also use magnetic beads covered with silica in a less automated procedure, while the automated NucliSens method uses standard silica. The QIAamp Viral RNA kit, QIAamp UltraSens method, and the MDx method from QIAGEN use a silica gel membrane to capture the nucleic acid in a manual or automated format.
View this table:
[in this window]
[in a new window]
|
TABLE 1. Description of extraction methods and volumes used at eight different sites to compare the recovery of SARS coronavirus nucleic acid from mock stool specimens
|
Quantitation by RT-PCR.
The extracted material was sent to a central site for quantitation
by a commercially available real-time RT-PCR assay (RealArt
HPA Coronavirus kit; artus GmbH, Hamburg, Germany) for detection
of SARS CoV nucleic acid. Five microliters of purified sample
was tested by the RealArt SARS coronavirus HPA RT-PCR (artus)
assay undiluted and at a 1:10 dilution in RNase-free water to
detect the presence of inhibitors. A LightCycler 1.2 or 2.0
instrument (Roche Diagnostics) was used for amplification. It
was previously determined by an evaluation of standards and
test samples that the results obtained with the two LightCycler
instruments were interchangeable (data not shown). The control
samples used to generate the standard curve were provided by
the manufacturer. An attempt was made to test the same sample
extracted by different methods in the same run. This was not
always possible, so crossing points for the standards were collected
for all runs and analyzed. Significant differences between runs
were not detected, suggesting that there was little interrun
variability (data not shown).
Data analysis.
For qualitative test results, the results of the extraction techniques were compared by the Cochran Q test for three or more repeated samples, followed by pairwise comparisons by the McNemar test (SPSS for Windows 11.5). For quantitative test results, robust multilevel-modeling linear regression models were constructed to determine the impacts of the extraction method, the concentration, and the 1:10 dilution on the copy number (the dependent variable) detected. The model fit was assessed. Concentrations and copy numbers were log transformed, and 0.001 was added to the results with a copy number of zero. Robust regression was performed with S-Plus for Windows, version 7.0 (Insightful Corporation, Seattle, Wash.). A P value <0.05 was taken as statistically significant.

RESULTS
Panels consisting of 20 mock-infected stool samples were extracted
by eight methods, including four manual methods, three automated
methods, and one semiautomated method, in seven different laboratories
(Table
1). The extracted material was sent to a central site
for quantitation by using a commercially available real-time
RT-PCR assay (RealArt) for the detection of SARS CoV RNA. The
samples were tested undiluted and at a 1:10 dilution to detect
the presence of inhibitors.
All unspiked samples or those spiked with other viral lysates (parainfluenza virus 3, influenza A virus, human coronaviruses 229E and OC43, respiratory syncytial virus, and adenovirus) tested negative by the real-time PCR, indicating a specificity of 100% (8/8) for the RT-PCR with all extraction methods, for a total of 64/64 negative results (8 sets of 8 samples).
Manual extraction methods.
For the stool specimens that were analyzed undiluted, the sensitivities of the manual methods ranged from 50% (6/12) for the QIAamp UltraSens method (140 µl; site 8) to 91.7% (11/12) for the QIAamp Viral RNA method (140 µl) at site 1 and 66.7% (8/12) at site 3 and from 83.3% (10/12) for the in-house guanidinium thiocyanate method at site 4 to 100% (12/12) for the Cortex Biochem Magazorb method at site 5 (Table 2). The QIAamp Viral RNA method (with 140 µl of sample), which was tested at two different laboratories, had a sensitivity of 91.7% (11/12) at one site and a sensitivity of 66.7% (8/12) at the other, although this difference was not statistically significant (McNemar test, P = 0.25). There did not appear to be a difference between the extraction of 140 µl and the extraction of 560 µl with the QIAamp Viral RNA kit, as the overall numbers of positive results was similar: 91.7% (11/12) and 83.3% (10/12), respectively (P = 1.0). The QIAamp UltraSens method with a starting volume of 140 µl used undiluted detected 6/12 (50.0%) positive samples, whereas 10/12 (83.3%) positive samples were detected at a 1:10 dilution. A similar pattern was seen when the QIAamp UltraSens kit was used with a 500-µl sample volume (undiluted, 8/12; 1:10 dilution, 10/12). Therefore, there was a slight increase in positivity rates when larger volumes of stool specimen were used. The additional dilution of 1:10 appeared to enhance the positivity rates by dilution of anticipated inhibitors for both volumes, although the increase was not statistically significant.
Automated and semiautomated extraction methods.
The sensitivities obtained by the automated methods (Table
3)
ranged from 100% (12/12) for both the bioMérieux NucliSens
automated extractor method (200 µl; site 7) and the bioMérieux
miniMag method (140 µl; site 1) to 75.0% (9/12) for the
QIAamp MDx, bioMérieux miniMag (200 µl; site 1),
and the bioMérieux NucliSens automated extractor (200
µl; site 4) methods with the undiluted specimen. The results
from the two sites performing the bioMérieux NucliSens
automated extractions showed markedly different sensitivities,
with site 4 having only 9/12 positive results with the undiluted
specimen and 6/12 positive results with the specimen diluted
1:10, whereas site 7 had 12/12 positive results with the undiluted
and the diluted specimens. Site 4 used half the volume of starting
material (100 µl, whereas site 7 used 200 µl) and
eluted their samples into 30 µl. It was identified following
investigation that the samples at site 2 had been extracted
following an additional freeze-thaw step, which may have caused
some of the RNA present to degrade, decreasing the template
available for amplification.
View this table:
[in this window]
[in a new window]
|
TABLE 3. Number of SARS CoV RNA copies recovered by four automated or semiautomated extraction methods performed at four sitesa
|
Comparison of numbers of copies of nucleic acid.
Recovery of SARS CoV RNA was also evaluated by comparing the
numbers of copies of the specific target detected. While the
copy numbers determined by quantitative PCR were not adjusted
for sample input volume, they demonstrate the absolute numbers
detected when the manufacturers' recommendations for extraction
were followed or when an increase in the sample volume was attempted
to improve sensitivity. Under these conditions, the semiautomated
bioMérieux miniMag method with a starting volume of 140
µl (Table
3, method A, site 1) and the Cortex Biochem
Magazorb 200 µl (Table
2, method B, site 5) had the highest
rate of recovery of SARS CoV target RNA, with approximate stepwise
10-fold decreases between dilutions. The in-house guanidinium
thiocyanate method with a 1:10 dilution showed a similar rate
of recovery (Table
2, site 3). To visualize the extraction efficiency
over 4 log units of viral RNA, the results were graphed according
to the log dilution of the copy numbers obtained for amplification
of 5 µl of extracted material and extraction method (Fig.
1A and B). The bioMérieux miniMag method had the highest
copy number overall for all dilutions, followed by the Cortex
Biochem Magazorb method. This was consistent with all concentrations
of template. More variability was seen between the methods at
the lower concentrations, presumably due to the effect of sampling.
Dilution of the purified RNA prior to testing appeared to improve
the RNA recovery for a number of methods. These sites used the
Roche MagNA Pure method, the in-house guanidinium thiocyanate
method, the QIAamp Viral RNA method (140 µl), and the
QIAamp MDx method. Following dilution, however, not all of these
methods were able to detect the lowest concentration of spiked
specimen (e.g., the QIAamp MDx and guanidinium thiocyanate methods).
Concentrations calculated with no dilution and 1:10 dilution.
All samples were tested at a 1:10 dilution to look for the presence
of inhibitors. Actual data are shown; hence, if no inhibitors
were present, a 10-fold drop in the calculated numbers should
have been seen between the undiluted and the diluted samples.
A number of the extraction methods, including the Roche MagNAPure
and the QIAamp UltraSens methods, showed increased copy numbers
and/or more positive results in tests with the diluted samples.
This suggests that inhibitors present in the undiluted sample
decreased the level of amplification and that this effect could
be diluted out. A 10-fold drop in the calculated concentrations
was seen with the Cortex Biochem Magazorb, bioMérieux
NucliSens automated extractor, and bioMérieux miniMag
methods (Tables
2 and
3). To illustrate the presence of inhibitors
based on the copy numbers obtained from real-time amplification,
analysis by robust regression (to account for outliers and correlation
among samples) of the log dilution versus the log-transformed
mean concentration of undiluted and diluted samples was performed.
Figure
2A and B was derived from quantitative counts obtained
from material extracted by the bioMérieux miniMag method
(140 µl) at site 1 and the QIAamp Viral RNA method (140
µl) at site 8, respectively. In Fig.
2A and B, the robust
linear regressions for the undiluted and diluted samples are
graphed as solid and dashed lines, respectively. In the regression
model for method A at site 1, each log concentration was associated
with an increased copy number of 0.98 ± 0.02 (versus
an expected increase of 1.0), and the samples diluted 1:10 had
an estimated concentration of 0.95 ± 0.04 (versus
an expected log concentration of 1.0). This is demonstrated
in Fig.
2A as two parallel lines, with the log concentration
of the undiluted samples being approximately 1 log greater than
that of the diluted samples. Hence, no inhibitors were detectable
with material extracted by the bioMérieux miniMag method.
By contrast, for the QIAamp Viral RNA method (140 µl)
extraction in Fig.
2B, the regressions found a weaker association
with the log concentration of 0.45 ± 0.8 (versus an expected
log concentration of 1.0), and the samples diluted 1:10 were
associated with a 0.40 ± 0.16 greater concentration (versus
an expected log concentration of 1.0;
P = 0.02). This
is demonstrated in Fig.
2B, in which the lines representing
the undiluted and the diluted samples are reversed from those
in Fig.
2A. The diluted specimens were associated with a higher
copy number, indicating partial amplification inhibition for
undiluted specimens. The internal control from the RT-PCR kit
was amplified by all of the assays, even though the presence
of discrepant results between the undiluted and the diluted
samples demonstrated the existence of inhibitory substances.
Effect of increasing the sample volume.
To assess whether increasing the input sample volume would decrease
the detection limit, two sample volumes were extracted by three
methods: the QIAamp Viral RNA method (140 µl and 560 µl),
the QIAamp UltraSens method (140 µl and 560 µl),
and the bioMérieux miniMag method (140 µl and 200
µl). Comparison of the quantitative numbers of the lower
and higher sample input volumes showed that increasing the sample
volume did not increase the numbers of target copies detected
by real-time PCR. In fact, with the miniMag extraction method,
the number of RNA copies was lower when a larger volume of specimen
was extracted (140 µl versus 200 µl).

DISCUSSION
The optimal method for the extraction of RNA for PCR detection
would ideally provide pure nucleic acid free of amplification
inhibitors and at the same time recover RNA quantitatively across
a range of concentrations. We evaluated different extraction
methods for the purification of RNA from stool specimens for
the optimal detection of SARS CoV RNA using a commercially available
quantitative real-time RT-PCR. Four 10-fold dilutions of SARS
CoV cell lysates were spiked into negative stool samples so
that the lowest concentration would be at or close to the assay
endpoint. Not all extraction methods were expected to be able
to recover detectable levels of RNA. The two methods that gave
the highest overall numbers of copies per 5 µl for all
samples were the miniMag (bioMérieux) and the Magazorb
(Cortex Biochem) methods. Both methods use magnetic beads coated
with silica in their methodologies. The Roche MagNA Pure method
also uses magnetic beads; however, it did not perform as well,
giving lower copy numbers and lower percentages of positivity,
often in the presence of amplification inhibitors. Without knowledge
of the proprietary ingredients of the kits, the reason for the
improved performance of the miniMag and Cortex Biochem Magazorb
methods compared with those of the other methods can only be
surmised but presumably is due to the composition of the magnetic
beads or the surface area available for the binding of nucleic
acid. The nature of the wash procedures, including the buffer
volume and composition, may also affect the purity and yield
of the nucleic acid. The internal control of the RealArt assay
was detected in tests with all undiluted and diluted samples;
however, the crossing point (the cycle threshold number) was
shifted later in some cases by as much as three cycles, which
represents a 1-log quantity of RNA. With such a high internal
control signal, weakly inhibited samples may still have a positive
internal control signal but the inhibition would be enough to
make a low-level-positive sample negative. This implies that
users should view the internal control signal cautiously and
should understand that the concentration of RNA in the internal
control may be higher than that found in some weakly positive
samples. Users may wish to examine crossing points for internal
controls to see if they have shifted and/or consider running
a dilution of the sample if they believe that inhibition is
possible.
The inclusion of replicates of dilutions at or near the detection limit in the panel, as previously described by our group for Chlamydia pneumoniae detection (2, 9), were meant to aggressively challenge the extraction methods. All methods have positive results for the samples at the higher RNA concentrations; however, as the concentration drops, differences between the methods can clearly be determined. The use of replicates minimizes the differences obtained over those obtained by the interpretation of results based on a single sample point, especially with samples that contain a low concentration of template and that are therefore susceptible to sampling bias. This strategy allowed us to demonstrate that there are differences in the ability of extraction methods to purify and recover RNA from stool samples that could be clinically significant. Automated extractors may not perform the best with this specimen type, and the advantages of high throughput may sacrifice performance. In our hands, a manual method and a semiautomated method performed the best for the quantitative recovery of RNA, suggesting that laboratories with limited resources and/or lower sample volumes can find manual extraction methods that do not require expensive instrumentation. For life-threatening infections (SARS and avian influenza), performance cannot be jeopardized, particularly when specimens that are taken early in the course of the infection and that may have low viral RNA copy numbers are tested.
In this study, five individual stool samples were pooled and a 10% suspension was generated; thus, a single sample matrix spiked with different concentrations of SARS CoV was used for all comparisons. One limitation of this approach was that it may not represent the specimen heterogeneity found in this sample type. Fortuitously, this particular sample matrix contained PCR amplification inhibitors, making it an excellent matrix for comparison of the abilities of various extraction methods to extract RNA and remove inhibitors. Performing quantitation with samples that were undiluted and diluted 1:10 allowed us to look for partial or complete inhibition. Only three of the extraction methods (the miniMag, Magazorb, and NucliSens automated extractor methods) clearly showed a 10-fold drop in counts from undiluted to diluted material, demonstrating a lack of inhibitors in the purified preparation. Some of the methods actually had higher numbers for the diluted sample than for the undiluted sample, suggesting the presence of a low level of amplification inhibitors. It is clear from this evaluation that ideal results, including assay sensitivity and quantitation, require an optimal extraction method. Not only were there differences in the number of positive samples detected following different extraction methods, but there were also considerable variations in the number of copies of SARS CoV template recovered by each extraction method. In this study the MiniMag extraction method (site 1) with the 104 dilution of viral lysate had a copy number of approximately 17,000, while the QIAamp MDx method demonstrated only
80 copies/5 ml. The assay was positive by both extraction methods with this concentration of target in the sample, demonstrating that reliable quantitative results can be achieved only when a good extraction method is married to a good amplification assay and linear results are obtained over a range. Care must be taken to ensure that a linear result is achieved over a large dynamic range of template concentration by using optimal extraction and amplification methods.
Poon et al. (8) were able to increase the sensitivity of the early detection of SARS CoV from nasopharyngeal aspirates from 22% to 80% by increasing the sample volume extracted (140 µl to 540 µl) and by incorporating a quantitative real-time RT-PCR for amplification. Wang et al. (10) were able to achieve a detection rate of 80% with 116 plasma samples from 44 SARS patients by increasing the amount of viral RNA input at three steps during the assay: (i) by increasing the sample input for extraction, (ii) by increasing the volume of purified nucleic acid in the RT reaction, and (iii) by using 100% of the cDNA from the RT step in the subsequent PCR. We could not verify a similar increase in sensitivity when we increased the sample input using three extraction methods. In contrast, the sensitivity remained the same or was decreased. This may be due to the difference in the type of sample tested and the load of cells and/or bacteria present in the sample. Too much nucleic acid in a sample may saturate the binding material used for extraction and exclude the desired target. Increasing the amount of starting material for some sample types will also increase the amount of inhibitors present in the purified eluate. High concentrations of nucleic acid are also known to interfere with amplification and prevent primers from finding the desired target nucleic acid.
This study identified the optimal methods for the extraction of SARS CoV RNA in stool samples. The results may be generalizable to other viruses present in stool samples; however, the physical properties of the virus (e.g., enveloped or nonenveloped) and the nature of the disease caused by the virus, which may result in variations in the sample matrix or differences in the virus copy numbers present, must be considered. Ideally, an evaluation of extraction methods should be undertaken for each virus and sample type being examined. Similar studies are warranted for optimization of the recovery of viral RNA and DNA in other specimens, in particular, respiratory specimens, for the detection of viruses which carry a high mortality rate, such as SARS CoV and avian influenza virus.

ACKNOWLEDGMENTS
We thank Martin Petric (British Columbia CDC, Vancouver, British
Columbia, Canada) for the SARS CoV (Tor-2 strain)-infected cell
lysate, Thomas F. Smith (Mayo Clinic, Rochester, Minn.) for
human respiratory coronavirus, and Dean Erdman (CDC, Atlanta,
GA) for allowing his laboratory to serve as an extraction site.
This study was funded in part by a grant from the Ontario Ministry of Health and Long-Term Care and the Canadian Institutes of Health Research.

FOOTNOTES
* Corresponding author. Mailing address: St. Joseph's Healthcare, L424, Microbiology, 50 Charlton Ave. East, Hamilton, ON L8N 4A6, Canada. Phone: (905) 522-1155, ext. 3270. Fax: (905) 521-6083. E-mail:
petricha{at}mcmaster.ca.


REFERENCES
1 - Chan, P. K. S., W.-K. To, K.-C. Ng, R. K. Y. Lam, T.-K. Ng, R. C. W. Chan, A. Wu, W.-C. Yu, N. Lee, D. S. C. Hui, S.-T. Lai, E. K. L. Hon, C.-K. Li, J. Y. Sung, and J. S. Tam. 2004. Laboratory diagnosis of SARS. Emerg. Infect. Dis. 10:825-831.[Medline]
2 - Chernesky, M., M. Smieja, J. Schachter, J. Summersgill, L. Schindler, N. Solomon, K. Campbell, L. A. Campbell, A. Cappuccio, C. Gaydos, S. Chong, J. Moncada, J. Phillips, D. Jang, B. J. Wood, A. Petrich, M. Hammerschlag, M. Cerney, and J. B. Mahony. 2002. Comparison of an industry-derived LCx Chlamydia pneumoniae PCR research kit to in-house assays performed in five laboratories. J. Clin. Microbiol. 40:2357-2362.[Abstract/Free Full Text]
3 - Johnson, G., S. Nelson, M. Petric, and R. Tellier. 2000. Comprehensive PCR-based assay for detection and species identification of human herpesviruses. J. Clin. Microbiol. 38:3274-3279.[Abstract/Free Full Text]
4 - Leung, W. K., K.-F. To, P. K. Chan, H. L. Chan, A. K. Wu, N. Lee, K. Y. Yuen, and J. J. Sung. 2003. Enteric involvement of severe acute respiratory syndrome-associated coronavirus infection. Gastroenterology 125:1011-1017.[CrossRef][Medline]
5 - Mahony, J. B., A. Petrich, L. Louie, X. Song, S. Chong, M. Smieja, M. Chernesky, M. Loeb, S. Richardson, and the Ontario Laboratory Working Group for the Rapid Diagnosis of Emerging Infections. 2004. Performance and cost evaluation of a commercial and six in-house conventional and real-time reverse transcription-PCR assays for detection of severe acute respiratory syndrome coronavirus. J. Clin. Microbiol. 42:1471-1476.[Abstract/Free Full Text]
6 - Monteiro, L., D. Bonnemaison, A. Vekris, K. G. Petry, J. Bonnet, R. Vidal, J. Cabrita, and F. Mégraud. 1997. Complex polysaccharides as PCR inhibitors in feces: Helicobacter pylori model. J. Clin. Microbiol. 35:995-998.[Abstract]
7 - Peiris, J. S. 2003. Severe acute respiratory syndrome. J. Clin. Virol. 28:245-247.[CrossRef][Medline]
8 - Poon, L. L. M., K. H. Chan, O. K. Wong, W. C. Yam, K. Y. Yuen, Y. Guan, Y. M. D. Lo, and J. S. M. Peiris. 2003. Early diagnosis of SARS coronavirus infection by real time RT-PCR. J. Clin. Virol. 28:233-238.[CrossRef][Medline]
9 - Smieja, M., J. B. Mahony, C. H. Goldsmith, S. Chong, A. Petrich, and M. Chernesky. 2001. Replicate PCR testing and probit analysis for detection and quantitation of Chlamydia pneumoniae in clinical specimens. J. Clin. Microbiol. 39:1796-1801.[Abstract/Free Full Text]
10 - Wang, H., Y. Mao, L. Ju, J. Zhang, Z. Liu, X. Zhou, Q. Li, Y. Wang, S. Kim, and L. Zhang. 2004. Detection and monitoring of SARS coronavirus in the plasma and peripheral blood lymphocytes of patients with severe acute respiratory syndrome. Clin. Chem. 50:1237-1240.[Free Full Text]
11 - Widjojoatmodjo, M. N., A. C. Fluit, R. Torensma, G. P. H. T. Verdonk, and J. Verhoef. 1992. The magnetic immuno polymerase chain reaction assay for direct detection of salmonellae in fecal specimens. J. Clin. Microbiol. 30:3195-3199.[Abstract/Free Full Text]
12 - Wilde, J., J. Eiden, and R. Yolken. 1990. Removal of inhibitory substances from human fecal specimen for detection of group A rotaviruses by reverse transcriptase and polymerase chain reaction. J. Clin. Microbiol. 28:1300-1307.[Abstract/Free Full Text]
Journal of Clinical Microbiology, August 2006, p. 2681-2688, Vol. 44, No. 8
0095-1137/06/$08.00+0 doi:10.1128/JCM.02460-05
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Mahony, J. B.
(2008). Detection of Respiratory Viruses by Molecular Methods. Clin. Microbiol. Rev.
21: 716-747
[Abstract]
[Full Text]
-
Dundas, N., Leos, N. K., Mitui, M., Revell, P., Rogers, B. B.
(2008). Comparison of Automated Nucleic Acid Extraction Methods with Manual Extraction. J. Mol. Diagn.
10: 311-316
[Abstract]
[Full Text]
-
Ginocchio, C. C., Manji, R., Lotlikar, M., Zhang, F.
(2008). Clinical Evaluation of NucliSENS Magnetic Extraction and NucliSENS Analyte-Specific Reagents for Real-Time Detection of Human Metapneumovirus in Pediatric Respiratory Specimens. J. Clin. Microbiol.
46: 1274-1280
[Abstract]
[Full Text]