A number of manual, semiautomated, and automated systems are reported to produce acceptable identification (ID) results for S. pneumoniae, beta-hemolytic streptococci, and enterococcal species (26, 27). These systems, however, were shown not to be sufficiently accurate in identifying streptococci of the viridans group (13, 20), organisms of complex taxonomy (2, 10, 26). The performance of some automated systems has been evaluated with regard to catalase-negative, gram-positive cocci (7, 13, 14, 17, 24). Reproducibility and accuracy of results, turnaround time, availability of data for epidemiological monitoring, and cost-effectiveness constitute the main reasons supporting the choice of automated systems.
Becton Dickinson (BD Diagnostic Systems, Sparks, MD) has introduced the Phoenix automated microbiology system for ID and antimicrobial susceptibility testing (AST) of human pathogenic bacteria, including enterobacteria, nonfermenting gram-negative bacteria, staphylococci, and enterococci (5, 8, 11, 25). Recently, the SMIC/ID-2 panel, dedicated to ID and AST of streptococcal species, was launched (15, 18). This study was designed to evaluate the performance of the Phoenix system for identification of streptococcal and enterococcal isolates at the species level.
|
|
|---|
Phoenix system procedures. The Phoenix system uses different panels for gram-positive cocci. The SMIC/ID-2 panel is dedicated to streptococci and the PMIC/ID-14 panel to enterococci and staphylococci. All panels include two separate sections: wells on the left contain ID substrates, and wells on the right side are dedicated to AST. Panel inoculation was performed according to the manufacturer's instructions. Both panel sections were inoculated, but only ID results have been taken into consideration for this study. After overnight culture, bacteria were suspended in the ID broth. Turbidity was adjusted to a 0.5 McFarland standard by using the CrystalSpec Nephelometer (Becton Dickinson). Panels were then sealed, logged, loaded into the instrument, and incubated at 35°C. Kinetic, colorimetric, and fluorescent signals were automatically collected by the instrument every 20 min until results were completed.
Comparator biochemical ID method. Two different API ID systems (bioMérieux, Marcy l'Étoile, France) were used to identify streptococcal and enterococcal isolates at the species level. The API 20 Strep system was used for beta-hemolytic streptococci. The rapid ID 32 Strep system was used for enterococci and non-beta-hemolytic streptococci. Inoculation, reading, and interpretation of panels were performed according to the manufacturer's instructions.
Data analysis and resolution of discrepancies. Isolates that were equally identified at the species level by both the API and the Phoenix systems were considered to be correctly identified and included in the "concordant ID" category. Due to the inability of the Phoenix system to discriminate between Enterococcus casseliflavus and Enterococcus gallinarum, the classification of an isolate as E. casseliflavus/E. gallinarum by the Phoenix system was considered concordant when the isolate was identified by the API system as either E. casseliflavus or E. gallinarum. Isolates with discordant species ID (i.e., an ID produced by the Phoenix system that differed from that obtained with the API system) were retested using both the Phoenix and the API systems. When discrepant results persisted, bacterial ID was investigated by molecular methods.
Amplification and sequencing of the 16S rRNA gene. Bacterial DNA was extracted from pure cultures by using a QIAamp DNA mini kit (QIAGEN, Basel, Switzerland). DNA eluted in Tris-EDTA was stored at 80°C. ABI 2400 thermal cyclers and PCR reagents were from Applied Biosystems (Foster City, CA). AmpliTaq Gold polymerase, PCR buffer II, standard deoxynucleoside triphosphate mixture, and the universal 16S rRNA gene primers 8f (5'-GAGAGTTTGATCCTGGCTCAG-3') and 1492r (5'-TACGGCTACCTTGTTACGACT-3') were used to produce a 1,498-bp amplicon (23). Amplification products were purified using a Mini Elut PCR purification kit (QIAGEN) and directly sequenced using an ABI 310 genetic analyzer (Applied Biosystems). Sequences for both DNA strands were determined, each by using the product of a different PCR as a template. Analysis and comparison of sequence data were carried out at the BLAST interface (http://www.ncbi.nlm.nih.gov/BLAST/) and ClustalW interface (http://www.ebi.ac.uk/clustalw/) websites.
Amplification and sequencing of housekeeping genes of viridans group streptococci. Samples were amplified with degenerate primers specific for the internal fragments of the zwf (encoding glucose-6-phosphate dehydrogenase) and gki (encoding glucose kinase) streptococcal genes (21). Two primer pairs were used: 5'-CCG(T/G)ATCGACCATTA(T/C)CTTGG(T/C)AAGG-3' and 5'-TC(A/T)GTCAG(T/A)CGTTTACCTGT(A/G)CGGA-3' for the zwf gene and 5'-GGCATTGGAATGGGATCACCAGG-3' and 5'-CCGATAA(C/T)TCCAGCGTCATTTCC-3' for the gki gene. Amplicons of 453 bp (zwf) and 624 bp (gki) were directly sequenced.
Amplification of housekeeping genes in enterococci. The ID of E. faecium and E. faecalis isolates was confirmed by amplification of a fragment internal to the ddl gene encoding a D-Ala-D-Ala ligase (4, 6). The reaction mixture contained 250 ng of DNA as a template, 50 pmol of each primer, 200 pmol per liter of each deoxynucleoside triphosphate (dATP, dCTP, dGTP, and dTTP), 10 mM Tris-HCl (pH 8.8), 1.5 mM MgCl2, 50 mM KCl, and 2 U of AmpliTaq Gold. Upon electrophoresis on a 2% agarose gel, ethidium bromide-stained DNA fragments were visualized under UV light with a Kodak CF440 camera (NEN Life Science Products, Boston, MA).
Quality controls. The following strains were included in each run: S. pneumoniae ATCC 49619, S. agalactiae ATCC 13813, and E. faecalis ATCC 29212. The identification results obtained with the above-mentioned reference strains were consistently satisfactory.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Results of biochemical identification of streptococci and enterococci by the Phoenix system, with the API system used as a comparatora
|
Analysis of discrepancies. Species ID of 20 discordant isolates was investigated by molecular methods. As shown in Table 2, sequencing of the 16S rRNA gene assessed that one S. mitis isolate was not identified by either the API or the Phoenix system. One S. pyogenes isolate (not identified by the Phoenix system) was correctly identified by the API system. Five isolates (which had been assigned to the S. mitis or Streptococcus salivarius group by the API system) did belong to the S. sanguinis group (Streptococcus parasanguinis, n = 2; S. sanguinis, n = 2; and Streptococcus gordonii, n = 1). The above-mentioned five isolates were correctly classified at the species level by the Phoenix system, not by the API system. One discordant isolate (resolved by 16S rRNA gene sequencing as S. gordonii) was not identified by the Phoenix system and was identified only at the group level by the API. Sequencing of the 16S rRNA and the housekeeping zwf and gki genes failed to resolve five additional isolates at the species level. The most probable molecular ID for those five isolates appeared to be S. mitis, Streptococcus oralis, or S. pneumoniae (species belonging to the S. mitis group) (26).
|
View this table: [in a new window] |
TABLE 2. Analysis of 20 discordant isolatesa
|
The performance of the Phoenix and API systems with regard to discordant isolates is summarized in Table 3. Of 13 streptococci, 9 were correctly identified at the species or group level by the Phoenix system and 5 by the API system. Three isolates could not be identified by either system. The correct IDs for seven discordant enterococci were given by the API system, not by the Phoenix system. Of the streptococci that could not be resolved at the species level by molecular methods, 4/5 were identified at the group level by both the API and the Phoenix systems. Overall, taking into consideration species IDs given by molecular methods, the accuracy of the Phoenix system in identifying streptococci rose from 89.9% to 93.8%. For enterococci, accuracy of ID at the species level remained at 90%.
|
View this table: [in a new window] |
TABLE 3. Analysis of 20 discordant isolates
|
|
|
|---|
The Phoenix automated system did agree with the API system for 89.9% of streptococcal IDs. Upon resolution of discrepancies, accuracy for streptococci rose to 93.8%. The performance of the new SMIC/ID-2 panel dedicated to streptococci was excellent for beta-hemolytic streptococci (49/50) and S. pneumoniae (45/46). Only one S. pyogenes isolate and one S. pneumoniae isolate were not identified. It should be noted that the latter isolate (reported by the API system as S. pneumoniae) was ultimately identified as S. mitis by molecular methods. This brings the accuracy for S. pneumoniae to 100% and underlines difficulties that may be encountered in the biochemical identification of streptococcal isolates by commercial methods (3, 16, 22).
Taken together, the results confirm the documented ability of automated systems in identifying beta-hemolytic streptococci and S. pneumoniae (18, 24). Kanemitsu et al. (18) reported that the Phoenix system performed satisfactorily with regard to beta-hemolytic streptococci (>90% concordance with a manual biochemical test supplemented by hemolysis data and serological grouping) and behaved less brilliantly with S. pneumoniae (85.9% concordance). Better performances with the Phoenix system were reported by Hirakata et al. (15); concordance with the comparator (phenotypic tests and serological grouping) was >90% for S. pneumoniae and >95% for beta-hemolytic streptococci.
With regard to viridans group streptococci, the performance of automated systems has been reported as problematic; only 55% (6/11) of S. bovis isolates and 40% (6/15) of viridans group streptococci were correctly identified at the species level by the VITEK 2 system (13). On the other hand, the cited Japanese studies on Phoenix panels evaluated IDs of viridans streptococci only at the group level (S. anginosus group or S. mitis group) (15). The performance of the Phoenix SMIC/ID-2 panel for species ID of viridans streptococci was evaluated for the first time by this study. The results for the Phoenix system were in agreement either with the API system or with molecular methods for 26/31 viridans streptococci (83.9%). Discrepancies between the Phoenix and the API systems were encountered especially within the S. mitis group, possibly due to close genetic relations among members of this group (19). Four of eight discordant isolates belonging to the S. mitis group were not resolved by molecular methods. The remaining four isolates were correctly identified exclusively by the Phoenix system (S. parasanguinis, n = 2; S. sanguinis, n = 2). Thus, the Phoenix system appeared to correctly identify 8/9 members of the S. sanguinis group.
Among enterococci, correct IDs were achieved in 90% of cases by the Phoenix system. Discrepancies were limited to E. faecalis and E. faecium. Problems in identifying enterococci with automated systems have already been reported. For instance, the VITEK 2 system failed to identify substantial numbers (9% to 37%) of E. faecium and E. faecalis isolates (1, 7, 12). The latter isolates were most frequently identified as E. casseliflavus/E. gallinarum. Similarly, the Phoenix system has been reported to misidentify E. faecium and E. faecalis as E. casseliflavus/E. gallinarum (5, 11). The present results show that automated ID of enterococci remains a problem. In a clinical laboratory, however, the simple motility test usually allows for discrimination of E. casseliflavus and E. gallinarum from other enterococci (27), thus improving ID accuracy.
In conclusion, the Phoenix system appears a reliable tool for identification of clinically relevant streptococcal and enterococcal species. The new SMIC/ID-2 panel proved particularly effective for beta-hemolytic streptococci and pneumococci. Though not perfect, ID performance with viridans group streptococci appeared to be superior to those of currently available systems.
This work was supported by grants from the Italian Ministry Education, University and Scientific Research (MIUR, Rome, Italy), and the Italian Ministry of Health (ISS, Rome, Italy).
|
|
|---|