Journal of Clinical Microbiology, January 2007, p. 231-233, Vol. 45, No. 1
0095-1137/07/$08.00+0 doi:10.1128/JCM.01243-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Department of Medical Microbiology and Infectious Diseases, University of Manitoba, Winnipeg, Manitoba,1 National Microbiology Laboratory, Public Health Agency of Canada, Winnipeg, Manitoba, Canada2
Received 16 June 2006/ Returned for modification 11 August 2006/ Accepted 25 October 2006
|
|
|---|
|
|
|---|
Two panels of sera were used for the analysis (n = 94). The low-avidity panel consisted of 49 sera collected from acute rubella cases as previously described (15). The high-avidity panel consisted of 53 sera from pregnant women undergoing routine prenatal rubella screening.
An in-house rubella virus IgG avidity assay based on the method developed and described by Thomas and Morgan-Capner (11, 12) was used as the gold standard avidity assay. Briefly, a 1:10 dilution of rubella virus antigen (BioGenesis, Poole, United Kingdom) was prepared in phosphate-buffered saline and sonicated in a bath style sonicator for 5 min. The diluted antigen was added to coating buffer (0.015 M sodium carbonate, 0.035 M sodium bicarbonate, pH 9.6) to a final dilution of 1:1,000. Fifty microliters was added to each well of a Nunc Polysorb microtiter plate (VWR, Mississauga, Canada), which was sealed and incubated overnight at 4°C. Control and patient sera were diluted 1:200 in 2% normal goat serum (NGS) (Invitrogen, Burlington, Canada) in phosphate-buffered saline plus 0.05% Tween (PBST). The plates were washed three times with PBST before 50 µl of the diluted serum was inoculated into duplicate wells of the microtiter plates. The plates were then incubated at room temperature for 1 hour in a humidified chamber. After aspiration of the dilute serum from the plate, 100 µl of 2% NGS-PBST was added to the wells (untreated wells) in parallel with 100 µl of 35 mM diethylamine (Sigma-Aldrich, Oakville, Canada) in 2% NGS-PBST (treated wells). The plates were incubated at room temperature for 5 min and then aspirated. This cycle was repeated a total of three times and then followed by six washes with PBST. Fifty microliters of a 1:2,500 dilution of goat anti-human IgG with horseradish peroxidase conjugate (Jackson, West Grove, PA) was added per well. The plates were reincubated for 1 h in a humidified chamber. After the plates were washed three more times with PBST, 50 µl of o-phenylenediamine dihydrochloride substrate and buffer (Sigma-Aldrich, Oakville, Canada) in deionized water was added to each well. The plates were incubated in the dark for 15 min before 50 µl 2 M sulfuric acid was added. The optical densities (OD) were read on a Vmax Kinetic Microplate Reader (Molecular Devices, Sunnyvale, CA) at 490 nm using SoftMAX Pro 2.6.1 software (Molecular Devices, Sunnyvale, CA). Avidity indices were calculated using the following formula: avidity index = (OD of diethylamine-washed well/OD of buffer-washed well) x 100 (%).
Figure 1 shows the results of the in-house rubella virus IgG avidity assay. The low-avidity panel clearly yielded low-avidity results (<50%), and the high-avidity panel clearly yielded high-avidity results (>60%), thus defining the in-house assay as the gold standard for the purpose of this study.
![]() View larger version (7K): [in a new window] |
FIG. 1. Performance of the in-house gold standard rubella virus IgG avidity assay using the low-avidity (diamonds) and high-avidity (squares) panels. The high-avidity (>60%) and low-avidity (<50%) cutoff values are indicated by the dashed horizontal lines.
|
The high- and low-avidity serum panels were run on each of the commercial rubella virus avidity assays (Fig. 2). Replicate testing was performed on separate days. The mean of replicate runs was used for this analysis. The weighted kappa statistic was used to assess the level of agreement between the gold standard and each commercial assay (Table 1). A weighted kappa of greater than 0.8 was considered excellent, 0.6 to 0.8 good, 0.4 to 0.6 fair, and less than 0.4 poor.
![]() View larger version (23K): [in a new window] |
FIG. 2. Scatterplots of commercial rubella virus IgG avidity assays with respect to the in-house gold standard assay. (a) Adaltis. (b) Behring. (c) Diesse. (d) Euroimmun. (e) Radim. High- and low-avidity cutoff values are indicated for the in-house gold standard assay (vertical dashed lines) and for the commercial assays (horizontal dashed lines).
|
|
View this table: [in a new window] |
TABLE 1. Weighted kappa values for commercial rubella virus avidity assays compared to in-house gold standard
|
0.40 in the untreated well in order for results to be interpretable. This may in part account for the greater number of incalculable results for the Adaltis kit (reflected in the n of 66 in the complete data set in Table 1). Assays producing measurable results were included in the calculation of the weighted kappa. The results of the comparison of each kit to the gold standard are presented in Table 1 and Fig. 2. The weighted kappa statistics for the Euroimmun and Radim assays were 0.95 and 0.94, respectively, demonstrating excellent correlation with the gold standard. The weighted kappa values for the Adaltis and Diesse kits were 0.28 and 0.33, respectively, indicated poor correlation with the gold standard. The kappa value for the Behring kit, 0.68, indicated good correlation with the gold standard. Since incalculable results were not included in this analysis, each weighted kappa was based on a different sample size. To determine whether the results were affected by these slightly different sample sizes, the entire analysis was repeated on a reduced common data set of 66 samples, for which the in-house gold standard assay and each commercial kit produced a measurable avidity result. The kappa values for the Euroimmun and Radim assays (0.95 and 0.94, respectively) were statistically significantly better than those of the other assays using the z test (P < 0.01) for both the full and reduced data sets. In conclusion, the Euroimmun and Radim rubella virus IgG avidity assays performed well and can be considered reliable commercial assays.
Published ahead of print on 8 November 2006. ![]()
|
|
|---|
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»