Journal of Clinical Microbiology, March 2007, p. 897-901, Vol. 45, No. 3
0095-1137/07/$08.00+0 doi:10.1128/JCM.02022-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Département de Microbiologie-EA3882,1 Service de Gynécologie, CHU Brest, France,3 Laboratoire de Virologie,2 Service de Gynécologie-Obstetrique,4 Centre Antivénérien,5 Centre de Protection Maternelle et Infantile,6 Laboratoire de Cytologie, CHU Angers, France7
Received 30 September 2006/ Returned for modification 13 November 2006/ Accepted 5 January 2007
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
Efficient HPV culturing remains elusive, and the clinical performance of serological assays is still poor. Thus, diagnosis of HPV infection is based almost entirely on molecular tools, including liquid hybridization (e.g., hybrid capture), Southern and dot blot hybridization with HPV type-specific probes, type-specific PCR, and general-primer PCR. Several general PCR primers have been developed to detect a broad spectrum of HPV genotypes. The majority of large studies to date have been performed with the MY 09/11, the GP5+/6+, and the SPF10 general primer sets allowing the amplification of the L1 HPV region. Various methods have been described for detection and identification of HPV genotypes after amplification with general PCR primers, such as the Amplicor HPV assay combined with linear array (Roche Diagnostics) or the SPF10-Line probe assay (LiPA; Innogenetics) showing similar results (10, 15, 18, 19, 23, 24).
In this study, we propose a new, highly sensitive, real-time general PCR that will allow quantification and typing of more than 50 HPV genotypes. This method can also be used for urine samples, permitting mass screening of HPV genital infections.
| MATERIALS AND METHODS |
|---|
|
|
|---|
DNA isolation. DNA was extracted from 200 µl of cervical samples using a QIAamp DNA mini kit (QIAGEN, Courtaboeuf, France), as recommended by the manufacturer. Briefly, sample lysis was obtained by proteinase K digestion at 56°C for 10 min, and the lysate was then loaded on the QIAamp column. After two washes, DNA was eluted with 200 µl of elution buffer provided in the kit. For DNA extraction from urine, a QIAamp viral RNA mini kit from QIAGEN was used. As recommended by the manufacturer, 1 ml of urine was centrifuged for 1 h at 23,000 x g, and the pellet was resuspended with 200 µl of sterile phosphate buffer. After lysis, the suspension was loaded onto the QIAamp column; DNA was then washed twice and eluted with 50 µl of elution buffer. Extracted DNA was kept at 80°C before PCR analysis.
Plasmid and cell controls. HPV type 6 plasmid (pHP6) was previously constructed in the Angers virology laboratory using the pMos Blue vector (Amersham, France). A 73-base amplified DNA in the L1 region was introduced in the unique EcoRV restriction site of pMos Blue. The insertion sequence was controlled by sequencing pHP6 using the Ceq2000 capillary system (Beckman), and BLAST analysis was performed on the NCBI website. Plasmid concentration was measured by DNA spectrophotometric analysis at 260 nm (160 ng/µl or 5 x 1010 DNA copies/µl).
The SiHa cell line, kindly provided to the Brest virology laboratory by Marianne Coste (Virology Laboratory, Nantes, France), is considered to contain one to two copies of HPV16 per cell (average of 1.5 copies according to our calculation). These cells were cultured in minimal essential medium and 10% fetal calf serum, trypsinized at confluence, and pelleted; cells were then resuspended in sterile phosphate buffer to get 107 HPV DNA copies per ml. Five 10-fold serial dilutions in phosphate buffer from this suspension were done. DNA was extracted from these pure and diluted cell suspensions with the QIAamp DNA mini kit as for cervical samples.
We used pHP6 and SiHa cells to assess the sensitivity and the reproducibility of our method.
Real-time PCR. HPV DNA amplification and quantification were carried out in the Mx4000 (Stratagene Europe, Amsterdam, The Netherlands) and in the LightCycler (Roche Diagnostics, Meylan, France), used, respectively, in the Angers and Brest virology laboratories. An SYBR green protocol was assessed on these instruments with modified SPF10 general primers (Table 1) allowing detection of about 40 HPV genotypes (9). Modification with 4 more bases at the 5' end of the original SPF primers adjusted to several HPV genotypes allowed hybridization at 60°C and a two-amplification step protocol at 60 and 95°C for real-time PCR. The addition of a biotin in the 5' end of reverse primers allowed direct typing of HPV-positive samples using the LiPA. Forward and reverse modified SPF primers were pooled at 2.5 pmol/µl. Five microliters of extracted DNA from samples was added to SYBR green master mix (Stratagene for Mx4000 or Roche Diagnostics for LightCycler) with forward and reverse modified SPF primers at 100 nM in a final volume of 25 µl for Mx4000 or 20 µl for the LightCycler. Thermal cycling was one step at 95°C for 10 min, followed by 45 cycles at 95°C for 30 s and 60°C for 1 min using the Mx4000 or 95°C for 15 s and 60°C for 30 s using the LightCycler. A melting curve was added between 55 and 95°C. The expected melting temperature for HPV L1 SPF amplicons was around 72°C (± 1°C) using Mx4000 and 77°C (± 1°C) using LightCycler.
|
Series of 10-fold diluted SiHa cells and diluted pHP6, corresponding to, respectively, 5, 50, 500, 5,000, and 50,000 DNA copies of HPV16 and 25, 250, 2,500, 25,000, and 250,000 DNA copies of HPV6 were run in parallel to plot the standard curves for HPV DNA quantification in samples. Triplicate dilution series were included as standards in each run. The HPV copy number in samples was evaluated using regression analysis. Corresponding CT values were reported to the standard curve allowing sample results expressed as negative (no CT) or positive, and in the latter, quantitatively in log10 DNA copies/ml of cervical medium or urine. All assays were performed in duplicate on the Mx4000 and in a single test on the LightCycler.
Optimization of our protocols on the Mx4000 and LightCycler was obtained by evaluation of primer concentrations in a range of 100 to 300 nM, MgCl2 concentration for LightCycler master mix from 3 to 5 mM, and thermal hybridization temperature between 52 and 62°C on Mx4000 in 10-fold diluted pHP6 and SiHa cells as DNA standards in triplicate.
Genotyping using reverse hybridization by the INNO-LiPA HPV genotyping system. Twenty microliters of the biotinylated real-time PCR product from positive samples was added to the denaturation solution and incubated with one strip in hybridization solution from the INNO-LiPA kit (Innogenetics, Ghent, Belgium). Each membrane strip contained 25 oligonucleotides specific for 25 different HPV types (namely, types 6, 11, 16, 18, 31, 33, 34, 35, 39, 40, 42, 43, 44, 45, 51, 52, 53, 54, 56, 58, 59, 66, 68/73, 70, and 74). The strips were washed twice under stringent conditions and then incubated with an alkaline phosphatase-labeled streptavidin conjugate for color development by the addition of 5-bromo-4-chloro-3-indolylphosphate and nitroblue tetrazolium in substrate buffer. After a final washing and drying, the strips were visually interpreted using a grid as previously described (10).
Statistical analysis.
Qualitative parameters were analyzed by using the Fisher exact test, and quantitative parameters were analyzed by using the Mann-Whitney nonparametric test and Pearson's correlation factor. The sensitivity of HPV detection in urine compared to cervical samples was expressed in percentages, with 95% exact confidence intervals. Cohen's kappa constant was calculated to evaluate the measure of kappa agreement (
value) for HPV detection between urine and cervix. A P value higher than 0.05 and a
value higher than 0.60 were considered significant. All statistical analysis was performed using the software SPSS, version 10.1, for Windows (SPSS Inc., Chicago, IL).
| RESULTS |
|---|
|
|
|---|
Development of real-time PCR to amplify and quantify HPV DNA. Mx4000 and LightCycler protocols gave very similar results (Table 2). Threshold levels of 50 and 95% of positive samples were evaluated using pHP6 and SiHA cells diluted 10-fold and 4-fold (Table 2). The lowest detection threshold was defined at about 5 HPV DNA copies. Linearity assays were obtained up to 8 log10 DNA copies/ml. The reproducibility of inter- and intra-assay results upon 10 runs was below 5%. The slope was excellent (3.22), showing high efficiency (98%). Results for pHP6 and SiHa cells containing, respectively, HPV6 and -16 types were similar (CT at 31.65 versus 31.37 for 500 HPV DNA copies and CT at 28.63 versus 28.28 for 5,000 HPV DNA copies).
|
HPV-positive samples were found in 150 cervical specimens out of 333 (45%) and 66 urine samples out of 177 (37%). Good agreement was obtained for HPV detection in cervical and urine samples, with a kappa test at 93% (Table 3). However, HPV viral load was 50-fold higher in cervical samples than in urine (respectively, 5.00 ± 1.73 versus 3.77 ± 1.32 log/ml; P < 0.0001). Levels of viral load in cervical specimens were correlated to levels in urine (r = 0.81; P = 0.009). Discordant results between positive cervical and negative urine samples more often showed low viral loads (below 3 log/ml). No correlation was found between viral load in both samples and levels of cells (41% of cervical specimens and 62% of urine samples were not correlated to cell levels estimated by GAPDH CT levels; r = 0.64 and 0.54, respectively; P > 0.05).
|
| DISCUSSION |
|---|
|
|
|---|
Therefore, we have adapted the previously described SPF10 PCR method (9) to real-time PCR, which shows higher sensitivity in the detection of more HPV infections from almost 57 different genotypes, compared to two other general PCR-based methods using the GP5+/6+ and MY09/11 primer sets. These primers (SPF, GP, and MY) allowed amplification of different overlapping fragments in the HPV L1 region between nucleotides 6500 and 7000 from HPV with, respectively, 65, 150, and 450 bp. The smallest fragment obtained with the SPF10 primers was the most appropriate for real-time PCR thermal conditions, since only small PCR fragments are required. Also, a type-specific assay is possible using the SPF10 amplicons with biotinylated reverse primers on the LiPA system developed by Innogenetics (10). We showed in this study that the use of biotinylated primers in real-time PCR did not affect amplification and SYBR green detection. We also optimized the SPF10 primers to allow a 60°C hybridization temperature as the optimal thermal conditions for real-time PCR.
A highly sensitive method may be required in order to realize the highest negative predictive value for detecting HPV-associated cervical carcinoma (i.e., no detected HPV is related to no risk of developing a carcinoma). Amplification detection assays, such as PCR-based methods, are more sensitive than liquid hybridization tests such as the Digene Hybrid Capture II (HC II) assay, as demonstrated previously for several viral molecular markers, such as cytomegalovirus DNA, hepatitis B DNA, and more recently HPV DNA (15). Indeed, Perrons et al. (15) showed that the HC II assay could detect only 56% of positive HR-HPV DNA using the SPF10 PCR method tested in parallel, whereas the latter could detect all HC II-positive HR-HPV. They also demonstrated that lower sensitivities of the HC II assay were obtained for Pap smears classified as normal or as atypical squamous cells of undetermined significance (respectively, 28 and 45% compared to 55 and 69% using the SPF10 PCR method), related perhaps to lower viral load in these groups of patients. The HC II assay is reported to have an analytical sensitivity at 1 pg of HPV DNA, corresponding to about 5,000 DNA copies, and to be less sensitive than PCR methods which allowed cutoff levels of about 5 to 500 copies (12). However, PCR methods may be impeded by inhibitors, as shown in cervical lavage samples (11), urethral swab specimens (22), or urine samples (3). It is important to evaluate the efficacy of the PCR using internal controls and to use the best DNA extraction procedure. Therefore, we used the GAPDH DNA cell control to assess the presence of cells in samples and elimination of inhibitors, which have been found in 7 to 19% of endocervical samples and about 50% of urethral and urine samples. GAPDH DNA has been found to be a stable marker for cell counting estimation (8). Other cell markers, such as ß-globin, were shown to be insensitive to some inhibitors in 8% of cervical lysates (11). Appropriate DNA extraction procedures have been shown to eliminate these inhibitors, as demonstrated by Brinkman et al. for HPV extraction in urine samples (3). Those authors found that urine could inhibit HPV DNA amplification 106-fold more than water by using a type 16 HPV diluted plasmid. They showed that by using the silica-based (1) viral DNA/RNA extraction method from QIAGEN, as we used in this study, they were able to reduce the inhibitor effect more than 105-fold.
A reliable HPV DNA quantification method is required for viral load measures. High viral load seems to be indicative of viral persistence and disease development, whereas low viral load is more often associated with viral clearance (5, 21). Although the HC II assay has not been validated as a quantitative test, it was shown in these studies that a high ratio of HPV to relative light units was related to the risk of developing HGSIL. However, more-precise methods have to be evaluated for this purpose. The recent availability of real-time PCR methods allows such an approach (17). However, clinical viral load thresholds have to be defined. Using a specific HPV type 16 quantitative method, Pretet et al. (17) showed that a viral load higher than 107 copies per ml may contribute to lesion evolution towards cervical intraepithelial neoplasia 2 to 3, corresponding to more than 100 pg/ml using the HC II test. We also observed that a higher viral load in patients (over 105 copies per ml) was related to HGSIL. Variation in the clinical threshold emphasizes the need of standardization of the methods with a control panel (18). The HC II assay is the only U.S. Food and Drug Administration-approved method, and the lack of a reference method makes the constitution of a control panel difficult. We also showed in this study that viral load was not correlated to cell levels. Therefore, we do not agree with Pretet and collaborators, who chose to normalize HPV copy numbers according to sample cellularity assessed by albumin PCR (17). Those authors found good correlations between their method and the HC II assay, whether HPV DNA was expressed in copies per ml or per 103 cells. However, we did not find a correlation when HPV DNA was expressed in copies per ml or per 102 cells. Indeed, we observed samples, mostly urine but also cervical specimens, with high viral load and few cells. Normalization of the HPV value might contribute to under- or overestimation of the viral load.
Finally, it is important to express HPV results in terms of specific genotype. Indeed, the recent study from Perrons et al. (15) showed that among followed HPV-positive patients assessed by the SPF10-LiPA method (60% of the included patients), 28% were positive with the same genotype at 6 months and 7% were positive with a different genotype; the first situation would be considered a persistent infection whereas the second would be seen as a "resolved-acquired" infection. Both situations were considered persistent infections by using the HC II assay. These investigators also observed, as did others (4, 16), cross-reactivity between HR- and LR-HPV; 13 out of 20 LR-HPVs assessed by the SPF10-LiPA system were found to be HR-HPV using the HC II assay. Cross-hybridization has been found with at least 22 different LR-HPV types with the current HR-HPV HC II probes (16).
In conclusion, we developed a highly sensitive and quantitative general HPV DNA real-time PCR method that allows mass screening of patients for HPV infection. This method allows specific genotyping using the LiPA system. Our method also seems to be sensitive enough to allow HPV detection in urine samples with high performance, owing to improvement of the whole PCR process, as suggested by the recent study of Daponte et al. (6),. The ongoing longitudinal and prospective multicenter PapU study should give us the opportunity to validate this approach by using this method in a larger population and in the new age of vaccination against HPV infections.
| ACKNOWLEDGMENTS |
|---|
We thank Laura Vernoux from Innogenetics and Fabienne Tardieu from Stratagene Europe for technical support. We also thank Kevin Erwin, a native-English-speaking medical translator who proofread the manuscript.