Previous Article | Next Article 
Journal of Clinical Microbiology, June 2007, p. 1712-1717, Vol. 45, No. 6
0095-1137/07/$08.00+0 doi:10.1128/JCM.02385-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.
Multicenter Evaluation of the New Abbott RealTime Assays for Quantitative Detection of Human Immunodeficiency Virus Type 1 and Hepatitis C Virus RNA
M. Schutten,1*
D. Peters,2
N. K. T. Back,3
M. Beld,3
K. Beuselinck,4
V. Foulongne,5
A.-M. Geretti,6
L. Pandiani,7
C. Tiemann,8 and
H. G. M. Niesters1
Department of Virology, Erasmus MC, Rotterdam, The Netherlands,1
InnovirGroup of Companies, Gardenview 2047, South Africa,2
Department of Medical Microbiology, Academic Medical Center, Amsterdam, The Netherlands,3
Laboratory of Molecular Diagnostics, University Hospital Gasthuisberg, Katholieke Universiteit Leuven, Leuven, Belgium,4
Laboratory of Virology, Montpellier University Hospital, 80 Ave. A. Fliche, 34295 Montpellier, France,5
Department of Virology, Royal Free Hospital and Royal Free and University College Medical School, London NW3 2PF, United Kingdom,6
Laboratoires Marcel Mérieux, BP 7322, 69357 Lyon Cedex 07, France,7
LABCON-OWL, Labor Dr. Krone und Partner, Herford, Germany8
Received 27 November 2006/
Returned for modification 26 January 2007/
Accepted 22 March 2007

ABSTRACT
The analytical performances of the new Abbott RealTime hepatitis
C virus (HCV) and human immunodeficiency virus type 1 viral
load assays were compared at nine laboratories with different
competitor assays. These included the Abbott LcX, Bayer Versant
bDNA, Roche COBAS Amplicor, and Roche COBAS TaqMan assays. Two
different protocols used during the testing period with and
without a pre-m1000 RNA isolation spin were compared. The difference
proved to be nonsignificant. A uracil-
N-glycosylase (UNG) contamination
control option in the HCV test for previous Roche COBAS Amplicor
users was evaluated. It proved to decrease amplicon carryover
by 100-fold independent of the amplicon input concentration.
The protocol including UNG proved to overcome problems with
false-positive negative controls. Comparison with other assays
revealed only minor differences. The largest difference was
observed between the Abbott HCV RealTime assay and the Roche
COBAS Amplicor HCV Monitor version 2.0 assay.

INTRODUCTION
Hepatitis C virus (HCV) and human immunodeficiency virus type
1 (HIV-1) are major causes of mortality in both developing and
developed countries. For both viruses, relatively effective
therapies have been available in developed countries for quite
some time. HCV and HIV-1 viral loads are important parameters
in patient management both before initiating therapy and during
therapy. The decline of the HCV viral load during the first
3 months of therapy is, for instance, a strong indicator of
the final outcome of therapy (
9,
23). Moreover, in many countries,
the HCV viral load in infected health care workers determines
whether they are allowed to perform surgical procedures. HIV-1
viral load is monitored during therapy, and a viral load above
a certain threshold, which may differ per treating physician,
requires a switch in medication (
6,
13).
The first tests that were described for determining the viral load were based on target amplification techniques like reverse transcriptase PCR (RT-PCR) and nucleic acid sequence-based amplification (NASBA). These tests had readout on agarose gel or readout with enzymatic detection of the amplicon after amplification with biotinylated primers (1, 3, 15, 17). Subsequently, signal amplification techniques were developed for the quantitative detection of HCV and HIV-1 (5, 16). Those techniques were improved and commercialized. The suppliers of the most widely used systems at the moment are Roche and Abbott, for RT-PCR-based systems (the COBAS Amplicor and LcX systems, respectively); bioMerieux (formerly Organon Technica), with a NASBA-based technique; and Bayer, with the Versant bDNA system being a signal amplification technique (4, 18, 19). Each technique has its own advantages and disadvantages. The relatively small dynamic range is a disadvantage that all current assay formats have in common. All assay formats are also rather sensitive to contamination, especially at the lower limit of detection (21). Handling of the sample after target amplification is a major cause of contamination for the NASBA- and RT-PCR-based techniques. Recently, amplification techniques that allow real-time detection of the amplicon have been developed for NASBA and RT-PCR. These new techniques make postamplification processing obsolete (14, 21). The major advantage of real-time RT-PCR and NASBA techniques is the lower sensitivity to contamination. These techniques are also less time-consuming and in general have a higher dynamic range. A serious drawback of most real-time techniques is the detection of amplicon with fluorescently labeled probes that are sensitive to point mutations within the target sequence. This is especially of concern with highly variable targets like HCV and HIV-1.
Abbott has recently CE (Conformité Européenne) marked its new HCV and HIV-1 viral load tests based on real-time RT-PCR (7, 8, 22). A major theoretical advantage of the real-time technique used is the probe system, which allows target detection by probe hybridization at low temperatures. This potentially reduces problems with genotype-dependent underquantification due to probe-target mismatches. Furthermore, the primers used for HIV-1 are located in a highly conserved region of the genome also used in the Abbott LcX HIV-1 assay. Before CE marking, Abbott initiated the Early Access Programme (EAP), involving 21 sites in Europe and 1 site in South Africa. This paper describes the validation of these HCV and HIV-1 viral load assays on the basis of data generated during the EAP.

MATERIALS AND METHODS
Collaborating sites.
Out of the 21 sites initially selected by Abbott Molecular to
join the prerelease EAP, 9 sites participated in the joint evaluation
by submitting data on patient samples. The assays and comparator
assays run at the different sites are summarized in Table
1.
Parts of the data from sites 4 and 5 have been published as
part of separate studies (
7,
8).
Patient materials and quality control materials.
Patient plasma samples were run either fresh (site 7) or after
storage at 80°C. In addition, all sites ran samples
from the 2005 quality control program from the QCMD (
www.QCMD.org).
Four laboratories selected for the EAP who were not part of
this joined evaluation ran QCMD samples and submitted data on
these samples (Leeds, United Kingdom [A. Hale], Lille, France
[L. Bocket], Lisbon, Portugal [R. Camacho], and Manchester,
United Kingdom [M. Guiver]).
Abbott RealTime HIV-1 and HCV viral load assays.
The Abbott RealTime HIV-1 and HCV viral load assays were performed according to the instructions provided by Abbott Molecular. In short, HIV-1 RNA and HCV RNA were isolated from 1.0 and 0.5 ml of plasma, respectively, with the Abbott m1000 nucleic acid extraction system. The lower limits of detection reported by Abbott for the use of these isolation volumes were 40 HIV-1 RNA copies per ml and 12 HCV IU per ml. Fifty microliters of eluted RNA was subsequently mixed with 50 µl Mastermix. Real-time RT-PCR was run on the m2000rt RealTime PCR system from Abbott. Data from the comparator assays were generated during routine diagnostics for patient management.
Data analysis.
Samples below the limit of detection for both the Abbott RealTime assays and the comparator assays were not included in the study analysis. Samples below the limit of detection in one assay but positive in the other were discarded in the regression analysis. Data were analyzed with Sigmaplot 2001 for Windows.

RESULTS AND DISCUSSION
Performance of the Abbott RealTime system at the EAP sites.
Several problems were encountered, which were solved during
the EAP. First, it was found that the Abbott RealTime HCV and
HIV-1 assays underquantified some samples that had been stored
at 80°C (data not shown). Spinning the plasma sample
before nucleic acid extraction could circumvent this "inhibitory"
effect (data not shown). It was hypothesized that in certain
samples, fibrin clotting after thawing might result in a loss
of RNA-coated magnetic beads during the extraction procedure
on the m1000 system. Part of the sample run during the EAP was
therefore performed using an alternative isolation protocol.
This protocol included a 5-min spin of the plasma at 2,000
x g after thawing but before m1000 extraction. Data obtained with
and without spinning were compared (Fig.
1). The difference
between the two protocols proved not to be significant (
P value
of 0.87 by
t test). It was therefore decided that the difference
in protocols should be ignored for further data analysis, except
for individual samples.
The second and probably more significant problem encountered
was that virtually all laboratories that had previously used
the Roche COBAS Amplicor HCV viral load test had significant
problems with false-positive results on negative controls. The
amplicon of the Roche COBAS Amplicor HCV test overlaps completely
with the Abbott RealTime HCV viral load test. The Roche COBAS
Amplicor assay is an open system after RT-PCR in the detection
phase. Since significant amounts of amplicon are released in
the laboratory, this system requires the dUTP/uracil-
N-glycosylase
(dUTP/UNG) contamination control system. The Abbott RealTime
HCV viral load system is a closed system post-RT-PCR, and consequently,
the generated amplicon will never be released in the laboratory
under normal circumstances. Abbott chose not to implement the
dUTP/UNG system in their assay since the dUTP/UNG contamination
control system may lower the sensitivity of the assay. This
is in part due to residual UNG activity after the 10-min 95°C
inactivation step normally used. It was hypothesized that the
Roche COBAS Amplicor HCV Monitor amplicons contaminated the
Abbott RealTime HCV viral load assay in laboratories that had
used the former test. A protocol that incorporated UNG without
the use of dUTP to circumvent the problem of amplicon carryover
from other systems like the COBAS Amplicor assay was developed.
To test the efficiency of this protocol, an amplicon from the
COBAS Amplicor HCV test with dUTP incorporated was quantified
using the Abbott RealTime HCV viral load assay without UNG.
This amplicon contained 4.2
x 10
12 copies per ml. Tenfold dilutions
from 10
6 to 10 copies per test were analyzed in quadruplicate
by the Abbott RealTime HCV viral load assay with and without
UNG (Fig.
2). UNG decreased the concentration by 100-fold irrespective
of the amount of amplicon input. With the use of UNG, no further
false-positive negative controls were encountered at sites where
the Roche COBAS Amplicor HCV assay was used. This was anticipated,
since UNG has been used successfully in the COBAS Amplicor assay
for many years now.
With these changes relative to the original protocol incorporated,
the HCV and HIV-1 RealTime viral load tests were generally perceived
as being easy and convenient to use.
Comparison of the Abbott RealTime HCV and HIV-1 viral load tests with comparators.
The Abbott RealTime HIV-1 viral load assay was compared with the Roche COBAS Amplicor HIV Monitor version 1.5, the Roche COBAS TaqMan, the Bayer Versant bDNA version 3.0, and the Abbott LcX assays. Table 2 shows the following results: (i) the number of samples tested, (ii) regression coefficient (r2), (iii) the mean difference between the Abbott viral load and the viral load measured with the comparator (mean log10 Abbott mean log10 comparator), (iv) the regression curve equation, and (v) the number of samples where the Abbott RealTime HIV viral load assay gave an underestimation (Abbott < comparator) or overestimation (comparator < Abbott) of more than 1 log10 plus the mean difference. Correlations between the Abbott RealTime HIV-1 viral load test and the comparators were generally good (r2 > 0.80). The Abbott RealTime HIV-1 viral load assay gave a lower viral load relative to those of all the comparators. No major differences were observed in the numbers of samples that differed more than 1 log10 relative to the comparator assays. Overall, the Abbott RealTime HIV-1 viral load assay scored well relative to the comparators, and only minor differences were observed. For subtypes B, C, and CRF02-AG, sufficient samples (65, 55, and 22 samples, respectively) could be tested for a comparison although not enough to compare the Abbott RealTime test separately with each comparator. Relative to the comparators, the Abbott RealTime HIV-1 viral load test quantified subtype B and C samples accurately. Mean differences (Abbott minus comparator) of the log10 results were 0.31 and 0.02, respectively. The Abbott RealTime HIV-1 assay generated higher viral load values; however, for subtype CRF02-AG, samples compared to the comparator assays (the mean difference [Abbott minus comparator] of the log10 results was 0.27). These data corroborate data from a recent report on the performance of the Abbott RealTime HIV-1 viral load test showing better performance on CRF02-AG recombinants than the comparator assays (12).
The Abbott RealTime HCV viral load assay was compared with the
Roche COBAS Amplicor HCV version 1.5, the Roche COBAS Taqman,
the Bayer bDNA version 3.0, and the Abbott LcX assays. Table
2 shows the following results: (i) the number of samples tested,
(ii) regression coefficient (
r2), (iii) the mean difference
between the Abbott viral load and the viral load measured with
the comparator (mean log
10 Abbott mean log
10 comparator),
(iv) the regression curve equation, and (v) the number of samples
where the Abbott RealTime HCV viral load assay gave an underestimation
(Abbott < comparator) or overestimation (comparator <
Abbott) of more than 1 log
10 plus the mean difference. For HCV,
the number of data points tested with the bDNA assay was too
low for proper comparison. The regression equation of the comparison
with the COBAS Amplicor HCV test differed significantly from
similarity (
a = 1 and
b = 0). This proved to be due to higher
values obtained with COBAS Amplicor assay at the low range and
lower values in the high range. One sample that was below the
limit of detection in the Abbott assay gave >10
5 IU/ml in
the comparator assay (site 2, Roche COBAS Amplicor v1.5). The
Abbott RealTime assay gave a lower viral load than all the comparators.
No major differences were observed in the numbers of samples
that differed more than 1 log
10 relative to the comparator assays.
Overall, the Abbott real-time HCV viral load assay scored well
relative to the comparators, and only minor differences were
observed. The number of data points generated on genotypes 2,
3, and 4 was rather low, and sufficient data were generated
for comparison for genotype 1 only. The mean difference of the
log
10 results between the Abbott assay and the comparators for
HCV genotype 1 samples was minor (0.18 log
10).
Performance of the Abbott RealTime HCV and HIV-1 viral load assays on the 2005 QCMD panels.
The QCMD quality control panels for the year 2005 were run at 11 EAP laboratories for further validation of the Abbott RealTime assays. Five sites ran the HCV panel, and eight sites ran the HIV-1 panel.
Three laboratories scored all samples correct in the qualitative analysis of the HCV panel. One laboratory scored the negative sample as positive but below the lower limit of detection. This was most probably due to contamination, since this laboratory ran the panel without using UNG, and the laboratory that made the QCMD HCV 2005 panel uses the COBAS Amplicor assay (M. Schutten, unpublished data). One laboratory reported the inhibition of one sample and one sample with a target concentration of 1.7 x 103 IU per ml as being negative.
For HIV-1, all 10 laboratories had a 100% correct score on all samples for the qualitative analysis. The quantitative data generated by the laboratories with the Abbott RealTime HCV and HIV-1 kits were averaged and compared with the target concentration (what the QCMD intended to put into the sample) and with the mean of the predistribution test results performed by the QCMD reference laboratories (Fig. 3A and B). With respect to the predistribution testing results, it should be noted that for HCV and HIV-1, testing was done mainly with the COBAS Amplicor Monitor assays. For samples 2 and 8 of the HCV panel, only one quantitative result was available because the other predistribution tests were below the limit of detection. For both samples, one Versant HCV RNA version 3.0 test and one COBAS Amplicor HCV Monitor test proved to be false negative. For the HIV-1 panel, sample 2 had only two predistribution results because two Roche COBAS Monitor version 1.5 results were below the limit of detection. Apart from these predistribution test difficulties, the mean of the Abbott RealTime HCV and HIV-1 results and the target concentration and the predistribution mean results were relatively close.
Recent problems with the introduction of a TaqMan-based system
have shown that the development of a robust quantitative real-time-detected
RT-PCR for highly variable targets like HCV and HIV-1 is not
an easy task (
2,
10-
12,
20). From the data presented, it can
be concluded that real-time RT-PCR systems based on hybridization
probes rather than hydrolysis probes bear promise for the future.
In conclusion, the RealTime HCV and HIV-1 viral load tests from
Abbott are convenient, easy-to-use assays with good performance
characteristics. It should be noted that laboratories that use
the Roche COBAS Amplicor HCV Monitor assay are recommended to
use the adapted protocol with UNG to prevent false-positive
results due to amplicon carryover.

ACKNOWLEDGMENTS
We thank A. Hale, L. Bocket, R. Camacho, and M. Guiver for the
data generated on the QCMD panels. We also thank the QCMD Neutral
Office for providing the data relating to the QCMD 2005 HCV
and HIV Proficiency Programmes. The m1000 and m2000rt systems,
including consumables and reagents for this study, were kindly
provided by Abbott Molecular.
Analysis and presentation of the QCMD data were performed by the authors of this paper only and are their interpretation of the QCMD data provided. The paper does not express or constitute QCMD proficiency results and/or reports for these programs.

FOOTNOTES
* Corresponding author. Mailing address: Department of Virology, Erasmus MC, Dr. Molewaterplein 40, 3015 GD Rotterdam, The Netherlands. Phone: (31) 10-4633431. Fax: (31) 10-4633441. E-mail:
m.schutten{at}erasmusmc.nl 
Published ahead of print on 4 April 2007. 

REFERENCES
1 - Abbott, M. A., B. J. Poiesz, B. C. Byrne, S. Kwok, J. J. Sninsky, and G. D. Ehrlich. 1988. Enzymatic gene amplification: qualitative and quantitative methods for detecting proviral DNA amplified in vitro. J. Infect. Dis. 158:1158-1169.[Medline]
2 - Caliendo, A. M., A. Valsamakis, Y. Zhou, B. Yen-Lieberman, J. Andersen, S. Young, A. Ferreira-Gonzalez, G. J. Tsongalis, R. Pyles, J. W. Bremer, and N. S. Lurain. 2006. Multilaboratory comparison of hepatitis C virus viral load assays. J. Clin. Microbiol. 44:1726-1732.[Abstract/Free Full Text]
3 - Compton, J. 1991. Nucleic acid sequence-based amplification. Nature 350:91-92.[CrossRef][Medline]
4 - de Baar, M. P., M. W. van Dooren, E. de Rooij, M. Bakker, B. van Gemen, J. Goudsmit, and A. de Ronde. 2001. Single rapid real-time monitored isothermal RNA amplification assay for quantification of human immunodeficiency virus type 1 isolates from groups M, N, and O. J. Clin. Microbiol. 39:1378-1384.[Abstract/Free Full Text]
5 - Dewar, R. L., H. C. Highbarger, M. D. Sarmiento, J. A. Todd, M. B. Vasudevachari, R. T. Davey, Jr., J. A. Kovacs, N. P. Salzman, H. C. Lane, and M. S. Urdea. 1994. Application of branched DNA signal amplification to monitor human immunodeficiency virus type 1 burden in human plasma. J. Infect. Dis. 170:1172-1179.[Medline]
6 - EACS Euroguidelines Group. 2003. European guidelines for the clinical management and treatment of HIV-infected adults in Europe. AIDS 17(Suppl. 2):S3-S26.[CrossRef]
7 - Foulongne, V., B. Montes, M. N. ot-Rousseau, and M. Segondy. 2006. Comparison of the LCx human immunodeficiency virus (HIV) RNA quantitative, RealTime HIV, and COBAS AmpliPrep-COBAS TaqMan assays for quantitation of HIV type 1 RNA in plasma. J. Clin. Microbiol. 44:2963-2966.[Abstract/Free Full Text]
8 - Garcia-Diaz, A., G. S. Clewley, C. L. Booth, W. Labett, N. McAllister, and A. M. Geretti. 2006. Comparative evaluation of the performance of the Abbott real-time human immunodeficiency virus type 1 (HIV-1) assay for measurement of HIV-1 plasma viral load following automated specimen preparation. J. Clin. Microbiol. 44:1788-1791.[Abstract/Free Full Text]
9 - Gavier, B., M. A. Martinez-Gonzalez, J. I. Riezu-Boj, J. J. Lasarte, N. Garcia, M. P. Civeira, and J. Prieto. 1997. Viremia after one month of interferon therapy predicts treatment outcome in patients with chronic hepatitis C. Gastroenterology 113:1647-1653.[CrossRef][Medline]
10 - Gelderblom, H. C., S. Menting, and M. G. Beld. 2006. Clinical performance of the new rRoche COBAS TaqMan HCV Test and High Pure System for extraction, detection and quantitation of HCV RNA in plasma and serum. Antivir. Ther. 11:95-103.[Medline]
11 - Giraldi, C., A. Noto, R. Tenuta, F. Greco, D. Perugini, M. Spadafora, A. M. Bianco, O. Savino, and A. Natale. 2006. A comparative evaluation between real time Roche COBas TAQMAN 48 HCV and bDNA Bayer Versant HCV 3.0. New Microbiol. 29:243-250.[Medline]
12 - Gueudin, M., J. C. Plantier, V. Lemee, M. P. Schmitt, L. Chartier, T. Bourlet, A. Ruffault, F. Damond, M. Vray, and F. Simon. 2007. Evaluation of the Roche Cobas TaqMan and Abbott RealTime extraction-quantification systems for HIV-1 subtypes. J. Acquir. Immune Defic. Syndr. 44:500-505.[CrossRef][Medline]
13 - Hammer, S. M., M. S. Saag, M. Schechter, J. S. Montaner, R. T. Schooley, D. M. Jacobsen, M. A. Thompson, C. C. Carpenter, M. A. Fischl, B. G. Gazzard, J. M. Gatell, M. S. Hirsch, D. A. Katzenstein, D. D. Richman, S. Vella, P. G. Yeni, and P. A. Volberding. 2006. Treatment for adult HIV infection: 2006 recommendations of the International AIDS SocietyUSA panel. Top. HIV Med. 14:827-843.[Medline]
14 - Heid, C. A., J. Stevens, K. J. Livak, and P. M. Williams. 1996. Real time quantitative PCR. Genome Res. 6:986-994.[Abstract/Free Full Text]
15 - Kievits, T., B. van Gemen, D. van Strijp, R. Schukkink, M. Dircks, H. Adriaanse, L. Malek, R. Sooknanan, and P. Lens. 1991. NASBA isothermal enzymatic in vitro nucleic acid amplification optimized for the diagnosis of HIV-1 infection. J. Virol. Methods 35:273-286.[CrossRef][Medline]
16 - Lau, J. Y., G. L. Davis, J. Kniffen, K. P. Qian, M. S. Urdea, C. S. Chan, M. Mizokami, P. D. Neuwald, and J. C. Wilber. 1993. Significance of serum hepatitis C virus RNA levels in chronic hepatitis C. Lancet 341:1501-1504.[CrossRef][Medline]
17 - Menzo, S., P. Bagnarelli, M. Giacca, A. Manzin, P. E. Varaldo, and M. Clementi. 1992. Absolute quantitation of viremia in human immunodeficiency virus infection by competitive reverse transcription and polymerase chain reaction. J. Clin. Microbiol. 30:1752-1757.[Abstract/Free Full Text]
18 - Murphy, D. G., L. Cote, M. Fauvel, P. Rene, and J. Vincelette. 2000. Multicenter comparison of Roche COBAS AMPLICOR MONITOR version 1.5, Organon Teknika NucliSens QT with Extractor, and Bayer Quantiplex version 3.0 for quantification of human immunodeficiency virus type 1 RNA in plasma. J. Clin. Microbiol. 38:4034-4041.[Abstract/Free Full Text]
19 - Ross, R. S., S. Viazov, S. Sarr, S. Hoffmann, A. Kramer, and M. Roggendorf. 2002. Quantitation of hepatitis C virus RNA by third generation branched DNA-based signal amplification assay. J. Virol. Methods 101:159-168.[CrossRef][Medline]
20 - Sarrazin, C., B. C. Gärtner, D. Sizmann, R. Babiel, U. Mihm, W. P. Hofmann, M. von Wagner, and S. Zeuzem. 2006. Comparison of conventional PCR with real-time PCR and branched DNA-based assays for hepatitis C virus RNA quantification and clinical significance for genotypes 1 to 5. J. Clin. Microbiol. 44:729-737.[Abstract/Free Full Text]
21 - Schutten, M., and H. G. Niesters. 2001. Clinical utility of viral quantification as a tool for disease monitoring. Expert Rev. Mol. Diagn. 1:153-162.[CrossRef][Medline]
22 - Swanson, P., S. Huang, V. Holzmayer, P. Bodelle, J. Yamaguchi, C. Brennan, R. Badaro, C. Brites, K. Abravaya, S. G. Devare, and J. Hackett, Jr. 2006. Performance of the automated Abbott RealTime HIV-1 assay on a genetically diverse panel of specimens from Brazil. J. Virol. Methods 134:237-243.[CrossRef][Medline]
23 - Tong, M. J., L. M. Blatt, J. G. McHutchison, R. L. Co, and A. Conrad. 1997. Prediction of response during interferon alfa 2b therapy in chronic hepatitis C patients using viral and biochemical characteristics: a comparison. Hepatology 26:1640-1645.[CrossRef][Medline]
Journal of Clinical Microbiology, June 2007, p. 1712-1717, Vol. 45, No. 6
0095-1137/07/$08.00+0 doi:10.1128/JCM.02385-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Labbett, W., Garcia-Diaz, A., Fox, Z., Clewley, G. S., Fernandez, T., Johnson, M., Geretti, A. M.
(2009). Comparative Evaluation of the ExaVir Load Version 3 Reverse Transcriptase Assay for Measurement of Human Immunodeficiency Virus Type 1 Plasma Load. J. Clin. Microbiol.
47: 3266-3270
[Abstract]
[Full Text]
-
Greengrass, V. L., Plate, M. M., Steele, P. M., Denholm, J. T., Cherry, C. L., Morris, L. M., Hearps, A., Crowe, S. M.
(2009). Evaluation of the Cavidi ExaVir Load Assay (Version 3) for Plasma Human Immunodeficiency Virus Type 1 Load Monitoring. J. Clin. Microbiol.
47: 3011-3013
[Abstract]
[Full Text]
-
Pyne, M. T., Konnick, E. Q., Phansalkar, A., Hillyard, D. R.
(2009). Evaluation of the Abbott Investigational Use Only RealTime HIV-1 Assay and Comparison to the Roche Amplicor HIV-1 Monitor Test, Version 1.5. J. Mol. Diagn.
11: 347-354
[Abstract]
[Full Text]
-
Sloma, C. R., Germer, J. J., Gerads, T. M., Mandrekar, J. N., Mitchell, P. S., Yao, J. D. C.
(2009). Comparison of the Abbott RealTime Human Immunodeficiency Virus Type 1 (HIV-1) Assay to the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test: Workflow, Reliability, and Direct Costs. J. Clin. Microbiol.
47: 889-895
[Abstract]
[Full Text]
-
Holguin, A., Lopez, M., Molinero, M., Soriano, V.
(2008). Performance of Three Commercial Viral Load Assays, Versant Human Immunodeficiency Virus Type 1 (HIV-1) RNA bDNA v3.0, Cobas AmpliPrep/Cobas TaqMan HIV-1, and NucliSens HIV-1 EasyQ v1.2, Testing HIV-1 Non-B Subtypes and Recombinant Variants. J. Clin. Microbiol.
46: 2918-2923
[Abstract]
[Full Text]