Journal of Clinical Microbiology, August 2007, p. 2521-2528, Vol. 45, No. 8
0095-1137/07/$08.00+0 doi:10.1128/JCM.02117-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Cell-Free Science and Technology Research Center,1 Venture Business Laboratory, Ehime University, Matsuyama, Ehime 790-8577, Japan,5 Department of Parasitology, Kangwon National University College of Medicine, Chunchon 200-701, Korea,2 Department of Entomology, Armed Forces Research Institute of Medical Sciences, Bangkok 10400, Thailand,3 Vector Borne Disease Training Center, Pra Budhabat, Saraburi 18120, Thailand4
Received 16 October 2006/ Returned for modification 6 December 2006/ Accepted 31 May 2007
|
|
|---|
|
|
|---|
Currently, the conventional method for the diagnosis of malaria is microscopic examination of thin and/or thick blood smears. Although light microscopy has a relatively high sensitivity and specificity and also provides parasite density, the stage, and species differentiation, this method is labor-intensive, requires well-trained experts, and may result in therapeutic delays. To improve the speed and precision of malaria diagnosis in areas where standard laboratory diagnosis is not available, researchers have developed malaria rapid diagnostic tests based on antigen-capture immunochromatographic technologies (12, 16). When they are in good condition, some products can achieve sensitivity for the detection of P. falciparum infection similar to that of expert microscopists (3, 10). However, the sensitivity can vary between products (14), and a species-specific product is available only for P. falciparum. Very long observation times and considerable expertise are required for a correct diagnosis by microscopy under several circumstances: when the level of parasitemia is low, during mixed infection, after drug treatment, and during the chronic phase of the infection. Therefore, this situation can lead to false-negative results or unreliable species determinations (1).
Subsequently, molecular methods based on DNA amplification, such as nested PCR and real-time quantitative PCR, were developed for the diagnosis of malaria. Compared to microscopy, these methods have demonstrated higher sensitivity, detecting one to five parasites/µl of blood, and greater specificity for mixed infection (7, 19, 22, 24-27). However, the long turnaround time, high cost, and availability only in well-equipped laboratories render this technology inadequate for routine diagnosis in hospital laboratories and field clinics in areas where malaria is endemic (4).
Recently, a new, rapid, simple, and sensitive technique called loop-mediated isothermal amplification (LAMP) was developed (17). LAMP is a nucleic acid amplification method that relies on autocycling strand-displacement DNA synthesis performed with Bst DNA polymerase. The amplification products are stem-loop DNA structures with several inverted repeats of the target and structures with multiple loops. The principal merit of this method is that no denaturation of the DNA template is required (15), and thus, the LAMP reaction can be conducted under isothermal conditions (ranging from 60 to 65°C). LAMP requires only one enzyme and four primers that recognize six distinct target regions. The method produces a large amount of amplified product, resulting in easier detection, such as detection by visual judgment of the turbidity or fluorescence of the reaction mixture (13).
Several investigators have reported on LAMP methods for the rapid identification of Plasmodium, Trypanosoma, and Babesia and have commended the usefulness of the LAMP assay (5, 8, 21, 29). Poon et al. estimated that the cost of running a LAMP assay is about 1/10 of that of the normal PCR method for P. falciparum detection (21). The biggest reduction in cost and time came from simple sample preparation, with no requirement for previous DNA extraction (6). Briefly, simple heating of the infected blood at 99°C for 10 min was enough to prepare the DNA template for LAMP without any inhibition of the reaction (21). However, to date, the use of LAMP for the detection of malaria parasites for clinical diagnosis has been validated only with acute falciparum malaria patients (21). Although P. falciparum is the most important cause of severe disease, its geographic distribution overlaps those of P. vivax, P. malariae, and P. ovale; and therefore, a method that allows the rapid detection and identification of all four species infecting humans would be preferable.
In the present study, we describe a LAMP for the clinical detection and identification of P. falciparum, P. vivax, P. malariae, and P. ovale. The technique was evaluated with blood samples obtained from field clinics. The results of LAMP were compared to those of conventional microscopy and nested PCR.
|
|
|---|
Conventional microscopy. Thick blood smears were examined under x1,000 magnification by microscopists with extensive experience in the identification of malaria parasites. The parasite density per 500 leukocytes was counted and was then calculated as the number of parasites per microliter by assuming a leukocyte count of 7,000/µl. The initial thick film was considered negative if no parasites were seen after 500 leukocytes were counted.
DNA extraction. The DNA template for the nested PCR and LAMP was prepared as described previously (20). Twenty-five to 50 µl of human blood was blotted as a single spot and dried on filter paper. A single blood spot from each filter paper was excised and then incubated 4 h at room temperature and/or overnight at 4°C in 1 ml of 0.5% saponin in phosphate-buffered saline (PBS). The filter paper was washed for 30 min in PBS at 4°C and transferred into new tubes containing 200 µl of 5% Chelex-100 (Bio-Rad, Hercules, CA), and the tubes were vortexed for 30 s. The mixture was incubated at 56°C for 15 min, vortexed for 30 s, heated at 100°C for 15 min to elute the DNA, vortexed, and centrifuged (10,000 x g for 5 min). The supernatant was either used immediately in the reaction or stored in aliquots at –20°C.
Nested PCR. For nested PCR, the species-specific nucleotide sequences of the 18S rRNA genes of P. falciparum, P. vivax, P. malariae, and P. ovale were amplified as described previously (7), with slight modifications. In the first PCR, 2 µl of template DNA (corresponding to approximately 0.25 to 0.5 µl of blood) was added to a 20-µl PCR mixture that consisted of 0.4 µM each universal primer (P1 forward primer [5'-ACGATCAGATACCGTCGTAATCTT-3'] and P2 reverse primer [5'-GAACCCAAAGACTTTGATTTCTCAT-3']), 200 µM each deoxynucleoside triphosphate, 25 mM MgCl2, 1x PCR Gold Buffer II (50 mM KCl, 15 mM Tris-HCl, pH 8.0), and 0.25 U AmpliTaq Gold DNA polymerase. The DNA amplification was carried out under the following conditions: 94°C for 10 min and then 35 cycles at 92°C for 30 s, 60°C for 1.5 min, and 72°C for 1 min, followed by a final extension at 72°C for 5 min. The first PCR product was diluted 20-fold in sterile water. One microliter of this solution was used in the second amplification. The second PCR was performed at 94°C for 10 min and then 20 cycles at 92°C for 30 s, 60°C for 1.5 min, 72°C for 1 min, followed by a final extension at 72°C for 5 min with the P1 forward primer in combination with each species-specific reverse primer (P. falciparum, 5'-CAATCTAAAAGTCACCTCGAAAGATG-3'; P. vivax, 5'-CAATCTAAGAATAAACTCCGAAGAGAAA-3'; P. malariae, 5'-GGAAGCTATCTAAAAGAAACACTCATAT-3'; P. ovale, 5'-ACTGAAGGAAGCAATCTAAGAAATTT-3'). The amplified products were visualized in 2% agarose gels stained with ethidium bromide. The expected band sizes were approximately 160 bp for the first PCR product and approximately 110 bp for the second one. To prevent cross-contamination, different sets of pipettes and distinct work areas were used for DNA template preparation, PCR mixture preparation, and DNA amplification. Moreover, 1 uninfected blood sample was included for every 10 samples processed.
LAMP conditions. The LAMP primer sets for P. falciparum were described previously (21). The remaining Plasmodium genus- and species-specific LAMP primer sets were designed on the basis of the genus- and the species-specific nucleotide sequences of the 18S rRNA genes of P. falciparum, P. vivax, P. malariae, and P. ovale by use of the LAMP primer design software PrimerExplorer V3 (http://primerexplorer.jp/e/). The location and nucleotide sequence of each primer are shown in Fig. 1 and Table 1, respectively. The LAMP reaction was performed with a Loopamp DNA amplification kit (Eiken Chemical Co., Ltd., Tokyo, Japan). The reaction mixtures (25 µl) contained 1.6 to 2.4 µM each primers FIP and BIP, 0.2 µM each primers F3 and B3c, 0.8 µM each primers LPF and LPB, 2x reaction mixture (12.5 µl), Bst DNA polymerase (1 µl), and 1 to 2 µl of DNA sample (corresponding to approximately 0.125 to 0.5 µl of blood). The LAMP reaction was performed at 60°C for 100 min, and then the enzyme was inactivated at 80°C for 2 min.
![]() View larger version (50K): [in a new window] |
FIG. 1. Location and sequence of LAMP targets and priming sites for Plasmodium genus (A) and four Plasmodium species (B). (A) The locations of the priming sites by the Plasmodium genus-specific primer set in the reference sequence (GenBank accession no. M19173.1) are indicated by arrows. (B) Partial sequence alignment of the 18S rRNA genes of four human malaria parasites, P. falciparum (Pf; GenBank accession no. M19173.1), P. vivax (Pv; GenBank accession no. U03079), P. malariae (Pm; GenBank accession no. M54897), and P. ovale (Po; GenBank accession no. L48986), along with the species-specific primer annealing sites.
|
|
View this table: [in a new window] |
TABLE 1. Primer sets used for amplification of 18S rRNA genes in LAMP
|
Diagnostic threshold of LAMP results. LAMP was monitored through nephelometric analysis by recording the scattering light intensity reflected from a light source (wavelength, 660 nm) every 6 s with the help of the Loopamp real-time turbidimeter (RT-160C). The threshold value was defined as the value halfway between the mean maximum differential value (which represents a maximum velocity of the turbidity increment) – 3 standard deviations (SDs) for positive samples and the mean maximum differential value + 3 SDs for negative samples. Threshold values were as follows: for the Plasmodium genus, 0.0041; for P. falciparum, 0.0036; for P. vivax, 0.0040; for P. malariae, 0.0043; and for P. ovale, 0.0045. The threshold time represented the time for the turbidity to increase to the threshold value. Most of the positive samples tested multiple times showed positivity within 1 h. Therefore, a sample with a turbidity greater than or equal to the threshold value within 1 h, as determined with the turbidimeter, was considered positive.
Positive control plasmid DNA and sequencing. For sensitivity assessment, plasmids containing the target region of the 18S rRNA gene were constructed for each species for use in the LAMP reaction. The target DNA sequence was amplified with two LAMP primers (primers F3 and B3c) by PCR with ExTaq DNA polymerase (Takara Bio Inc., Otsu, Japan) and was then cloned into the pCR2.1-TOPO TA cloning vector (Invitrogen, Carlsbad, CA), according to the manufacturer's instructions. Plasmid DNA purification was performed with a QIAprep Miniprep kit (QIAGEN, Hilden, Germany). The nucleotide sequences were determined with a BigDye Terminator sequencing kit (Applied Biosystems, Foster City, CA) by using an automated DNA sequencer (ABI PRISM 310 genetic analyzer; Applied Biosystems), according to the manufacturer's instructions. The resulting sequences were aligned by using the 18S rRNA sequences for the four species of Plasmodium deposited in GenBank to confirm that the target sequences were correct.
Analytical sensitivity and specificity of LAMP. To establish the minimum copy number (lower detection limit) of the target gene sequence detectable by LAMP, positive control plasmid DNAs were used. The standard curve for LAMP was constructed by using 10-fold serial dilutions of plasmid DNA (106 copies to 1 copy) to sterile water. For each standard, the copy number was plotted against the threshold time. The resulting plots were analyzed by linear regression, and the statistical significance of the r2 values was analyzed by analysis of variance. Probabilities of less than 0.05 were considered statistically significant. The specificities of the genus- and the species-specific LAMP assays were evaluated with each control plasmid DNA and P. falciparum genomic DNA (gDNA) purified from strain NF54, P. vivax gDNA purified from strain Sal-I, P. malariae gDNA purified from strain Uganda, P. ovale LS type gDNA purified from a Thai isolate, P. ovale CDC type gDNA purified from a CDC strain, P. knowlesi gDNA purified from strain H, and P. yoelii gDNA purified from strain 17XNL.
Clinical sensitivity and specificity. The clinical sensitivity and specificity of the Plasmodium LAMP were calculated by using 121 whole-blood samples and microscopy as the reference standard method. Sensitivity was calculated as (number of true positives)/(number of true positives + number of false negatives), and specificity was calculated as (number of true negatives)/(number of true negatives + number of false positives).
|
|
|---|
![]() View larger version (41K): [in a new window] |
FIG. 2. Sensitivities of Plasmodium genus- and species-specific real-time LAMP assays performed with serial dilutions of plasmid DNA (106 copies to 1 copy per reaction) containing an 18S rRNA gene. (A) Amplification with a Plasmodium genus-specific primer set. One representative result of four replicates is shown. Samples contained a plasmid harboring the P. falciparum 18S rRNA gene. (B) Plot of the mean threshold time of the Plasmodium genus-specific LAMP. The error bars represent the standard errors of the mean values from four replicates. The plot of the mean threshold time against the log of the input DNA fit a linear function (r2 = 0.96). (C) Amplification with the P. falciparum species-specific primer set. One representative result of four replicates tested with the plasmid harboring the P. falciparum 18S rRNA gene is shown. (D) Plot of the mean threshold time of the P. falciparum species-specific LAMP from four replicates, which fit a linear function (r2 = 0.91). (E) Amplification with a P. vivax species-specific primer set. One representative result of four replicates tested with plasmid harboring P. vivax 18S rRNA gene is shown. (F) Plot of the mean threshold time of the P. vivax species-specific LAMP from four replicates, which fit a linear function (r2 = 0.95). (G) Amplification with a P. malariae species-specific primer set. One representative result of four replicates tested with a plasmid harboring the P. malariae 18S rRNA gene is shown. (H) Plot of the mean threshold time of the P. malariae species-specific LAMP from four replicates, which fit a linear function (r2 = 0.84). (I) Amplification with a P. ovale species-specific primer set. One representative result of four replicates tested with a plasmid harboring the P. ovale 18S rRNA gene is shown. (J) Plot of the mean threshold time of the P. ovale species-specific LAMP from four replicates, which fit a linear function (r2 = 0.90). Genus, genus Plasmodium LAMP; Pf, P. falciparum LAMP; Pv, P. vivax LAMP; Pm, P. malariae LAMP; Po, P. ovale LAMP; O.D., optical density.
|
![]() View larger version (72K): [in a new window] |
FIG. 3. Restriction analysis of Plasmodium genus- and species-specific LAMP products amplified from plasmid DNA containing each target 18S rRNA gene. The digestion products were run on a 3% agarose gel. Lane M, DNA ladder marker; lane 1, genus Plasmodium LAMP product; lane 2, DdeI digestion of genus Plasmodium product (123-, 44-, and 20-bp bands were expected); lane 3, P. falciparum LAMP product; lane 4, HpyCH4V digestion of P. falciparum product (130- and 79-bp bands were expected); lane 5, P. vivax LAMP product; lane 6, HpyCH4V digestion of P. vivax product (121- and 77-bp bands were expected); lane 7, P. malariae LAMP product; lane 8, HpyCH4V digestion of P. malariae product (142- and 84-bp bands were expected); lane 9, P. ovale LAMP product; lane 10, AluI digestion of P. ovale product (152- and 69-bp bands were expected).
|
|
View this table: [in a new window] |
TABLE 2. Detailed comparison of microscopy, nested PCR, and LAMP for malaria parasite detection and species identification
|
LAMP yielded results in about 25.7 ± 4.9 min (mean ± SD; range, 19.4 to 52.9 min) for detection of the genus Plasmodium in the 67 samples that were positive by microscopy. The detection times were 31.7 ± 4.8 min (range, 25.8 to 44.9 min) for 17 P. falciparum samples, 30.6 ± 5.2 min (range, 25.4 to 46.6 min) for 37 P. vivax samples, 34.8 ± 4.8 min (range, 30.5 to 46.6 min) for 10 P. malariae samples, and 36.1 ± 6.8 min (range, 29.9 to 49.8 min) for 8 P. ovale samples. The average copy numbers of the target genes in the clinical samples were calculated on the basis of the linear regression formulas from the analysis of serial dilutions of plasmid DNA (Fig. 2B, D, F, H, and J). The average log copy numbers in clinical samples were 3.95 for the Plasmodium genus, 5.02 for P. falciparum, 3.78 for P. vivax, 4.10 for P. malariae, and 3.17 for P. ovale.
|
|
|---|
Since 96.7% of the results of LAMP were consistent with those of nested PCR and the sensitivity (98.5%) and specificity (94.3%) of both LAMP and nested PCR compared with the results of microscopy were equally high, we can propose that LAMP is as reliable as nested PCR for the clinical detection of four species of malaria parasites. Among five samples positive for both P. falciparum and P. vivax by microscopy, one was positive only for P. vivax by both nested PCR and LAMP. The level of P. falciparum parasitemia (approximately 0.01%) was lower than that of P. vivax parasitemia (approximately 0.05%) in this sample. Since the level of parasitemia was within the level detectable by nested PCR or LAMP, there may be another reason why this sample was negative for P. falciparum by both nested PCR and LAMP, such as the use of a smaller amount of blood or a lower efficiency of extraction of DNA from this specimen. Even though the LAMP results were in good agreement with those of nested PCR, there were some nonconcordant results. Among 12 samples positive for P. malariae by microscopy, 1 was negative by nested PCR and another 1 was negative by the species-specific LAMP. However, these two samples were successfully amplified by the Plasmodium genus-specific LAMP. One explanation is that the target sequences of the rRNA gene of P. malariae have variant sequences (9); thus, new primer sets for both nested PCR and LAMP would be required to target regions that are universally conserved among P. malariae variants for reliable diagnosis. Among the samples found to contain P. ovale by both LAMP and PCR, microscopy had detected P. ovale in five of them, P. vivax in two samples, and P. malariae in the last sample. In contrast, both LAMP and nested PCR could successfully detect P. ovale in all eight samples, including both the LS type (six samples) and the CDC type (two samples), as confirmed by the P. ovale type-specific PCR method (28) (data not shown). On the basis of the comparable results between the LAMP and the nested PCR methods, as described above, the overall specificities of LAMP and nested PCR may be greater than that of microscopy.
The LAMP method can be used to quantify parasites in infected blood, although it is not a perfect assay for quantification. To quantify the parasites in infected blood by LAMP, we applied the average threshold time for the clinical blood samples to the formula of linear regression for each genus- and species-specific analysis. While there was a statistically significant linear correlation between the threshold time and the log of the initial copy number of template DNA, the linear correlation was rather poor for low copy numbers of template DNA (Fig. 2B, D, F, H, and J). Possible explanations are that (i) the longer incubation time before the detectable amplification caused by the lower template copy number makes the enzyme activity lower and (ii) the lower template DNA concentration, which decreases the annealing efficiency of the primers, may cause a delay in the threshold time. Although further improvement is needed, the statistically significant negative correlations between the threshold time and the parasite count by microscopy were obtained for the genus-specific LAMP-positive samples (n = 67), the samples positive for P. vivax only (n = 30), and the samples positive for P. falciparum only (n = 12) by using Spearman's rank correlation test (genus-specific LAMP, P = 0.0002; P. vivax-specific LAMP, P < 0.0001; P. falciparum-specific LAMP, P = 0.0034).
Diagnosis by LAMP does not require expensive reagents for DNA extraction, a turbidimeter, a thermal cycler, or skilled technicians. The template can be prepared by direct heat treatment of blood samples, without time-consuming and expensive DNA extraction with a commercial kit (21). Moreover, LAMP requires only a simple incubator, such as a heat block or a water bath that provides a constant temperature of 60°C, which makes it more economical and practical than nested PCR and real-time PCR. The white turbidity that results from magnesium pyrophosphate accumulation as a by-product of DNA amplification can be detected with the naked eye or a turbidimeter (13). Alternatively, inspection for amplification can be simply be performed with the naked eye by using SYBR green I, which turns green in the presence of amplified DNA. The results obtained by use of SYBR green I were consistent with those deduced by use of the real-time turbidimeter (18). Since the turbidity assay can be carried out in a closed system, the risk of contamination is lower than that when agarose gel electrophoresis is used. This is an additional advantage of LAMP for clinical use (2, 21, 23). Screening by genus-specific LAMP and then confirmation by a species-specific LAMP may provide a simple and reliable test for epidemiological surveillance. However, the routine screening for malaria parasites in both clinical laboratories and malaria clinics in areas where malaria is endemic may require a species-specific LAMP. Background epidemiological information will be helpful for local governments to decide how to apply LAMP for the diagnosis of malaria in each area.
In conclusion, the LAMP methods developed in this study can be useful for clinical diagnosis and active surveillance of malaria parasites in countries where malaria is endemic because it has a sensitivity and a specificity similar to those of nested PCR, requires minimal laboratory facilities, and is simpler and less expensive to perform than nested PCR. The on-site performance of LAMP at malaria clinics in the field will be required for further evaluation of this technique for the rapid diagnosis of malaria.
This work was supported in part by Grants-in-Aid for Scientific Research 16406009 and 18390129 and Scientific Research on Priority Areas grant 16017273 from the Ministry of Education, Culture, Sports, Science and Technology of Japan and a Grant-in-Aid of the Ministry of Health, Labor and Welfare (grant H17-Sinkou-ippan-019), Japan, and was partially supported by U.S. Military Infectious Diseases Research Program.
Published ahead of print on 13 June 2007. ![]()
|
|
|---|