To be clinically useful, an assay for ganciclovir resistance must be highly accurate, and detect multiple mutations associated with resistance. Furthermore, it must be performed rapidly, such that the results are available in a clinically relevant time frame. In this regard, the gold standard method for mutation detection, sequencing of the viral genome, has been limited by the need to amplify virus in culture before PCR and sequencing can be performed (9). Since the growth of CMV in culture is extremely slow, this typically results in delays of 3 to 4 weeks until the assay is completed, at which time the clinical utility of the result may be limited.
To address the need for a rapid HCMV resistance assay suitable for use directly on clinical specimens, we first sought to optimize the efficiency of PCR amplification of the UL97 region containing the most common mutations. The most commonly used PCR primers for this region, described by Lurain et al. (9), generally are used for PCR on viral isolates that have been amplified by tissue culture and fail to successfully amplify direct patient samples (unpublished observations). These primers amplify a 921-bp region of UL97 (from codons 400 to 707), and we reasoned that the efficiency of amplification by these primers might be limited by their length. Our hypothesis was supported by the results of Spector and coworkers, who used seven overlapping primer sets to amplify and sequence the relevant region of UL97 directly from plasma from 10 patients with known ganciclovir-resistant CMV infections (10, 11), and by the results of Jabs and coworkers, who used a nested PCR approach to amplify this region for sequencing directly from patient materials (6, 7). Since the use of seven primers sets would be impractical for most clinical laboratories and to avoid the need for nested PCR, we designed two optimized primer sets allowing amplification of this region: forward primer 5'-TGG CCG ACG CTA TCA AAT TT-3' and reverse primer 5'-CCC AGC GCC GAC AGC TCC GAC-3', amplifying codons 439 to 557, and forward primer 5'-ATG TCG GAG CTG TCG GCG-3' and reverse primer 5'-CGA CAC GAG GAC ATC TTG-3', amplifying codons 550 to 645. Amplification by these primers was performed on a Roche LightCycler instrument, using the Roche FastStart High-Fidelity PCR system, with the inclusion of SYBR green to allow real-time monitoring of accumulation of PCR product. The program for cycling was as follows: denaturation at 95°C for 2 min, followed by 45 cycles of 95°C for 20 s, 55°C for 30 s, and 72°C for 30 s. Final elongation was done at 72°C for 4 min, with cooling at 37°C for 1 min. Both primer sets efficiently amplified HCMV, with an efficiency comparable to our clinical HCMV quantitation assay (2) (data not shown). The resulting amplicons were then sequenced by using the ABI Prism BigDye terminator cycle sequencing kit on an Applied Biosystems 3730 DNA analyzer, and the sequence was analyzed by using Applied Biosystems SeqScape software (Fig. 1). The assay could be completed and reported on the next working day after receipt of the specimen.
![]() View larger version (28K): [in a new window] |
FIG. 1. Detection of CMV UL97 mutations. The figure shows a typical sequencing result and analysis from a clinical specimen showing infection with a mixture of two viral variants (arrows). seq., sequence.
|
We next compared the performance of our rapid assay to that of established assays based on analysis of culture-expanded specimens. Stored culture fluid, obtained from CMV-positive clinical samples that had been sent to an outside commercial reference laboratory for phenotypic and/or genotypic drug resistance testing, was recultured and the supernatant tested with our new assay. A subset of these samples had been tested in-house for genotypic drug resistance by the primer set described by Lurain et al. (9). Of 20 samples compared to the commercial test, 18 showed agreement, while for two samples a mutation (Q449K) was identified by our rapid assay that is not reported by the commercial reference laboratory assay. Similarly, of 15 samples compared to culture and the in-house assay using the primer set of Lurain et al., 14 showed agreement. In one specimen, the new method identified one mutation (D605E) that was not detected by the traditional assay. In a recent study evaluating this issue, there was >90% agreement between UL97 sequences directly amplified from clinical specimens and those from culture isolates (7). In the case of discrepancies, it is likely that the results of an assay performed directly on patient specimens more closely reflect the clinically relevant viral species.
Finally, we monitored the performance of our rapid assay in clinical use over the period from July 2004 to December 2006. At our institution, the assay is used mainly in the management of patients receiving solid organ or stem cell transplants. Of 143 specimens submitted for resistance testing, usable sequence was obtained from 120 (84%). The most common cause of assay failure was the submission of samples with very low or undetectable viral levels. Of the six samples that had detectable virus but could not be sequenced, five had viral levels of <200 copies/ml, and the other had 1,300 copies/ml. The 120 samples successfully sequenced were from 71 different individuals. Overall, only 9 of 71 patients had wild-type sequences at the nucleotide level compared to the reference AD169 strain, whereas 25 of 71 had silent nucleotide mutations that did not result in amino acid substitution. The most common mutations seen in the samples were the silent mutations D456D, G557E, G579G, and L634L. Multiple mutations were found in the same sample ca. 30% of the time and probably indicate the presence of distinct CMV genotypes in our samples (3). Of the patients with amino acid mutations, 25 of 37 (68%) had at least one mutation previously reported to be associated with ganciclovir resistance. An additional 12 of 37 patients had a single amino acid mutation with uncertain impact on resistance: D605E, H469Y, and Q449K (8, 3, and 1 patient, respectively). Ten patients (14%) had evidence of mixed infection with more than one sequence. Of the 22 patients with samples from more than one time point, 5 (23%) had changes in the UL97 sequence over time. A single patient had virus with different UL97 sequences detected from urine versus cerebrospinal fluid. The mutations detected in all 120 specimens are described in Table 1.
|