Journal of Clinical Microbiology, August 2007, p. 2698-2700, Vol. 45, No. 8
0095-1137/07/$08.00+0 doi:10.1128/JCM.00759-07
Copyright © 2007, American Society for Microbiology. All Rights Reserved.
Quality Control of Direct Molecular Diagnostics for Methicillin-Resistant Staphylococcus aureus
Alex van Belkum,1*
Hubert G. M. Niesters,2,
William G. MacKay,3 and
Willem B. van Leeuwen1
Department of Medical Microbiology and Infectious Diseases, Unit Research and Development,1
Department of Virology, Erasmus MC, 's-Gravendijkwal 230, 3015 CE Rotterdam, The Netherlands,2
the Neutral Office, Quality Control for Molecular Diagnostics, Block 4, Kelvin Campus, West of Scotland Science Park, Glasgow G20 OSP, Scotland3
Received 10 April 2007/
Returned for modification 12 May 2007/
Accepted 11 June 2007

ABSTRACT
Ten samples containing various amounts of methicillin-resistant
Staphylococcus aureus (MRSA), methicillin-susceptible
S. aureus,
methicillin-resistant
Staphylococcus epidermidis (MRSE), and
combinations thereof were distributed to 51 laboratories for
molecular diagnostics testing. Samples containing 10
2 to 10
3 MRSA cells were frequently reported to be negative. MRSE samples
were scored as negative by all commercial tests but by only
two out of three in-house tests.

TEXT
Methicillin-resistant
Staphylococcus aureus (MRSA) requires
timely detection in order to prevent infection and nosocomial
transmission (
1,
4). Classical microbiologists routinely use
culture-based enrichment assays as the basis for detection and
subsequent identification. The inclusion of an elevated salt
concentration and specific antibiotics in the growth medium
allows for highly specific detection (
8). However, culture requires
prolonged incubation periods, and in general, confirmatory assays
are required upon positive culture. Molecular assays may offer
benefits over more traditional culture-based assays, such as
reduced time to identification and better specificity and sensitivity.
The commercial diagnostics industry has introduced a range of
molecular assays in recent years. However, there is little information
available on their performance, and some studies suggest that
they may suffer from reduced sensitivity as a result of sample
inhibition (
3). Specificity may also be an issue as the presence
of methicillin-resistant coagulase-negative staphylococci, including
methicillin-resistant
Staphylococcus epidermidis (MRSE), may
result in false-positive results. Given the number of commercially
available test systems, the clinical impact of MRSA, and the
need for timely diagnostics, a multicenter external quality
assessment (EQA) study on the efficiency and efficacy of molecular
testing for MRSA was initiated by Quality Control for Molecular
Diagnostics (QCMD) (
www.qcmd.org).
The QCMD MRSA EQA panel samples were distributed in October 2006 to 51 participating laboratories from 11 countries, along with detailed instructions on how to process the panel samples. Participants were given 6 weeks to report their results back to the QCMD Neutral Office by using an online data collection system.
The QCMD 2006 MRSA panel consisted of five samples containing various amounts of MRSA, three samples containing various amounts of staphylococci other than S. aureus, and one sample containing Escherichia coli (Table 1). The contents of the samples were quantified on the basis of culture and molecular testing results. The IDI-MRSA test (Becton Dickinson) was employed in combination with the Sigma plant DNA isolation kit according to the manufacturers' instructions.
Out of the 51 participants, 46 (90%) responded. Nonrespondents
indicated technical problems (
n = 2) or "test under development"
(
n = 2) as the reason for not returning results. Overall, 58
data sets were returned, 55 of which included qualitative data
only; three labs reported both qualitative and quantitative
data. All participants received the expected results following
the close of the program. Subsequently, the QCMD Neutral Office
analyzed the data, which was released to participants in the
form of a detailed EQA final report.
Most of the real-time data were generated with the Roche LightCycler system (n = 14), Roche LightCycler 2.0 (n = 6), and the Corbett Research Rotor-Gene 3000 (n = 5) and Applied Biosystems systems (the ABI 7500 real-time PCR system [n = 6], the ABI PRISM 7000 sequence detection system [n = 5], and the ABI PRISM 7900 sequence detection system [n = 1]). Two Bio-Rad machines were included, and the Cepheid Smart Cycler II system, Roche LightCycler 480, and the Stratagene MxP3000 real-time system each provided a single data set. For the in-house conventional PCRs, the Eppendorf MasterCycler, the MWG AG Biotech Primus 96, and the Perkin-Elmer 9600 were used. The diversity of the equipment covers the spectrum commercially available systems quite well.
QCMD used a simple scoring system for qualitative EQA data, which was as follows: 2 points for a correct result and 0 points for all other results (including "not determined" and "equivocal"). The results obtained are summarized in Table 2. Results for the panel sample with the highest number of MRSA cells (MRSA06-08) were reported correctly in 97% of the data sets. This finding indicates that the usage of molecular tests to follow up positive cultures is reliable: all the tests can be used for culture confirmation. However, the samples containing smaller amounts of MRSA (MRSA06-09 and MRSA06-06) had levels that were below the limit of detection of most participants' assays. Only 12 and 52% of the data sets included correct results for these samples. MRSE sample MRSA06-07 was correctly reported as negative for MRSA by all commercial PCR tests but incorrectly reported as positive in 33% (1 of 3) of data sets generated with conventional in-house assays and 21% (9 of 43) of data sets generated using real-time in-house PCR. The methicillin-susceptible S. aureus (MSSA)-containing sample MRSA06-10 was scored incorrectly by both commercial and in-house tests (17% of conventional PCR tests and 9% of real-time PCR tests), indicating moderate specificity. These levels of false positives underscore the need for improved specificity of these MRSA tests. Even the E. coli specimen could not be adequately tested (results were indeterminate in 7% [4 of 58] of data sets). Table 3 summarizes the performance scores for the various PCR platforms. The performance of the commercial conventional PCRs and the in-house real-time PCRs can be considered satisfactory. However, only three participants attained the maximum score of 20.
Quality control of molecular diagnostics tools is important
in order to maintain high-quality clinical care in medical institutions.
Multicenter studies on the sensitivity and specificity of several
bacterial and viral DNA tests have been performed in the past,
and the outcomes of such studies have provided important information
on performance (
2,
5,
6,
7). Here we report on an EQA study
of one of the most important nosocomial bacterial pathogens,
MRSA. In conclusion, the molecular identification of MRSA by
using samples with high CFU counts is reliable and can be implemented
in the laboratory setting with confidence. Essentially, all
testing formats perform equally well. However, for direct molecular
diagnostics, we have to conclude that the present array of tests
do not meet the clinical quality criteria. The sensitivity of
many tests is (too) low, and the specificity needs to be improved.
The reasons for poor assay sensitivity may be grounded in the
common practice of preenriching samples by culture before confirmation
by molecular assays. Preenrichment may lead to concentrations
of MRSA in excess of 10
10 CFU/ml, which is greater than the
concentrations of MRSA likely to be encountered in an unmodified
patient sample (and those in this EQA panel). Some assays may
have been designed specifically for this higher target concentration
range, leading to poor levels of sensitivity. The preenrichment
approach also reduces one of the key improvements offered by
molecular assays, which is more rapid diagnosis. The sensitivity
of molecular assays for the detection of MRSA will improve only
if the molecular diagnostics community moves away from preenrichment
and tests directly from the clinical sample. Performance would
also be improved by the provision of proper quality control
materials and international standards for MRSA testing.
Although the stage has been set for direct clinical detection of MRSA, our data here indicate that the present testing systems are insufficient. And clinical practice may even be worse: here we used clean samples, and in the case of samples such as urine, blood, and sputa, inhibitory compounds within these samples may deteriorate the test performance. These data clearly support the need for improvements in the molecular detection of MRSA.

ACKNOWLEDGMENTS
The QCMD program is organized in collaboration with the European
Society for Clinical Virology and the European Society for Clinical
Microbiology and Infectious Diseases. Panels were produced in
the Department of Medical Microbiology and Infectious Diseases
of Erasmus MC, Rotterdam, The Netherlands. Nothing in this report
may be reproduced without permission of the QCMD Executive Office.

FOOTNOTES
* Corresponding author. Mailing address: Department of Medical Microbiology and Infectious Diseases, Unit Research and Development, Erasmus MC, 's-Gravendijkwal 230, 3015 CE Rotterdam, The Netherlands. Phone: 00-31-10-4635813. Fax: 00-31-10-4633875. E-mail:
a.vanbelkum{at}erasmusmc.nl 
Published ahead of print on 20 June 2007. 
Present address: Department of Medical Microbiology, Section Clinical Virology Hanzeplein 1, University Medical Center Groningen, P.O. Box 30 001, 9700 RB Groningen, The Netherlands. 

REFERENCES
1 - Brown, D. F., D. I. Edwards, P. M. Hawkey, D. Morrison, G. L. Ridgway, K. J. Towner, M. W. Wren, and the Joint Working Party of the British Society for Antimicrobial Chemotherapy, Hospital Infection Society, Infection Control Nurses Association. 2005. Guidelines for the laboratory diagnosis and susceptibility testing of methicillin resistant Staphylococcus aureus (MRSA). J. Antimicrob. Chemother. 56:1000-1018.[Abstract/Free Full Text]
2 - Donoso Mantke, O., S. W. Aberle, T. Avsic-Zupanc, M. Labuda, and M. Niedrig. 2007. Quality control assessment for the PCR diagnosis of tick-borne encephalitis virus infections. J. Clin. Virol. 38:73-77.[CrossRef][Medline]
3 - Hodgson, J., M. Zuckerman, and M. Smith. 2007. Development of a novel internal control for a real time PCR for HSV DNA types 1 and 2. J. Clin. Virol. 38:217-220.[CrossRef][Medline]
4 - Mellmann, A., A. W. Friedrich, N. Rosenkotter, J. Rothganger, H. Karch, R. Reintjes, and D. Harmsen. 2006. Automated DNA sequence based early warning system for the detection of methicillin-resistant Staphylococcus aureus outbreaks. PLoS Med. 3:e33.[CrossRef][Medline]
5 - Noordhoek, G. T., S. Mulder, P. Wallace, and A. M. Van Loon. 2004. Multicentre quality control study for detection of Mycobacterium tuberculosis in clinical samples by nucleic acid amplification methods. Clin. Microbiol. Infect. 10:295-301.[CrossRef][Medline]
6 - Templeton, K. E., C. B. Forde, A. M. Loon, E. C. Claas, H. G. M. Niesters, P. Wallace, and W. F. Carman. 2006. A multi-centre pilot proficiency programme to assess the quality of molecular detection of respiratory viruses. J. Clin. Virol. 35:51-58.[CrossRef][Medline]
7 - Valentine-Thon, E. 2002. Quality control in nucleic acid testing: where do we stand? J. Clin. Virol. 25(Suppl.):S13-S21.
8 - Wertheim, H., H. A. Verbrugh, C. Van Pelt, P. De Man, A. Van Belkum, and M. C. Vos. 2001. Improved detection of methicillin-resistant Staphylococcus aureus using phenyl mannitol broth containing aztreonam and ceftizoxime. J. Clin. Microbiol. 39:2660-2662.[Abstract/Free Full Text]
Journal of Clinical Microbiology, August 2007, p. 2698-2700, Vol. 45, No. 8