Journal of Clinical Microbiology, September 2007, p. 2863-2871, Vol. 45, No. 9
0095-1137/07/$08.00+0 doi:10.1128/JCM.00981-07
Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Department of Pathology, University of Iowa Carver College of Medicine, Iowa City, Iowa,1 Robert Wood Johnson Medical School, New Brunswick, New Jersey,2 UCLA Healthcare, Los Angeles, California,3 Lab Science of Arizona, Tempe, Arizona4
Received 10 May 2007/ Returned for modification 13 June 2007/ Accepted 16 July 2007
|
|
|---|
|
|
|---|
2 µg/ml) and the overall rate of multidrug resistance (i.e., resistance to at least three classes of antimicrobial agents) is 22% (6). International surveillance of S. pyogenes in 2002 and 2003 found an overall macrolide resistance rate of 11% with higher rates in Spain (23%) and Italy (25%) (1), while U.S. rates approached 7% (12). For S. agalactiae, the reported rates of resistance to erythromycin and clindamycin in the United States during the previous decade were 16 to 20% and 7 to 9%, respectively (10, 11). U.S. surveillance of viridans group streptococcal bloodstream isolates in the 1990s reported that 32 to 56% were not susceptible to penicillin and 38 to 46% were resistant to erythromycin (4, 5). Because of the emergence of antimicrobial resistance among streptococci, the availability of accurate and convenient antimicrobial susceptibility testing (AST) methods in clinical microbiology laboratories is essential. This study evaluated a new BD Phoenix Automated Microbiology System STREP AST panel (BD Diagnostic Systems, Sparks, MD) in four clinical microbiology laboratories by comparing its performance to the Clinical and Laboratory Standards Institute (CLSI) reference broth microdilution method for the susceptibility testing of streptococci.
(This research was presented in part at the 43rd Interscience Conference on Antimicrobial Agents and Chemotherapy, 14 September 2003, Chicago, IL [abstr. D215].)
|
|
|---|
7 days, never frozen], 25% recent [stored for 8 to 59 days, never frozen], and 40% stock) were collected and tested at one of the four clinical laboratory sites (University of Iowa Health Care, Iowa City; Robert Wood Johnson University Hospital, New Brunswick, NJ; UCLA Healthcare, Los Angeles, CA; or Lab Science of Arizona, Tempe). In addition, 135 challenge isolates from BD representing a wide range of resistance phenotypes were divided among three study sites for testing. All organisms were tested following subculture on Trypticase soy agar with 5% sheep blood and 18 to 24 h of incubation at 35°C in an atmosphere of 5 to 7% CO2. |
View this table: [in a new window] |
TABLE 1. Phoenix identification of the 2,013 streptococci evaluated in this study
|
5 x 105 CFU/ml per tube) and onto Trypticase soy agar with 5% sheep blood as a purity check. A 1.5-ml aliquot of the ID broth was transferred to 12.5 ml of sterile diluent for inoculation of reference broth microdilution trays. Within 30 min, each side of the two Phoenix AST panels and the right side of the combination panels were inoculated with a tube of AST-S broth. The left side of the combination panel was inoculated with the remaining Phoenix ID broth. Panels were placed in the Phoenix instrument within 30 min of inoculation. Reference method. The CLSI broth microdilution method (2) was performed with frozen microdilution trays containing cation-adjusted Mueller-Hinton broth with 2 to 5% lysed horse blood prepared by PASCO Laboratories (Wheat Ridge, CO). The reference panels were thawed at room temperature and used within 2 h. Within 15 min of standardization of the Phoenix ID broth, 1.5 ml was transferred to 12.5 ml of sterile diluent. After mixing, the diluted suspension was poured into a seed tray and a new disposable multipronged device was used to inoculate each panel. The final concentration of bacteria in the microdilution tray was ca. 5 x 105 CFU/ml. A purity check was performed with the remaining seed tray inoculum. The panels were incubated in ambient air for 20 to 24 h at 35°C prior to the visual reading of endpoints.
Quality control. Four quality control (QC) strains were tested daily with the reference method and the Phoenix system. These strains were S. pneumoniae ATCC 49619, S. pneumoniae BD 3951, S. pneumoniae BD 4218, and S. mitis BD 3992.
Antimicrobials tested. The antimicrobial agents (and dilutions) tested in the Phoenix panels were cefepime (0.03 to 4 µg/ml), cefotaxime (0.015 to 4 µg/ml), ceftriaxone (0.015 to 4 µg/ml), clindamycin (0.015 to 4 µg/ml), erythromycin (0.015 to 16 µg/ml), gatifloxacin (0.06 to 8 µg/ml), levofloxacin (0.25 to 16 µg/ml), linezolid (0.25 to 16 µg/ml), meropenem (0.03 to 2 µg/ml), penicillin (0.015 to 32 µg/ml), tetracycline (0.06 to 16 µg/ml), trimethoprim-sulfamethoxazole (TMP-SMX; 0.06 to 16 µg/ml), and vancomycin (0.06 to 32 µg/ml). The broth microdilution reference panels included the same antimicrobial dilutions, with the exception that the linezolid range extended to 32 µg/ml. The clindamycin wells of the Phoenix panels included a low level of inducer (erythromycin).
Evaluation of results. CLSI interpretive criteria were used to interpret each MIC as susceptible (S), intermediate (I), or resistant (R) (nonmeningitis breakpoints were used for pneumococci) (3). For each antimicrobial agent tested, the Phoenix system MIC and Phoenix interpretative category of susceptible (S), intermediate (I), or resistant (R) was compared to the broth microdilution results to determine rates of essential agreement (EA; MIC within 1 log2 dilution), categorical agreement (CA; interpretative category of S, I, or R agreement), very major (VM) errors (false susceptible with rates determined by using the number of resistant organisms as the denominator), major errors (false resistance with rates determined by using the number of susceptible isolates as the denominator), and minor errors (I by the reference method and S or R by the Phoenix system or I by the Phoenix system and S or R by the reference method). The Phoenix MICs used for comparison were taken directly from the instrument without expert system (BDExpert) interpretation.
For discrepancy resolution of clinical isolates with major or VM errors, the organism-drug combination was retested in duplicate by both methods. All three results (initial and two repeat results) were used to determine the majority result. The majority result became the final result. Results were excluded from analysis when Phoenix or reference test results were not available, when duplicate isolates of the same species had been tested from the same patient and the same body site, when isolate purity was questioned, when there was no growth in either the Phoenix or reference panel, and when QC results were out of control.
Macrolide resistance phenotype. The double-disk diffusion test was performed on erythromycin-resistant and clindamycin-susceptible S. agalactiae clinical isolates (D-zone test) with 15-µg erythromycin and 2-µg clindamycin disks placed 12 mm apart (3). Isolates with blunting of the inhibition zone around the clindamycin disk adjacent to the erythromycin disk (D-zone positive) were considered to have inducible clindamycin resistance.
|
|
|---|
|
View this table: [in a new window] |
TABLE 2. Comparison of Phoenix to reference method MICs and interpretive category errors for S. pneumoniae
|
|
View this table: [in a new window] |
TABLE 3. Comparison of Phoenix to reference method MICs and interpretive category errors for beta-hemolytic streptococcus groups A, C, and G
|
|
View this table: [in a new window] |
TABLE 4. Comparison of Phoenix to reference method MICs and interpretive category errors for S. agalactiae and other streptococci
|
For the VGS clinical isolates (Table 5), the CA was below 90% for only one antimicrobial agent (penicillin, 13.1% minor errors). Discrepancy testing of VGS clinical isolates resolved 9 of the 27 VM errors, but the final VM error rate for 7 agents exceeded 1.5%. The major error rate for VGS was 0 to 3% after discrepancy testing resolved 6 of the 21 major errors. Among the 19 clinical isolates identified as S. bovis and S. acidominimus (other streptococci, Table 4), there was a single VM error (100%, meropenem) and one major error (5.6%, clindamycin).
|
View this table: [in a new window] |
TABLE 5. Comparison of Phoenix to reference method MICs and interpretive category errors for viridans group streptococci
|
|
View this table: [in a new window] |
TABLE 6. Times required to obtain results for 2,013 streptococcal clinical and challenge isolates
|
|
|
|---|
90% CA,
90% EA,
1.5% VM errors,
3% major errors) (7). In this multicenter study, the Phoenix system generated accurate susceptibility test results for most of the streptococcal species tested. The advantages of using the Phoenix instrument rather than a manual method for testing streptococci in a clinical laboratory are the generation of results in a shorter time period and labor savings associated with automated reading and interpretation of MICs. The BDExpert system analyzes susceptibility profiles for unusual results, and potentially erroneous results are flagged for verification. Reporting of results is predicated on CLSI guidelines (3). BDExpert interpretation was not evaluated in this study. Of the Streptococcus species included in this study, pneumococci are the most frequently tested in clinical microbiology laboratories because of the significant rate of resistance to multiple antimicrobial agents now recognized with this pathogen. The provision of accurate susceptibility results in a timely manner can minimize the time patients receive inappropriate therapy and limit the use of needlessly broad-spectrum agents. The Phoenix is the second automated AST instrument to offer a panel for S. pneumoniae testing that may be reported after a relatively short incubation period (<16 h).
The Phoenix results for the clinical isolates of S. pneumoniae compared favorably to those obtained by the reference method, with few exceptions (Table 2). Although the VM error rates of clinical isolates for cefotaxime (4%) and ceftriaxone (4.5%) exceeded 1.5%, these rates represented a failure to detect one resistant isolate; further, the instrument accurately detected resistance for isolates included in the challenge set that were cefotaxime (n = 14) or ceftriaxone (n = 13) resistant. The VM error rates which exceeded 1.5% for challenge isolates (erythromycin, tetracycline, TMP-SMX) were also each due to one isolate. The clinical isolate collection included three- to fourfold more isolates resistant to erythromycin, tetracycline, and TMP-SMX than the challenge set, yet they were detected without difficulty (0 to 0.9% VM error rates). Overcalling of resistance was a rare event, but minor error rates exceeded 10% for challenge isolates with four agents (cefepime, ceftriaxone, levofloxacin, and meropenem). However, the high EAs (98.9 to 100%) for those four drugs suggest that the errors were a consequence of MICs for isolates falling close to the interpretative breakpoints.
An evaluation of the VITEK 2 card for susceptibility testing of S. pneumoniae isolates reported major and VM error rates that are similar to the Phoenix error rates in the present study for the six common antimicrobial agents cefotaxime, ceftriaxone, erythromycin, penicillin, tetracycline, TMP-SMX, and vancomycin (8). Minor error rates in the VITEK 2 study obtained with TMP-SMX (16.9 to 19.1%), cefotaxime (11.9 to 19.8%), and penicillin (9.5 to 16.7%) for clinical and challenge pneumococcal strains (8) were higher than those found in this Phoenix evaluation. A more rigorous evaluation of VITEK 2 for detecting fluoroquinolone resistance (challenge set of 196 pneumococci with 66 gatifloxacin resistant or intermediate) reported 13.3% minor errors and one (1.7%) VM error for gatifloxacin (9). There are obvious limitations associated with the comparison of results from studies that tested different isolates; however, head-to-head evaluations of the VITEK 2 and Phoenix instruments with common collections of streptococci have not been described in the literature.
Although S. pyogenes and S. agalactiae are important pathogens, these species are predictably susceptible to penicillin, the therapeutic agent of choice. The M100-S17 CLSI document states that there is no clinical need to perform susceptibility testing of penicillins and other ß-lactams with these organisms (3). However, testing of the susceptibility of these organisms to clindamycin and erythromycin is recommended when the patient is allergic to ß-lactam agents (13). The Phoenix system performed well when determining the susceptibility of beta-hemolytic streptococci to erythromycin, with no VM errors, 0 to 1.1% major errors, and 0.7 to 5.9% minor errors. The Phoenix system had no VM errors and acceptable minor error rates (0.7 to 5.9%) when testing clindamycin against beta-hemolytic streptococci.
The detection of inducible clindamycin resistance among beta-hemolytic streptococci by the CLSI reference method requires the performance of the manual D-zone test on erythromycin-resistant, clindamycin-susceptible isolates (3). An evaluation of the VITEK 1 and VITEK 2 systems for the detection of erythromycin and clindamycin resistance among 304 S. agalactiae isolates also concluded that the double-disk diffusion method was needed to ensure accurate results (14). In this study, 23 of the 24 S. agalactiae clinical isolates with major errors for clindamycin were available for additional testing, and all had positive D-zone test results (inducible clindamycin resistance). Of the 25 S. agalactiae isolates that were erythromycin resistant and clindamycin susceptible by both the Phoenix system and the reference method, only 1 was D-zone test positive. Although further evaluation is necessary with more strains, it appears that the Phoenix system's use of an inducer (a small amount of erythromycin in clindamycin wells) reliably defines isolates of S. agalactiae with inducible clindamycin resistance as being clindamycin resistant or intermediate, potentially eliminating the need for D-zone testing of erythromycin-resistant and clindamycin-susceptible S. agalactiae isolates.
The need to critically evaluate all susceptibility results (not only those generated by an automated instrument) was demonstrated by the VM errors remaining after discrepancy testing for two S. pyogenes isolates reported as resistant to linezolid by the reference method only. Since linezolid resistance has not been previously reported in this species, the isolates were sent to a reference laboratory, where testing by the CLSI broth microdilution method in duplicate and Etest revealed linezolid susceptibility.
The lowest rates of concordance of the Phoenix results with the reference method occurred for the VGS isolates (Table 5). Additional studies with more VGS strains are needed to further assess the performance of the Phoenix system for antimicrobial agents with VM error rates above 1.5% or minor error rates exceeding 10%.
Our findings suggest that clinical microbiology laboratories may rely on the Phoenix system for accurate susceptibility testing of S. pneumoniae and beta-hemolytic streptococci. The shorter incubation time required by the Phoenix system in comparison to reference methods has the potential to enhance patient care.
Financial support for this study came from BD Diagnostic Systems (Sparks, MD).
S.S.R. has received research funding from Becton Dickinson, Abbott Laboratories, and Pfizer. G.V.D. has received research support during the past 12 months from Abbott Laboratories, Bayer Pharmaceutical Co., Astra-Zeneca, Sanofi-Aventis, and Merck & Co. D.A.B. has received research support from Dade Behring. M.P.W. received research support from Becton Dickinson, Dade Behring, and Bayer during the study period. The rest of us had no conflict of interest.
Published ahead of print on 25 July 2007. ![]()
|
|
|---|