Journal of Clinical Microbiology, January 2008, p. 171-176, Vol. 46, No. 1
0095-1137/08/$08.00+0 doi:10.1128/JCM.00877-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Department of Molecular Cell Biology,1 Department of Parasitology, Leiden University Medical Center, P.O. Box 9600, 2300 RC Leiden, The Netherlands2
Received 26 April 2007/ Returned for modification 8 August 2007/ Accepted 10 October 2007
|
|
|---|
|
|
|---|
Diagnosis of the infection is classically based on the detection of parasite eggs in urine or in feces. However, this method has several disadvantages. The number of excreted eggs is often low and shows a high day-to-day fluctuation. Therefore, stool or urine examination needs to be repeated several times. Alternatively, detection of antibodies is a highly sensitive and specific method to diagnose schistosomiasis. High antibody responses are generally seen with travelers originating from areas where schistosomiasis is not endemic. However, in immigrant travelers with a life-long history of exposure, antibody responses are mostly moderate to low. Some may even become serologically negative, while still excreting viable eggs. In addition, antibody levels are not associated with the actual worm burden and remain unaffected by treatment of the infection. Consequently, serology mostly gives straightforward answers for patients tested within months after their first exposure, but data are difficult to interpret for those who have a history of previous infection (23). A sensitive, serum- or urine-based test demonstrating active Schistosoma infection would be valuable in these cases.
Assays for the detection of Schistosoma circulating antigens (adult worm gut-associated antigens) seem very promising, as serum levels of circulating anodic antigen (CAA) are related to actual worm burden and rapidly decrease following drug treatment (reviewed by van Lieshout et al. [23]). The current monoclonal antibody-based enzyme-linked immunosorbent assay (ELISA) determines serum CAA levels for all human Schistosoma species with virtually 100% specificity (7); its value in (sero)epidemiological studies dealing with populations with moderate- to high-intensity infections has been demonstrated previously (17). However, the CAA-ELISA still lacks sensitivity when testing light infections, e.g., in the group of international travelers (24). Moreover, due to the relatively high complexity of the test, it lacks robustness if performed only occasionally for single case identification. This may hamper implementation of the CAA-ELISA within clinical routine diagnostic settings.
In this report we describe the development of a rapid test strip designed to detect CAA in single serum samples by utilizing the same genus-specific monoclonal antibodies as the current standard CAA-ELISA. Sensitivity of the test is increased by up-converting phosphor technology (UPT) using instrument-assisted assay analysis (5). The important advantages of UPT reporters compared to other (conventionally applied) fluorescent labels include high sensitivity (no autofluorescence of other biological materials), long shelf life, permanent record (no fading), and low costs (28). Various bioassays, including lateral flow (LF) devices, that demonstrate the usefulness of UPT reporters have been developed (11, 12, 14, 15, 21, 27, 29).
We analyzed the diagnostic performance of the novel UPT-LF assay for the detection of CAA with a set of well-characterized serum samples originating from different areas of schistosomiasis endemicity, as well as a series of serum samples from schistosomiasis-suspected individuals submitted to our diagnostic laboratory for serological analysis. The latter group comprised a heterogeneous mixture of Dutch citizens (region where schistosomiasis is not endemic) that may have been infected during recent travel and immigrants with an unclear history of exposure. Schistosoma infections are expected to be very rare in this population. The correlation between the results of the routine antibody test, the standard CAA-ELISA, and the new UPT-LF assays was determined.
|
|
|---|
![]() View larger version (44K): [in a new window] |
FIG. 1. Illustration of the UPT-LF CAA test strip and UPT-LF CAA assay. The immunoassay was performed in microtiter plate wells. After binding of the antigen to the CAA-specific UPT reporter, lateral flow was initiated by placing LF strips in the microtiter plate wells. The LF strips have a width of 4 mm and a total length of 7.8 cm and contain a CAA-specific test line and a UPT control line. Upon completion of the flow, the LF strips were interrogated with infrared light, revealing the deposition of the UPT reporter along the strip. Examples of scans obtained with positive and negative UPT test line signals are shown.
|
CAA reporter conjugate.
Mouse monoclonal anti-CAA antibody 147 (LUMC, Parasitology) was coupled to 400 nm UPT reporter particles (OraSure Technologies, Inc., Bethlehem, PA) as described earlier (4), utilizing a conjugation load of 25 µg antibody per mg reporter particles. The resulting UPTM
CAA conjugate is stable for at least 6 months in a refrigerator at 4°C. Directly before use, the stock solution of the conjugate was homogenized, and a desired amount was sonicated (1 min, water bath sonicator, 100 W) and diluted in assay buffer (see below) to 1 ng/µl. The UPT assay buffer used in these assays is a buffer developed for direct testing of serum and plasma samples (6): 100 mM HEPES, pH 7.2, 270 mM NaCl, 0.5% (wt/vol) Tween 20, 1% (wt/vol) bovine serum albumin.
UPT-LF CAA assay.
A schematic illustration of the assay is depicted in Fig. 1. In detail, the UPT-LF CAA assay consists of four steps: (i) mixing of 100 µl assay buffer containing 1 ng/µl UPTM
CAA conjugate with 10 µl CAA test sample in a microtiter plate well or 1.5-ml tube; (ii) incubation in an orbital shaker (1,200 rpm) at 37°C for 60 min; (iii) addition of the mixture to the sample pad of a CAA-specific UPT-LF strip placed in a well of a microtiter plate (sample pad down), allowing chromatography to continue until strips are dry (between 30 min and 1 h); and (iv) scanning of the strips by using a modified Packard Fluorocount meter (4). Test line signals were normalized to control line signals of each individual strip; the results are expressed as ratio signals (UPT value). The cutoff threshold for the UPT-LF CAA assay was determined with 30 Dutch blood bank donors without a known history of schistosomiasis.
CAA-ELISA. CAA concentrations were determined in serum as described previously (16). Serial dilutions of the trichloroacetic acid (TCA)-soluble fraction of Schistosoma adult worm antigen (AWA-TCA) were assayed simultaneously on each ELISA plate to calculate CAA concentrations. AWA-TCA contains approximately 3% (wt/wt) CAA. The lower limit of detection (LOD) of the CAA-ELISA is 10 pg CAA/ml.
Sample pretreatment and standards. All serum samples tested by CAA-ELISA or UPT-LF were pretreated with TCA to remove interfering proteins and to dissociate immune complexes (8). After TCA extraction, the samples were subjected to a neutralization step, resulting in a 1:4 dilution compared to the original serum. A standard serial dilution of AWA-TCA was used to determine the analytical sensitivity of the UPT-LF assay.
Characterized samples from areas of schistosomiasis endemicity. For an initial laboratory evaluation of the UPT-LF assay, four sets (10 serum samples each), representing different Schistosoma infection levels, were selected. These well-defined samples were collected earlier as part of epidemiological studies on the prevalence and intensity of S. mansoni infections. Ten low-reactive serum samples were selected from banked specimens from a study performed at Saramacca, Surinam, an area with low S. mansoni transmission (22). The CAA concentrations as determined by ELISA ranged from 190 pg/ml to 7.5 ng/ml (median, 1.9 ng/ml). Ten moderate- and ten high-reactive serum samples were selected from banked specimens from a study performed at Mwanza, Tanzania (13), an area with moderate to high transmission. The moderate-intensity-infection set had CAA concentrations ranging from 3.5 to 28.5 ng CAA/ml (median, 11.1 ng/ml), whereas in the high-intensity-infection set the CAA concentrations ranged from 34.5 to 114.0 ng CAA/ml (median, 66.2 ng/ml). The specificity of the assay was tested with 10 serum samples selected from banked specimens from a study performed in an area of Senegal where schistosomiasis is not endemic (16). No Schistosoma eggs were found in the latter group following extensive stool and urine examination, and all samples tested negative in the CAA-ELISA.
Schistosomiasis-suspect samples from areas of schistosomiasis nonendemicity.
A second evaluation study of the UPT-LF assay was performed with 166 serum samples sent to our reference laboratory by general practitioners or other diagnostic laboratories within The Netherlands. All samples originated from individuals living within The Netherlands and were sent with the request of Schistosoma antibody testing because of suspected schistosomiasis. This group represents recent travelers or immigrants from a region of schistosomiasis endemicity. Routine testing implied screening of serum by two "in-house" antibody assays, an immunofluorescence assay for anti-adult worm antibodies (IFA-AWA) and an ELISA for anti-soluble egg antigen antibodies (ELISA-SEA) (23). IFA-AWA titers of
16 and ELISA-SEA titers of
32 were considered positive. Based on serology outcome, the group was divided into an antibody-negative group (n = 43) and an antibody-positive group (n = 123). Samples were tested in the CAA-ELISA for comparison with the UPT-LF assay.
|
|
|---|
![]() View larger version (15K): [in a new window] |
FIG. 2. Analytical sensitivity of the UPT-LF assay. (A) Comparison of UPT-LF with ELISA. (B) LOD of UPT-LF and indication of positive cutoff threshold (UPT-LF value of 0.081) and negative cutoff threshold (UPT-LF value of 0.053) for CAA detection in serum TCA extracts.
|
0.081 (highest negative plus 2 SD), and the cutoff threshold below which a sample was designated "CAA negative" was set to
0.053 (average plus 2 SD). Samples resulting in ratio values between 0.053 and 0.081 were designated "potentially positive." The positions of the positive and negative cutoff thresholds determined with TCA extracted sera are indicated in Fig. 2B. Evaluation of UPT-LF with a set of characterized samples from areas of schistosomiasis endemicity. An evaluation of the UPT-LF assay with a set of 40 characterized epidemiological serum samples indicated excellent correlation with infection status as determined by the CAA-ELISA (Fig. 3). The UPT values of the 10 samples from the Surinam area of low schistosomiasis endemicity ranged from 0.30 to 3.3 (median, 2.1). The lowest measured signal within this group was far above the positive cutoff threshold of 0.081. The sera from Tanzania with moderate and high intensities of infection ranged in UPT value from 3.0 to 5.8 (median, 4.7) and from 3.5 to 6.4 (median, 4.8), respectively. The UPT value of the Schistosoma-negative control group from an area of schistosomiasis endemicity in Senegal ranged from 0.010 to 0.026 (median, 0.015), well below the negative cutoff threshold of 0.053.
![]() View larger version (17K): [in a new window] |
FIG. 3. Evaluation of the UPT-LF CAA assay. Low, moderate, high, and controls each represent a set of 10 defined serum samples from different areas of schistosomiasis endemicity. Antibody negatives (n = 43) and antibody positives (n = 123) represent a selection of 166 collected serum samples of suspected imported cases.
|
|
View this table: [in a new window] |
TABLE 1. UPT-LF and ELISA analysis of 166 clinical serum samples suspected of schistosomiasisa
|
|
|
|---|
Recently a rapid immunochromatography test strip for the detection of Schistosoma circulating cathodic antigen in urine has been introduced. This simple-to-use field test has shown its value in epidemiological surveys, particularly in remote areas of S. mansoni endemicity (19, 20). The applied immunochromatography or LF format is suitable for single case diagnosis. In this paper, we present a UPT-LF assay for detection of CAA that was initially designed for lab-based applications. Serum assays are appropriate and readily available for this setting, whereas CAA detection in general has a higher sensitivity and specificity than circulating cathodic antigen detection (16). The UPT-LF assay, as with the ELISA, is performed on TCA extracted serum samples and utilizes the same mouse monoclonal antibody to build a CAA immunosandwich (7). The ELISA detection format is replaced with an LF-based detection of CAA-bound UPT-reporter particles.
Analytical sensitivity: comparison of UPT-LF with ELISA. The analytical sensitivity of UPT-LF indicated an LOD of 0.5 pg CAA per ml, more than 10-fold better than the 10 pg/ml for the ELISA. Assuming a CAA molecular mass between 10 and 100 kDa and a 10-µl sample volume, this translates to a detection level between 106 and 105 target molecules per LF strip (in agreement with LODs as determined in other UPT-LF assays [5, 29]).
The dynamic range of the UPT-LF assay is 4 orders of magnitude, spanning CAA concentrations of 0.5 pg/ml through 500 pg/ml. The 0.5-pg/ml value represents the LOD of the UPT-LF assay whereas 500 pg/ml indicates a UPT plateau value. It is possible to perform quantitative measurements above 500 pg/ml by increasing the number of UPT reporter particles in the assay; this, however, negatively affects the LOD. In the very few cases (in regions where schistosomiasis is not endemic) that UPT-LF diagnosis indicates concentrations above 500 pg/ml and further quantification is required, e.g., for determination of accurate drug treatment, the ELISA is applicable up to concentrations of 30,000 pg/ml. Alternatively, the UPT-LF analysis could be repeated simply with a dilution of the TCA-treated sample.
The UPT-LF cutoff threshold value used in this study was determined from samples obtained from 30 Dutch blood bank donors. Although UPT values were presented as ratio values (allowing interassay comparison [4]), an indeterminate group (potentially positive) was included because assays were performed with different batches of LF strips. All LF strips were produced manually in small batches of 40 strips. The average UPT value of 0.036 (median, 0.034) with an SD of 0.0087, determined for the blood bank donors, indicates that a negative cutoff threshold of 0.053 is not unrealistic. When using this 0.053 threshold, only one of the 30 blood bank samples and one of the serology-negative samples scored CAA positive. A precise assay cutoff threshold needs to be determined in future with a higher number of negative controls (from regions of schistosomiasis endemicity as well as nonendemicity) by using strips from large production batches.
Evaluation of the UPT-LF assay. A retrospective analysis of four defined sets of epidemiological samples demonstrated 100% diagnostic agreement between the CAA-ELISA and the UPT-LF. For the three Schistosoma-positive serum sets, this result was not surprising, as all samples were selected based on CAA-ELISA results and the UPT-LF assay obviously demonstrated a better analytical sensitivity. Although the UPT-LF assay initially was not designed to function as a quantitative assay, excellent correlations were seen in these three serum sets between the UPT ratio and the serum CAA concentration as determined by the CAA-ELISA. Only the differentiation between moderate- and high-intensity-infection sets was less pronounced with the UPT-LF assay than with the CAA-ELISA. This is a consequence of reaching a plateau value in the UPT ratio above 500 pg CAA per ml. In addition, the specificity of the UPT-LF assay was found to be high, as samples collected in a region of Senegal where schistosomiasis is not endemic all tested clearly negative.
The performance of the UPT-LF assay was analyzed in 166 serum samples referred to our diagnostic laboratory because of suspected schistosomiasis. These samples were tested routinely by antibody serology, which is the standard procedure to identify a Schistosoma infection in a setting where schistosomiasis is not endemic (25). The UPT-LF assay identified 27% more CAA-positive samples than the ELISA and 140% more cases when including the potentially positive group. With respect to the results obtained with the negative controls, the majority of the samples from the potentially positive group are likely to be truly positive. Similarly to previous publications indicating antibody detection to be a poor indicator of an active Schistosoma infection (25), UPT values did not correlate with absolute serology values. Although more active cases were identified by the UPT-LF assay than by the ELISA, the majority of antibody-positive individuals had no detectable CAA levels. Based on previous data, there is no indication that lack of detectable CAA serum levels is related to the Schistosoma species involved (23). Unfortunately, data regarding geographical location and time period of possible exposure, microscopic stool or urine examination, and clinical signs of infection were mostly not available. The antibody-positive, CAA-negative group may include several individuals where adult worms are not yet fully developed (very recent infection), cases with only minimal antigen production (low number of parasites), or individuals who may have lost their infection. Additional studies are planned to further explore the clinical diagnostic value of the UPT-LF assay, evaluating anamnestic data and following patients after antischistosomal treatment.
Conclusions. The UPT-LF assay is a straightforward, easy-to-use single-sample test for detection of CAA in TCA extracted serum samples. It shows higher analytical sensitivity and better accuracy than the CAA-ELISA when testing clinical samples. Although precise determination of the UPT-LF cutoff threshold value awaits a study with a wider selection and higher number of negative controls and more controlled LF strip production (large batches), it can be concluded that the UPT-LF assay is of potential value in laboratory diagnosis of Schistosoma infections.
Part of this work was supported by NIH grant UO1-DE-017855.